MRI glossary
Key MRI terms explained: what each one means, why it’s used, and a worked example. Use the contents to jump to a category, or read everything in order.
On this page
Physics & basicsSequences & acronymsAdvanced imagingAcquisition & reconstructionArtifactsContrastSafetyClinical & navigationPhysics & basics
- T1 (longitudinal relaxation)
- The time constant for spins to realign with the main magnetic field after excitation.
- Why: Tissues recover at different rates, giving the contrast that distinguishes fat, fluid, and soft tissue on T1-weighted images.
- Example: On a T1-weighted brain scan, fat and subacute blood appear bright while CSF is dark.
- T2 (transverse relaxation)
- The time constant for loss of phase coherence among spins in the transverse plane.
- Why: Fluid and many pathologies have long T2 and appear bright, making T2 weighting sensitive to oedema and disease.
- Example: A T2-weighted image shows CSF and joint effusions as bright white.
- T2* (T-two-star)
- Transverse decay including local field inhomogeneity, faster than true T2.
- Why: Sensitive to iron, calcium, and blood products because these distort the local field.
- Example: A T2*/GRE or SWI sequence blooms around cerebral microbleeds.
- TR (repetition time)
- Time between successive excitation pulses.
- Why: Controls T1 weighting: short TR increases T1 contrast; long TR reduces it.
- Example: A short TR (~500 ms) is chosen for T1-weighted spin echo.
- TE (echo time)
- Time from excitation to readout of the echo.
- Why: Controls T2 weighting: long TE increases T2 contrast.
- Example: A long TE (~90 ms) yields a heavily T2-weighted image.
- Flip angle
- The angle through which the net magnetisation is tipped by the RF pulse.
- Why: Tunes contrast and signal, especially in gradient-echo; low angles allow rapid imaging.
- Example: A 10–15° flip angle is used in fast 3D gradient-echo T1 sequences.
- k-space
- The raw spatial-frequency data matrix that is Fourier-transformed into the image.
- Why: Its centre encodes contrast and its periphery encodes detail; how it is filled defines speed and artefact behaviour.
- Example: Partial-Fourier acquisition samples just over half of k-space to shorten scan time.
- SNR (signal-to-noise ratio)
- Ratio of signal to background noise in an image.
- Why: The core quality trade-off: higher field, larger voxels, and more averages raise SNR at a cost in time or resolution.
- Example: Doubling voxel volume roughly doubles SNR but halves spatial detail.
Sequences & acronyms
- Spin echo (SE) / Turbo & Fast SE (TSE/FSE)
- A 90° excitation followed by a 180° refocusing pulse; TSE/FSE uses multiple refocusing pulses per TR.
- Why: Refocusing corrects field inhomogeneity for clean T1/T2 contrast; TSE/FSE speeds this up dramatically.
- Example: T2 TSE of the spine acquires several lines per excitation, cutting scan time from minutes to under a minute.
- Gradient echo (GRE)
- Echo formed by gradient reversal instead of a 180° pulse, usually with a low flip angle.
- Why: Fast, and sensitive to susceptibility; the basis of most 3D and dynamic imaging.
- Example: Breath-hold GRE captures the arterial phase of a liver study.
- MPRAGE / BRAVO / 3D T1 TFE
- Vendor names for an inversion-prepared 3D gradient-echo T1 sequence (Siemens MPRAGE, GE BRAVO, Philips 3D TFE).
- Why: Produces isotropic sub-millimetre T1 volumes that reformat in any plane, the workhorse for brain volumetry and neuronavigation.
- Example: A post-contrast MPRAGE is loaded into a surgical navigation system to plan a tumour resection.
- FLAIR (fluid-attenuated inversion recovery)
- A T2 sequence with an inversion pulse that nulls CSF signal.
- Why: Suppressing bright CSF makes periventricular and cortical lesions conspicuous.
- Example: FLAIR reveals MS plaques adjacent to the ventricles that T2 would obscure.
- STIR (short tau inversion recovery)
- Inversion recovery with a TI chosen to null fat.
- Why: Robust fat suppression even where the field is uneven; highlights oedema and marrow disease.
- Example: STIR of the ankle shows bone-marrow oedema as bright signal against suppressed fatty marrow.
- SPACE / CUBE / VISTA (3D TSE)
- Vendor names for isotropic 3D turbo-spin-echo (Siemens SPACE, GE CUBE, Philips VISTA).
- Why: A single high-resolution volume reformats into any plane, reducing the number of 2D series needed.
- Example: A 3D SPACE FLAIR of the brain is reformatted sagittally and coronally from one acquisition.
- DWI (diffusion-weighted imaging) & ADC
- Measures random water motion; the ADC map quantifies it, separating true restriction from T2 shine-through.
- Why: Restricted diffusion flags acute stroke, abscess, and hypercellular tumour.
- Example: Acute infarct is bright on high-b DWI and dark on ADC.
- SWI (susceptibility-weighted imaging)
- High-resolution GRE combining magnitude and phase to maximise susceptibility contrast.
- Why: Exquisitely sensitive to blood products, calcium, and small veins.
- Example: SWI detects microhaemorrhages in diffuse axonal injury after trauma.
Advanced imaging
- MRA & TOF (MR angiography / time-of-flight)
- Vascular imaging; time-of-flight exploits inflowing unsaturated blood to appear bright without contrast, while contrast-enhanced MRA uses gadolinium.
- Why: Visualises arteries and veins non-invasively to assess stenosis, aneurysm, or occlusion. TOF needs no contrast, useful when it is contraindicated.
- Example: A 3D TOF circle-of-Willis study detects an unruptured intracranial aneurysm without any injection.
- Perfusion (DSC / DCE / ASL)
- Measures tissue blood flow: DSC tracks a gadolinium bolus through susceptibility signal loss, DCE tracks T1 enhancement, and ASL magnetically labels arterial blood without contrast.
- Why: Quantifies vascularity to grade tumours, assess stroke penumbra, and distinguish recurrence from treatment change.
- Example: Elevated cerebral blood volume on DSC perfusion suggests high-grade glioma rather than radiation necrosis.
- MR spectroscopy (MRS)
- Measures the chemical composition of tissue, displaying metabolite peaks rather than an image.
- Why: Adds biochemical information: metabolite ratios help characterise tumours and metabolic disease.
- Example: A raised choline-to-NAA ratio in a brain lesion points toward an aggressive tumour.
- DTI & tractography
- Diffusion tensor imaging measures the direction of water diffusion to map white-matter fibre tracts, which tractography renders in 3D.
- Why: Shows the relationship of eloquent tracts to a lesion, guiding safe surgical approach.
- Example: Tractography displays the corticospinal tract displaced by a tumour so the surgeon can avoid it.
- MRCP (MR cholangiopancreatography)
- A heavily T2-weighted sequence that makes static fluid in the biliary and pancreatic ducts bright.
- Why: Non-invasively maps the ducts to detect stones, strictures, and anatomical variants without ERCP.
- Example: MRCP shows a filling defect in the common bile duct from an obstructing gallstone.
- Phase-contrast & flow imaging
- Uses velocity-encoding gradients so signal phase is proportional to flow velocity.
- Why: Quantifies direction and speed of flow (blood or CSF) for shunts, valves, and hydrocephalus.
- Example: Phase-contrast imaging measures CSF flow across the aqueduct to assess a suspected obstruction.
Acquisition & reconstruction
- 2D vs 3D acquisition
- 2D excites one slice at a time; 3D excites a whole slab and phase-encodes through it.
- Why: 3D gives thin, contiguous, isotropic voxels that reformat in any plane, essential for navigation and fine anatomy; 2D is faster per slice and often higher-contrast.
- Example: A 3D isotropic 1 mm T1 volume can be resliced axially, sagittally, and coronally without re-scanning.
- Isotropic voxel
- A voxel with equal dimensions in all three directions (e.g. 1×1×1 mm).
- Why: Reformats in any plane with no loss of resolution, the prerequisite for high-quality multiplanar and 3D rendering.
- Example: An isotropic MPRAGE feeds cleanly into surgical planning and CT-fusion software.
- Parallel imaging (SENSE / GRAPPA / ASSET)
- Uses multiple receiver coil elements to undersample k-space and reconstruct faster.
- Why: Shortens scans or shrinks echo trains, trading some SNR for speed.
- Example: An acceleration factor of 2 halves the phase-encoding steps of a breath-hold sequence.
- Fat suppression
- Techniques (spectral saturation, Dixon, STIR) that remove or separate fat signal.
- Why: Prevents bright fat from masking enhancement or oedema and reduces chemical-shift artefact.
- Example: Dixon reconstruction outputs in-phase, opposed-phase, water-only, and fat-only images from one acquisition.
- Image fusion / co-registration
- Spatially aligning two datasets (e.g. MRI with CT, or MRI with PET) into a common coordinate frame.
- Why: Combines MRI soft-tissue detail with CT bony or metabolic data; underpins surgical navigation and radiotherapy planning.
- Example: A 3D MRI tumour volume is fused with a planning CT so bone and lesion appear in one workspace.
Artifacts
- Motion artifact
- Ghosting or blurring caused by patient movement, breathing, or pulsation during acquisition.
- Why: Recognising it prevents mis-reading and guides mitigation (breath-holds, triggering, faster sequences).
- Example: Respiratory motion smears the liver edge along the phase-encode direction.
- Chemical-shift artifact
- Fat and water resonate at slightly different frequencies, mis-mapping fat along the frequency-encode axis.
- Why: Explains dark/bright borders at fat–water interfaces and can be reduced with wider bandwidth or fat suppression.
- Example: A black-and-white rim outlines the kidney against surrounding fat.
- Susceptibility artifact
- Signal loss and geometric distortion from local field disruption, often near metal or air.
- Why: Warns of metal implants and is exploited by SWI; mitigated with SE-based or metal-artefact-reduction sequences.
- Example: Dental amalgam blooms and distorts the adjacent facial anatomy.
- Aliasing (wrap-around)
- Anatomy outside the field of view folds onto the opposite side of the image.
- Why: Recognising it allows correction by enlarging FOV or oversampling.
- Example: A patient’s arm wraps into the abdomen on a small-FOV scan.
Contrast
- Gadolinium-based contrast agent (GBCA)
- A paramagnetic agent that shortens T1, brightening enhancing tissue.
- Why: Highlights vascular structures, breakdown of the blood–brain barrier, inflammation, and tumour.
- Example: A meningioma enhances avidly on post-gadolinium T1 imaging.
- Macrocyclic vs linear GBCA
- Two structural classes of gadolinium chelate; macrocyclic agents hold gadolinium more stably.
- Why: Stability governs safety, especially the risk of gadolinium retention, driving agent choice.
- Example: A macrocyclic agent is preferred for a patient needing repeated contrast studies.
- Dynamic contrast-enhanced (DCE) imaging
- Rapid repeated imaging during and after contrast injection to capture enhancement over time.
- Why: Enhancement kinetics help characterise lesions as benign or malignant.
- Example: A breast lesion with rapid wash-in and wash-out raises suspicion of malignancy.
Safety
- MR Safe / Conditional / Unsafe
- Standardised labels (ASTM F2503) for how an item behaves in the MR environment. MR Safe: nonmetallic, nonconducting and nonmagnetic, with no known hazards in any MR environment. MR Conditional: may be used in the MR environment only when the manufacturer’s specified conditions are met. MR Unsafe: poses unacceptable risks in the MR environment (ACR Manual on MR Safety, 2026, pp. 78, 80, 81).
- Why: Determines whether an implant can be scanned and under what conditions, a core screening decision.
- Example: An MR Conditional pacemaker may be scanned only within specified field and SAR limits.
- SAR (specific absorption rate)
- RF energy deposited in tissue, measured in W/kg, causing heating.
- Why: SAR limits manage whole-body and local heating, but they do not prevent burns. Loops, skin-to-skin contact and badly placed conductors can burn a patient even in Normal Operating Mode, so careful positioning and insulating pads are needed whatever the SAR reading (ACR Manual on MR Safety, 2026, pp. 63, 64, 69). High-SAR sequences may be modified for at-risk patients.
- Example: The scanner lengthens TR to keep a TSE sequence within SAR limits for a patient with an implant.
- dB/dt (rate of change of the gradient field)
- How fast the magnetic field at a point changes as the gradients switch, in tesla per second (T/s). Slew rate is a different quantity: how fast a gradient ramps up, in tesla per metre per second (T/m/s). dB/dt rises with slew rate and varies with position in the bore, so it can be highest away from isocentre (ACR Manual on MR Safety, 2026, pp. 10, 70; RANZCR MRI Safety Guidelines v3.0, s. 3.2).
- Why: Rapid switching can cause peripheral nerve stimulation; limits protect patient comfort and safety.
- Example: A fast diffusion sequence is throttled to avoid stimulating the patient’s peripheral nerves.
- Projectile (missile) effect
- Ferromagnetic objects being violently pulled toward the magnet bore.
- Why: A leading cause of serious MRI accidents; drives strict zoning and ferromagnetic screening.
- Example: A steel oxygen cylinder brought into the scan room is drawn into the bore.
Clinical & navigation
- Fiducial markers
- Reference points (skin markers or anatomical landmarks) used to register imaging to the patient.
- Why: Registration accuracy of the navigation system depends on well-placed fiducials.
- Example: Adhesive scalp fiducials visible on the STEALTH scan are matched to the patient’s head at the start of surgery.
- MRI–CT fusion for planning
- Co-registering a 3D MRI with a planning CT to combine soft-tissue and bony/dose information.
- Why: Surgery and radiotherapy need MRI lesion detail aligned to CT geometry used by planning and navigation systems.
- Example: A radiotherapy plan overlays the MRI-defined tumour on the CT used for dose calculation.
Educational reference only, not a substitute for local imaging protocols, device labelling, or clinical judgement. Vendor sequence names are trademarks of their respective manufacturers.
More in Physics and foundations
Next in Physics and foundations: MRI high-yield summary
Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.