MRI question bank: Free sampler

MRI question bank: Free sampler

30 exam-style single-best-answer questions across MRI physics, safety, anatomy, artifacts, and contrast. FRCR / ABR / ARRT / MRSO style.

On this pagePhysicsSafetyAnatomyArtifactsContrastSources

How to use: pick an answer to see the explanation. This is a free sampler. The full question bank, with topic sets and progress tracking, is coming with Pro.

Physics

Start the questions (12)

1 Which value is closest to the resonant frequency of hydrogen protons at 1.5 T?

Correct: C

The Larmor frequency equals the gyromagnetic ratio times B0: 42.58 MHz/T × 1.5 T ≈ 63.9 MHz. About 43 MHz is the value at 1 T, 128 MHz at 3 T, and 21 MHz at about 0.5 T.

2 Simple fluid such as CSF is bright on T2-weighted images because it has a:

Correct: A

Free water has a long T2, so its transverse magnetisation decays slowly and much signal remains at the long TE of a T2-weighted image, while short-T2 tissues such as muscle have faded. A short T1 makes tissue bright on T1-weighted images; CSF has a long T1, which is why it is dark there.

3 Spin–lattice (T1) relaxation describes:

Correct: C

After excitation, spins hand absorbed energy to their surroundings, the lattice, and longitudinal magnetisation regrows exponentially towards equilibrium with time constant T1. Loss of phase between spins, the tempting option, is T2; extra dephasing from field inhomogeneity makes up T2*; tipping by an RF pulse is excitation, not relaxation.

4 What is the main job of the 180° pulse in a spin-echo sequence?

Correct: D

The 180° pulse flips the phase each spin has gathered, so spins dephased by a static, uneven field come back into phase at TE and the signal follows T2 rather than T2*. Dephasing from spin–spin interactions is random and cannot be reversed, so T2 decay remains. Tipping magnetisation into the transverse plane is the job of the 90° pulse.

5 A T2-weighted spin-echo series is repeated with two signal averages instead of one. Compared with the first series, the SNR is:

Correct: B

Signal adds directly with each average but random noise adds only as its square root, so SNR rises with the square root of the number of averages: √2 ≈ 1.41 (Graessner; Gruber; Schmitz). Twice the SNR is the tempting answer, but that needs four averages. Scan time doubles. About 0.7 is the effect of doubling the bandwidth.

6 The centre of k-space mainly determines:

Correct: D

Low spatial frequencies near the centre of k-space carry most of the signal and set overall contrast; the outer rows carry the high spatial frequencies that define edges and fine detail. FOV depends on the spacing between k-space samples, and slice position on the RF pulse frequency, not on which part of k-space is filled.

7 In a STIR sequence, the inversion time is chosen to null the signal from:

Correct: A

STIR times the excitation for the moment fat’s recovering magnetisation crosses zero after the inversion pulse, so fat gives no signal. Fat has a short T1, so this TI is short. Nulling cerebrospinal fluid needs a much longer TI and is the job of FLAIR; nulling flowing blood uses a separate black-blood preparation.

8 A protocol is moved unchanged from a 1.5 T to a 3 T scanner. Which change should the technologist expect?

Correct: D

The fat–water frequency difference is about 220 Hz at 1.5 T and 440 Hz at 3 T, so at the same bandwidth per pixel the shift doubles; Huang’s example goes from 0.9 to 1.8 mm. RF power for a given flip angle rises with the square of the field, about fourfold, susceptibility effects grow, and SNR roughly doubles.

9 Restricted diffusion, as in an acute infarct, appears on DWI and the ADC map as:

Correct: B

Water in the infarct moves less, so it keeps signal at high b (bright DWI) and the calculated ADC is low (dark map). High DWI with a high ADC is the tempting choice, but that pattern is T2 shine-through, not restriction.

10 At 1.5 T, at about which TE are fat and water first back in phase after excitation?

Correct: D

Fat and water differ by about 220 Hz at 1.5 T, so they drift apart and realign every 1/220 s, about 4.5 ms: opposed at about 2.2–2.3 ms and back in phase at about 4.4–4.6 ms. At 3 T the cycle halves, so 1.1 and 3.4 ms are opposed-phase echoes there and 2.3 ms is in phase.

11 Why are periventricular plaques easier to see on FLAIR than on T2-weighted fast spin echo?

Correct: A

FLAIR uses a long inversion time to null CSF and a long TE to keep T2 weighting, so lesions beside the ventricles and sulci are no longer hidden by bright fluid. STIR is the inversion method that nulls fat, black-blood preparations null flowing blood, and white matter keeps its signal on FLAIR.

12 Time-of-flight MRA makes vessels bright mainly through:

Correct: D

Rapid repeated pulses saturate stationary tissue in the slab, while blood flowing in has not yet been excited and arrives with full magnetisation, so it is bright. Bipolar phase encoding is phase contrast, gadolinium T1 shortening is contrast-enhanced MRA, and the T2/T1 ratio drives balanced SSFP angiography.

Safety

13 Which zone is the actual magnet room where the scanner resides?

Correct: C

In the ACR four-zone model, Zone IV is the MR scanner room within its walls and contains the projectile area. Zone III is the access-controlled area around it, including the operator console; Zone II is the interface used for waiting and screening; Zone I is open to the public.

14 Under MHRA guidance, which concerns need particular attention when a neonate is scanned?

Correct: B

MHRA lists neonates and infants among patients with compromised thermoregulation, so RF heating and room conditions need care, and names neonates as a group of particular concern for acoustic noise, with dedicated neonatal hearing protection. Nerve stimulation and static-field sensory effects are not singled out for neonates, and MRI uses no ionising radiation.

15 RANZCR recommends mapping with a hand-held meter, using the maker’s site plans. What is being mapped?

Correct: C

RANZCR recommends mapping the fringe field with a hand-held gaussmeter alongside the maker’s site plans, paying attention to areas around the scan room and anywhere the public could meet fields above 0.5 mT (RANZCR’s 5 gauss line; IEC 60601-2-33 Ed 4.0 (2022) and the ACR Manual on MR Safety (2026) use the 0.9 mT, 9 gauss line for the MR Environment); maps are kept and shown to staff. Sound levels need a sound level meter, and RF output is reported by the scanner itself.

16 What produces the loud knocking heard during image acquisition?

Correct: D

MHRA explains that switching currents in the gradient coils, inside the strong static field, exert forces that make the coils vibrate like a loudspeaker, the Lorentz force described by Wu et al. This is why patients are warned about the noise and given hearing protection. The RF coil and the cryogen system are not the source.

17 What happens when an MRI magnet quenches?

Correct: B

In a quench the superconducting windings become resistive, the very large current is dissipated as heat and the remaining liquid helium boils off rapidly, so the gas must be vented outside; MHRA describes the rapid boiling and pressure rise. Shim, RF amplifier and gradient faults degrade or stop scanning but do not end superconductivity.

18 What distinguishes an item labelled MR Safe from one labelled MR Conditional?

Correct: D

MR Safe means no known hazards in all MR environments and is reserved for non-metallic, non-conducting, non-magnetic items such as a plastic or cotton object; RANZCR uses the ASTM definition. Testing up to 3 T describes MR Conditional conditions, reprogramming applies to some active implants, and a metal item can at best be MR Conditional.

19 Why are hair extensions asked about at MRI screening?

Correct: A

MHRA notes that some hair extensions contain metal, so they should be tested with a hand-held magnet before the MR Environment and their heating risk considered. Ferromagnetic hair grips are among the objects everyone removes before entering. The hazard is the metal, not glue or dye.

20 Why do MR units run regular drills for events such as a quench or a cardiac arrest in Zone IV?

Correct: C

The ACR Manual (2026) advises regular training and drills on emergency magnet-off procedures and mock code exercises with debriefing, because these are high-stress events with MR-specific hazards. The CAMRT CPR guideline notes resuscitation skills deteriorate over time and recommends yearly recertification, which drills do not replace.

Back to contents

Anatomy

21 On a sagittal lumbar spine MRI of an adult, at about which vertebral level does the conus medullaris normally end?

Correct: B

In adults the conus most often ends at about L1, with a normal range from about T12 to L2 in a large MRI series. S2 is the usual end of the thecal sac, not the cord, and L3 is at or beyond the lower limit reported for adults.

22 On a T1-weighted brain image, CSF appears dark mainly because CSF has:

Correct: D

With a short TR, tissues with a long T1 regain little longitudinal magnetisation before the next excitation and give low signal, so CSF is dark while short-T1 fat is bright. CSF has a long T2 and plenty of water protons, which is why it is bright on T2-weighted images.

23 On axial liver MRI, which structure lies in the plane that divides the liver into right and left halves for surgical planning?

Correct: B

Cantlie’s line runs from the gallbladder fossa to the inferior vena cava along the middle hepatic vein, dividing the right and left hemilivers. The right hepatic vein divides the right liver into anterior and posterior sectors, and the falciform ligament separates segment IV from segments II and III within the left liver.

24 A colloid cyst at the roof of the third ventricle blocks the interventricular foramina. On T2 images, which ventricles are expected to enlarge?

Correct: A

The interventricular foramina (of Monro) connect each lateral ventricle to the third ventricle, which drains through the aqueduct into the fourth. A colloid cyst at the foramina traps CSF in both lateral ventricles, so they dilate, while the fourth ventricle beyond the block does not. The case reports cited describe exactly this lateral ventricular dilatation.

Back to contents

Artifacts

25 An axial T2-weighted abdominal series uses anterior–posterior phase encoding and left–right frequency encoding. Along which image direction will breathing ghosts repeat?

Correct: C

Breathing changes the signal between phase-encoding steps taken one TR apart, so the error spreads as ghosts along the phase axis, here anterior–posterior. Each frequency-encoded echo lasts milliseconds, too short for breathing to matter, so ghosts do not run left–right. Swapping the two directions moves the ghosts rather than removing them.

26 Aliasing (wrap-around) occurs when:

Correct: B

The phase-encoding steps define a field of view; tissue beyond it is undersampled, cannot be told apart from tissue inside, and folds onto the opposite side. Readout oversampling prevents it in the frequency direction at no time cost. A FOV larger than the body part removes wrap, a narrow bandwidth causes chemical shift, and motion causes ghosting.

27 Diffusion-weighted EPI of the brain shows warping and signal loss in the frontal lobes just above the sinuses. Which effect is this?

Correct: C

Air in the sinuses and the nearby bone and brain differ sharply in susceptibility, so the local field is distorted. Single-shot EPI reads all of k-space in one long gradient-echo train, so it warps and loses signal there; spin echo is less affected. Chemical shift displaces fat, wrap folds in anatomy outside the FOV, and Gibbs rings at edges.

28 On a short-TE proton density image, intermediate signal appears in the supraspinatus tendon where its fibres run about 55° to B0, and it fades on a long-TE T2-weighted image. The most likely cause is:

Correct: A

Tendon collagen is highly ordered; near 54.7° to B0 its T2 lengthens, so short-TE images show artefactual signal even in healthy tendon. That T2 stays far below the T2 of fluid, so the signal fades on long-TE images, which helps separate it from a tear. Susceptibility, chemical shift and partial volume behave differently.

Back to contents

Contrast

29 In a contrast-enhanced study, why is a T1-weighted series acquired before gadolinium as well as after it?

Correct: C

Fat, methaemoglobin, melanin and concentrated protein can be bright on T1 without contrast. Comparing or subtracting matched pre- and post-contrast T1 images separates true enhancement from this intrinsic brightness. Transmitter gain is calibrated separately, CSF nulling is FLAIR, and a pre-contrast series does not lower SAR.

30 Why does RANZCR’s guidance prefer macrocyclic gadolinium agents to linear ones?

Correct: A

RANZCR’s renal guideline ranks cyclic chelates such as gadoterate, gadoteridol and gadobutrol as more stable than linear ones, and retention, best documented in the dentate nucleus and globus pallidus, is greater with the less stable linear agents. Its MRI Safety Guidelines therefore prefer macrocyclic agents unless a linear agent offers clear clinical benefit.

PRO

Want the full bank?

This is a free 30-question sampler. The complete High Yield Imaging MRI question bank, with topic-specific sets across physics, safety, anatomy, artifacts, contrast and clinical protocols in FRCR / ABR / ARRT / MRSO style, is coming with Pro.

What Pro will include

Related
Back to contents

Sources

Sources (14)

14 sources listed

  1. ACR Manual on Contrast Media & ACR Manual on MR Safety (2026 edition)
  2. ACR MR safety zones and MRMD/MRSO/MRSE roles
  3. IEC 60601-2-33 operating modes/SAR/dB-dt
  4. Westbrook & Talbot, MRI in Practice
  5. McRobbie et al., MRI from Picture to Proton
  6. MHRA, Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, fifth edition, V5.0.3 (September 2026)
  7. RANZCR, MRI Safety Guidelines, Version 3.0 (2021)
  8. RANZCR, Guideline on the use of Gadolinium-containing MRI Contrast Agents in Patients with Renal Impairment, Version 3 (2019)
  9. CAMRT, Best Practice Guidelines 02.01.02, Maintenance of CPR certification (December 2019)
  10. Graessner J. The signal-to-noise indicator or how to navigate the ‘Bermuda triangle’. MAGNETOM Flash 1/2012:144-148
  11. Gruber B, Froeling M, Leiner T, Klomp DWJ. RF coils: a practical guide for nonphysicists. J Magn Reson Imaging 2018;48(3):590-604
  12. Schmitz BL, Aschoff AJ, Hoffmann MHK, Grön G. Advantages and pitfalls in 3T MR brain imaging: a pictorial review. AJNR Am J Neuroradiol 2005;26(9):2229-2237
  13. Huang SY, Seethamraju RT, Patel P, Hahn PF, Kirsch JE, Guimaraes AR. Body MR imaging: artifacts, k-space, and solutions. RadioGraphics 2015;35(5):1439-1460
  14. Wu Z, Kim YC, Khoo MCK, Nayak KS. Evaluation of an independent linear model for acoustic noise on a conventional MRI scanner and implications for acoustic noise reduction. Magn Reson Med 2014;71(4):1613-1620

¹H gyromagnetic ratio ≈ 42.58 MHz/T. Educational use only. Verify all safety limits and contrast dosing against current local protocols and latest guideline editions.

More in Question banks and mocks

Next in Question banks and mocks: Radiologist MRI mock exam

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.

Contents