MR Conditional equipment guide

MR Conditional equipment guide

Choosing ancillary equipment for the MRI suite that is labelled for MR use and good to work with: patient monitors, power injectors, infusion pumps, ventilators/anaesthesia machines, wheelchairs and stretchers. “MR Conditional” is not a quality rating; it defines the exact conditions under which a device may be used. Two conditional monitors can differ hugely in usability, so judge against the criteria below and always read the device’s own conditions of use.

On this pageHow to judge MR Conditional equipmentEquipment categories at a glance
Safety first: MR Safe / MR Conditional / MR Unsafe are ASTM F2503 labelling categories. An MR Conditional device may be used only within its stated conditions: field strength, spatial-gradient limit, and minimum distance from the bore. Verify the label and conditions for every item before it enters Zone IV. Use our implant & device safety checker and safety guidelines pages alongside this guide.
Related

How to judge MR Conditional equipment

Beyond the label, good equipment scores well on:

  • Conditions that fit your suite: rated to your field strength (1.5T / 3T), with a spatial-gradient and minimum-distance limit that works with your room layout. A monitor rated only to a large distance from the bore is awkward in a small room.
  • Fibre-optic, wireless signal paths: for SpO2 and data, fibre-optic probes and wireless telemetry reduce burn risk and cable clutter versus conductive leads.
  • Full parameter set: ECG (with MR gating/MHD-tolerant algorithm), NIBP + invasive pressures (ABP), EtCO2 capnography, SpO2, temperature; ideally on one module that travels with the patient.
  • In-room + control-room display: a slave/repeater display at the console so the trace is visible through loud sequences.
  • Battery life & workflow: long runtime, quick docking, easy cleaning, clear alarms visible when acoustic noise masks audible ones.
  • Interoperability: integrates with your ventilator/anaesthesia machine, injector and PACS/monitoring network.
  • Support & training: local service, spare fibre-optic probes/electrodes, and vendor training on safe placement.

Equipment categories at a glance

CategoryWhat to look for
Patient monitorsFibre-optic SpO2, wireless, full ABP/EtCO2/ECG, control-room slave display, MHD-tolerant gating
Power / contrast injectorsConditional to your field, in-room head with shielded controls, dual-head for gadolinium + saline, safe siting
Infusion / syringe pumpsConditional at a workable distance, long extension sets, clear pressor labelling, battery runtime
Ventilators / anaesthesiaMR Conditional machine, EtCO2/agent monitoring, gas-scavenging, works at stated distance
Wheelchairs / stretchers / transferNon-ferromagnetic build, tested MR Conditional, brakes, IV poles, hover/transfer boards
Ancillary (IV poles, O2 cylinders, laryngoscopes)Aluminium/brass non-magnetic O2 cylinders, MR Conditional IV poles, tested accessories
Availability and MR Conditional status vary by model, region and field strength; confirm the specific model’s conditions of use with the manufacturer.

More in Safety and contrast

Where next: Protocol recommendations

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.