MRI safety guidelines compared

MRI safety guidelines compared

A structured comparison of the principal MRI safety frameworks in current clinical use, with concise summaries, a side-by-side table, and citations to each primary source.

On this pageOverviewAt a glanceWhere they differ, and where they agreeReferenceRevisionSources
Key points
  • The IEC standard defines what a scanner may do, the ACR and MHRA set out how facilities operate around it, and ISMRM/ISMRT statements cover specific clinical scenarios.
  • The ACR uses four zones (I–IV), with Zone IV the scanner room itself; the MHRA instead asks each site to define an MR Controlled Access Area containing the MR Environment.
  • IEC 60601-2-33 Ed 4.0 (2022) and the ACR 2026 manual bound the MR Environment at 0.9 mT (9 G); RANZCR still uses 0.5 mT (5 G), so check your site plan.
  • IEC 60601-2-33 sets three operating modes; whole-body SAR is 2 W/kg in Normal mode and 4 W/kg in First-Level, with head SAR limited to 3.2 W/kg.
  • All the frameworks converge on ASTM implant labelling (MR Safe, MR Conditional, MR Unsafe), rigorous screening, control of the magnet environment, and prevention of RF burns and projectiles.
Explain it simply
  • One rulebook (IEC) sets what a scanner is allowed to do. Two others (ACR and MHRA) set how departments run around it, and expert groups (ISMRM/ISMRT) give advice for particular situations.
  • The ACR divides the area around the scanner into four zones, and Zone IV is the scanner room itself. The MHRA instead asks each site to mark out its own controlled area around the magnet.
  • The IEC and the ACR 2026 manual mark the edge of the magnet area at 0.9 mT (9 G); RANZCR still uses 0.5 mT (5 G). Check which line your site plan uses.
  • The IEC sets three levels of scanner operation and caps SAR, the radio-wave energy the body takes in: 2 W/kg at the normal level, 4 W/kg at the next level, and 3.2 W/kg for the head.
  • All the rulebooks agree on the basics: the standard implant labels (MR Safe, MR Conditional, MR Unsafe), careful screening, control of the area around the magnet, and preventing burns and flying metal objects.

A guide only: follow your departmental protocols and regional requirements.

Related

Overview

Reflects the ACR Manual on MR Safety (2026 edition), MHRA 5th ed. (2026), IEC 60601-2-33 (2022), and ISMRM/ISMRT consensus statements (2020–2022).

Four bodies shape everyday MRI safety practice, and they are complementary rather than competing. The IEC standard defines what a scanner may do; the ACR and MHRA documents set out how facilities should operate around it; and ISMRM/ISMRT consensus statements address specific clinical scenarios in depth.

01 · United States

ACR Manual on MR Safety

American College of Radiology · Best-practice manual · 2026 edition

The principal educational resource for operating a safe MRI service. The 2024 edition was a substantial revision, and the current 2026 edition followed (revision dated 1 January 2026). It establishes the four-zone site model, the MR personnel hierarchy (MRMD, MRSO, MRSE, with Level 1 and Level 2 staff), screening and “full stop / final check” procedures, implant management, emergency response, and, more recently, remote scanning and a patient pre-appointment guide. It is guidance for developing local policy, not a legal standard.

02 · United Kingdom

MHRA: Safety Guidelines for MRI Equipment

Medicines & Healthcare products Regulatory Agency · National guidance · 5th edition (2026)

UK good-practice guidance organised around the four physical hazards: static field, gradients, RF, and cryogens. It asks each site to define an MR Controlled Access Area that contains the MR Environment. IEC 60601-2-33 Edition 4.0 (2022) bounds the MR Environment at 0.9 mT (9 G), and the ACR Manual on MR Safety, 2026 edition, delimits it by the same 9 gauss line. In the UK, 0.5 mT (5 G) remains the legal action level for indirect effects under the Control of Electromagnetic Fields at Work Regulations 2016, to be considered in risk assessments for staff with implanted devices. It also sets out governance roles and referrer responsibilities, and describes adverse-incident reporting within the UK regulatory framework.

03 · International

IEC 60601-2-33

International Electrotechnical Commission · Equipment standard · 2022

The technical standard that governs how MR scanners are designed and labelled. It defines the three operating modes (Normal, First-Level Controlled, and Second-Level Controlled) and the exposure limits for SAR, gradient switching (dB/dt), static field, and acoustic noise on which the clinical guidelines depend. It is referenced by regulators, including the FDA.

04 · International

ISMRM / ISMRT Consensus Statements

Int. Society for Magnetic Resonance in Medicine & technologist section · Peer-reviewed guidance

Topic-specific recommendations published in the Journal of Magnetic Resonance Imaging. They address areas the general manuals treat only broadly: imaging patients with cardiac implantable electronic devices, 7 T safety, gadolinium deposition, safety governance, and scanning during a pandemic.

Back to contents

At a glance

  ACR Manual MHRA (UK) IEC 60601-2-33 ISMRM / ISMRT
RegionUnited StatesUnited KingdomInternationalInternational society
TypeBest-practice manualNational guidanceEquipment standardExpert consensus
Current version2026 ed.5th ed. (2026)20222020–2022
Legal statusVoluntaryNon-statutoryReferenced by regulatorsAdvisory
Access modelFour zones (I–IV)Locally defined controlled access areaFringe-field limit 0.9 mT (Ed 4.0, 2022)Defers to facility
PersonnelMRMD, MRSO, MRSE; Level 1/2Responsible Person, MRSE, Authorised PersonNot applicableRecommends roles
Exposure limitsApplies IEC limitsApplies IEC + EMF regsDefines the limitsWorks within IEC
ImplantsASTM labelling + appendixASTM labellingDefines test basisDevice-specific
Remote scanningAddressedNot a focusNot applicable—
Best used forUS policy and trainingUK policy and procurementLimit definitionsSpecific clinical decisions

Table scrolls horizontally on narrow screens.

Back to contents

Where they differ, and where they agree

Zones versus a locally defined controlled area
The ACR uses a four-zone architecture; the MHRA asks each site to define an MR Controlled Access Area that contains the MR Environment, which IEC 60601-2-33 Ed 4.0 (2022) bounds at 0.9 mT (9 G). RANZCR (v3.0), followed in Australia and New Zealand, still draws Zone III and the pacemaker perimeter at 0.5 mT (5 G). Both contours remain in live use and device labels and site plans may reference either. Always check which one your plan is drawn to. Both exist to keep unscreened people and ferromagnetic objects away from the magnet.
Who sets the numbers
IEC 60601-2-33 defines the SAR, dB/dt, and static-field limits and the three operating modes. The ACR and MHRA apply those limits and concentrate on organisation, people, and process.
Manual, standard, or paper
The ACR and MHRA documents are operational guidance; the IEC document is a technical equipment standard; the ISMRM and ISMRT statements are peer-reviewed papers on individual topics.
Governance titles
The ACR names an MRMD, MRSO, and MRSE; the MHRA names a Responsible Person, MRSE, and Authorised Person. The MR Safety Expert role is common to both.
Points of agreement
All converge on ASTM implant labelling, rigorous screening, control of the magnet environment, prevention of RF burns and projectile incidents, and cryogen safety.
An educational summary, not a substitute for the source documents. This page distils publicly available guidance for study and reference. Clinical practice and local policy should be based on the current, full versions of the primary documents cited below and on your institution’s own procedures. Figures such as SAR and field thresholds are representative of the standard; consult the source for exact, context-specific values.
Back to contents

Reference

01 · United States

ACR Manual on MR Safety

American College of Radiology · Best-practice manual · 2026 edition

Published by the ACR Committee on MR Safety, this is the successor to the ACR Guidance Documents on MR Safe Practices (a 2002 white paper, updated in AJR in 2004 and 2007 and in JMRI in 2013 and 2019). The 2024 edition was reorganised into 16 chapters and 5 appendices, and a peer-reviewed review in Radiology described it (Pedrosa et al., 2025). The 2026 edition added a patient pre-appointment guide, refined the Zone III and IV definitions, updated terminology to peripheral nerve stimulation, rewrote the SAR section, and tightened guidance on unconnected coils and hearing protection.

Framework and governance

  • Four safety zones (I–IV). Zone III is the MR controlled access area, covering the console and any region where the fringe field extends beyond the scanner room; Zone IV is the scanner room itself. Access is supervised by Level 2 MR personnel.
  • A defined personnel hierarchy: an MR Medical Director (MRMD), supported by an MR Safety Officer (MRSO) and MR Safety Expert (MRSE), with staff stratified as Level 1 and Level 2 MR personnel and a standing MR Safety Committee to review incidents and near-misses.
  • The supervising radiologist makes the risk–benefit decision and may delegate a function but not the underlying responsibility.

Core practices

  • Screening of patients and staff, ferromagnetic detection, pocketless attire, and the exclusion or tethering of MR Unsafe equipment from Zone IV.
  • A “full stop and final check”: a pre-scan time-out with routine and augmented checklists.
  • Implant and device management under the ASTM MR Safe / MR Conditional / MR Unsafe framework, with an appendix for unclear conditions.
  • Emergency response covering the emergency stop, power-off, and quench controls, mock-code drills, and removal of the patient from Zone IV before a resuscitation.
  • Policies for remote scanning (a Level 2 technologist onsite is recommended) and for special populations, including pregnancy, sedation and claustrophobia, and large body habitus.
02 · United Kingdom

MHRA: Safety Guidelines for MRI Equipment

Medicines & Healthcare products Regulatory Agency · National guidance · 5th edition (2026)

Issued by the UK’s medical device regulator, the current version is the 5th edition (2026). It is non-statutory good-practice guidance but carries regulatory weight, sitting alongside UK device regulation and the Control of Electromagnetic Fields at Work Regulations 2016. It covers procurement, installation, and clinical use.

Structure: the four hazards

  • Static magnetic field: projectile risk, biological effects, and implant and peripheral-equipment compatibility.
  • Gradient fields: dB/dt, peripheral nerve stimulation, and acoustic noise.
  • Radiofrequency fields: tissue heating, burns, and SAR.
  • Cryogens and quench: asphyxiation risk and quench-pipe integrity.

Governance and access

  • Named roles: an MR Responsible Person, an MR Safety Expert, MR Authorised Persons, and Operators.
  • An MR Controlled Access Area, set by each site to contain the MR Environment, with restricted entry and signage at all entrances. 0.5 mT (5 G) remains the UK legal action level for indirect effects. (IEC 60601-2-33 Ed 4.0 (2022) sets the fringe-field limit at 0.9 mT (9 G) and the ACR Manual on MR Safety, 2026 edition, delimits the MR Environment by the 9 gauss line; 0.9 mT lies inside 0.5 mT, so a site drawn at 0.5 mT remains conservative and the ACR states it needs no adjustment. Check which contour your own site plan uses.)
  • The referrer is responsible for identifying implants and contraindications before referral, supported by local screening.
  • Adverse incidents are reportable to the MHRA.
03 · International

IEC 60601-2-33

International Electrotechnical Commission · Equipment safety standard · 2022

The IEC particular standard for the basic safety and essential performance of MR equipment for medical diagnosis. It is a design standard (it governs how scanners are built, labelled, and constrained) and is the source of the exposure limits that clinical guidance references. National adoptions include EN IEC 60601-2-33.

Three operating modes

  • Normal: routine operation, with no setting expected to cause physiological stress.
  • First-Level Controlled: may cause physiological stress; requires deliberate operator action and medical supervision.
  • Second-Level Controlled: significant risk; requires ethics-committee approval for research use.

Representative limits

  • Whole-body SAR (6-minute average): 2 W/kg in Normal mode, rising to 4 W/kg in First-Level; head SAR is limited to 3.2 W/kg, with higher limits for local and extremity exposure.
  • Static field: up to 3 T in Normal mode; above 3 T and up to 8 T in First-Level; above 8 T is Second-Level.
  • Gradient output (dB/dt): limited to roughly 80% of the peripheral-nerve-stimulation threshold in Normal mode, and 100% in First-Level.
  • Fringe field: Edition 4.0 (2022) revised the contour that bounds the MR Environment from 0.5 mT (5 G) to 0.9 mT (9 G). The ACR Manual on MR Safety (2026) delimits the MR Environment by the 9 gauss line; because 0.9 mT is closer to the magnet than 0.5 mT, facilities that drew their controlled access area at 5 G do not need to adjust it.
  • Acoustic noise: where a scanner can exceed 99 dB(A), hearing protection is required and must bring the level at the ear below 99 dB(A); the manufacturer’s instructions state the minimum protection needed.

Core temperature: ICNIRP (2004) limits the rise in body core temperature to 0.5 °C in Normal mode and 1 °C in controlled mode.

04 · International

ISMRM / ISMRT Consensus Statements

ISMRM Safety Committee with the technologist section (ISMRT) · Peer-reviewed

Produced by the ISMRM Safety Committee with the ISMRT and published mainly in the Journal of Magnetic Resonance Imaging. Rather than a single manual, these are focused recommendations that complement the ACR and MHRA documents:

  • Cardiac implantable electronic devices: practical protocols for imaging patients with pacemakers and defibrillators, both MR Conditional and, in selected non-conditional cases, at 1.5 T (Vigen et al., 2021).
  • 7 T safety: best-practice recommendations for ultra-high-field clinical imaging.
  • MR safety management: recommended responsibilities and governance roles.
  • Radiographer and technologist practice issues in MR safety.
  • Gadolinium deposition in the brain: a summary of the evidence and recommendations.
  • Scanning during a pandemic: operational guidance from the COVID-19 period.
Back to contents

Revision

Quick-reference figures

ValueMeaning
4 zonesACR site model (I–IV); Zone IV is the scanner room
0.5 mT / 0.9 mT5 G, the UK legal action level for indirect effects and RANZCR’s Zone III and pacemaker line; and 9 G, the fringe-field limit set by IEC 60601-2-33 Ed 4.0 (2022) and the line by which the ACR Manual on MR Safety, 2026 edition, delimits the MR Environment. 9 G sits inside 5 G, so a 5 G boundary stays conservative
2 → 4 W/kgWhole-body SAR: Normal → First-Level mode (IEC)
3.2 W/kgHead SAR limit (IEC)
3 TUpper static field for Normal mode; 3–8 T is First-Level
99 dB(A)Where a scanner can exceed this, hearing protection is required and must bring the level at the ear below it (IEC 60601-2-33:2022)
3 modesIEC operating modes: Normal, First-Level, Second-Level
1.5 TField at which many non-conditional CIEDs are imaged under protocol (ISMRM)

Rapid recall

Which body defines the four-zone (I–IV) model?

The ACR, in its Manual on MR Safety. Zone III is the MR controlled access area; Zone IV is the scanner room. The MHRA instead asks each site to define an MR Controlled Access Area that contains the MR Environment.

Name the ACR MR safety leadership roles.

MR Medical Director (MRMD), MR Safety Officer (MRSO), and MR Safety Expert (MRSE), supported by Level 1 and Level 2 MR personnel and an MR Safety Committee.

What are the IEC operating modes and key SAR limits?

Normal (no physiological stress; whole-body SAR ≤2 W/kg), First-Level Controlled (supervised; whole-body SAR ≤4 W/kg, head ≤3.2 W/kg), and Second-Level Controlled (ethics approval required). Gradient output is capped at 80% of the PNS threshold in Normal mode and 100% in First-Level.

How is the MHRA guidance structured?

Around the four physical hazards (static field, gradients, RF, and cryogens) together with governance roles, the controlled access area, screening and referrer duties, and MHRA incident reporting. The current version is the 5th edition (2026).

What is the ACR “full stop and final check”?

A pre-scan time-out, analogous to a surgical time-out, using routine and augmented checklists to catch safety issues before the patient enters Zone IV.

Which document applies to a pacemaker or ICD patient?

The ISMRM CIED recommendations (Vigen et al., 2021), together with the device’s MR Conditional labelling and the local ACR- or MHRA-based policy.

What are the three ASTM implant labels?

MR Safe, MR Conditional, and MR Unsafe (ASTM F2503), used across the ACR, MHRA, and IEC frameworks.

What changed in the ACR Manual on MR Safety (2026)?

The 2026 edition added a patient pre-appointment guide, refined the Zone III and IV definitions, adopted peripheral-nerve-stimulation terminology, rewrote the SAR section, advised against unconnected coils in the bore, and recommended ear plugs plus ear muffs whenever possible while warning that muffs may be displaced. Remote scanning is addressed, with a Level 2 technologist onsite.

Put these guidelines into practice: use our free generic MRI departmental protocol templates (patient & staff safety, emergencies, QA, cleaning, governance) and the MR safety self-audit checklist to benchmark your department against ACR/MHRA requirements.

Back to contents

Sources

Sources (9)

9 sources listed

  1. American College of Radiology. MR Safety Resources: current Manual on MR Safety (2026 edition) and summary of changes.
  2. Pedrosa I, Altman DA, Dillman JR, et al. American College of Radiology Manual on MR Safety: 2024 Update and Revisions. Radiology 2025;315(1):e241405 (review of the 2024 edition).
  3. MHRA. Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, 5th edition (2026).
  4. International Electrotechnical Commission. IEC 60601-2-33:2022: MR equipment for medical diagnosis.
  5. Vigen KK, Reeder SB, Hood MN, et al. Recommendations for Imaging Patients With Cardiac Implantable Electronic Devices. JMRI, 2021.
  6. ISMRM / ISMRT. MR Safety resources and consensus statements.
  7. U.S. Food & Drug Administration. MRI (Magnetic Resonance Imaging).
  8. Royal Australian and New Zealand College of Radiologists. MRI Safety Guidelines, version 3.0 (2021).
  9. International Commission on Non-Ionizing Radiation Protection. Medical magnetic resonance (MR) procedures: protection of patients. Health Phys 2004;87(2):197–216.

More in Safety and contrast

Next in Safety and contrast: SAR reduction

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.