Structural MRI does two jobs in cognitive decline: rule out something treatable, and read the pattern of brain shrinkage that points toward a diagnosis.

On this pageFirst job: exclude treatable causesSecond job: recognise atrophy patternsWhat it does and doesn’t doSources

First job: exclude treatable causes

Before attributing decline to a neurodegenerative disease, MRI excludes potentially reversible or surgical causes: a tumour, subdural haematoma, normal-pressure hydrocephalus, or significant vascular disease. This alone justifies imaging in new cognitive impairment.

Coronal T1 brain MRI through the pons (MNI ICBM152 template) labelled: grey matter, pons, corpus callosum, lateral ventricle. Handwritten note: CSF: dark on T1.
MRI: MNI ICBM152 2009b template. Copyright (C) 1993-2004 Louis Collins, McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University. Labels added.

Second job: recognise atrophy patterns

Different dementias thin different regions, and the distribution is the clue:

PatternSuggests
Medial temporal / hippocampal atrophyAlzheimer’s disease
Frontal & anterior temporal atrophyFrontotemporal dementia
Small-vessel disease, strategic infarctsVascular contribution
Relatively preserved medial temporal, occipital changesDementia with Lewy bodies (supportive)
Visual rating scales (e.g. medial temporal atrophy / MTA, Koedam posterior, Fazekas white-matter) give a structured, reproducible way to report these patterns.

What it does and doesn’t do

MRI supports a clinical diagnosis and tracks change, but atrophy patterns overlap and are age-dependent; imaging is interpreted with the clinical picture and, increasingly, molecular biomarkers (PET, CSF). It is a pattern-recognition and exclusion tool, not a standalone diagnosis.

Scroll real scansBrain MRI atlas: real scans with the structures named on the image.

Sources

1 source listed

  1. Harper L et al. An algorithmic approach to structural imaging in dementia. J Neurol Neurosurg Psychiatry 2014;85:692–700.

Educational summary for clinicians; diagnosis is multidisciplinary. Not medical advice.

Vendor terminology

Reformat coronal-oblique T1 through hippocampi for MTA (Scheltens); optional volumetry.

✓SequencePlaneTRTETISlice/gapFOVMatrix/resOther
3D T1 IR-GREMPRAGEBRAVO / IR-FSPGR3D TFEFastFE 3Dcoronal-oblique reformatSag 3D20002–4~9001 iso240256MTA ⟂ hippocampus
FLAIRFLAIRFLAIRFLAIRFLAIRAx800012024005230256Vascular burden
T2 TSE/FSET2 TSET2 FSET2 TSET2 FSEAx4000100–5230320–
Susceptibility (SWI)SWISWANSWIpFSBB / SWIAx2720–40–2–3220320Microbleeds / CAA
DWI (SS-EPI)DWI (EPI)DWI (EPI)DWI (EPI)DWI (EPI/FASE)Ax3000min–4230160Exclude CJD/acute

ESNR working group minimum protocol.18

What each sequence is for

Sagittal 3D T1 IR-GRE, re-sliced coronal oblique
A 3D volume re-sliced straight across the hippocampi. It shows how much the hippocampus has shrunk (the MTA score), which supports Alzheimer’s disease, and the pattern of shrinkage elsewhere points to other dementias.
Axial FLAIR
Shows small-vessel damage as bright patches in the white matter, and old strokes, which can cause or add to memory problems (vascular dementia).
Axial T2
A general check for other causes of memory problems, such as a tumour, old strokes or enlarged fluid spaces.
Axial SWI
Tiny old bleeds show as black dots. Many dots near the brain surface point to cerebral amyloid angiopathy, where amyloid builds up in the walls of small vessels.
Axial DWI
Shows a recent small stroke, and the bright cortex or deep grey matter of Creutzfeldt–Jakob disease (CJD), a rare cause of rapidly worsening dementia.

References: ESNR working group dementia protocol (Neuroradiology 2019); Boston criteria v2.0 for cerebral amyloid angiopathy (Lancet Neurol 2022); ACR–ASNR–SPR practice parameter: MRI of the brain (2024).

All protocols

Grey isometric drawing: a patient lies on their back, head first at the MRI tunnel, the head inside a cage-like head coil with the shoulders up against it, a wedge under the knees, a pad between the ankles, the near arm by the side on a pad holding a call bell; a technologist stands near the foot end. Numbered markers 1 to 3.
Brain: head first, the head inside the head coil. Illustration only; follow your coil’s instructions.
  1. Head coil around the head
  2. Wedge under the knees
  3. Call bell in the hand
Position
Supine, head first, with a support under the knees.
Coil
head coildS Head coilHead/Neck 20 or 64 coil. Usually the body coil transmits and the head coil only receives; some head coils transmit as well.
Centre
The centre mark of the head coil.
Set up
  1. Slide the patient up until the shoulders touch the coil, so the headphones sit over the ears and the coil top closes cleanly.
  2. Earplugs in, headphones on, call bell in hand.
  3. Small head: pack foam pads either side so the head cannot roll.
  4. Straighten the head before the coil top goes on: nose up, not tilted or turned.
What goes wrong, and the fix
  • Head too big for the coil top with headphones on: change to slimmer passive ear defenders and take out the pad under the head; then check they can still hear you.
  • Head rolls or nods during the scan: add pads around the head and check comfort; a radial motion-correctingMultiVane XDBLADE T2 can reduce motion artefact.
  • Patient anxious in the closed coil: explain the scan, talk between sequences, and keep the call bell in reach.
Check on the first localisersurveylocalizer
  • The whole brain sits inside the coil’s signal on the sagittal and coronal localisersurveylocalizer.
  • The head is not rotated or tilted; if it is badly off, straighten it now rather than angling every sequence.
Explain it simply

The head goes inside a cage-like coil that picks up the signal. Pads stop it rolling, and earplugs and headphones protect the ears from the noise.

Why it matters: A straight, still head gives symmetrical images that are easier to compare side to side.

Sources: Department notes; Zaitsev 2015; ACR MR Safety 2026; MHRA 2026; Gruber 2018

All positioning

Systematic search order

  1. Symmetry first: compare left vs right hemispheres, then descend cranial → caudal.
  2. Grey–white differentiation & cortex: ribbon, sulci, gyral pattern.
  3. Deep structures: basal ganglia, thalami, internal capsules.
  4. Ventricles & CSF spaces: size, symmetry, ependymal margins.
  5. Midline & posterior fossa: corpus callosum, brainstem, cerebellum, 4th ventricle.
  6. Then the review areas (below), the peripheral belt that the parenchymal search skips.

Key structures × sequence

StructureBest onLooking for
Acute infarctDWI + ADCRestricted diffusion (bright DWI, dark ADC). Must-check on every brain.
Oedema / gliosis / demyelinationFLAIRPeriventricular & juxtacortical hyperintensity; sulcal FLAIR signal (SAH, meningitis).
Microbleeds / haemorrhage / calciumSWI / GREBlooming foci: amyloid, cavernoma, DAI, haemorrhagic mets.
Enhancing lesion / breakdownT1 post-GdTumour, abscess rim, leptomeningeal & dural enhancement.
Anatomy / marrow / fat planesT1Skull base marrow, clival signal, orbital fat.

Review areas

Work the periphery structure by structure; these are the documented blind spots:
  • Skull base & clivus: marrow replacement, chordoma, sphenoid lesions (T1).
  • Cavernous sinuses: asymmetry, convexity, ICA flow voids, CN III–VI (coronal post-Gd).
  • Internal auditory canals (IACs) & CPA: vestibular schwannoma, enhancement (thin post-Gd / heavily T2).
  • Pituitary & sella: microadenoma, stalk position, height/convexity of gland.
  • Extra-axial spaces: thin subdural/epidural collections that parallel the calvarium.
  • Vessels & flow voids: absent flow void (thrombosis/occlusion), aneurysmal flow void, dural venous sinuses.
  • Orbits & globes: optic nerves/sheaths, extra-ocular muscles, orbital apex.
  • Paranasal sinuses & mastoid air cells: fluid, mucosal disease, opacification.
  • Calvarium & scalp: lytic/blastic lesions, marrow signal, subgaleal collections.
  • Cervicomedullary junction & foramen magnum: tonsillar position (Chiari), cord signal at the bottom slice.
Never sign a brain without eyeballing DWI/ADC. An acute infarct can be invisible on T1/T2/FLAIR in the first hours, and missing it can change management.