MRI patient positioning: Exam by exam
For technologists, radiographers and students at the scanner. How to lie the patient, which coil, where to centre, how to pad and immobilise, what goes wrong and how to fix it, and what to check on the first localiser. Switch the terms to Philips or Siemens wording at the top.
On this page
Find your examBefore every examHead and spineShoulder and upper limbPelvis and lower limbBody and whole bodySourcesDotted terms change with the switch.Showing Philips wording. Names vary between software versions and coil generations; your console is the final word.Showing Siemens wording. Names vary between software versions and coil generations; your console is the final word.
- Centre the anatomy you are imaging at the isocentre and in the sensitive part of the coil, then check the first localisersurveylocalizer before you plan anything. Repeat it if the anatomy is off-centre.
- A comfortable, well-supported patient keeps still. Pad, support and explain before the table moves.
- Prevent RF burns every time: no skin-to-skin loops, no cable loops or cables on the skin, and pads between the patient and the bore.
- Hearing protection for every patient, and a call bell in the hand, before you leave the room.
Explain it simply
- The scanner sees best in the middle of the machine and right under the coil, so the body part has to sit there, and the first quick picture shows whether it does.
- Someone lying comfortably finds it easier to keep still, and keeping still keeps the pictures sharp.
- Skin touching skin, or wires touching skin, can heat up during the scan, so pads keep them apart.
- The scanner is very loud, so everyone wears ear protection and holds a buzzer to call for help.
Why it matters: getting the setup right the first time means a clear scan with no repeats.
A guide only: follow your departmental protocols and regional requirements, and the instructions for use of your coils and scanner.
Sources: ACR MR Safety 2026; Department notes
Find your exam
Before every exam
These habits apply to every examination on this page. Most image problems that start at the table are fixed at the table: centre well, pad well, keep cables and skin apart, and make sure the patient can hear you and call you.
Sources: Department notes; MHRA 2026; RANZCR MRI Safety 2021; ACR MR Safety 2026
Centre accurately, then check the first localiser
- Set up
- Put the anatomy of interest in the middle of the coil, not just inside it.
- Mark the landmark with the laser lightlight visorlaser-light localizer and send it to the middle of the bore. The positioning laser carries a laser hazard marking, so ask the patient not to look into the beam, and follow your scanner’s instructions.
- Run the localisersurveylocalizer and look at it before you plan anything.
- What goes wrong, and the fix
- Anatomy near the edge of the coil, or signal falling away at one end: bring the patient out, move the coil or the patient, and repeat the localisersurveylocalizer.
- The localisersurveylocalizer is too high or too low to see your landmarks: re-centre and repeat it rather than planning by guesswork.
- Spectral fat suppressionSPIR/SPAIR fat suppressionFat Sat/SPAIR fat suppression patchy at the edge of a large field of view: move the anatomy nearer the middle of the bore if the position allows, or use DixonmDIXONDixon or STIR.
- Check on the first localisersurveylocalizer
- The anatomy you need is in the middle of the coil’s signal, top to bottom and side to side.
- The landmarks you plan from are visible on all three planes.
- Nothing you need is in the dark fall-off zone at the coil’s edge.
- Good to know
- A receive coil is most sensitive close to its elements and loses signal with distance, so anatomy at the edge of a coil comes out dark and noisy. The magnetic field is most uniform at the middle of the bore (the isocentre). Off-centre anatomy makes chemically selective spectral fat suppressionSPIR/SPAIR fat suppressionFat Sat/SPAIR fat suppression less even and adds geometric distortion. If you use automatic slice planningSmartExamAutoAlign, check its slices against the same landmarks.
Explain it simply
The scanner sees best in the middle of the coil and the middle of the tunnel. A quick first picture shows whether the patient is in the right place before the long scans start.
Why it matters: Re-centring takes a minute; a repeat scan means another visit for the patient.
Sources: Department notes; Gruber 2018; Del Grande 2014; ACR Wrist PP 2022; ACR Elbow PP 2026; RANZCR MRI Safety 2021; PhIngeniaTD
Coil cable care
- Set up
- Look over the coil, cable and plug for cracks, kinks or damage before each use.
- Run the cable straight, with no loops, no twists and no tight bends.
- Keep the cable off the patient’s skin and never across the patient diagonally; pad it where it has to lie near them.
- Remove any coil you are not using; never leave an unplugged coil on the table.
- What goes wrong, and the fix
- Cable sharply bent at the coil or the socket: move the patient up or down the table until it lies straight.
- Cable has to pass near the skin: lay an insulating pad between cable and skin.
- Damaged cable, cover or plug: do not use the coil; report it.
- Check on the first localisersurveylocalizer
- Images show no unexpected dark band or missing coil elements; if they do, check the plug and cable before blaming the patient.
- Good to know
- Looped or crossed cables in the bore can pick up RF energy and heat. The guidelines ask for regular damage checks on all coils, cables and leads, and say an unplugged coil must never be in the scanner with the patient, because it can injure the patient or be damaged itself.
Explain it simply
The coil’s cable carries the signal out of the scanner. Kinks can damage it, and loops can warm up, so it stays straight and away from skin.
Why it matters: A cracked cable can spoil the images and burn the patient.
Sources: Department notes; MHRA 2026; RANZCR MRI Safety 2021; ACR MR Safety 2026
Pads, sponges, sandbags and a level coil
- Set up
- Ask the patient what hurts before they lie down, and set up for that.
- Support under the knees for supine exams and support for the arms, so nothing has to be held up by effort.
- Fill gaps between the body part and the coil with foam pads so it cannot move inside the coil.
- Use a sandbag on top of a limb or coil to hold the position.
- Keep a flexible or anterior coil level and parallel to the table, using pads under its edges.
- Use only pads, pillows and sandbags approved for the MR room.
- What goes wrong, and the fix
- Coil tilts or rocks on the body: pad under the low side until it sits level.
- Patient moves during the scan: go in, find out why, and improve support before changing sequences.
- Check on the first localisersurveylocalizer
- On the localisersurveylocalizer, the coil sits evenly over the anatomy with no wedge of fall-off on one side.
- Good to know
- Comfort and gentle immobilisation are the main defences against involuntary movement; time spent making the patient comfortable improves their cooperation. Pillows can hide springs and sandbags can contain metal pellets, so only labelled MR room equipment goes in. A tilted coil sits further from one side of the anatomy, and coil signal falls with distance.
Explain it simply
Pads hold the body still and take the strain off sore spots. A still, comfortable patient gives sharp pictures.
Why it matters: Motion blur is often fixed with a pad rather than a new sequence.
Sources: Department notes; ACR Knee PP 2025; ACR Shoulder PP 2024; MHRA 2026; Gruber 2018
RF burn prevention
- Set up
- Patient in a gown provided by the department, with metal and conductive clothing, and unneeded leads and electrodes, taken off. Remove a metal-backed medication patch if the prescriber agrees; if it has to stay on, warn the patient to report any warmth at once and consider a cold pack over it, as your policy directs; warn patients with tattoos, especially new ones, to report warmth.
- No skin-to-skin loops: hands apart and not clasped, legs uncrossed, and pads between the thighs, between the ankles, between the arms and the body, and in skin folds.
- No skin touching the bore: put the manufacturer’s insulating pads between the patient and the bore wall. A sheet alone is not enough.
- Cables and leads off the skin, straight, not crossing each other, and running along the length of the bore, down the middle and as far from the bore wall as possible.
- Tell the patient to report any warmth or tingling straight away, without waiting for the sequence to end.
- Patients who cannot feel or report heat (sedated, anaesthetised or unconscious patients, infants, reduced sensation, confusion, a language barrier): check every pad and cable twice before the table goes in, look at them between sequences, and follow your sedation and monitoring policy.
- Before the table goes in, look along both sides of the patient for skin contact with the bore, with each other or with cables.
- What goes wrong, and the fix
- Large patient touches the bore or the coil cable: add insulating pads, or reconsider the scan if contact cannot be prevented.
- Patient reports heat or burning: stop the scan, bring them out, look at the skin, tell the supervising radiologist, follow your local burns procedure, and report it through your incident system and any national scheme.
- Worn, flattened pads: replace them; compressed pads no longer keep enough distance.
- Check on the first localisersurveylocalizer
- No dark or bright artefact from a forgotten lead, patch or clothing item over the area being scanned.
- Good to know
- RF burns are the most commonly reported adverse event in MRI. A closed loop of tissue with a small contact point, such as a hand on a thigh or the thighs touching, can concentrate induced current and heat at the contact. Skin against the bore wall or a cable can burn by proximity. Burns may not hurt at first, and patients who cannot communicate are at higher risk. Your manufacturer sets the minimum pad thickness.
Explain it simply
The scanner’s radio waves can make loops of skin, metal or cable warm up. Pads keep skin apart and away from the tunnel and cables.
Why it matters: A burn can be deep and may only show after the patient has gone home.
Sources: Department notes; ACR MR Safety 2026; RANZCR MRI Safety 2021; MHRA 2026
Hearing protection and headphone fit
- Set up
- Earplugs for every patient, fitted properly: put them in yourself, or check them if the patient does it. Neonates and small infants get neonatal ear muffs instead, as your protocol directs.
- Headphones or ear defenders on top where the coil allows.
- In a head or head and neck coil, slide the patient up until the shoulders touch the coil so the headphones sit over the ears and the coil closes without pressing on them.
- Anyone staying in the room during scanning wears hearing protection too.
- At the start, ask the patient something they must answer, so you know they can hear you and you can hear them.
- What goes wrong, and the fix
- Head too big for headphones under the coil top: change to slimmer passive ear defenders and take out the pad under the head.
- Patient cannot hear the instructions: check the plugs and headphones are seated, raise the intercom volume, or speak the instructions yourself. Do not take the earplugs out.
- Headphones slip off during the scan: refit them, and pad them in place.
- Good to know
- Most sequences are loud enough that hearing protection is mandatory for all patients unless the sound level is documented below the threshold. Manufacturer headphones alone are often not enough, ear muffs can be displaced during the scan, and plugs and muffs together are recommended whenever possible. Hearing protection must still let the patient hear your instructions.
Explain it simply
The scanner is very loud, like a building site. Earplugs protect hearing, and headphones on top make it quieter still.
Why it matters: Noise damage to hearing can be permanent, and a patient who cannot hear you cannot follow breath-holds.
Sources: Department notes; RANZCR MRI Safety 2021; ACR MR Safety 2026; MHRA 2026
Call bell, comfort and contact
- Set up
- Give the call bell (squeeze ball) to every patient, explain when to use it, and have them test it.
- Check the intercom works both ways and that you can see the patient on the camera.
- Make sure their hands are on the table, not over the edge, before the table moves.
- Set the bore fan and lighting for the patient; turn the fan up if they get warm and take off extra blankets.
- Tell them how long the scan takes and talk to them between sequences.
- What goes wrong, and the fix
- Patient cannot use the bell (weak grip, confusion, sedation): find another way for them to signal, or have a screened staff member or companion stay in the room wearing hearing protection; keep continuous sight and sound contact, and monitor sedated patients as your sedation policy directs.
- Patient gets hot during a long exam: fan up, fewer covers, and check on them between sequences.
- Patient keeps moving: go in and fix the cause (pain, cold, a pressure point) before changing the protocol or switching to a radial motion-correctingMultiVane XDBLADE sequence.
- Good to know
- The guidelines require a way for the patient to signal distress at any time, and the patient is not left unattended. A screened companion may stay in the room, and sedated patients need monitoring. Blankets and covers should not stop the patient losing heat.
Explain it simply
The patient holds a squeeze ball to call for help. Knowing they can stop at any time helps them relax and keep still.
Why it matters: A patient who trusts the call bell is more likely to finish the scan.
Sources: Department notes; RANZCR MRI Safety 2021; ACR MR Safety 2026; MHRA 2026; Zaitsev 2015
Breath-hold coaching and breathing sensors
- Set up
- Explain the breath-holds before the patient goes in, and practise one or two on the table.
- Ask for the same size of breath every time, not too deep and not too shallow, so each hold puts the organs in the same place.
- Place the respiratory sensorrespiratory beltrespiratory belt where the breathing movement is greatest, under the anterior body coildS Torso coilBody 18 coil with a soft pad so the coil presses on it. If your system senses breathing through the coil or a camera instead of a belt, skip the belt and position as its instructions direct.
- Check the breathing trace is regular on the console before the first breath-hold sequence, and use it to start each hold once the trace is still.
- What goes wrong, and the fix
- Patient cannot hear or understand: check the hearing protection is seated and the volume is up, check for a language barrier, then switch from automatic breath-hold commandsAutoVoiceauto voice commands to giving the instructions yourself.
- Patient cannot hold long enough: split each hold into more, shorter holds, or lengthen the rest between holds.
- Breath-holds still fail: change to respiratory triggering or navigatorrespiratory navigator2D PACE navigator versions, or single-shot TSESSh TSEHASTE or radial motion-correctingMultiVane XDBLADE sequences.
- Flat or noisy breathing trace: move the respiratory sensorrespiratory beltrespiratory belt to where the belly or lower ribs move most and check the coil is pressing on it; with a built-in or camera sensor, check the coil or patient position its instructions require.
- Organs shift between consecutive breath-hold images: coach a more consistent breath and practise again.
- Check on the first localisersurveylocalizer
- The organs you need are inside the coil’s signal on the breath-hold localisersurveylocalizer.
- Good to know
- Breath-holding works, but it limits how long each acquisition can be, and several holds in a row do not always return the organs to the same place. The sensor must be placed so it registers the full range of breathing. In one volunteer study the diaphragm crept slowly upwards during long holds, especially after a very deep breath, a further reason to keep holds short. Whether to hold on breathing in or breathing out follows your protocol.
Explain it simply
Breathing moves the liver and kidneys up and down. Holding the breath the same way each time keeps them still and in the same place for every picture.
Why it matters: Breathing blur can hide a small liver lesion or a stone in a bile duct.
Sources: Department notes; Zaitsev 2015; Khot 2020; PhVitalEye; SHBioMatrix
Head and spine
Head and cervical work happens in a head or head and neck coil; thoracic, lumbar and pelvic spine work uses the spine coildS Posterior coilSpine 32 coil in the table with the anterior body coildS Torso coilBody 18 coil on top. For any spine level that has to be named, count from the top of the spine.
Sources: Department notes; ACR Spine PP 2023; Tins 2016
Brain

- Head coil around the head
- Wedge under the knees
- Call bell in the hand
- Position
- Supine, head first, with a support under the knees.
- Coil
- head coildS Head coilHead/Neck 20 or 64 coil. Usually the body coil transmits and the head coil only receives; some head coils transmit as well.
- Centre
- The centre mark of the head coil.
- Set up
- Slide the patient up until the shoulders touch the coil, so the headphones sit over the ears and the coil top closes cleanly.
- Earplugs in, headphones on, call bell in hand.
- Small head: pack foam pads either side so the head cannot roll.
- Straighten the head before the coil top goes on: nose up, not tilted or turned.
- What goes wrong, and the fix
- Head too big for the coil top with headphones on: change to slimmer passive ear defenders and take out the pad under the head; then check they can still hear you.
- Head rolls or nods during the scan: add pads around the head and check comfort; a radial motion-correctingMultiVane XDBLADE T2 can reduce motion artefact.
- Patient anxious in the closed coil: explain the scan, talk between sequences, and keep the call bell in reach.
- Check on the first localisersurveylocalizer
- The whole brain sits inside the coil’s signal on the sagittal and coronal localisersurveylocalizer.
- The head is not rotated or tilted; if it is badly off, straighten it now rather than angling every sequence.
Explain it simply
The head goes inside a cage-like coil that picks up the signal. Pads stop it rolling, and earplugs and headphones protect the ears from the noise.
Why it matters: A straight, still head gives symmetrical images that are easier to compare side to side.
Sources: Department notes; Zaitsev 2015; ACR MR Safety 2026; MHRA 2026; Gruber 2018
Cervical spine

- Head and neck coil
- Front of the coil over the neck
- Pad between arm and body
- Call bell in the hand
- Position
- Supine, head first, with a support under the knees and arms by the sides.
- Coil
- head and neck coildS HeadNeck coilHead/Neck 20 coil.
- Centre
- The middle of the neck, at the coil’s centre mark.
- Set up
- Shoulders down against the coil so the headphones fit and the lower neck is inside the coil.
- Earplugs and headphones, with the same big-head and small-head fixes as the brain.
- Pad under the knees and a pad between each arm and the body.
- Pad around the head if it can move inside the coil.
- What goes wrong, and the fix
- Head too big for the coil top: slimmer passive ear defenders and no pad under the head.
- Small head moving: foam pads either side.
- Lower cervical levels dark on the localisersurveylocalizer: bring the patient further up so the shoulders sit right against the coil, and repeat.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer covers from the skull base to the lung apices with good signal throughout.
- On the coronal the cord runs straight down the middle.
Explain it simply
The neck sits in a coil that wraps round the head and neck. Getting the shoulders right up to the coil keeps the bottom of the neck in the picture.
Why it matters: A routine cervical spine covers at least down to the C7-T1 disc, so the lower neck cannot fall off the edge of the coil.
Sources: Department notes; ACR Spine PP 2023; RANZCR MRI Safety 2021; MHRA 2026
Thoracic spine

- Coil strapped over the chest
- Arm in an arm holder
- Call bell in the hand
- Low, flat head pad
- Position
- Supine, feet first, with a support under the knees and arms by the sides in arm holders.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil over the chest.
- Centre
- Mid-chest (department practice), so the thoracic localisersurveylocalizer covers the whole region.
- Set up
- Lie the patient high enough on the table for the coil to connect without the cable straining.
- Keep the head support low so the upper back lies flatter, as far as comfort allows.
- Arms in arm holders with a pad between each arm and the body.
- Level the anterior body coildS Torso coilBody 18 coil with pads if it tilts on the chest.
- What goes wrong, and the fix
- Level cannot be counted: add a cervical localisersurveylocalizer, join it to the thoracic one as composed multi-stationcomposed (MobiView)composed (Composing) views, and count down from C2.
- Patient breathless or in pain lying flat: raise the head support a little and pad more.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer shows from C7 down to L1, and below the end of the cord if it ends lower.
- You can count the vertebrae from C2 on the joined view before naming any level.
- Good to know
- Counting down from the atlantoaxial joint is the reliable way to name a level; anatomical landmarks within the thoracic or lumbar spine alone are not reliable, and numerical variants can only be identified with whole-spine imaging.
Explain it simply
The thoracic spine has twelve similar-looking bones, so the scanner takes a long picture from the neck down so they can be counted from the top.
Why it matters: Naming the wrong vertebra can send a surgeon or injection to the wrong level.
Sources: Department notes; Tins 2016; ACR Spine PP 2023; RANZCR MRI Safety 2021
Lumbar spine

- Coil strapped over the abdomen
- Wedge under the knees
- Call bell in the hand
- Ear protection
- Position
- Supine, feet first, with a support under the knees.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil over the abdomen.
- Centre
- The lower costal margin, which sits at about L3, the middle of the lumbar spine (department practice).
- Set up
- Support under the knees for comfort.
- Hands apart, on the chest with a fold of gown or a pad under them, or by the sides with pads between arms and body; not clasped and not touching the thighs.
- Square the pelvis and shoulders on the table so the spine is not rotated.
- Centre the anterior body coildS Torso coilBody 18 coil over the same level as the landmark.
- What goes wrong, and the fix
- LocaliserSurveyLocalizer too high or too low to show T11 to S2: re-centre and repeat it; do not plan from guessed landmarks.
- Spine rotated on the axial localisersurveylocalizer: straighten the patient on the table if you can.
- Check on the first localisersurveylocalizer
- The sagittal localisersurveylocalizer shows from about mid T11 to S2.
- Look at the kidneys, liver and abdomen on the localisersurveylocalizer for anything unexpected, and tell the radiologist.
- Good to know
- The knee support is for comfort. A knee support is often said to flatten the lumbar curve, but studies disagree: one small study found no practical change, and a larger one found the curve slightly flatter with a bolster.
Explain it simply
The lower back is scanned lying on the back with a cushion under the knees. The first quick picture is checked to be sure the whole lower spine is in view.
Why it matters: A well-placed first picture means every later slice lands on the right disc.
Sources: Department notes; Murrie 2002; Simonovich 2020; MHRA 2026; RANZCR MRI Safety 2021
Whole spine

- Head and neck coil
- Flat coil over the trunk
- Arm in an arm holder
- Call bell in the hand
- Position
- Supine, head first, with a support under the knees and arms by the sides in arm holders.
- Coil
- head and neck coildS HeadNeck coilHead/Neck 20 coil without its base plate, with the spine coildS Posterior coilSpine 32 coil and the anterior body coildS Torso coilBody 18 coil.
- Centre
- Mid-chest (department practice).
- Set up
- Remove the base plate of the head and neck coil so the head is not lifted and the whole spine lies at one level.
- Shoulders against the coil for headphone fit, with the same big-head and small-head fixes as the brain.
- Arms in holders with a pad between each arm and the body.
- Cover the thoracic and lumbar spine with the anterior body coildS Torso coilBody 18 coil, level on the body.
- What goes wrong, and the fix
- Patient moves during one station: repeat that station without moving the planning box, so the stations still join.
- Stations do not line up: keep the box in exactly the same place for every sagittal sequence once the first has started.
- Check on the first localisersurveylocalizer
- Sagittal and coronal composed multi-stationcomposed (MobiView)composed (Composing) views run from the brainstem to the sacrum so the vertebrae can be counted and aligned.
- Signal is even across the junctions between coils.
- Good to know
- Counting from C2 on a whole-spine view is the reliable way to name levels and pick up numerical variants.
Explain it simply
The whole spine is too long for one picture, so the scanner takes it in two or more sections and joins them. The patient must lie flat and still so the pieces fit together.
Why it matters: Matching sections let the radiologist follow the spinal cord from top to bottom without a gap.
Sources: Department notes; Tins 2016; RANZCR MRI Safety 2021
Sacrum and coccyx

- Coil strapped over the pelvis
- Wedge under the knees
- Call bell in the hand
- Ear protection
- Position
- Supine, feet first, with a support under the knees.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil, placed as for a lumbar spine.
- Centre
- Midway between the iliac crest and the greater trochanter (department practice).
- Set up
- Support under the knees; pad under a sore coccyx if lying on it hurts.
- Hands apart, on the chest with a fold of gown or a pad under them, or by the sides with pads between arms and body; not resting on the thighs.
- Centre the anterior body coildS Torso coilBody 18 coil over the pelvis at the same level.
- What goes wrong, and the fix
- Lying on a painful coccyx makes the patient move: add padding for comfort and keep to the sequences needed.
- Coccyx near the bottom edge of the signal: move the patient so the centring point is right and repeat the localisersurveylocalizer.
- Check on the first localisersurveylocalizer
- Run a second localisersurveylocalizer angled to the sacrum: boxes along the sacrum on the sagittal and symmetrical through it on the coronal and axial.
- The whole sacrum and coccyx sit in the middle of the coil’s signal.
Explain it simply
The tailbone sits at the bottom of the spine. The patient lies on the back and the coil is centred low, between the hip bone and the top of the thigh bone.
Why it matters: The coccyx is small and easy to cut off the edge of the images.
Sources: Department notes; ACR Spine PP 2023; RANZCR MRI Safety 2021
Sacroiliac joints

- Coil strapped over the pelvis
- Wedge under the knees
- Call bell in the hand
- Ear protection
- Position
- Supine, feet first, with a support under the knees.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil, placed as for a lumbar spine.
- Centre
- Midway between the iliac crest and the greater trochanter (department practice).
- Set up
- Support under the knees.
- Lie the pelvis square on the table, not rotated.
- Hands apart, on the chest with a fold of gown or a pad under them, or by the sides with pads between arms and body; not resting on the thighs.
- What goes wrong, and the fix
- Pelvis rotated on the localisersurveylocalizer: straighten the patient; small rotations can then be planned out.
- Joints near the coil edge: re-centre and repeat the localisersurveylocalizer.
- Check on the first localisersurveylocalizer
- Run a second localisersurveylocalizer angled to the sacrum, with the boxes symmetrical through both joints on the coronal and axial.
Explain it simply
The sacroiliac joints join the bottom of the spine to the pelvis, one on each side. Both sides are scanned together so they can be compared.
Why it matters: Comparing the two joints is easier when the pelvis lies straight.
Sources: Department notes; RANZCR MRI Safety 2021
Lumbar pars defects

- Coil strapped over the abdomen
- Wedge under the knees
- Call bell in the hand
- Ear protection
- Position
- Supine, feet first, with a support under the knees, as for a lumbar spine.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil.
- Centre
- As for a lumbar spine: the lower costal margin.
- Set up
- Set up exactly as for a lumbar spine.
- Allow for a longer examination: the pars protocol adds a thin 3D volume after the routine lumbar spine.
- Explain the length to the patient and support the knees and arms well so they can stay still.
- What goes wrong, and the fix
- Patient moves during the long 3D sequence: improve support and comfort first, then repeat.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer covers the lumbar spine as normal, with the lower lumbar levels well inside the coil’s signal.
- Good to know
- The routine sagittal STIR is required for pars, and the protocol adds a thin 3D T1 gradient echo volume without fat suppression, reformatted into left and right obliques along each pars.
Explain it simply
The pars is a thin bridge of bone at the back of each vertebra. A crack in it is called a pars defect, and extra angled pictures show each bridge along its length.
Why it matters: The angled views show the pars along its length, where routine slices cut across it.
Sources: Department notes; Del Grande 2014
Shoulder and upper limb
The upper limb sits away from the middle of the bore when the arm is by the side, where the field is less uniform and spectral fat suppressionSPIR/SPAIR fat suppressionFat Sat/SPAIR fat suppression is harder to get even. For the elbow, wrist and hand, the department’s default is the prone “superman” position (arm overhead), which brings the joint nearer the isocentre; arm by the side is the fallback for patients who cannot hold it, and the guidelines note that the hand at the side is often more comfortable. Overhead: pad between the raised arm and the head, keep the coil and hand off the bore wall, and put the call bell in the other hand. Arm by the side: pads between the arm and the body, between the hand and the thigh, and between the arm and the bore.
Sources: ACR Elbow PP 2026; ACR Wrist PP 2022; ACR Fingers Toes PP 2023; Del Grande 2014; RANZCR MRI Safety 2021; MHRA 2026; ACR MR Safety 2026; Department notes
Shoulder

- Shoulder coil over the joint
- Sponge in the hand
- Pad between arm and body
- Pad between the ankles

- Shoulder coil over the joint
- Pad between arm and body
- Thumb up (not turned in), sponge in the hand
- Ear protection
- Position
- Supine, head first, affected arm by the side.
- Coil
- shoulder coildS Shoulder coilShoulder Shape 16 coil.
- Centre
- The shoulder joint, in the middle of the coil.
- Set up
- Arm by the side with the hand turned partly out (partial external rotation): not turned in, and not forced.
- Put a small sponge in the hand as a reminder to keep it turned out.
- Large patient: support the forearm on a rolled towel or pad so the arm stays turned out and the hand rests clear of the thigh.
- Pad between the arm and the body, and between the hand and the thigh.
- Where the bore allows, slide the patient towards the opposite side so the shoulder sits nearer the middle, with a pad between the other arm and the bore.
- What goes wrong, and the fix
- Coil will not fit over a large shoulder: roll the patient slightly onto the affected side.
- Breast tissue in the way of the coil: ask the patient to move the breast away from the affected side, then fit the coil.
- Hand drifts back to internal rotation: re-place the sponge and remind them.
- Check on the first localisersurveylocalizer
- On the axial localisersurveylocalizer the glenoid and humeral head are centred in the coil’s signal.
- On the coronal localisersurveylocalizer the coverage includes the AC joint and the supraspinatus tendon.
- Good to know
- Internal rotation can distort the supraspinatus and subscapularis tendons and should be avoided. Partial external rotation is the aim: forcing full external rotation can make fluid in the biceps tendon sheath look like a small cuff tear. Keep the turn comfortable rather than forced.
Explain it simply
The shoulder is scanned lying on the back with the arm by the side and the hand turned slightly outwards. That turn puts the tendons in a position the radiologist expects.
Why it matters: A rotated arm can make a healthy tendon look torn, or distort the tendons so a tear is harder to judge.
Sources: Department notes; ACR Shoulder PP 2024; Ashir 2020; Del Grande 2014; RANZCR MRI Safety 2021; MHRA 2026; ACR MR Safety 2026
Scapula

- Shoulder coil over the shoulder
- Thumb up, sponge in the hand
- Pad between the other arm and body
- Call bell in the other hand
- Position
- Supine, head first, affected arm by the side.
- Coil
- shoulder coildS Shoulder coilShoulder Shape 16 coil.
- Centre
- The body of the scapula, in the middle of the coil.
- Set up
- Set up as for a shoulder: hand turned partly out with a sponge in it, and for a large patient a rolled towel under the forearm so the hand rests clear of the thigh.
- Pads between the arm and the body and between the hand and the thigh.
- Large patient: roll slightly onto the affected side to fit the coil.
- Move breast tissue away from the affected side before fitting the coil.
- What goes wrong, and the fix
- Medial border of the scapula outside the coil’s signal: move the coil medially and repeat the localisersurveylocalizer.
- Scapulothoracic bursa queried: make sure the coverage reaches far enough medially.
- Check on the first localisersurveylocalizer
- A second localisersurveylocalizer angled to the blade of the scapula shows the whole blade inside the coil’s signal.
Explain it simply
The shoulder blade is a flat bone on the back of the ribs. The coil is placed so the whole blade, including its inner edge, is inside the picture.
Why it matters: Problems between the shoulder blade and ribs sit at the inner edge, which is easy to miss.
Sources: Department notes; ACR Shoulder PP 2024; RANZCR MRI Safety 2021
Pectoralis major

- Flat coil strapped on the upper back
- Pad between the raised arm and head
- Pad between arm and chest
- Call bell in the hand
- Position
- Prone, head first.
- Coil
- anterior body coildS Torso coilBody 18 coil on the back, with the spine coildS Posterior coilSpine 32 coil in the table under the chest.
- Centre
- The affected side of the chest, moved as close to the middle of the bore as the patient allows.
- Set up
- Move the affected side as far towards the isocentre as possible, then pad between the patient and the bore on the side that moved towards it, and recheck both sides.
- Affected arm down by the side and tucked in close, with an insulating pad between the arm and the chest wall.
- Unaffected arm up above the head if comfortable, with a pad between the raised arm and the head.
- Head turned to the comfortable side on a folded towel or a prone head support, face and airway clear, and the call bell in the hand by the side.
- Turn the bore fan up: patients often get hot in this position.
- What goes wrong, and the fix
- Patient too hot: fan up, fewer covers, and check on them between sequences.
- Spectral fat suppressionSPIR/SPAIR fat suppressionFat Sat/SPAIR fat suppression patchy over the chest: check the affected side is as central as it can go.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer covers the muscle from the clavicle and sternum to its insertion on the humerus.
- Good to know
- The field is less uniform away from the isocentre, which is why the affected side goes as central as possible. Lying face down puts the muscle directly on the spine coildS Posterior coilSpine 32 coil in the table, the coil elements closest to it, because a receive coil is most sensitive near its elements; the anterior body coildS Torso coilBody 18 coil on the back completes the set-up (department practice).
Explain it simply
The chest muscle is scanned lying face down, shifted so the injured side is near the middle of the scanner, where the pictures are best.
Why it matters: With the injured side central, fat suppression is more even across the muscle.
Sources: Department notes; Gruber 2018; Del Grande 2014; MHRA 2026; RANZCR MRI Safety 2021; ACR MR Safety 2026
Elbow

- Coil around the elbow, sandbag on top
- Sponge under the elbow
- Palm up, sandbag over the wrist
- Head turned, cheek on a towel
- Position
- Prone in the superman position (affected arm overhead), or lying on the affected side.
- Coil
- small wrap-around extremity coildS Small Extremity coilUltraFlex Small 18 coil. If the elbow is held flexed, use the shoulder coildS Shoulder coilShoulder Shape 16 coil (department practice) or a knee coildS Knee coilTx/Rx Knee 18 coil, as the guideline suggests for an adult who cannot straighten the elbow.
- Centre
- The elbow joint, in the middle of the coil.
- Set up
- Towel under the armpit; keep the shoulder flat, shuffling the patient back if needed.
- Small sponge or flannel under the elbow; extend the elbow as far as is comfortable.
- If the arm will not straighten fully, a small sponge under the forearm supports it slightly bent.
- Hand turned palm up (supinated), with a sandbag over the hand to hold it.
- Arm overhead: pad between the raised arm and the head, keep the hand and coil off the bore wall, and put the call bell in the other hand.
- Sponge or flannel over the arm, then close the coil with the larger flap over the smaller, overlapping. Sandbag over the coil.
- What goes wrong, and the fix
- Signal drop-off across the joint: the coil flaps are not overlapping; reclose with the larger flap over the smaller.
- Patient cannot hold the superman position: scan with the arm down by the side, with pads between the arm and the body, the hand and the thigh, and the arm and the bore; expect less even spectral fat suppressionSPIR/SPAIR fat suppressionFat Sat/SPAIR fat suppression, and use DixonmDIXONDixon or STIR.
- Elbow held flexed: change to the shoulder coildS Shoulder coilShoulder Shape 16 coil or a knee coildS Knee coilTx/Rx Knee 18 coil, with a pad between the skin and the coil.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer shows the joint centred, with about a third of the humerus and of the forearm bones either side.
- Good to know
- Arm overhead puts the elbow near the isocentre, where the field is most uniform; by the side it lies where the field is less uniform and chemically selective fat suppression suffers. The prone position is uncomfortable for many patients, and pronating the forearm can distort the collateral ligaments and tendons on the coronal images.
Explain it simply
For the elbow, the patient lies face down with the arm stretched above the head, like Superman flying. That puts the elbow in the middle of the scanner.
Why it matters: Even fat suppression makes small ligament and tendon tears easier to see.
Sources: Department notes; ACR Elbow PP 2026; Del Grande 2014; RANZCR MRI Safety 2021; MHRA 2026
Wrist

- Wrist coil around the wrist
- Wrist straight, in line with the forearm
- Sandbag over the forearm
- Pad between the arm and the head
- Position
- Prone in the superman position, affected arm overhead.
- Coil
- wrist coildS Wrist coilHand/Wrist 16 coil. Use the small wrap-around extremity coildS Small Extremity coilUltraFlex Small 18 coil if the hand does not fit or the wrist is in a cast.
- Centre
- The wrist joint, at the level of the tip of the ulnar styloid.
- Set up
- Wrist straight in line with the forearm, not bent to either side.
- Towel under the armpit, and a sponge or towel under the forearm.
- Small wrist: a small sponge over the hand to stop it moving in the coil.
- Sandbag over the forearm.
- Arm overhead: pad between the raised arm and the head, keep the hand and coil off the bore wall, and put the call bell in the other hand.
- What goes wrong, and the fix
- Wrist cannot be held straight: shuffle the whole body away from the side being scanned until the wrist lines up.
- Patient cannot lie in the superman position: arm by the side using the base plate that holds the wrist coil on its side, with pads between the arm and the body, the hand and the thigh, and the arm and the bore; rolling the patient partly onto their side brings the wrist nearer the middle.
- Superman uncomfortable: a soft pad under the chest can make it easier to hold.
- Check on the first localisersurveylocalizer
- On the coronal localisersurveylocalizer the radius and ulna run straight and the carpal bones are in the middle of the coil.
- Good to know
- Overhead positioning improves the evenness of fat suppression because the wrist is at the isocentre. A wrist angled steeply in the magnet can produce magic angle artefact in the tendons.
Explain it simply
The wrist goes in a small coil above the head while the patient lies face down. Keeping it straight shows the small wrist bones and ligaments clearly.
Why it matters: The wrist ligaments are tiny, and a bent wrist can make healthy tendons look abnormal.
Sources: Department notes; ACR Wrist PP 2022; Del Grande 2014; RANZCR MRI Safety 2021; MHRA 2026
Hand for rheumatoid arthritis

- Fingers and thumb taped together
- Thumb alongside the hand, not tucked under
- Hand flat on a long sponge
- Coil open; it closes before scanning
- Position
- Prone in the superman position, affected hand overhead.
- Coil
- small wrap-around extremity coildS Small Extremity coilUltraFlex Small 18 coil.
- Centre
- The middle of the hand and wrist.
- Set up
- Towel under the armpit; shoulder flat.
- Tape the fingers and thumb together gently so there are no air gaps, without bending or squeezing any joint.
- Hand and fingers straight, in line with the forearm, flat on a long straight sponge.
- Thumb alongside the hand, not tucked under the palm.
- Sponge over the hand, close the coil with the larger flap over the smaller and overlapping, then a sandbag over the arm.
- Arm overhead: pad between the raised arm and the head, keep the hand and coil off the bore wall, and put the call bell in the other hand.
- What goes wrong, and the fix
- Spectral fat suppressionSPIR/SPAIR fat suppressionFat Sat/SPAIR fat suppression patchy between the fingers: re-tape to close the air gaps, or use DixonmDIXONDixon.
- Thumb tucked under: bring it into line, or you will need extra coronal coverage.
- Signal drop-off: the coil flaps are not overlapping.
- Patient cannot lie in the superman position: arm by the side, with pads between the arm and the body, the hand and the thigh, and the arm and the bore.
- Check on the first localisersurveylocalizer
- On the coronal localisersurveylocalizer the hand and wrist lie flat and straight, with all the joints in one plane.
- Good to know
- Air spaces against soft tissue make the local field uneven, which spoils chemically selective fat suppression; taping also immobilises the fingers. With the thumb adducted alongside the fingers, magic angle artefact in the thumb tendons is reduced. Some services image both hands together in the “prayer” position for inflammatory arthritis.
Explain it simply
The fingers are taped together so there is no air between them. Air next to skin upsets the scanner’s fat-removal trick, so taping gives cleaner pictures.
Why it matters: Patchy fat suppression can hide joint inflammation or mimic it.
Sources: Department notes; ACR Fingers Toes PP 2023; ACR Ankle PP 2026; Del Grande 2014; ACR Wrist PP 2022; RANZCR MRI Safety 2021; MHRA 2026
Finger

- Tape round the fingers and thumb
- Coil open; it closes before scanning
- Coil reaches past the fingertips
- Sponge under the forearm
- Position
- Prone in the superman position, affected hand overhead.
- Coil
- small wrap-around extremity coildS Small Extremity coilUltraFlex Small 18 coil.
- Centre
- The joint or segment of the finger under question.
- Set up
- Towel under the armpit; shoulder flat.
- Tape the fingers and thumb together gently with no air gaps, without deforming any joint.
- Hand and fingers as straight as possible.
- Sponge over the hand, coil flaps overlapping with the larger over the smaller, sandbag over the arm.
- Arm overhead: pad between the raised arm and the head, keep the hand and coil off the bore wall, and put the call bell in the other hand.
- What goes wrong, and the fix
- Signal drop-off: the flaps are not overlapping.
- Patient cannot lie in the superman position: arm by the side, with pads between the arm and the body, the hand and the thigh, and the arm and the bore.
- Glomus tumour queried: make sure the fingertip is well inside the coil.
- Check on the first localisersurveylocalizer
- Second localisersurveylocalizer angled to the anatomy or lesion in question; coverage follows the indication.
Explain it simply
The finger is taped to its neighbours to keep it still and straight, and to remove air gaps that spoil the pictures.
Why it matters: Small finger structures blur with even slight movement.
Sources: Department notes; ACR Fingers Toes PP 2023; ACR Ankle PP 2026; Zaitsev 2015; RANZCR MRI Safety 2021; MHRA 2026
Thumb

- Hand on its side, thumb on top
- Sponge in the palm
- Four fingers taped, thumb left free
- Coil open; it closes before scanning
- Position
- Prone in the superman position, affected hand overhead.
- Coil
- small wrap-around extremity coildS Small Extremity coilUltraFlex Small 18 coil.
- Centre
- The thumb joint in question (IP or MCP joint).
- Set up
- Towel under the armpit; shoulder flat.
- Hand on its side with a sponge in the palm.
- Tape the fingers together to remove air gaps; leave the thumb slightly abducted and lined up with the length of the bore, supported by the sponge, with the MCP joint in its natural shape.
- Sponge over the hand, coil flaps overlapping with the larger over the smaller, sandbag over the arm.
- Arm overhead: pad between the raised arm and the head, keep the hand and coil off the bore wall, and put the call bell in the other hand.
- What goes wrong, and the fix
- Signal drop-off: the flaps are not overlapping.
- Patient cannot lie in the superman position: arm by the side, with pads between the arm and the body, the hand and the thigh, and the arm and the bore.
- Check on the first localisersurveylocalizer
- Second localisersurveylocalizer angled to the joint in question (IP or MCP joint).
- Good to know
- When the thumb is the main interest, the guideline aligns it with the main magnetic field, slightly abducted, which may need a towel roll or foam pad in the palm.
Explain it simply
The hand rests on its side with a sponge in the palm, which lines the thumb up with the scanner.
Why it matters: Lining the thumb up gives clean views along its ligaments and tendons.
Sources: Department notes; ACR Fingers Toes PP 2023; RANZCR MRI Safety 2021; MHRA 2026
Pelvis and lower limb
Lower limb examinations are done feet first. The ankle and foot coil cards share one set-up; the details that differ are the foot position and the centring point. For every lower limb card: legs uncrossed, a pad between the legs and between the other leg and the coil, and a pad between the skin and the coil, especially if the coil transmits as well as receives.
Sources: Department notes; RANZCR MRI Safety 2021; ACR MR Safety 2026
Pelvis and hip

- Coil over the pelvis
- Folded towel between the thighs
- Sponge under the knee
- Hands apart on pads, bell in hand
- Position
- Supine, feet first, small sponges under the knees and a folded towel between the thighs.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil over the pelvis.
- Centre
- The level of the greater trochanters (department practice).
- Set up
- Small sponges under the knees.
- A towel or insulating pad between the thighs so they do not touch.
- Hands apart, resting on the chest in arm holders with a fold of gown or a pad between the hands and the chest; not clasped and not touching the thighs.
- Lie the pelvis square on the table and centre the anterior body coildS Torso coilBody 18 coil level over it.
- What goes wrong, and the fix
- Pelvis rotated on the localisersurveylocalizer: straighten the patient before planning.
- Thighs or hands touching skin to skin: add pads before the table goes in.
- Check on the first localisersurveylocalizer
- A second localisersurveylocalizer aligned symmetrically through both hips using the acetabula, with the boxes at matching angles.
- Good to know
- Pads between the thighs, and between the arms and the body, break skin-to-skin loops that can concentrate RF heating at small contact points.
Explain it simply
The hips are scanned lying on the back with a towel between the thighs. The towel stops the legs touching, which helps prevent a skin burn from the radio waves.
Why it matters: A straight pelvis lets the radiologist compare the two hips side by side.
Sources: Department notes; RANZCR MRI Safety 2021; ACR MR Safety 2026; MHRA 2026
Hamstring origin

- Coil over pelvis and upper thighs
- Folded towel between the thighs
- Sponge under the knee
- Hands apart on pads, bell in hand
- Position
- Supine, feet first, small sponges under the knees and a folded towel between the thighs.
- Coil
- spine coildS Posterior coilSpine 32 coil with the anterior body coildS Torso coilBody 18 coil over the pelvis and upper thighs.
- Centre
- The level of the greater trochanters (department practice).
- Set up
- Set up as for a pelvis and hip: knee sponges, towel between the thighs, hands apart on the chest in arm holders with a pad or fold of gown under them.
- Use this protocol when the question is the hamstring origin, or the pain is clearly at the ischial tuberosity.
- If a second station may be needed, check the anterior body coildS Torso coilBody 18 coil can be moved down to cover the thigh as far as the knee.
- What goes wrong, and the fix
- High signal runs below the bottom of the field of view: add a second station, moving the anterior body coildS Torso coilBody 18 coil down so it covers to the knee.
- Torn tendon end not on the images: extend the coverage until it is included.
- Check on the first localisersurveylocalizer
- A second localisersurveylocalizer centred on the ischia, symmetrical through both of them (or through the hips if the ischia are hard to see).
Explain it simply
The hamstrings attach to the sitting bone at the bottom of the pelvis. If the tendon has torn and pulled down the thigh, the scan is extended to find the torn end.
Why it matters: A scan that stops above the torn end has to be repeated.
Sources: Department notes; RANZCR MRI Safety 2021; ACR MR Safety 2026; MHRA 2026
Knee

- Knee coil around one knee
- Supports under the other leg
- Call bell in the hand
- Ear protection
- Position
- Supine, feet first, knee straight or nearly straight in the coil.
- Coil
- knee coildS Knee coilTx/Rx Knee 18 coil.
- Centre
- The apex (lower tip) of the patella at the centre of the coil.
- Set up
- Place the apex of the patella at the middle of the coil.
- Flex the knee no more than about 10 degrees (department practice), resting comfortably.
- Pad the knee inside the coil so it cannot move.
- Legs uncrossed, with a pad between the legs and between the other leg and the coil.
- Mild external rotation of the leg is often comfortable.
- What goes wrong, and the fix
- Signal drop-out in the femur or the tibia and fibula: the knee is not centred in the coil; move it and repeat the localisersurveylocalizer.
- Large knee a tight fit: use a slightly larger coil if one is available, rather than forcing the knee in.
- Large effusion: add axial slices so all of the fluid is covered.
- Check on the first localisersurveylocalizer
- On the localisersurveylocalizer the joint line is in the middle of the coil, with the fibular neck visible at the lower end.
- The fibula is visible on the second localisersurveylocalizer for planning.
- Good to know
- The guideline describes the knee as completely or nearly completely extended, with gentle immobilisation and comfort measures to reduce involuntary motion, and a coil sized closely to the knee.
Explain it simply
The knee sits in a tube-shaped coil with the bottom of the kneecap in the middle. A common reason for a poor knee scan is the knee sitting too high or too low in the coil.
Why it matters: A knee centred in the coil shows the whole joint evenly, top to bottom.
Sources: Department notes; ACR Knee PP 2025; Gruber 2018; Lefevre 2016; RANZCR MRI Safety 2021; ACR MR Safety 2026
Ankle

- Foot and ankle coil on one foot
- Wedge under the knee and calf
- Pad between the lower legs
- Block between other foot and coil

- Boot-shaped foot and ankle coil
- Wedge under the knee
- Block between other foot and coil
- Position
- Supine, feet first, with a support under the knees.
- Coil
- foot and ankle coildS FootAnkle coilFoot/Ankle 16 coil.
- Centre
- The ankle joint, in the middle of the coil.
- Set up
- Support under the knees.
- Foot at a right angle (neutral) so the heel rests on the base plate.
- Close the coil top without it pressing on the shin.
- Check the coil cable has no sharp bend; move the patient up the table if it is kinked.
- Legs uncrossed, with a pad between the legs and between the other leg and the coil.
- What goes wrong, and the fix
- Coil top presses on the leg: reduce the knee flexion and take the sponge out of the coil.
- Still pressing: put a small foam sponge under the heel instead, or a flannel on the leg, so the coil does not dig in.
- Syndesmosis injury: extend the axial coverage upwards until the coil signal runs out.
- Check on the first localisersurveylocalizer
- On the localisersurveylocalizer the talus is in the middle of the coil, with the plantar fascia and the ankle joint both in view on the sagittal.
- Good to know
- DixonmDIXONDixon gives more reliable fat suppression than chemically selective methods where the off-centre position and the irregular air and soft tissue edges of the ankle and foot make the field uneven.
Explain it simply
The foot goes into a boot-shaped coil with the heel down on its base. Getting the heel down and the coil off the shin keeps it still and comfortable.
Why it matters: A coil pressing on the shin is uncomfortable, and an uncomfortable patient moves.
Sources: Department notes; ACR Ankle PP 2026; MHRA 2026; RANZCR MRI Safety 2021
Achilles tendon

- Foot and ankle coil on one foot
- Wedge under the knee and calf
- Pad between the lower legs
- Block between other foot and coil

- Foot and ankle coil
- Heel resting down in the coil base
- Wedge under the knee
- Block between other foot and coil
- Position
- Supine, feet first, with a support under the knees.
- Coil
- foot and ankle coildS FootAnkle coilFoot/Ankle 16 coil. Use the anterior body coildS Torso coilBody 18 coil instead when a complete rupture is suspected.
- Centre
- The Achilles tendon, in the middle of the coil.
- Set up
- Set up as for the ankle: foot at a right angle with the heel on the base plate, coil top off the shin, cable without sharp bends.
- Before you start, check the request and look at the back of the leg for the signs of a complete rupture.
- Suspected complete rupture: use the ankle coil base plate without its top, swap the thick sponge for a thin one so the leg lies more level with the ankle, rest the heel on the base plate, and lay the anterior body coildS Torso coilBody 18 coil over the whole ankle and calf, padded level. If the base plate contains coil elements, keep it plugged in or use a plain positioning pad instead; never leave an unplugged coil part in the bore.
- What goes wrong, and the fix
- Dent where the tendon should be, with a lump in the calf from the retracted tendon: change to the anterior body coildS Torso coilBody 18 coil set-up so the whole tendon is covered.
- Tendon still attached at its insertion: scan in the foot and ankle coildS FootAnkle coilFoot/Ankle 16 coil, and add anterior body coildS Torso coilBody 18 coil imaging afterwards if needed.
- High signal runs above the ankle coil’s coverage: finish the ankle study, then image the calf with the anterior body coildS Torso coilBody 18 coil.
- Patient cannot flex the ankle because of the rupture: accept the position and adjust the coverage.
- Check on the first localisersurveylocalizer
- On the localisersurveylocalizer the whole tendon, from the calcaneus up to the muscle junction or the torn end, is inside the coil’s signal.
- Good to know
- The guideline notes that a large Achilles tear may need a sagittal field of view of about 22 cm, which can be more than an ankle coil covers.
Explain it simply
If the tendon has snapped, its top end springs up into the calf. A longer, flexible coil laid over the calf makes sure both ends are in the picture.
Why it matters: A study that misses the upper torn end has to be repeated in a different coil.
Sources: Department notes; ACR Ankle PP 2026; ACR MR Safety 2026; RANZCR MRI Safety 2021
Foot

- Foot and ankle coil on one foot
- Wedge under the knee and calf
- Pad between the lower legs
- Block between other foot and coil

- Foot and ankle coil
- Coil base under the heel
- Block between other foot and coil
- Position
- Supine, feet first, with a support under the knees.
- Coil
- foot and ankle coildS FootAnkle coilFoot/Ankle 16 coil.
- Centre
- The toes for the forefoot; the region in question for the midfoot.
- Set up
- Support under the knees; foot at a right angle (neutral) so the heel rests on the base plate.
- Close the coil top without it pressing on the leg.
- Check the coil cable has no sharp bend; move the patient up the table if it is kinked.
- Legs uncrossed, with a pad between the legs and between the other leg and the coil.
- What goes wrong, and the fix
- Coil top presses on the leg: reduce the knee flexion and take the sponge out of the coil; add a sponge or flannel on the leg if it still touches.
- Pain spread across the forefoot or midfoot: start with a large field of view DixonmDIXONDixon sagittal, then angle the later sequences to what it shows.
- Problems in both the ankle and the foot: consider two separate studies so each is angled correctly.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer shows the forefoot or midfoot region in the middle of the coil, with the toes included when they are part of the question.
Explain it simply
The foot goes into the same boot-shaped coil as the ankle, pushed further in so the front of the foot sits in the middle.
Why it matters: Foot bones lie at many angles, so a well-centred first picture makes the angled sequences easier to plan.
Sources: Department notes; ACR Ankle PP 2026; RANZCR MRI Safety 2021
Great toe

- Toes under the coil top
- Foot relaxed, not held upright
- Heel resting on the base plate
- Position
- Supine, feet first, with a support under the knees.
- Coil
- foot and ankle coildS FootAnkle coilFoot/Ankle 16 coil.
- Centre
- The great toe.
- Set up
- Leave the foot relaxed rather than held at a right angle; this makes it easier to rest the great toe against the base plate.
- Lift the coil so the great toe sits more centrally in it.
- Check the coil cable has no sharp bend; move the patient up the table if needed.
- Legs uncrossed, with a pad between the legs and between the other leg and the coil.
- What goes wrong, and the fix
- Toe near the edge of the coil: lift or shift the coil until the toe is central, and repeat the localisersurveylocalizer.
- Plantar plate or sesamoids in question: expect a much flatter angle than for the joints or bone.
- Check on the first localisersurveylocalizer
- Second localisersurveylocalizer angled to the anatomy in question (MTP or IP joint, plantar plate, sesamoids, tendons).
Explain it simply
For the big toe the foot rests loosely, and the coil is lifted so the toe is right in the middle.
Why it matters: The big toe joint is small, so centring it well gives sharper detail.
Sources: Department notes; ACR Fingers Toes PP 2023; Gruber 2018; RANZCR MRI Safety 2021
Body and whole body
Body examinations add breathing and bowel movement to the usual positioning tasks. Coach the breath-holds and place the respiratory sensorrespiratory beltrespiratory belt before the patient goes in (see Breath-holds and breathing sensors). Preparation and medicines are checked for contraindications and given by authorised staff as your protocol directs.
Sources: Department notes; Zaitsev 2015; Avesani 2024
MRCP, pancreas and liver

- Coil over the upper abdomen
- Arms above the head, hands apart
- Pads between arms and head
- Wedge under the knees
- Position
- Supine, feet first, with a support under the legs; arms above the head on a support, hands apart and not touching the head, with a pad between the arms and the head and the hands padded off the bore; bring them down if the hands go numb or tingle.
- Coil
- anterior body coildS Torso coilBody 18 coil over the upper abdomen, with the spine coildS Posterior coilSpine 32 coil in the table.
- Centre
- The upper abdomen, so the coil covers the whole liver and biliary system.
- Set up
- Check the patient has fasted as your protocol directs; if not, tell the radiologist and record it.
- Place the respiratory sensorrespiratory beltrespiratory belt over the upper abdomen with a soft sponge, so the anterior body coildS Torso coilBody 18 coil presses on it; with a built-in or camera sensor, position as its instructions direct.
- Check the breathing rate and trace are regular on the console before the first breath-hold or respiratory triggering sequence.
- Explain the breath-holds and practise them: the same size of breath every time.
- If you use a navigatorrespiratory navigator2D PACE navigator, place it across the dome of the right hemidiaphragm, about a third in lung and two thirds in liver (department practice).
- What goes wrong, and the fix
- Not coping with breath-holds: shorten each hold or lengthen the rest between them.
- Still not coping: change from automatic breath-hold commandsAutoVoiceauto voice commands to giving the instructions yourself, and start each hold once the breathing trace and the coil have stopped moving.
- Still failing: change to respiratory triggering versions, or your department’s fast protocol.
- Arms cannot stay above the head: bring them down by the sides, with pads between the arms and the body.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer shows the whole liver and biliary tree inside the coil’s signal; a long liver may need the coil moved down.
- Good to know
- Fasting reduces fluid in the stomach and duodenum, reduces bowel movement and helps the gallbladder fill, so fluid in the bowel does not overlap the ducts. Some services also give a negative oral agent to darken stomach and duodenal fluid. Contrast for liver protocols is given as your protocol and radiologist direct.
Explain it simply
The bile ducts show up because they contain still fluid. An empty stomach and steady breathing keep other fluid and movement from getting in the way.
Why it matters: A small stone in a bile duct can hide behind fluid in the bowel or a breathing blur.
Sources: Department notes; Griffin 2012; Zaitsev 2015; RANZCR MRI Safety 2021; ACR MR Safety 2026; PhVitalEye; SHBioMatrix
Female pelvis

- Coil over the pelvis
- Cushion under the legs
- Hands apart, bell in one hand
- Ear protection
- Position
- Supine, feet first, with a pillow or sponge under the legs.
- Coil
- anterior body coildS Torso coilBody 18 coil over the pelvis, with the spine coildS Posterior coilSpine 32 coil in the table. Add a second anterior array when the abdomen is included.
- Centre
- The pelvis; higher when abdominal sequences are part of the protocol.
- Set up
- If your protocol uses an antispasmodic or contrast, the patient is checked for contraindications and the drug is given by authorised staff as your protocol directs; place any IV line before positioning.
- A small pillow under the legs keeps the anterior body coildS Torso coilBody 18 coil parallel to the table.
- A strap across the pelvis can limit breathing movement, if the patient can tolerate it.
- Place the respiratory sensorrespiratory beltrespiratory belt on the abdomen when abdominal breath-hold sequences are included, or use your system’s built-in or camera sensor.
- When the abdomen is included, set the coil high enough and the arms above the head on a support, hands apart and not touching the head, with a pad between the arms and the head and the hands padded off the bore; bring them down if the hands go numb or tingle.
- For a pelvis-only scan, rest the hands apart on the chest with a fold of gown under them, or by the sides with pads between the arms and the body; keep the hands off the thighs.
- What goes wrong, and the fix
- Uterus or bowel blurred on T2 images: check whether the antispasmodic was given, and time the key sequences while it is working.
- Abdominal sequences with dark areas at the top: the coil is too low; move it up and repeat the localisersurveylocalizer.
- Breath-holds failing on the abdominal single-shot TSESSh TSEHASTE or 3D T1 gradient echoTHRIVEVIBE sequences: see Breath-holds and breathing sensors.
- Check on the first localisersurveylocalizer
- The localisersurveylocalizer covers the whole uterus and both ovaries in the coil’s signal, and the abdomen to the diaphragm when it is part of the protocol.
- Good to know
- An antispasmodic reduces motion artefact from bowel and uterine peristalsis; its effect lasts a shorter time given intravenously than intramuscularly, which is why the timing of the key sequences matters. Bladder filling and fasting advice differs by indication, so follow your protocol for the clinical question.
Explain it simply
The bowel and womb move a little on their own. A medicine that relaxes them for a short time, given just before the scan, keeps the pictures sharp.
Why it matters: A blurred uterus is hard to measure or assess.
Sources: Department notes; Avesani 2024; Thomassin-Naggara 2025; RANZCR MRI Safety 2021; ACR MR Safety 2026
Whole-body STIR

- Head and neck coil
- Two flat coils over the body
- Pads between arms and body
- Wedges and pads at the knees
- Position
- Supine, head first, legs straight with the knees pointing up.
- Coil
- head and neck coildS HeadNeck coilHead/Neck 20 coil without its base plate, with two anterior body coildS Torso coilBody 18 coil arrays and the spine coildS Posterior coilSpine 32 coil in the table, plus the further arrays your system’s whole-body set-up uses for the legs and feet.
- Centre
- As your system’s whole-body set-up directs; the coverage runs from the top of the head to the feet.
- Set up
- Take the base plate off the head and neck coil.
- Arms by the sides, held in with arm supports and sandbags, with insulating pads between the arms and the body.
- Small wedge sponges or a rolled towel under the legs, knees pointing straight up, and sandbags at the feet.
- Insulating pads between the thighs, knees and ankles so the legs do not touch, and between the arms and the bore in a larger patient.
- Place the respiratory sensorrespiratory beltrespiratory belt over the lower chest or upper abdomen, where breathing movement is greatest, or use your system’s built-in or camera sensor.
- Check that no bare skin touches a coil edge or cable, and turn the bore fan up.
- What goes wrong, and the fix
- Patient warm during the long exam: fan up, fewer covers, and check on them between stations.
- Top of the head or the feet cut off the composed multi-stationcomposed (MobiView)composed (Composing) views: reposition so the full length fits, then repeat the localisersurveylocalizer.
- Head and neck coil base plate left on: the head is lifted and the spine does not lie level; take it off.
- Check on the first localisersurveylocalizer
- Coronal and sagittal composed multi-stationcomposed (MobiView)composed (Composing) views run from the top of the head to the feet.
- The stations overlap, with no gap between them.
- Good to know
- STIR is used because it suppresses fat evenly over a large field of view and away from the isocentre, where chemically selective methods struggle. With many coils and cables on the body, check for skin contact before the table goes in.
Explain it simply
The whole body is scanned in sections from head to toe. Lots of coils cover the body, so the team puts pads under coil edges and cables so they never touch bare skin.
Why it matters: A missing section means a repeat scan, so the full length must be covered.
Sources: Department notes; Del Grande 2014; RANZCR MRI Safety 2021; MHRA 2026; ACR MR Safety 2026
Sources
Practical steps come from departmental scanning notes written by working technologists, rewritten and checked for this page. Reasons and vendor terms cite published sources. Guidelines are cited in their current editions: the ACR Manual on MR Safety (2026) and the RANZCR MRI Safety Guidelines (version 3.0, 2021). Sources were checked in November 2026.
- American College of Radiology, ACR Committee on MR Safety. ACR Manual on MR Safety, 2026 edition. Reston, VA: ACR; 2026. https://www.acr.org/Clinical-Resources/Clinical-Tools-and-Reference/Radiology-Safety/MR-Safety
- American College of Radiology, ASNR, SABI, SSR. ACR-ASNR-SABI-SSR Practice Parameter for the Performance of Magnetic Resonance Imaging (MRI) of the Adult Spine. Revised 2023 (Resolution 6). https://gravitas.acr.org/PPTS/GetDocumentView?docId=4
- American College of Radiology, SABI, SPR, SSR. ACR-SABI-SPR-SSR Practice Parameter for the Performance of Magnetic Resonance Imaging (MRI) of the Wrist. Revised 2022 (Resolution 35). https://gravitas.acr.org/PPTS/GetDocumentView?docId=147
- American College of Radiology, SPR, SSR. ACR-SPR-SSR Practice Parameter for the Performance and Interpretation of Magnetic Resonance Imaging (MRI) of the Ankle and Hindfoot. Revised 2026 (Resolution 6). https://gravitas.acr.org/PPTS/GetDocumentView?docId=178
- American College of Radiology, SPR, SSR. ACR-SPR-SSR Practice Parameter for the Performance and Interpretation of Magnetic Resonance Imaging (MRI) of the Elbow. Revised 2026 (Resolution 7). https://gravitas.acr.org/PPTS/GetDocumentView?docId=180
- American College of Radiology, SPR, SSR. ACR-SPR-SSR Practice Parameter for the Performance and Interpretation of Magnetic Resonance Imaging (MRI) of the Fingers and Toes. Revised 2023 (Resolution 14). https://gravitas.acr.org/PPTS/GetDocumentView?docId=10
- American College of Radiology, SPR, SSR. ACR-SPR-SSR Practice Parameter for the Performance and Interpretation of Magnetic Resonance Imaging (MRI) of the Knee. Revised 2025 (Resolution 7). https://gravitas.acr.org/PPTS/GetDocumentView?docId=90
- American College of Radiology, SPR, SSR. ACR-SPR-SSR Practice Parameter for the Performance and Interpretation of Magnetic Resonance Imaging (MRI) of the Shoulder. Revised 2024 (Resolution 18). https://gravitas.acr.org/PPTS/GetDocumentView?docId=45
- Ashir A, Lombardi A, Jerban S, Ma Y, Du J, Chang EY. Magnetic resonance imaging of the shoulder. Pol J Radiol 2020;85:e420-e439. doi:10.5114/pjr.2020.98394
- Avesani G, Panico C, Nougaret S, Woitek R, Gui B, Sala E, et al. ESR Essentials: characterisation and staging of adnexal masses with MRI and CT: practice recommendations by ESUR. Eur Radiol 2024;34(12):7673-7689. doi:10.1007/s00330-024-10817-1
- Del Grande F, Santini F, Herzka DA, Aro MR, Dean CW, Gold GE, Carrino JA. Fat-suppression techniques for 3-T MR imaging of the musculoskeletal system. RadioGraphics 2014;34(1):217-233. doi:10.1148/rg.341135130
- Griffin N, Charles-Edwards G, Grant LA. Magnetic resonance cholangiopancreatography: the ABC of MRCP. Insights Imaging 2012;3(1):11-21. doi:10.1007/s13244-011-0129-9
- Gruber B, Froeling M, Leiner T, Klomp DWJ. RF coils: a practical guide for nonphysicists. J Magn Reson Imaging 2018;48(3):590-604. doi:10.1002/jmri.26187
- Khot R, McGettigan M, Patrie JT, Feuerlein S. Quantification of gas exchange-related upward motion of the liver during prolonged breathholding: potential reduction of motion artifacts in abdominal MRI. Br J Radiol 2020;93(1106):20190549. doi:10.1259/bjr.20190549
- Lefevre N, Naouri JF, Herman S, Gerometta A, Klouche S, Bohu Y. A current review of the meniscus imaging: proposition of a useful tool for its radiologic analysis. Radiol Res Pract 2016;2016:8329296. doi:10.1155/2016/8329296
- Ma C, Li J, Chen Y, et al. “Automated Fast Liver MR Scan: A Preliminary Evaluation.” MAGNETOM Flash 69 (3/2017). https://marketing.webassets.siemens-healthineers.com/1800000004387135/36f10e71d530/magnetom-flash-69_automated-fast-liver-mr-scan-a-preliminary-evaluation_lu_1800000004387135.pdf
- Medicines and Healthcare products Regulatory Agency. Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, fifth edition (revision 5.0.3). London: MHRA; first published July 2026, revised September 2026. https://www.gov.uk/government/publications/safety-guidelines-for-magnetic-resonance-imaging-equipment-in-clinical-use
- Murrie VL, Wilson H, Hollingworth W, Antoun NM, Dixon AK. Supportive cushions produce no practical reduction in lumbar lordosis. Br J Radiol 2002;75(894):536-538. doi:10.1259/bjr.75.894.750536
- Philips FieldStrength, “How liver MR imaging gets boost from latest techniques” (University of Bonn, G. M. Kukuk), June 2015. https://www.philips.si/healthcare/education-resources/publications/fieldstrength/liver-mri-with-latest-methods-university-bonn
- Philips Ingenia Release 5 Technical Description, 03/2014. Operator documentation, not publicly available.
- Philips Release 5.6 Parameters manual, 2020. Operator documentation, not publicly available.
- Philips, “Ingenia 1.5T Evolution: Boost your MR performance, and confidence”, brochure, 2022. https://www.documents.philips.com/assets/20230113/208c77f49d034bc6bf41af8901420f6a.pdf
- Philips, dS FootAnkle 16ch datasheet, 2016. https://www.documents.philips.com/assets/20171026/3ce739b4709445acaa29a81800f407ea.pdf
- Philips, dS HandWrist 16ch datasheet, 2018. https://www.documents.philips.com/assets/20180511/8c61b82c569444668980a8dd0081d23b.pdf
- Philips, dS HeadNeckSpine 1.5T datasheet (Breeze), 2021. https://www.documents.philips.com/assets/20220215/f9b0a4be0b514fe99a89ae3d00fe0e27.pdf
- Philips, dS Shoulder 16ch datasheet, 2016. https://www.documents.philips.com/assets/20171026/485de3b40af34e948a54a81800f40823.pdf
- Philips, dS Small Extremity 16ch datasheet. https://www.documents.philips.com/assets/20171026/d7cff10b07d84b82beb4a81800f4084d.pdf
- Philips, dStream Flex S datasheet, 2015. https://www.documents.philips.com/assets/20170523/d406743263c3443d9a93a77c015acf69.pdf
- Philips, dStream Head 32ch coil product page. https://www.philips.co.uk/healthcare/product/HCNMRB375A/dstream-head-32ch-coil-mr-coil
- Philips, dStream HeadNeckSpine datasheet, 2018. https://www.documents.philips.com/assets/20180718/81f4030a6c85415ba3aba92100afd056.pdf
- Philips, dStream Knee 16 coil product page. https://www.philips.ch/healthcare/product/HCNMRB236A/dstream-knee-16-coil-mr-coil
- Philips, dStream Knee 16ch datasheet (1.5T), 2018. https://www.documents.philips.com/assets/20180911/ec8ff8cba0ab45ef8994a95800daae58.pdf
- Philips, dStream T/R Knee 16ch datasheet (3.0T), 2018. https://www.documents.philips.com/assets/20180718/138b83998e544b48b22ba92100aff66b.pdf
- Philips, dStream Torso coil product page. https://www.philips.co.uk/healthcare/product/HCNMRB101/dstream-torso-coil-mr-coil
- Philips, dStream WholeBody datasheet, 2018. https://www.documents.philips.com/assets/20180718/fb599167437340a7bf0da92100affaf4.pdf
- Philips, dStream Wrist 8ch datasheet, 2020. https://www.documents.philips.com/assets/20200827/378f34feb69e4914b8b1ac2400d35645.pdf
- Philips, FieldStrength, Issue 60, 2025. https://www.documents.philips.com/assets/20250918/70f69962d1354b36b542b35c00bd35a1.pdf
- Philips, VitalEye datasheet, 2019. https://www.documents.philips.com/assets/20180528/8426153593cd4a9aa1dda8ee00bc1aa6.pdf
- Royal Australian and New Zealand College of Radiologists. MRI Safety Guidelines, Version 3.0. Approved 30 September 2021 (current version; PDF re-posted June 2026 with identical text). Sydney: RANZCR. https://www.ranzcr.com/wp-content/uploads/2026/06/MRI-Safety-Guidelines-V3-3.pdf
- Siemens Healthineers Academy, BioMatrix Respiratory Sensors (MAGNETOM Free.Max, 2023). https://academy.siemens-healthineers.com/_/en-us/biomatrix-respiratory-sensors
- Siemens Healthineers Academy, Off-center imaging guide (syngo MR XA31A/XA70A examples), 2024. https://academy.siemens-healthineers.com/_/en-us/off-center-imaging-usa
- Siemens Healthineers, “MRI Acronyms: Cross-Vendor Comparisons” v2, Order No. MR-20-NAM-1350, 08.2020. https://marketing.webassets.siemens-healthineers.com/7c27289490b774c4/7494aa896362/siemens-healthineers_mri_mri-acronyms_cross-vendor-comparison_v2.pdf
- Siemens Healthineers, AutoAlign Head LS page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/clinical-applications/autoalign-head-ls
- Siemens Healthineers, BioMatrix Technology page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/technologies-and-innovations/biomatrix-technology
- Siemens Healthineers, Foot/Ankle 16 page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/foot-ankle-16
- Siemens Healthineers, Hand/Wrist 16 page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/hand-wrist-16
- Siemens Healthineers, Head 32 coil page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/head-32
- Siemens Healthineers, Knee Dot Engine page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/clinical-applications/knee-dot-engine
- Siemens Healthineers, MAGNETOM Avanto Knee T2 TSE FS SAG protocol printout, syngo MR B17. https://marketing.webassets.siemens-healthineers.com/1800000000015075/2a146f51b292/MAGNETOM_Avanto_Knee_T2_TSE_FS_SAG_1800000000015075.pdf
- Siemens Healthineers, MAGNETOM Flow. Platform Operator Manual, Breast BI 7, syngo MR XA70, Print No. MR-04014X.623.25.03.02. https://content.doclib.siemens-healthineers.com/rest/v1/view?document-id=1046455
- Siemens Healthineers, MAGNETOM Vida page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/3t-mri-scanner/magnetom-vida
- Siemens Healthineers, Shoulder Large 16 / Shoulder Small 16 page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/shoulder-16-coil-kit
- Siemens Healthineers, Spine Matrix Coil page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/spine-matrix-coil
- Siemens Healthineers, Tx/Rx Knee 18 page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/tx-rx-knee18
- Siemens Healthineers, UltraFlex Large 18 / UltraFlex Small 18 page. https://www.siemens-healthineers.com/en-us/magnetic-resonance-imaging/options-and-upgrades/coils/ultra-flex-large-small-18
- Simonovich A, Nagar Osherov A, Linov L, Kalichman L. The influence of knee bolster on lumbar spinal stenosis parameters on MR images. Skeletal Radiol 2020;49(2):299-305. doi:10.1007/s00256-019-03287-w
- Thomassin-Naggara I, Dolciami M, Chamie LP, Guerra A, Bharwani N, Freeman S, Rousset P, et al. ESUR consensus MRI for endometriosis: protocol, lexicon, and compartment-based analysis. Eur Radiol 2025;35(11):7272-7286. doi:10.1007/s00330-025-11611-3
- Tins BJ, Balain B. Incidence of numerical variants and transitional lumbosacral vertebrae on whole-spine MRI. Insights Imaging 2016;7(2):199-203. doi:10.1007/s13244-016-0468-7
- Unpublished departmental MRI scanning notes (Philips 1.5 T and 3 T; one Siemens protocol), 2022 to 2023, written by working technologists including the site’s founder.
- Zaitsev M, Maclaren J, Herbst M. Motion artifacts in MRI: a complex problem with many partial solutions. J Magn Reson Imaging 2015;42(4):887-901. doi:10.1002/jmri.24850