MR calculators
MR physics, contrast and renal, reporting and safety calculators for radiologists and MR technologists
Educational estimates only. Verify every value against product labelling, current guidelines, and your local protocol before clinical use. Physics calculators are idealised. Reference ranges are typical adult values, not diagnostic thresholds. Renal / contrast guidance follows the ACR Manual on Contrast Media (2026); safety thresholds follow the ACR Manual on MR Safety (2026).
Renal and contrast
Age, weight and creatinine entered once, then all three results
Shared inputs
eGFR
CKD-EPI 2021Race-free renal screen before a gadolinium-based contrast agent (GBCA). Both sexes shown.
Creatinine clearance
Cockcroft-GaultWeight-based renal estimate still used for some drug/contrast protocols. Both sexes shown.
Gadolinium dose
Pick the agent; its concentration and labelled dose auto-fill. Always confirm the current package insert.
Field strength
One B₀ drives both results
Shared input
In-phase / opposed-phase TE
Fat–water Dixon echo timing by field strength (fat taken as one peak 3.5 ppm from water; the main methylene peak sits about 3.4 ppm from water). Fixed to ¹H fat and water: the nucleus and shift below apply only to the Larmor result.
Larmor frequency & chemical shift
Resonant frequency of a nucleus and the Hz value of a ppm shift.
Tissue and timing
All three read a tissue T1. Leave TR blank in TI to null to assume TR much greater than T1. The panel TR fills Ernst, relative signal and scan time. Presets are typical 1.5 T values, and measured relaxation times vary with the method (27). White matter and cortical grey matter T1 are means of three in vivo studies (27,28,29), CSF T1 of two (29,30), and muscle T1 and T2 of two (31,32); muscle T1 differs by about a third between their methods. Fat T1 is one group’s measurement (31), also given in a review (26). Liver lies within the normal in vivo range at 1.5 T (35). Blood T1 was measured in vitro (36) and varies with haematocrit and oxygenation. CSF T2 is approximate.
Ernst angle
Flip angle that maximises spoiled-GRE signal for a given TR and T1.
Inversion time to null a tissue
STIR / FLAIR / LGETI that nulls a chosen tissue in an inversion-recovery sequence.
Relative signal intensity
≈1.5 TIdealised signal (0–1) for SE, IR, or spoiled GRE with a chosen tissue. Fat is not a preset here: its measured T2 depends strongly on the method (31,33,34), so enter a value for your sequence.
Protocol, time and SNR
Voxel volume feeds the SNR before field. Scan time takes one NEX and parallel R; SNR takes a before and an after pair.
Pixel / voxel size
In-plane resolution and voxel volume from FOV, matrix and slice thickness.
Scan time (2D)
Acquisition time from TR, phase-encode steps, averages, turbo factor and parallel imaging.
Relative SNR, before and after
How SNR changes when you alter voxel size, averages, bandwidth or parallel factor.
Other
MR arthrography dilution
Dilute GBCA to ≈2 mmol/L for intra-articular injection (≈1:250 of 0.5 M stock).
ADC from DWI
Monoexponential ADC from two b-values and their signal intensities.
Ellipsoid volume & PSA density
Prostate / lesion volume and PSA density (flagged >0.15).
Cardiac function & BSA
Stroke volume, ejection fraction, output and body-surface-indexed volumes.
Fetal weight from MR volumetry
Estimated weight from segmented fetal body volume (best >22 weeks).
Fringe field, distance to a target line
Far-field 1/r³ falloff anchored to one point from the magnet’s published fringe-field plot. Not a substitute for the vendor fringe-field map when siting a magnet.
Note on the controlled-access line
Historically the 5-gauss (0.5 mT) line defined the controlled-access boundary. IEC 60601-2-33:2022 revised the fringe-field limit for the controlled-access area to 9 gauss (0.9 mT), and the ACR Manual on MR Safety (2026) delimits the MR Environment by the 9 gauss line. The 5 G line remains a conservative reference for legacy devices and passers-by with cardiac implants. Always defer to local policy and the vendor’s certified fringe-field map. See 19,20,21
References
References (36)
36 sources listed
Primary sources for the formulas, clinical thresholds, and reference ranges above.
- Inker LA, Eneanya ND, Coresh J, et al. New creatinine- and cystatin C–based equations to estimate GFR without race. N Engl J Med. 2021;385(19):1737–1749. doi:10.1056/NEJMoa2102953
- Delgado C, Baweja M, Crews DC, et al. A unifying approach for GFR estimation: recommendations of the NKF-ASN Task Force on Reassessing the Inclusion of Race in Diagnosing Kidney Disease. Am J Kidney Dis. 2022;79(2):268–288. doi:10.1053/j.ajkd.2021.08.003
- Cockcroft DW, Gault MH. Prediction of creatinine clearance from serum creatinine. Nephron. 1976;16(1):31–41. doi:10.1159/000180580
- American College of Radiology Committee on Drugs and Contrast Media. ACR Manual on Contrast Media (2026). Reston, VA: American College of Radiology; 2026. acr.org
- American College of Radiology, SPR, SSR. ACR–SPR–SSR Practice Parameter for the Performance and Interpretation of Magnetic Resonance Imaging (MRI) Arthrography. Reston, VA: ACR; rev. 2021. acr.org
- Ernst RR, Anderson WA. Application of Fourier transform spectroscopy to magnetic resonance. Rev Sci Instrum. 1966;37(1):93–102. doi:10.1063/1.1719961
- Dixon WT. Simple proton spectroscopic imaging. Radiology. 1984;153(1):189–194. doi:10.1148/radiology.153.1.6089263
- Ridgway JP. Cardiovascular magnetic resonance physics for clinicians: part I. J Cardiovasc Magn Reson. 2010;12(1):71. doi:10.1186/1532-429X-12-71
- Biglands JD, Radjenovic A, Ridgway JP. Cardiovascular magnetic resonance physics for clinicians: Part II. J Cardiovasc Magn Reson. 2012;14(1):66. doi:10.1186/1532-429X-14-66
- de Bazelaire CMJ, Duhamel GD, Rofsky NM, Alsop DC. MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology. 2004;230(3):652–659. doi:10.1148/radiol.2303021331
- Le Bihan D, Breton E, Lallemand D, Aubin ML, Vignaud J, Laval-Jeantet M. Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging. Radiology. 1988;168(2):497–505. doi:10.1148/radiology.168.2.3393671
- Turkbey B, Rosenkrantz AB, Haider MA, et al. Prostate Imaging Reporting and Data System Version 2.1 (PI-RADS v2.1). Eur Urol. 2019;76(3):340–351. doi:10.1016/j.eururo.2019.02.033
- Nordström T, Akre O, Aly M, Grönberg H, Eklund M. Prostate-specific antigen (PSA) density in the diagnostic algorithm of prostate cancer. Prostate Cancer Prostatic Dis. 2018;21(1):57–63. doi:10.1038/s41391-017-0024-7
- Schulz-Menger J, Bluemke DA, Bremerich J, et al. Standardized image interpretation and post-processing in cardiovascular magnetic resonance — 2020 update. J Cardiovasc Magn Reson. 2020;22:19. doi:10.1186/s12968-020-00610-6
- Kawel-Boehm N, Hetzel SJ, Ambale-Venkatesh B, et al. Reference ranges (“normal values”) for cardiovascular MRI in adults and children: 2020 update. J Cardiovasc Magn Reson. 2020;22:87. doi:10.1186/s12968-020-00683-3
- Mosteller RD. Simplified calculation of body-surface area. N Engl J Med. 1987;317(17):1098. doi:10.1056/NEJM198710223171717
- Baker PN, Johnson IR, Gowland PA, et al. Fetal weight estimation by echo-planar magnetic resonance imaging. Lancet. 1994;343(8898):644–645. doi:10.1016/S0140-6736(94)92638-7
- Bouachba A, Massoud M, Nasr B, et al. Normative range of MRI-based fetal body volume and association with ultrasonographically estimated fetal weight at 16–36 weeks. Ultrasound Obstet Gynecol. 2025. doi:10.1002/uog.29234
- American College of Radiology Committee on MR Safety. ACR Manual on MR Safety (2026 edition). Reston, VA: American College of Radiology; 2026. acr.org
- International Electrotechnical Commission. IEC 60601-2-33:2022 Medical electrical equipment, Part 2-33: Particular requirements for MR equipment for medical diagnosis. Geneva: IEC; 2022. iec.ch
- American College of Radiology. ACR Manual on MR Safety: What’s New? Summary of changes from the 2024 manual to the 2026 edition (Chapter 3, Zone III: MR Environment delimited by the 9 gauss line). Reston, VA: American College of Radiology; 2026. acr.org
- Kidney Disease: Improving Global Outcomes (KDIGO) CKD Work Group. KDIGO 2024 Clinical Practice Guideline for the Evaluation and Management of Chronic Kidney Disease. Kidney Int. 2024;105(4S):S117–S314. doi:10.1016/j.kint.2023.10.018
- Sener RN. Diffusion MRI: apparent diffusion coefficient (ADC) values in the normal brain and a classification of brain disorders based on ADC values. Comput Med Imaging Graph. 2001;25(4):299–326. doi:10.1016/S0895-6111(00)00083-5
- Society for Cardiovascular Magnetic Resonance. SCMR reference values (“normal values”) in cardiovascular magnetic resonance: 2025 update. J Cardiovasc Magn Reson. 2025. journalofcmr.com
- Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:17. doi:10.1186/s12968-020-00607-1
- Del Grande F, Santini F, Herzka DA, Aro MR, Dean CW, Gold GE, Carrino JA. Fat-suppression techniques for 3-T MR imaging of the musculoskeletal system. RadioGraphics. 2014;34(1):217–233. doi:10.1148/rg.341135130
- Dieringer MA, Deimling M, Santoro D, et al. Rapid parametric mapping of the longitudinal relaxation time T1 using two-dimensional variable flip angle magnetic resonance imaging at 1.5 Tesla, 3 Tesla, and 7 Tesla. PLoS One. 2014;9(3):e91318. doi:10.1371/journal.pone.0091318
- Wright PJ, Mougin OE, Totman JJ, et al. Water proton T1 measurements in brain tissue at 7, 3, and 1.5 T using IR-EPI, IR-TSE, and MPRAGE: results and optimization. MAGMA. 2008;21:121–130. doi:10.1007/s10334-008-0104-8
- West J, Blystad I, Engström M, Warntjes JBM, Lundberg P. Application of quantitative MRI for brain tissue segmentation at 1.5 T and 3.0 T field strengths. PLoS One. 2013;8(9):e74795. doi:10.1371/journal.pone.0074795
- Rooney WD, Johnson G, Li X, et al. Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn Reson Med. 2007;57:308–318. doi:10.1002/mrm.21122
- Han E, Gold G, Stainsby J, Wright G, Beaulieu C, Brittain J. In-vivo T1 and T2 measurements of muskuloskeletal tissue at 3T and 1.5T. Proc Intl Soc Mag Reson Med. 2003;11:450. cds.ismrm.org
- Varghese J, Scandling D, Joshi R, et al. Rapid assessment of quantitative T1, T2 and T2* in lower extremity muscles in response to maximal treadmill exercise. NMR Biomed. 2015;28(8):998–1008. doi:10.1002/nbm.3332
- Jordan CD, Saranathan M, Bangerter NK, Hargreaves BA, Gold GE. Musculoskeletal MRI at 3.0 T and 7.0 T: a comparison of relaxation times and image contrast. Eur J Radiol. 2013;82:734–739. doi:10.1016/j.ejrad.2011.09.021
- Edden RAE, Smith SA, Barker PB. Longitudinal and multi-echo transverse relaxation times of normal breast tissue at 3 Tesla. J Magn Reson Imaging. 2010;32(4):982–987. doi:10.1002/jmri.22306
- Meloni A, Carnevale A, Gaio P, et al. Liver T1 and T2 mapping in a large cohort of healthy subjects: normal ranges and correlation with age and sex. MAGMA. 2024;37(1):93–100. doi:10.1007/s10334-023-01135-6
- Stanisz GJ, Odrobina EE, Pun J, et al. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med. 2005;54(3):507–512. doi:10.1002/mrm.20605
MR Calculators · High Yield Imaging · Educational estimates. Verify against labelling, the ACR Manuals, and local protocol before clinical use.
More in Calculators
Next in Calculators: CT and radiotherapy calculators
Everything on Tools: Tools hub. Or search the whole reference.
Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.