Ultrasound QA toolkit

Ultrasound QA toolkit

Ultrasound systems drift and transducers degrade (dead elements, lens cracks, delamination) long before the failure is obvious on a clinical image. A structured quality-assurance (QA) programme catches that decline, protects diagnostic accuracy and patient safety, and records of equipment testing or maintenance are required, each in its own form, for ACR, AIUM and IAC accreditation (see Accreditation programmes). This toolkit sets out why QA matters, the routine test set with methods, frequencies and tolerances, the phantoms and testers you need, the accreditation programmes compared, and the governance/records that tie it together, with printable log and checklist blocks.

On this pageWhy QARoutine QA testsPhantoms & testersAccreditationGovernance & recordsSources

Test methods, frequencies and tolerances summarised from AAPM ultrasound QC guidance, ACR Ultrasound Accreditation and QC requirements, and IAC. Tolerances and frequencies vary by programme, equipment and jurisdiction. Always follow your medical-physics service, the accrediting body’s current requirements and the manufacturer’s instructions for use.

Key points
  • Inspect each transducer’s lens, housing, cable and connector at each use or daily; any crack, bulge, exposed conductor or fluid ingress means removal from service, because a breached lens risks shock or burns and cannot be reliably disinfected.
  • Acceptance testing at installation sets a baseline reference image for every transducer; routine QA is about detecting change from that baseline, so re-baseline after any repair or probe replacement.
  • Check for dead elements by scanning in air (or against a flat reflector) and inspecting the reverberation pattern for vertical drop-out lines; investigate any persistent drop-out band.
  • Give every trended parameter an action limit and an owner. In one three-tier scheme, drift near tolerance means repeat and investigate; an out-of-tolerance result, cracked lens, electrical safety failure or dead-element cluster beyond local limits means remove from use, log and escalate.
  • A test that is not recorded, trended and acted on is not QA: keep an equipment inventory with baselines, dated QA logs, a fault and action log, decontamination and traceability records, and audit minutes.
Explain it simply
  • Check each probe’s surface, body, cable and plug at each use or daily. A crack, bulge, bare wire or fluid inside means removing it from use: it could shock or burn and can’t be properly disinfected.
  • When equipment is installed, test it and save reference images for every probe. Later checks look for changes from those images, so make new reference images after any repair or probe swap.
  • To find failed parts of a probe, scan with nothing touching it and look for dark vertical stripes in the pattern on screen. Look into any stripe that keeps showing up.
  • Each check needs a limit and an owner. For example: near the limit, retest and investigate; a failed result, cracked lens, electrical fault or too many dead elements, remove it from use, log and report it.
  • A check that isn’t written down, tracked over time and acted on doesn’t count. Keep an equipment list with reference images, dated check records, a fault log, cleaning and tracing records, and review-meeting notes.

A guide only: follow your departmental protocols and regional requirements.

Related

Why QA matters

Ultrasound is often thought of as a “safe, self-calibrating” modality because the operator sees the image live, but a failing probe is often not obvious. Transducers and systems develop faults in everyday use: when 151 transducers in clinical use at one large centre were tested, about one in six failed an electronic transducer test and a quarter had a physical flaw. QA exists to find these faults objectively, before they affect a patient’s scan.1,2

How transducers and systems fail

Dead / weak elements

Individual piezoelectric elements fail electrically. A few dead elements produce a vertical drop-out band in the image and degrade lateral resolution and sensitivity along that line. Element loss is a recognised transducer fault and a reason to retire a probe; in two surveys of probes in clinical use, lens delamination and cable breaks were the commonest faults and element failures were unusual.20,22

Lens cracks & wear

The acoustic lens/matching layer is soft and cracks, delaminates or abrades with handling and cleaning. A cracked lens causes artifact and sensitivity loss, and is also an infection-control and electrical-safety breach (fluid ingress to a patient-contact surface).

Delamination

Separation of the lens, matching or backing layers from the element stack. Produces reverberation artifact, dead zones and reduced penetration. Often invisible externally but obvious on a uniformity phantom.

Cable & connector faults

Flexing fractures conductors; bent connector pins drop channels. Intermittent drop-out, flicker or whole-sector loss. A frequent cause of “the machine is broken” call-outs that trace to a cable, not the system.

System / monitor drift

Beam-former, gain-calibration and above all display drift shift grey-scale rendering over time. A miscalibrated or ambient-light-swamped monitor makes correctly-acquired data look wrong.

Image-quality drift

The net result: reduced depth of penetration, degraded resolution, poorer low-contrast (cyst) detectability and non-uniformity, each of which lowers diagnostic confidence and raises the chance of a missed or mischaracterised lesion.

A cracked lens is not just an image problem A cracked or delaminated transducer lens breaches the sealed patient-contact surface. It allows fluid ingress that can cause electrical leakage/shock and burn hazard, and it can no longer be reliably disinfected (an infection-control failure). Any transducer with visible lens damage should be removed from service pending assessment, regardless of how the image looks.2
The three drivers of a QA programme
DriverWhat it protectsWhat QA delivers
Diagnostic qualityAccurate diagnosis: penetration, resolution, contrast and measurement accuracyObjective, trended evidence that image quality is stable and within tolerance; early detection of drift
Patient safetyFreedom from electrical, burn and cross-infection hazardPhysical inspection catches cracked lenses, frayed cables and worn housings before they harm a patient
Accreditation & medico-legalACR / AIUM / IAC status, funding eligibility, defensibilityEquipment QC records are required for accreditation (ACR: an annual survey summary for each unit with the application; AIUM: the most recent QA report for each machine; IAC: the method and frequency of equipment maintenance recorded) and are a traceable record for governance and litigation2,10,11
Acceptance → routine → the “reference” baseline QA is a lifecycle. Acceptance testing at installation confirms the system meets specification and establishes a baseline reference image for every transducer. Routine QA thereafter compares against that baseline. QA is fundamentally about detecting change from baseline, so a system with no acceptance baseline has nothing to trend against. Re-baseline after any repair or probe replacement.
Who does QA?

A workable programme is layered: daily/each-use visual and physical checks by the sonographer or radiographer at the point of care; periodic routine QC tests, which the ACR says are typically done by trained sonographers or equipment service engineers; and an annual survey of every unit and transducer in clinical use by a qualified medical physicist or designee, which ACR accreditation requires. UK guidance from BMUS recommends daily, weekly and monthly sonographer checks within a programme that includes more comprehensive testing at longer intervals, which may be provided by medical physics. The exact split and cadence is set locally and by the accrediting body (Routine QA tests and Accreditation programmes).1,2,23

Back to contents

Routine QA tests

The routine test set breaks into two layers: frequent, tool-free checks anyone can do (visual/physical inspection, in-air reverberation for element drop-out), and periodic phantom-based measurements that quantify image quality against baseline. The table below is the working reference: method, who, how often and the action tolerance. Treat the figures as representative: exact tolerances and frequencies are set by your medical-physics service and accrediting body.1,2,3

Show the full section

In a tissue-mimicking QA phantom, embedded targets test distance accuracy (vertical & horizontal pin arrays), axial and lateral resolution (closely-spaced pin pairs), grey-scale / dynamic range (stepped-echogenicity targets) and anechoic-object detection (cyst targets at increasing depth, checked for edge sharpness and absence of fill-in). Depth of penetration is read as the deepest visible speckle. Target geometry varies by phantom model.

QA test reference table

TestMethodFrequency / whoTolerance / action
Physical / visual inspection Inspect transducer face, lens, housing, cable and connector, and system controls/wheels/locks. Look for lens cracks, delamination, bulges, frayed cable, bent pins. BMUS (recommended): transducers daily, cables, controls and system weekly · sonographer. ACR: in the required annual survey (medical physicist or designee) and in optional routine QC, at least semiannually2,23 Any crack, bulge, exposed conductor or fluid ingress → remove from service. Zero tolerance for lens/housing breach.
Transducer element testing Scan in air (or against a flat reflector) and inspect the reverberation pattern for vertical drop-out lines; or use a dedicated element tester (e.g. FirstCall). BMUS (recommended): daily look at reverberation images for drop-out; element drop-out test if a fault is seen, and monthly for phased arrays · sonographer. ACR: element testers optional, not a replacement for required tests2,23 Investigate any persistent drop-out band. Element-loss limits are set locally; a cluster of adjacent dead elements typically fails.
Image uniformity Scan a uniform region of the tissue-mimicking phantom; assess for horizontal/vertical shadows, dead zones or bright/dark bands across the field. ACR: required in the annual survey (medical physicist or designee); optional routine QC at least semiannually (typically trained sonographers or service engineers). BMUS (recommended): daily in-air reverberation check, not a phantom test2,23 No new non-uniformity vs baseline. New dark bands suggest dead elements/delamination.
Depth of penetration (max depth of visualisation) At maximum output and focus set deep, with gain raised, read the deepest depth at which speckle/echoes are still visible in the phantom. ACR: system sensitivity (maximum depth of visualisation) required in the annual survey (medical physicist or designee). BMUS (recommended): monthly air-reverberation sensitivity check by sonographers, not a phantom test2,23 No fall from baseline beyond the limit your programme sets for that probe and setting. A fall in penetration is a sensitive early sign of probe/system decline.
Axial resolution Image closely-spaced resolution pins aligned along the beam axis; find the smallest pin separation resolvable in depth. Set locally. ACR: spatial resolution is optional in the annual survey (medical physicist or designee)2 No degradation vs baseline (better axial resolution at higher frequency is expected).
Lateral resolution Image pins perpendicular to the beam at set depths; find the smallest separation resolvable side-to-side (worsens with depth and away from focus). Set locally. ACR: spatial resolution is optional in the annual survey (medical physicist or designee)2 No degradation vs baseline at matched depth/focus.
Distance / calliper accuracy Measure known pin separations vertically (axial/depth) and horizontally (lateral) with the on-screen callipers. Set locally. ACR: geometric accuracy is optional in the annual survey (medical physicist or designee), and in routine QC only for mechanically scanned transducers2 Within your programme’s limits, which differ between programmes: for example, the AAPM TG-1 defect levels are 2 mm or 2% vertically (axial) and 3 mm or 3% horizontally (lateral), and some local programmes use 5%.
Anechoic-target / cystic detection Image the anechoic (cyst-mimicking) targets at increasing depth; assess edge sharpness, correct anechoic centre and absence of fill-in. Set locally. ACR: contrast resolution (anechoic and low-contrast targets) is optional in the annual survey (medical physicist or designee)2 Smallest cyst resolved and fill-in ≥ baseline. New fill-in or missing small cysts indicates loss of low-contrast/resolution performance.
Grey-scale / dynamic range & low-contrast Image stepped-echogenicity (contrast) targets; confirm the expected number of grey levels/steps is distinguishable and low-contrast targets are seen. Set locally. ACR: contrast resolution is optional in the annual survey (medical physicist or designee)2 Same number of grey-scale steps and low-contrast targets visible as at baseline.
Image uniformity & artifacts (full-field) Full-field review for reverberation, comet-tail, drop-out and dead-zone artifacts across the image. ACR: required in the annual survey (medical physicist or designee); optional routine QC at least semiannually (typically trained sonographers or service engineers)2 No new artifacts vs baseline reference image.
Monitor / display QC Assess viewing-monitor grey-scale rendering, contrast, brightness and ambient light (SMPTE / greyscale pattern); relevant to reporting workstations too. ACR: scanner display required in the annual survey (medical physicist or designee) and in optional routine QC at least semiannually. BMUS (recommended): daily brightness, contrast and grey-bar checks by sonographers2,23 All grey-scale steps distinguishable; display within calibration; control ambient light.
Doppler performance (where used) On a Doppler flow phantom: velocity accuracy, sample-volume/gate position, colour box registration, sensitivity/penetration and flow direction. Set locally · IAC (vascular, echo): equipment kept in good operating condition, with the method and frequency of maintenance recorded (Vascular Testing 1.2.2B; Adult Echocardiography 2.3A)11 Velocity error within programme tolerance; correct colour registration and direction.
Electrical safety Leakage-current and protective-earth testing of the system and transducers to IEC 62353 (recurrent testing of medical electrical equipment), by your biomedical engineering service. Interval set by your biomedical engineering service or local policy · biomedical engineer. BMUS (recommended): sonographers check monthly that testing is not due or overdue23 Within the IEC 62353 limits your biomedical engineering service applies; fail → out of service.
Tolerances and frequencies are representative, not universal Distance-accuracy and penetration tolerances differ between programmes (for example, AAPM TG-1 uses 2 mm or 2% vertically and 3 mm or 3% horizontally for distance accuracy), and the exact figures, and which tests are mandatory at what cadence, are set by your accrediting body (ACR/IAC) and your medical-physics service. Do not adopt these numbers in place of your local protocol.2,3
The minimum viable programme (ACR-aligned) Where a full physics programme is not in place, the practical floor is: (1) daily physical inspection of every transducer in use23; (2) regular in-air element check for drop-out; (3) periodic phantom image-quality review vs baseline; and (4) monitor QC, all documented and trended, with a defined escalation for failures. ACR lists routine QC as optional: it is strongly recommended and should be done at least semiannually, but is no longer required for accreditation submission. The annual survey of every unit and transducer in routine clinical use by a qualified medical physicist or designee is required: a signed report must be provided to the physician directing the practice, the original report must be kept by the facility, and a summary is submitted with the accreditation application.2

Test equipment & phantoms

QA measurements are only as good as the test object behind them. Ultrasound QA uses tissue-mimicking phantoms (grey-scale/B-mode), Doppler flow phantoms and dedicated element testers. Each answers a different question.6,7

Tissue-mimicking (grey-scale) phantoms

A tissue-mimicking phantom is a block of material whose speed of sound (~1540 m/s) and attenuation mimic soft tissue, with embedded targets. It is the workhorse of B-mode QA. Major vendors include CIRS, Gammex/Sun Nuclear and ATS (CIRS).

Embedded targetWhat it measures
Background tissue-mimicking gelDepth of penetration (max depth of visualisation); overall sensitivity and uniformity
Vertical pin groupAxial (depth) distance/calliper accuracy
Horizontal pin groupLateral (horizontal) distance/calliper accuracy
Closely-spaced pin pairsAxial and lateral spatial resolution
Anechoic (cystic) targets, varied size/depthCyst detectability, edge sharpness, fill-in (low-contrast performance)
Stepped grey-scale / contrast targetsGrey-scale, dynamic range and low-contrast object detectability
Phantoms need care too Tissue-mimicking gel phantoms can desiccate, develop air bubbles or grow voids if allowed to dry, freeze or overheat, any of which corrupts results. Keep the scanning surface sealed/hydrated per the manufacturer, store within the specified temperature range, and inspect before use. A degraded phantom will manufacture false “failures”.
Doppler flow phantoms

Grey-scale phantoms cannot assess flow. A Doppler flow phantom pumps a blood-mimicking fluid through a vessel at a known velocity within tissue-mimicking surround, letting you verify:

  • Velocity accuracy: does the measured velocity match the set flow?
  • Sample-volume / gate placement and spectral display
  • Colour-flow box registration, sensitivity and correct flow direction
  • Doppler penetration: deepest depth flow is still detected

Velocity measurements drive diagnosis in vascular testing: IAC-accredited vascular services that perform extracranial cerebrovascular (carotid) testing must have internal carotid stenosis criteria that state how velocity measurements are used (IAC Vascular Testing 3.5.2.1A).11

Element / transducer testers

Dedicated tools drive each element and read its response to detect dead or weak elements more objectively than the in-air method. One example is FirstCall (Sonora/Unisyn), which produces a per-element report. Handheld/impedance testers also flag element and cable faults. These are especially valuable for a fleet, where systematic element mapping catches decline early and supports repair-vs-replace decisions.

Tissue-mimicking phantom

B-MODEPenetration, resolution, distance accuracy, cyst detection, grey-scale/uniformity. The core QA test object.

Doppler flow phantom

DOPPLERVelocity accuracy, colour registration, flow sensitivity/penetration. Essential for vascular/echo QA.

Element tester (FirstCall)

TRANSDUCERPer-element drive/response mapping for objective dead-element detection and fleet surveillance.

Display test pattern

MONITORSMPTE / grey-scale pattern for viewing-monitor and reporting-workstation QC.

Right tool, right question Match the test object to the failure you are hunting: a tissue-mimicking phantom finds image-quality and geometric drift; a Doppler phantom finds flow/velocity faults; an element tester localises the exact channels that have died; and a display pattern rules out “the data is fine, the monitor is wrong”. Owning only a grey-scale phantom leaves Doppler and element faults unmonitored.
Back to contents

Accreditation programmes

In many jurisdictions QA is required for accreditation, and accreditation is often required for reimbursement. The main programmes overlap heavily on principle (documented QC, qualified personnel, image-quality review) but differ in scope and detail. ACR, AIUM and IAC accreditation each run on a 3-year cycle; the UK is organised differently (see the UK row).9,10,11

Show the full section
Programme Scope Clinical image-quality review Personnel & CME Equipment QC & documentation
ACR Ultrasound (US)9 General, OB, gynae, vascular, paediatric modules. Apply for all modules routinely performed at the site. Submit clinical images for each module sought; every submission must meet the ACR Exam Requirements. For clinical image testing, use of a phantom test object is optional at this time.4 Physician/interpreting-provider qualifications + sonographer credentialing; ongoing CME. Annual survey required: every unit and transducer in routine clinical use, at least annually, by a qualified medical physicist or designee (up to 14 months between surveys); a summary for each unit, and corrective action for any failed test, go with the application. Routine QC is optional: strongly recommended, at least semiannually, typically by trained sonographers or equipment service engineers, and not submitted for accreditation.2
AIUM Practice Accreditation (US)10 Accreditation of a practice (one or more sites following the same policies, protocols and diagnostic criteria) in the specialties it applies for. Case studies for each specialty sought, meeting AIUM’s case submission requirements for that specialty, which are based on the relevant AIUM Practice Parameter, and scored by two reviewers (a third if their scores disagree); final reports must meet the AIUM Practice Parameter for Documentation. Designated Director of Ultrasound (licensed medical provider meeting AIUM training guidelines); sonographers certified (ARDMS or ARRT) in each specialty they perform, with each additional specialty registered within one accreditation cycle (3 years); at least one registered sonographer for each specialty performed. Equipment kept in optimal operating condition, with documented QA testing at least annually, or more often if issues arise, meeting or exceeding AIUM’s Routine Quality Assurance of Clinical Ultrasound Equipment (Version 2.0). The most recent QA report for each machine goes with the application; QA records are kept for 3 years or longer.
IAC: Vascular Testing (US)11 Vascular-lab accreditation (carotid, venous, arterial, etc.). Case studies submitted for peer review; strong emphasis on measurement/velocity accuracy and reporting. Medical director + technical staff; sonographer credentialing requirement; CME. Written QI programme (1.1C); equipment kept in good operating condition, with the method and frequency of maintenance recorded (1.2.2B); its guideline says the schedule should be frequent enough for accurate data (p. 25).
IAC: Echocardiography (US)11 Adult/paediatric TTE, TEE, stress echo. Case review of studies and reports against IAC echo Standards. Medical director, sonographer credentialing, CME. Written QI programme (1.1C); instrumentation kept in good operating condition, with the method and frequency of maintenance recorded (2.3A, 2.3.1A).
UK / Europe (IPEM, national)5 Governed through clinical-governance/quality frameworks rather than a single US-style accreditation body; IPEM’s QA report (Report 102, 2010) is withdrawn for review; BMUS (2014) and EFSUMB give technical QA guidance;23 since July 2024 the Royal College of Radiologists and the College of Radiographers run the QSI Quality Mark, which members of the QSI Hub can apply for when they are confident they meet all the applicable standards; its review runs on a 3-year cycle.21 Local governance + peer review; national imaging-quality frameworks. Professional registration (e.g. HCPC), recognised training, CPD. Sonographer checks to BMUS guidance (daily and weekly checks, monthly tests with a simple test tool) within a programme that includes more comprehensive testing at longer intervals, which may be provided by medical physics; documented programme and records.23
The common thread Whatever the badge, an accreditation reviewer looks for the same four things: (1) qualified, credentialed, CME-current personnel; (2) clinical images/reports that meet the programme’s quality criteria; (3) a documented equipment-QC programme with records; and (4) a functioning QA/QI process that closes the loop on problems. Build the QA programme in Routine QA tests and Governance & records and you are building for all of them at once.
Confirm current requirements before you rely on them Accreditation requirements change: for example, ACR no longer requires semiannual QC for accreditation submission, and IAC republishes its Standards regularly (the current IAC Vascular Testing Standards were published on 3 August 2026). Always work from the accrediting body’s current published Standards and requirements, not a summary.4,11

Governance & records

A test that is not recorded, trended and acted on is not QA. Governance is what turns individual measurements into a defensible, closed-loop programme: logs, action limits, escalation, decontamination records and audit. This is exactly what an accreditation visit inspects.2,9

Action limits & escalation

Every trended parameter needs a defined action limit and an owner. A simple three-tier scheme works:

LevelTriggerAction
PASSWithin tolerance of baselineLog result, continue in service, watch the trend.
INVESTIGATEDrift approaching tolerance, or a new minor artifactRepeat test, check technique/phantom, log, notify QA lead, schedule physics review.
FAIL / STOPOut of tolerance; cracked lens; failed electrical safety; dead-element clusterRemove transducer/system from clinical use, tag out, log the fault, escalate to physics/engineering, re-baseline after repair.
Cross-link: infection control is part of QA governance Physical inspection overlaps directly with reprocessing: a cracked or delaminated probe cannot be reliably disinfected and must leave service. Your QA records and your decontamination/traceability records should reference each other. See the reprocessing and Spaulding-classification detail on the Ultrasound Safety & Bioeffects page (Infection control & governance section).
Records to keep
  • Equipment inventory: every system and transducer with serial number and acceptance-test baseline reference image.
  • QA test logs: dated results per test per transducer, with the phantom/tester used and the operator.
  • Fault & action log: failures, investigations, repairs, and re-baseline dates; tag-out records.
  • Decontamination / reprocessing records: HLD cycles and probe-to-patient traceability (medico-legal for endocavity/interventional probes).
  • Training & competency: who is authorised to perform which QA task; credentialing and CME/CPD.
  • Audit & review minutes: periodic QA/QI meeting that reviews trends, incidents and accreditation status.
Printable QA log & checklists
Daily transducer & system inspection Printable

Site: ______________ System: ______________ Serial: ______________ Week commencing: ____________

  • Transducer face & lens intact (no cracks, bulges or delamination)
  • Housing intact; no exposed conductor or fluid ingress
  • Cable & strain relief undamaged; connector pins straight
  • In-air reverberation check: no vertical drop-out band
  • Controls, wheels, brakes & monitor arm functional
  • Viewing monitor clean; grey-scale steps distinguishable; ambient light controlled
  • Probe reprocessing status confirmed before first patient

Result: PASS / INVESTIGATE / FAIL · Signed: ______________ Date: __________

Periodic phantom QA log Printable
TestBaselineResultTolerance (example; use your programme’s limits)Pass / Fail
Depth of penetration (cm)within your programme’s limit
Axial distance accuracy<3% or <3 mm
Lateral distance accuracy<3% or <3 mm
Axial resolution= baseline
Lateral resolution= baseline
Smallest anechoic target seen≥ baseline
Grey-scale steps visible= baseline
Uniformity / artifactsno new artifact

Transducer: __________ Phantom model: __________ Performed by: __________ Date: __________ Physics reviewed: __________

Governance in one line A defensible ultrasound QA service needs four written pillars: an equipment inventory with baselines, a logged routine-QA programme with action limits and escalation, decontamination/traceability records that cross-reference the physical-inspection log, and a periodic audit/QI review.
Back to contents

Sources

Sources (23)

23 sources listed

  1. Sipilä O, Mannila V, Vartiainen E. Quality assurance in diagnostic ultrasound. Eur J Radiol 2011;80(2):519–525. (Transducer element tests, phantom measurements and visual checks found partly different faults; all three appeared necessary.)
  2. American College of Radiology. Quality Control: Ultrasound and Breast Ultrasound (revised 2025). ACR Accreditation Support. (Transducer inspection, QC requirements.)
  3. Mannila V, Sipilä O. Phantom-based quality assurance measurements in B-mode ultrasound. Acta Radiol Short Rep 2013;2(8):2047981613511967. (Penetration, resolution, distance accuracy; applies the AAPM TG-1 distance defect levels, 2 mm or 2% vertical and 3 mm or 3% horizontal.)
  4. American College of Radiology. Complete Accreditation Information: Ultrasound (revised 4-9-2025), with Clinical Image Testing: Ultrasound (revised 4-17-2026). ACR Accreditation Support. Read 22 November 2026.
  5. Institute of Physics and Engineering in Medicine. IPEM Report 102: Quality Assurance of Ultrasound Imaging Systems. IPEM; 2010, withdrawn for review (IPEM resources list, read 21 November 2026). (UK technical QA guidance.)
  6. American Association of Physicists in Medicine. Real-Time B-Mode Ultrasound Quality Control Test Procedures (AAPM Report No. 65 / TG-1). AAPM. (Phantom-based QC test procedures.)
  7. American Association of Physicists in Medicine. Quality Assurance of Ultrasound Imagers: Procedures, Expectations and Philosophies. AAPM educational course. (Phantoms, testers, expectations.)
  8. American Association of Physicists in Medicine. AAPM Task Group 128: Quality assurance tests for prostate brachytherapy ultrasound systems. AAPM. (B-mode QA of transrectal ultrasound for prostate brachytherapy: grey scale, resolution, distance, area and volume accuracy, template alignment; no Doppler tests.)
  9. American College of Radiology. ACR Ultrasound Accreditation. ACR. (Modules, clinical image submission, QC documentation. The ACR accreditation process flowchart gives a 3-year certificate.)
  10. American Institute of Ultrasound in Medicine. Standards and Guidelines for the Accreditation of Ultrasound Practices. AIUM; reapproved 20 April 2026; with the AIUM Accreditation Application Manual (updated 19 November 2025), Accreditation FAQs and What is AIUM Accreditation? Read 22 November 2026. (Director of Ultrasound, sonographer certification, case studies, equipment QA at least annually, reaccreditation every 3 years.)
  11. Intersocietal Accreditation Commission. IAC Standards & Guidelines for Vascular Testing Accreditation (published 3 August 2026); IAC Standards & Guidelines for Adult Echocardiography Accreditation (published 1 April 2025); and IAC Standards & Guidelines for Pediatric and Congenital Echocardiography Accreditation (published 1 April 2025). IAC. (Standards: Technical Director and technical staff credentials, written QI programme (1.1C), vascular case-study submission (Section 3A guidelines, p. 18). IAC Vascular Testing and Echocardiography programme pages, read 22 November 2026: facilities first conduct a self-evaluation by reviewing the IAC Standards, then apply with case studies; application fees cover each three-year accreditation cycle.)
  12. Nyhsen CM, Humphreys H, Koerner RJ, et al. Infection prevention and control in ultrasound — best practice recommendations (ESR Ultrasound Working Group). Insights Imaging 2017;8:523–535. (Probe damage and reprocessing cross-link.)
  13. Goodsitt MM, Carson PL, Witt S, et al. Real-time B-mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1. Med Phys 1998;25(8):1385–1406.
  14. Sun Nuclear / Gammex. Ultrasound tissue-mimicking QA phantoms: product & test-object documentation. (Grey-scale phantom targets.)
  15. CIRS (Computerized Imaging Reference Systems). Multipurpose and Doppler flow ultrasound phantoms. CIRS. (Tissue-mimicking and Doppler flow test objects.)
  16. Sonora Medical / Unisyn. FirstCall transducer testing: per-element performance testing. (Dead-element detection and fleet surveillance.)
  17. International Electrotechnical Commission. IEC 60601-1 / 60601-2-37: medical electrical equipment safety, including diagnostic ultrasound. IEC. (Type-test safety standards; in-service testing follows IEC 62353.)
  18. Tsapaki V, Tsalafoutas IA, Triantopoulou SS, Triantopoulou C. Development and implementation of a quality control protocol for B-mode ultrasound equipment. J Ultrasound 2022;25(2):155–165. (Practical QC protocol, tolerances, in-service testing.)
  19. International Electrotechnical Commission. IEC 62353:2014 Medical electrical equipment — Recurrent test and test after repair of medical electrical equipment. (In-service electrical safety testing.)
  20. Mårtensson M, Olsson M, Segall B, et al. High incidence of defective ultrasound transducers in use in routine clinical practice. Eur J Echocardiogr 2009;10(3):389–394. (Transducer fault survey.)
  21. Royal College of Radiologists and College of Radiographers. Quality Standard for Imaging (QSI). (UK Quality Standard for Imaging. QSI Quality Mark Successes: scheme established jointly by the College of Radiographers and the Royal College of Radiologists in July 2024. The QSI Quality Mark: the review runs on a three-year cycle. Read 22 November 2026.)
  22. Mårtensson M, Olsson M, Brodin LÅ. Ultrasound transducer function: annual testing is not sufficient. Eur J Echocardiogr 2010;11(9):801–805. (299 transducers in 13 Stockholm clinics; 27.1% defective a year after passing a test.)
  23. Dudley N, Russell S, Ward B, Hoskins P; BMUS QA Working Party. BMUS guidelines for the regular quality assurance testing of ultrasound scanners by sonographers. Ultrasound 2014;22(1):8–14. (Three levels of sonographer QA, daily, weekly and monthly, within a programme that includes more comprehensive testing at longer intervals.)

This page is an educational summary for imaging professionals and students. It is not clinical, regulatory or accreditation advice and does not replace your medical-physics service, the accrediting body’s current published Standards, employer’s procedures, local infection-control policy, or the equipment manufacturer’s instructions for use. QA test lists, tolerances and frequencies, and accreditation requirements, are periodically revised and vary by programme, equipment and jurisdiction. Always confirm against the current primary source before relying on a value in practice. © High Yield Imaging.

More in Foundations and safety

Next in Foundations and safety: Ultrasound glossary

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.

Contents