A small-animal (dog & cat) companion to our human MRI glossary: plain definitions of the physics, sequences, positioning, artifacts, contrast, safety and clinical terms you meet in canine and feline MRI.

On this pagePhysics & basicsSequencesAcquisitionPositioning & anaesthesiaArtifactsContrastSafetyClinical & indications

Physics & basics

T1 (longitudinal relaxation)

The time constant for protons to realign with the main magnet after a pulse. T1-weighted images show fat bright and fluid dark, giving crisp anatomy.

Why: T1-weighted sequences are the anatomical backbone of a canine brain study and the pre-contrast baseline for gadolinium enhancement.

Example: On a dog head MRI, subcutaneous fat and the marrow of the calvaria appear bright on T1; cerebrospinal fluid in the ventricles is dark.

T2 (transverse relaxation)

The time constant for loss of signal in the transverse plane. On T2-weighted images fluid and most pathology are bright.

Why: Most disease in small animals (oedema, inflammation, cysts, tumour) is conspicuous because it holds water and lights up on T2.

Example: A cat with a hippocampal lesion shows bright T2 signal; spinal cord oedema above a disc extrusion is bright on sagittal T2.

T2* (T2-star)

Signal decay that also includes local magnetic field inhomogeneity. T2*-weighted gradient-echo images are very sensitive to blood products and mineral.

Why: Old or acute haemorrhage and mineralisation cause signal “blooming” that is easily missed on standard sequences.

Example: Microhaemorrhages in a dog with suspected vascular event show dark blooming foci on a T2* GRE brain sequence.

TR (repetition time)

The interval between successive excitation pulses. A long TR reduces T1 weighting; a short TR increases it.

Why: TR is one of the main dials that sets image contrast. Multiplied across slices, it drives how long an anaesthetised patient must stay still.

Example: A T2-weighted feline spine acquisition uses a long TR to let tissues recover fully between pulses.

TE (echo time)

The time from excitation to when the echo (signal) is read. A long TE increases T2 weighting.

Why: Choosing TE tunes whether fluid appears bright or dark, which changes how easily you see a lesion.

Example: A long-TE T2 sagittal of a Dachshund spine makes cerebrospinal fluid bright, outlining the cord and any disc-associated compression.

Flip angle

How far the radiofrequency pulse tips the magnetisation away from alignment, measured in degrees. Larger angles give more T1 weighting in gradient-echo imaging.

Why: Flip angle balances signal against scan time and contrast, which is important on the lower-field magnets common in veterinary practice.

Example: A 3D T1 gradient-echo of a dog brain uses a small flip angle to preserve tissue contrast at speed.

SNR (signal-to-noise ratio)

The strength of true tissue signal relative to background noise. Higher SNR means cleaner, more diagnostic images.

Why: Small patients, small anatomy and lower-field magnets all challenge SNR, so vet protocols carefully trade resolution against averaging.

Example: Imaging a cat’s tiny pituitary may need extra signal averages, and thus a longer scan, to lift SNR to a readable level.

Sequences & acronyms

SE / TSE / FSE (spin echo & turbo/fast spin echo)

Workhorse sequences that use a refocusing pulse. TSE/FSE collect several echoes per excitation, so they are much faster than plain spin echo.

Why: Faster acquisition means shorter anaesthesia and fewer motion artifacts, a real welfare advantage in animals.

Example: T2 TSE is the standard fluid-sensitive sequence for a canine brain or spine survey.

GRE (gradient echo)

A sequence that refocuses signal with gradients rather than a 180° pulse. Fast, and sensitive to susceptibility effects.

Why: GRE underpins fast 3D anatomy and, in its T2* form, is the go-to for detecting haemorrhage and mineral.

Example: A GRE sequence flags a small bleed within a suspected canine glioma that a T2 TSE alone might obscure.

3D T1 (MPRAGE / BRAVO)

A high-resolution, isotropic 3D T1 gradient-echo acquisition (vendor names MPRAGE, BRAVO, etc.) that can be reformatted in any plane.

Why: Thin, gap-free slices improve detection of small lesions and support surgical or radiotherapy planning in neuro-oncology cases.

Example: A post-contrast 3D T1 of a dog with a meningioma is reformatted sagittally and dorsally to map the tumour margins.

FLAIR (fluid-attenuated inversion recovery)

A T2-based sequence that nulls (darkens) free cerebrospinal fluid, so lesions next to fluid stand out.

Why: Distinguishing bright pathology from bright CSF is essential in the brain, where lesions often sit near the ventricles or sulci.

Example: Periventricular inflammation in a dog with meningoencephalitis stays bright on FLAIR while the ventricular CSF turns dark.

STIR (short tau inversion recovery)

An inversion-recovery sequence that suppresses fat, making oedema and marrow lesions conspicuous. Robust on lower-field and less-uniform magnets.

Why: Reliable fat suppression is invaluable for musculoskeletal and spinal work in patients where chemical fat-sat can fail.

Example: STIR reveals bright bone-marrow oedema in a dog with discospondylitis when the vertebral bodies otherwise look normal.

3D TSE (SPACE / CUBE / VISTA)

Single-slab, high-resolution 3D turbo spin echo (vendor names SPACE, CUBE, VISTA) yielding isotropic T2 or fluid-sensitive data.

Why: One volume can be reformatted in every plane, valuable for nerve roots and cord anatomy without extra acquisitions.

Example: A 3D TSE myelographic-style sequence shows the shape of a cat’s spinal cord and nerve roots around a lesion.

DWI / ADC (diffusion-weighted imaging)

DWI measures the microscopic movement of water; the ADC map quantifies it. Restricted diffusion (bright DWI, dark ADC) indicates dense or cytotoxic tissue.

Why: Restricted diffusion helps flag acute infarction, abscess and highly cellular tumour. Always read DWI together with the ADC map to avoid “T2 shine-through” traps.

Example: A dog with a fibrocartilaginous embolism shows restricted diffusion in the affected spinal cord segment on DWI/ADC.

T2* / GRE for haemorrhage

Gradient-echo sequences weighted for susceptibility, exquisitely sensitive to blood-breakdown products and mineral, which “bloom” dark.

Why: Haemorrhage changes prognosis and differential in brain and cord disease and can be invisible on routine sequences.

Example: Prior bleeding within a canine choroid plexus tumour is confirmed as blooming low signal on T2* GRE.

Acquisition & patient factors

2D vs 3D acquisition

2D collects a stack of individual slices with gaps; 3D excites a whole volume and can be reformatted in any plane, usually with thinner slices.

Why: 3D catches small lesions between 2D slice gaps but takes longer, a trade-off that matters under anaesthesia.

Example: A tiny trigeminal nerve sheath tumour in a dog may be seen only on a thin 3D sequence, not on 4 mm 2D slices.

Isotropic voxel

A voxel that is equal in all three dimensions (e.g. 0.8 × 0.8 × 0.8 mm), so reformats in any plane look as sharp as the original.

Why: Isotropic 3D data lets one acquisition serve every plane, saving anaesthetic time versus repeating orthogonal 2D scans.

Example: An isotropic 3D T1 of a cat brain is reformatted transverse, sagittal and dorsal from a single volume.

Parallel imaging

A technique using multi-channel coils to skip data lines and reconstruct the image, shortening scan time at some cost to SNR.

Why: Faster scans mean lighter anaesthetic load, though the SNR penalty is felt more keenly on lower-field vet magnets.

Example: Parallel imaging trims a canine spine study by minutes, helping keep total anaesthesia time down.

Fat suppression

Methods (chemical/spectral fat-sat, or STIR) that darken fat so that enhancement, oedema and pathology are not masked by bright fat.

Why: Fat is abundant in small-animal anatomy; suppressing it is often what makes a lesion visible.

Example: Fat suppression on a post-contrast dog orbit study confirms an enhancing retrobulbar mass against darkened orbital fat.

Low-field vs high-field

Field strength (measured in tesla, T) sets baseline SNR. High-field magnets (1.5–3 T) give more signal and speed; low-field magnets (roughly 0.2–0.5 T) give less but cost far less to run.

Why: Many veterinary practices operate lower-field systems; low-field needs longer scans or thicker slices to reach diagnostic quality, and favours robust sequences like STIR.

Example: A 0.3 T practice magnet still diagnoses a large canine disc extrusion well, but a subtle brainstem lesion may be clearer at 1.5 T.

Field strength & why vet magnets are often low-field

The tesla rating of the magnet. Higher field boosts SNR and enables advanced techniques; lower field reduces purchase, siting and running costs.

Why: Cost, siting flexibility and adequate diagnostic performance for common indications make 0.2–1.0 T systems widespread in veterinary settings, while referral centres favour 1.5 T.

Example: A first-opinion referral hospital chooses a 0.5 T magnet for everyday IVDD and brain work, referring complex neuro-oncology cases to a 1.5 T centre.

Back to contents

Positioning & anaesthesia

General anaesthesia (essential in vet MRI)

Animals cannot be asked to lie still for minutes at a time, so nearly all veterinary MRI is performed under general anaesthesia to eliminate movement.

Why: Even slight motion ruins the long acquisitions MRI requires; anaesthesia guarantees immobility and a controlled, monitorable patient.

Example: A fractious cat is fully anaesthetised and intubated before a brain MRI so that no motion blurs the images.

Sternal vs dorsal recumbency

Patient orientation in the bore: sternal (lying on the breastbone, belly down) or dorsal (lying on the back). Choice depends on region, coil and access.

Why: Consistent, symmetrical positioning simplifies interpretation and comparison with the contralateral side, and affects coil coupling and comfort.

Example: A dog head study is often run in sternal recumbency with the nose extended to keep the brain straight in the coil.

Apnoea / breath-hold via ventilator

Briefly pausing ventilation (induced apnoea using the anaesthetic ventilator) to freeze respiratory motion during body sequences.

Why: Breathing motion badly degrades abdominal and thoracic images; a controlled apnoeic pause is the veterinary equivalent of a patient breath-hold.

Example: The anaesthetist induces a short apnoea while a dynamic liver sequence is acquired in a dog to avoid respiratory blur.

Patient warming / hypothermia risk under GA

Anaesthetised animals lose heat readily, and MRI rooms are kept cool for the equipment; active warming and temperature monitoring protect the patient.

Why: Hypothermia prolongs recovery and can cause complications, especially in small or thin patients such as cats and toy breeds.

Example: A small cat is wrapped and placed on a warm-air blanket, with core temperature checked throughout a lengthy spinal study.

Back to contents

Artifacts

Motion (respiratory / cardiac)

Ghosting and blurring caused by patient movement, most often breathing and heartbeat during body imaging.

Why: Motion is the commonest cause of a non-diagnostic study; controlled anaesthesia, apnoea and gating are the countermeasures.

Example: Respiratory ghosts smear across a dog abdomen scan until a breath-hold apnoea is introduced.

Chemical shift

A misregistration artifact at fat–water boundaries, seen as a bright or dark rim in the frequency-encode direction.

Why: It can mimic or obscure a real margin, e.g. around the kidney or a fat-containing lesion, and is worse at higher field.

Example: A dark chemical-shift line rims a cat’s kidney at the fat interface and should not be mistaken for a capsular lesion.

Susceptibility (microchips, implants)

Signal loss and geometric distortion around metal, caused by local magnetic field disruption from identification microchips, orthopaedic implants or dental metal.

Why: A microchip near the shoulders or neck can distort adjacent anatomy; knowing its location helps distinguish artifact from disease.

Example: A blooming void over the dorsal neck of a dog corresponds to its identification microchip, not a lesion.

Aliasing / wrap-around

Anatomy outside the field of view “wraps” onto the opposite side of the image when the field of view is smaller than the patient.

Why: It can superimpose limbs or body wall onto the region of interest; enlarging the field of view or oversampling removes it.

Example: A dog’s forelimb wraps into a cervical spine image until the phase field of view is widened.

Magic-angle

Artefactually increased signal in tendons and ligaments oriented near 55° to the main field on short-TE sequences, mimicking pathology.

Why: It can falsely suggest tendinopathy; recognising the angle-dependence and checking a long-TE sequence avoids over-diagnosis.

Example: Bright signal in a dog’s shoulder tendon at a critical angle fades on the long-TE T2, confirming magic-angle rather than a tear.

Back to contents

Contrast

Gadolinium (GBCA) in dogs & cats

Gadolinium-based contrast agents shorten T1, brightening tissues where they accumulate, typically areas of blood-brain-barrier breakdown or hypervascularity.

Why: Enhancement patterns sharpen the differential for brain, spinal and soft-tissue disease and help define surgical margins.

Example: A canine meningioma enhances strongly and uniformly on post-contrast T1, often with a dural-tail-type margin.

Contrast dosing note

A standard clinical dose is typically around 0.1 mmol/kg intravenously, but agent, concentration and indication vary.

Why: Correct dosing balances diagnostic enhancement against cost and safety; underdosing can hide lesions.

Example: A 20 kg dog receives its calculated dose. Always verify the figure against the product label and your local protocol (see our MR calculators). Confirm every dose before administration.

Dynamic enhancement

Rapid, repeated imaging during and after contrast injection to capture how a lesion takes up and washes out gadolinium over time.

Why: Temporal enhancement behaviour helps characterise vascular lesions and pituitary or hepatic masses.

Example: Dynamic imaging of a cat pituitary shows the normal gland enhancing before an adjacent microadenoma, aiding detection.

Back to contents

Safety

MR Safe / Conditional / Unsafe

Standard ASTM F2503 labels for objects and implants: MR Safe (no known hazard in any MR environment; reserved for non-conducting, non-metallic, non-magnetic items such as plastic), MR Conditional (may enter only under the stated conditions), MR Unsafe (a hazard in any MR environment).

Why: Any implant or device on the patient (or on staff) must be checked before entering the room to prevent injury.

Example: A dog’s stainless orthopaedic plate is checked against its labelling before the study proceeds.

Microchip artifact & safety

Identification microchips create local susceptibility artifact near their site, which can be marked at 1.5 T. Published evidence on heating and movement is limited: one small study (three dogs at 1.5 T) found no migration and only very subtle tissue change. An anaesthetised animal cannot report heating, so monitor the patient and follow local protocol.

Why: Knowing the chip is present and where it sits prevents mistaking its signal void for disease, particularly on neck and shoulder studies.

Example: A signal void over a cat’s dorsal neck is correctly attributed to its microchip rather than a soft-tissue mass.

SAR & heating

Specific absorption rate (SAR) is the radiofrequency energy deposited in tissue, which produces heat; the scanner limits it automatically.

Why: Combined with an anaesthetised animal’s impaired thermoregulation, heating deserves attention, especially near conductive implants or leads.

Example: A long, RF-intensive sequence on a dog is monitored for patient temperature to keep within safe thermal limits.

Projectile (missile) effect

The strong static field can violently attract ferromagnetic objects, turning scissors, oxygen cylinders or clamps into dangerous projectiles.

Why: Anaesthetic and surgical equipment around veterinary patients is a real risk; only kit labelled MR Safe or MR Conditional enters the room.

Example: An MR Conditional anaesthetic trolley and non-ferrous instruments are used so nothing is pulled toward the bore.

Monitoring under GA in the bore

Continuous anaesthetic monitoring (capnography, pulse oximetry, ECG, temperature) using MR Conditional equipment while the patient is deep inside the magnet.

Why: The patient is out of arm’s reach and partly hidden, so remote MR Conditional monitoring is the only way to detect problems promptly.

Example: A fibre-optic pulse oximeter and MR Conditional capnograph track a cat’s vitals from the control room during scanning.

Back to contents

Clinical & indications

Intracranial indications

Common brain reasons for MRI include neoplasia (e.g. meningioma, glioma), inflammatory disease such as MUO/meningoencephalitis, and structural work-up of seizures/epilepsy.

Why: MRI is the modality of choice for the brain, distinguishing tumour from inflammation from vascular disease to guide treatment.

Example: A dog with new-onset seizures has MRI to look for a mass, encephalitis or a structural anomaly before starting long-term therapy.

Spinal indications

Frequent spinal reasons include intervertebral disc herniation (IVDD), fibrocartilaginous embolism (FCE) and acute non-compressive nucleus pulposus extrusion (ANNPE).

Why: MRI localises the lesion, separates compressive from non-compressive disease and informs whether surgery is indicated.

Example: A Dachshund with acute paraparesis is imaged to confirm a compressive disc extrusion suitable for surgical decompression.

Musculoskeletal indications

MRI evaluates soft-tissue orthopaedic disease such as cranial cruciate ligament injury, meniscal tears and shoulder tendon/ligament pathology.

Why: It shows ligament, tendon, cartilage and marrow that radiographs cannot, refining diagnosis of lameness.

Example: A dog with chronic forelimb lameness has shoulder MRI revealing a biceps/supraspinatus tendinopathy.

Brain protocol basics

A typical canine/feline brain study pairs multiplanar T2 and FLAIR with a T2* GRE, plus pre- and post-contrast T1 (ideally 3D), and often DWI/ADC.

Why: This combination characterises most lesions (oedema, haemorrhage, enhancement pattern and diffusion) in one anaesthetic episode.

Example: An enhancing extra-axial mass with a broad dural base on post-contrast 3D T1 supports a meningioma diagnosis.

Spinal cord / IVDD imaging

Sagittal and transverse T2, T1 and often STIR localise disc extrusions, cord compression and cord signal change; DWI helps identify FCE/ANNPE.

Why: Accurate localisation and grading of compression directly determine surgical versus conservative management and prognosis.

Example: Transverse T2 shows a lateralised extradural disc compressing the cord on one side in a chondrodystrophic dog.

MUO (meningoencephalitis of unknown origin)

An umbrella term for non-infectious inflammatory brain disease of dogs (including GME and necrotising encephalitides), a signature canine MRI diagnosis.

Why: MRI plus cerebrospinal fluid analysis is central to diagnosis; recognising the typical multifocal T2/FLAIR-bright, variably enhancing pattern guides immunosuppressive treatment.

Example: A young small-breed dog shows multifocal T2/FLAIR-hyperintense lesions with patchy contrast enhancement, supporting MUO after infectious causes are excluded.

For education only. Not a substitute for professional veterinary judgement, product labelling, or your practice’s own MRI and anaesthesia protocols. Contrast doses, safety labelling and clinical management must always be verified against the manufacturer’s instructions and current standard of care.