Veterinary MRI anaesthesia, safety and patient-prep checklist

Veterinary MRI anaesthesia, safety and patient-prep checklist

A practical high-yield companion for veterinarians and veterinary students.

On this pageBefore the scanAnaesthesia & monitoringMR safety zones & staffGadolinium contrastStudent high-yield
How to use this page. This is a practical aid to be worked through alongside your practice’s local SOPs, anaesthetic protocols and the specific product labelling for your contrast agent and equipment. It does not replace clinical judgement or on-site supervision. Tick-boxes are for your working session only and are not saved.
Key points
  • Before the patient enters the room, remove collars, tags and harnesses, locate the microchip (expect local artefact) and check any implant or metallic foreign body against its device label; if MR status is unknown, treat with caution and seek advice.
  • Treat the magnet as always on: it can pull ferromagnetic objects violently into the bore, so screen every item and person and keep instruments, steel oxygen cylinders and phones out unless MR Safe, or MR Conditional within labelled conditions.
  • Clinical small-animal MRI is done under general anaesthesia, not sedation alone, because acquisition is long and very sensitive to motion, and in-room monitoring must be MR Conditional, covering at least SpO₂, ETCO₂, ECG, temperature and blood pressure.
  • Route long monitoring cables without conductive loops or skin contact to avoid burns, and monitor temperature throughout with a warming or heat-retention plan suited to the MR room, as scanner rooms are cool and anaesthesia impairs thermoregulation.
  • In a patient emergency, move the patient out of the scanner room and resuscitate there, never bringing ferrous crash equipment to the bore; quench only as a last resort and never for a cardiac arrest.
Explain it simply
  • Before the scan, remove collars, tags and harnesses, find the microchip (it distorts the image nearby) and check any implant or metal object’s label. If its MRI status is unknown, be cautious and get advice.
  • Treat the magnet as always on: it can pull magnetic metal objects into it with deadly force. Check every person and item; keep tools, steel oxygen cylinders and phones out unless their label allows them in.
  • Dogs and cats have MRI under general anaesthesia, not sedation alone, as any movement spoils the pictures. In-room monitors must be MR Conditional and track at least oxygen, exhaled carbon dioxide, heart trace, temperature and blood pressure.
  • Lay monitor cables without loops and away from the skin to avoid burns. Check temperature throughout and keep the animal warm with equipment suitable for the scanner room, which is cool.
  • In an emergency, take the animal out of the scanner room and treat it there; never bring iron-containing emergency kit to the magnet. Shutting down the magnet is a last resort, never for a stopped heart.

A guide only: follow your departmental protocols and regional requirements.

Back to contents

Before the scan

Complete before the patient is anaesthetised and before it enters the scanner room.

Clinical & consent

Metal & MR-safety screen of the patient

Prep for anaesthesia & scanning

Back to contents

Anaesthesia & monitoring

Why general anaesthesia is the standard

Why: MRI acquisition is long and very sensitive to motion, so clinical small-animal MRI is done under general anaesthesia (not sedation alone) to keep the patient still. Awake scanning has been reported only in research with dogs specially trained for it. Follow your anaesthetist’s protocol.

MR Conditional monitoring

All monitoring inside the room must be MR Conditional. Standard minimum: SpO₂, ETCO₂ (capnography), ECG, temperature and blood pressure.

  • Route long cables/fibre-optic leads carefully; avoid conductive loops and skin-cable contact (burn risk).
  • Know where the emergency magnet-off (quench) button is and the narrow situations in which it is used (see Emergency response below). Position the patient, monitors and staff so the anaesthetist can see the patient and the monitor display continuously, through the control-room window or a camera.
  • Keep infusion pumps and non-conditional gear outside the marked fringe-field contour unless certified for the room, and at the distance the device’s own labelling states.

Positioning by study

  • Brain: usually sternal (ventral) recumbency in a head coil, head straight and symmetrical. Spine: often dorsal recumbency on a spine coil (lateral in some large or deep-chested dogs), spine straight in the coil. Follow the coil and site protocol.
  • Some body/limb studies: dorsal recumbency as dictated by the region and coil.
  • Pad pressure points; keep the region of interest at isocentre and within the coil.

Breathing & body imaging

For thoracic/abdominal imaging, respiratory motion blurs images; a brief apnoea / ventilator breath-hold during acquisition (per your anaesthetist) reduces artefact.

Hypothermia & recovery

  • Scanner rooms are cool and anaesthesia impairs thermoregulation, so most patients lose heat, but some (large or long-coated dogs, and after alpha-2 agonists) warm up during scanning. Monitor temperature throughout, reduce heat loss with insulation or MR-suitable warming, and adjust if the patient warms.
  • Plan a warm, quiet recovery area with continued monitoring until extubation and sternal recumbency.

No specific anaesthetic drug doses are given here. Follow your anaesthetist and your practice’s anaesthetic protocol.

Back to contents

MR safety zones & staff

The projectile (missile) effect

Why: On superconducting and permanent-magnet scanners the static field is always on, even when the scanner is idle or the power is off. Only resistive magnets can be switched off, so treat the magnet as on unless your site’s signage and procedures say otherwise. Ferromagnetic objects can be pulled violently into the bore, causing fatal injury and equipment damage. Every item entering the room must be ferromagnetic-screened.

The fringe-field contour

  • The controlled access area is bounded by a fringe-field contour marked on the site plan. IEC 60601-2-33 Edition 4.0 (2022) sets that limit at 0.9 mT (9 G) and the ACR Manual on MR Safety (2026) delimits the MR Environment by the 9 gauss line; the older 0.5 mT (5 G) contour lies further from the magnet, is still named in much national guidance, and a site drawn to it is being more conservative. The ACR states such sites need no adjustment.
  • Keep non-conditional equipment, and people with pacemakers, cochlear implants or infusion pumps, outside it. Device labelling and your site plan may quote different contours. Check both.

No loose ferrous items

Keep out of the room unless labelled MR Safe, or MR Conditional and used within the conditions on its label: scissors, haemostats/clamps, kidney dishes, needles, steel oxygen cylinders, laryngoscopes, stethoscopes, pens, mobile phones, keys and jewellery. Use equivalents labelled MR Safe, or MR Conditional and used within their labelled conditions.

Staff screening

  • Screen every person entering: cardiac pacemakers/ICDs, cochlear implants, infusion pumps, aneurysm clips, metallic foreign bodies (especially intra-ocular), recent surgery.
  • Empty pockets before each entry; treat the room as always-magnetised.
  • Anyone who stays in the room during scanning wears hearing protection (RANZCR MRI Safety Guidelines v3.0 (2021), s. 12.1; MHRA 5th edition (2026), s. 5.3.9).

Emergency response

  • In a patient emergency, move the patient out of the scanner room to a safe area and resuscitate there. Do not bring ferrous crash equipment to the bore.
  • Quench as a last resort: it vents cryogen, is hazardous and does not stop the field instantly. Use it when someone is pinned or trapped by a ferromagnetic object and cannot be freed quickly and safely, or when firefighters must enter the magnet room; never for a cardiac arrest.
Back to contents

Gadolinium contrast

Indication

Gadolinium-based contrast improves detection and characterisation of many lesions: neoplasia, inflammatory/infectious disease (e.g. MUO, meningitis), disruption of the blood–brain barrier, and vascular/soft-tissue lesions.

Dosing

  • Dose and volume per the product label and your practice protocol; concentration and licensed dose vary between agents, and use in animals is usually off-label or under the veterinary prescribing cascade.
  • Image promptly after injection per your sequence protocol; note timing for consistent enhancement.

Tip: use the site’s MR Calculators page to work through weight-based contrast volume and related conversions.

Cautions

  • Renal impairment: confirm renal function beforehand; use with caution and minimise dose where function is reduced.
  • Extravasation: ensure the IV catheter is patent before injecting; watch the site and manage extravasation promptly.
  • Record agent, dose, volume and time in the anaesthetic/imaging record.
Back to contents

Student high-yield

Concise facts to anchor small-animal neuro/MRI reporting. Always integrate with signalment, neuro-localisation and CSF.

Inflammatory

MUO

Meningoencephalitis of unknown origin is a common cause of meningoencephalitis in dogs, classically in young adult small and toy breeds. Multifocal T2/FLAIR-hyperintense lesions with variable enhancement.

Neoplasia

Meningioma

Commonest primary brain tumour in both dogs and cats. Extra-axial, broad dural base, strong homogeneous enhancement. In cats often smoothly marginated and sometimes incidental.

Disc

Hansen type I vs II

Type I = extrusion (chondrodystrophic breeds, acute, nucleus pulposus). Type II = protrusion (chronic, older large breeds, annular bulge).

Peracute

FCE & ANNPE

Fibrocartilaginous embolism and acute non-compressive nucleus pulposus extrusion: peracute, often non-painful after onset, often lateralising, non-compressive; pain at onset and spinal hyperaesthesia are more common with ANNPE. Focal intramedullary T2 hyperintensity, often over an intervertebral disc space.

Exclusion

Degenerative myelopathy

Typically MRI-negative, a diagnosis of exclusion. Older dogs of at-risk breeds (e.g. German Shepherd, Boxer, Pembroke Welsh Corgi); associated with the SOD1 mutation. Rule out compressive disease first.

Malformation

Chiari-like / syringomyelia

Classic in Cavalier King Charles Spaniels: caudal fossa overcrowding with cerebellar herniation and a fluid-filled syrinx in the cervical cord. Look for phantom scratching/pain.

Sequence

DWI / ADC

Diffusion-weighted imaging with ADC maps helps identify acute ischaemia (restricted diffusion) and abscess (restricted central pus vs cystic tumour).

Sequence

STIR at low field

STIR is the preferred fat-suppression technique at low field (more homogeneous than spectral fat-sat), useful for oedema, bone and orbit.

Neoplasia

Glioma vs meningioma

Gliomas are intra-axial (brachycephalic breeds; variable enhancement: none, ring-like, heterogeneous or solid) versus the extra-axial, dural-based, strongly enhancing meningioma.

Feline

Cat brain tumours

Meningioma dominates in cats and can be multiple; lymphoma is an important differential. Correlate with FeLV/FIV status and CSF.

Vascular

Haemorrhage on MRI

Signal evolves with age of blood (deoxy→met→haemosiderin). Use T2*/GRE (or SWI) to detect microbleeds and susceptibility.

Principle

Integrate everything

MRI is never read in isolation. Always combine images with signalment, neuro-localisation and CSF analysis to build the differential list.

Educational use only. This guide is a learning and practical aid for qualified veterinary professionals and students. It is not a substitute for your practice’s local anaesthetic and MR-safety protocols, manufacturer product labelling, the MR conditions of each device, or clinical supervision. Always verify drug and contrast doses against current product labels and follow your own SOPs.

More in Veterinary imaging

Next in Veterinary imaging: Veterinary MR calculators

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.

Contents