What you’ll be able to do after this lesson
- Run the first sixty seconds of a cardiac arrest inside a scanner room correctly.
- Explain to an arriving code team why they are being stopped at a door.
- Tell the emergency stop, the power off and the quench button apart.
- Recognise a quench venting failure and act on it.
- Say what happens in a fire, and why the fire service waits.
The rule that inverts everything
In every other part of the hospital, the resuscitation comes to the patient. In Zone IV, the patient goes to the resuscitation. The site’s emergency wall poster puts it in one line: the code does not come to the patient; the patient comes to the code.
The ACR states it as policy: in arrest or other medical emergency within Zone IV requiring intervention, the patient must be removed immediately to a predetermined, magnetically safe location, and trained MR personnel should immediately begin basic life support. Facilities must have immediate access to an MR Conditional gurney when the MR table cannot be undocked.
One nuance stops this becoming dogma. If moving the unstable patient out would be delayed by more than a few seconds, prioritise patient safety and start compressions and rescue breaths while still in Zone IV. The equipment rule does not change: MR Unsafe items such as automated external defibrillators are never brought in. Once resuscitation has moved to Zone III, the Zone IV door is closed and secured so arriving responders do not walk in.
The site’s poster sequence is the practical version:
- Recognise and call. Confirm the arrest; call the resus team and state clearly: “MRI, Zone IV.”
- Start BLS where the patient is. Do not wait.
- Evacuate onto the MR-safe trolley and out to the designated resuscitation area, continuing compressions.
- Guard the door. One person stops the crash cart, defibrillator and code team entering Zone IV.
- Defibrillate outside, once clear of Zone IV.
- Full ALS in the resuscitation area. Note any contrast and any sedation given.
Door-guarding is a real job with a name on it, decided before the emergency. A crash team running toward a patient is doing exactly what it is trained to do; the person at the door is the only thing between that training and a defibrillator entering a magnet. The ACR also recommends emergency calls from an MR facility be forwarded to a designated MR personnel member, on site before responders.
Three buttons, three different things
Staff confuse these, and the consequences differ enormously.
Emergency table stop. Immediately stops scanning and table motion. Found on or near the console and on the bore cover or table. Used when something is caught on the table and further movement could injure the patient. The magnet stays on.
Emergency power off. Cuts power to the suite and computer room, including any uninterruptible supply. Used for fire, flooding or voltage accidents; resetting may need an electrician and the main breaker. The magnet stays on.
Emergency magnet off — the quench button. Ramps the field down by quenching a superconducting magnet, dissipating enormous energy as the cryogen boils off and vents. Deliberate, costly, and the only one of the three that removes the field.
If either of the first two is used and the patient must come out quickly, someone has to know the manual method for undocking the table.
When to quench
The ACR’s general principle: Level 1 and Level 2 MR personnel should activate a quench whenever there is imminent or potential risk of injury or death to anyone in Zone IV — entrapment by a metal object attracted by the magnet, or a projectile event — and removing them safely is not feasible. When in doubt, the manual says, quench.
Equally explicit is what is not a quench situation: a medical code. The patient can be removed from Zone IV for care, so an arrest does not justify quenching. Nor should anyone try to pull a ferromagnetic object off the magnet — that can put the trapped person at greater risk.
A quench is also considered when emergency responders must enter Zone IV without screening. A Level 2 person then monitors it to judge when entry is safe; the field can take more than a minute to dissipate.
Quench venting failure
Helium normally exhausts outside through the quench pipe. If venting fails, cryogenic gas enters the scanner room — signalled by white cloud or fog around or above the scanner. The hazards are cold injury and asphyxiation: helium is less dense than air and displaces oxygen from the top down. The ACR’s steps:
- If the patient is ambulatory, advise them to leave Zone IV at once, walking as low to the floor as possible to keep their head below the accumulating gas.
- Turn on the suite exhaust fan immediately, and open the scan room door to ventilate Zone IV; consider opening doors to other zones to ventilate Zone III.
- If entering to rescue someone, stay low. Assume the magnet is still at field and use MR Conditional equipment if a trolley is needed.
- Once the rescue is complete, close the Zone IV door to stop helium flowing into Zone III, evacuate and let the gas dissipate. Vendor personnel advise when re-entry is safe.
Over-pressure can jam the door — another reason to evacuate rather than investigate.
Fire
Assume in any Zone IV fire that the field is present, fully operational and dangerous. Free access by firefighters with air tanks, axes and crowbars could be lethal.
Facilities must keep clearly marked MR Conditional or MR Safe fire extinguishing equipment inside Zones III or IV; conventional extinguishers are restricted from both. The ACR offers RACE: Rescue persons in danger if safe to do so; Activate the nearest alarm and call with specific information; Contain by closing all doors including fire doors, using an MR Conditional extinguisher if safe; Evacuate if instructed by the fire department.
If a Zone IV fire cannot be contained by sprinklers or a conditional extinguisher and the fire service needs access, the quench is engaged to avoid serious injury to responders. If the fire is in Zone II or III, close and lock the doors to Zone IV.
None of this behaves correctly at two in the morning unless it has been rehearsed. The ACR recommends mock code training with debriefing; if your unit has not drilled evacuation-to-resus in the last year, that is worth raising.
On the ward
You are the ward nurse who came down with an inpatient. Mid-scan, the radiographer calls the arrest. You are the only other clinical person in the control room.
Do the two things that are yours. First, help move the patient — the table comes out, the patient goes onto the MR-safe trolley, compressions continue as you go to the designated area. Second, take the door. When the code team arrives with the trolley and defibrillator, you stop them: “Do not bring that in. The patient is coming out to you, in the bay on your left.”
Then give the information nobody else has: what contrast was given and when, what sedation is running, and what came off the patient before the scan.
Check yourself
If someone called an arrest in the scanner room you visit most, do you know where the designated resuscitation area is, and where the MR Conditional trolley lives? Could you say “do not bring that in” to a consultant running toward you? Nothing here is stored.
Key takeaways
- The patient is removed from Zone IV to a predetermined safe location and resuscitated there; the cart, defibrillator and team never enter.
- If evacuation would be delayed more than a few seconds, start compressions and ventilation in Zone IV — MR Unsafe equipment, including AEDs, still stays out.
- Close and secure the Zone IV door once resuscitation moves to Zone III; guarding it is an assigned role.
- Emergency table stop halts scanning and table motion; emergency power off cuts suite power; only the quench removes the field.
- Quench for imminent risk of injury or death, such as entrapment — not for a medical code, and never try to pull the object free.
- In a venting failure: get out low, run the exhaust, open the door, stay low if re-entering, then close the door and let the gas dissipate.
- Only MR Conditional or MR Safe extinguishers enter Zones III and IV; if the fire service must enter Zone IV, the magnet is quenched first.
References
- ACR Manual on MR Safety, 2026 edition https://www.acr.org/Clinical-Resources/Clinical-Tools-and-Reference/Radiology-Safety/MR-Safety
- highyieldimaging.com — MRI Nursing & Emergency Cards https://highyieldimaging.com/mri-nursing-emergency-cards/
- highyieldimaging.com — MRI for Radiology Nurses https://highyieldimaging.com/mri-for-radiology-nurses/
- highyieldimaging.com — MRI Screening, Patient Prep & Positioning https://highyieldimaging.com/mri-screening-positioning/