What you’ll be able to do after this lesson
- Ask screening questions that surface what a form misses.
- Name the items that are removed at the Zone IV door even after screening is complete.
- Describe what a “full stop and final check” is and why an escorted inpatient gets the augmented version.
- Explain the hearing protection requirement in terms a patient or a companion will accept.
- Say what happens when a patient cannot give their own history.
The form is a prompt for a conversation
Screening is the highest-value safety task in MRI, and much of it happens away from the scanner — on your ward, in a prep bay, in a phone call from a referrer. The ACR expects conscious, non-emergency patients to be screened twice, at least once by Level 2 MR personnel, with a written or electronic questionnaire completed before they are introduced to Zone III. If a safety concern is identified, entry to Zone III waits until it is resolved.
A form on its own is not screening. Patients under-report, misremember and misunderstand. “I don’t have anything metal” is routinely said by people with stents, clips, coils and dental hardware, because they do not think of those as metal. Better questions are open ones: “Have you ever had any operation, anywhere on your body?” Then probe visible scars and cross-check old records or prior imaging.
The topics that must be covered every time:
- Cardiac devices: pacemaker, ICD, loop recorder, abandoned leads, epicardial wires.
- Neurostimulators, cochlear implants, programmable shunts, drug pumps.
- Aneurysm clips, coils, stents, heart valves — what, where, and when implanted.
- Metallic foreign bodies, especially orbital metal. A grinding, welding or metalworking history means an orbital radiograph before scanning if there is any doubt; a ferromagnetic orbital fragment can move and blind the patient.
- Shrapnel, bullets, piercings; recent surgery.
- Transdermal patches — some have metallic backing, a burn risk and a drug-delivery problem.
- Tattoos and permanent make-up (rare heating; ask the patient to report any hot spot immediately).
- Pregnancy status, breastfeeding if contrast is planned, and renal function if gadolinium is likely.
One rule prevents a lot of harm: never accept “I had an MRI before” as clearance for a device. Conditions depend on the exact model, field strength, scanner and body part imaged. Every implant is checked against its current labelling, every time.
When the patient cannot answer
Unconscious, sedated, intubated, confused and very young patients are treated as unscreened until proven otherwise. The ACR describes a layered approach: consult the electronic record including any implanted-devices module and prior imaging; have family or guardians complete a written screening questionnaire and a verbal interview; and, where no reliable history can be obtained and the scan cannot wait, inspect visually for scars and obvious implants and — if recent radiographs, CT or MRI of the core anatomy are not available — obtain plain radiography of head and neck, chest, abdomen and pelvis, and the contiguous upper arms and thighs.
Language and capacity matter as much. Use a professional interpreter rather than a family member, use translated forms, and confirm understanding by teach-back. Document who screened and how. Children may be unreliable historians: where policy allows they are screened twice by Level 2 personnel, once with the parent present and once separately.
The full stop and final check
Immediately before a patient crosses from Zone III into Zone IV, the MR technologist performs a “full stop and final check”: patient identity and visual inspection, the examination and any contrast risk assessment, that screening was properly performed, that implanted or worn devices have been prepared, programmed or removed, and that nothing about the patient’s status has changed in Zone III.
Hearing aids are a specific trap. If they have been left in so the patient can be spoken to in Zone III, strict attention is needed to make sure they come out and are stored before Zone IV — as must external insulin pumps and continuous glucose monitors.
Escorted, hospitalised, sedated or interventional patients get the augmented version: a verbal review by the supervising Level 2 technologist, acknowledged by a second MR personnel team member, modelled on the surgical time-out. It adds screening of any support staff entering Zone IV, visual inspection of the patient and transport equipment for concealed hazards — steel oxygen cylinders, unconnected ECG electrodes and leads, RFID tags in hospital linen — identification of the line to be used for contrast, and confirmation that anything needing tethering in Zone IV is secured before the patient goes in.
What stops at the door
Patients change into facility-supplied MR Safe pocketless garments. Ordinary clothing can contain metallic microfibre, and pockets hide things.
MR Unsafe transport equipment such as ward wheelchairs and gurneys may enter Zone III only when MR personnel judge it necessary for patient care, supervised by designated staff who know the equipment, and physically secured or tethered at all times. It does not enter Zone IV.
Anything portable entering Zone IV must carry MR Conditional or MR Unsafe labelling, and MR Conditional items are obtained in preference. Equipment not needed for the immediate care of the patient should not be in Zone IV while a patient is in the room — the working principle is that the patient is last in and first out with respect to other objects.
Comfort items from home — pillows, soft toys — are discouraged and admitted only case by case after screening. Service animals are not recommended in Zone IV, as they cannot be screened or given hearing protection.
Hearing protection
Gradient switching is loud. The FDA treats systems capable of exceeding 99 A-weighted decibels with hearing protection in place as a significant risk, and IEC 60601-2-33:2022 requires protection to bring the level below that threshold. Protection frequently needs to exceed 28 to 30 dB given the loudness of contemporary magnets and sequences.
Two points from the 2026 edition are new enough that people miss them. Muffs may be displaced during the examination, offering inadequate protection — so both ear plugs and ear muffs are recommended whenever possible. And fit is checked, not assumed: staff verify placement, fit and function before the examination, while making sure the patient can still hear instructions. If a patient refuses hearing protection, the unit should have a process for discussing the risk, and may consider cancelling.
It is recommended for accompanying companions in Zone IV too, and it is not optional for sedated or anaesthetised patients — they cannot tell you it has fallen out.
On the ward
Mrs Ferreira, 78, is going for an urgent MRI head. She is drowsy after a fall, gives inconsistent answers, and her daughter has gone home. Her notes mention “heart surgery, 1990s”. She is wearing hearing aids so you can talk to her, a fentanyl patch, and a glucose sensor on her arm.
Nothing here is a reason to give up, but nothing here is screened either. Ring the daughter and take a verbal screening history. Flag “heart surgery” to the MRI team early — they may want an operative note or a chest radiograph. Remove the patch and document it, so the analgesia gap is planned rather than discovered. Tell the team about the glucose sensor and the hearing aids so they are removed and stored at the final check, not found halfway down the bore.
Then hand over in the order the MRI team needs: who screened, what could not be confirmed, what is attached to the patient, and what you have already removed.
Check yourself
Think of the last patient you sent or escorted to MRI. Could you have answered every item on that screening list from what you knew — and if not, who would you have rung? Nothing here is stored.
Key takeaways
- Conscious non-emergency patients are screened twice, at least once by Level 2 personnel, with a written form plus a real conversation.
- Open questions beat tick boxes; a previous MRI is never clearance for a device.
- Patients who cannot give a history are treated as unscreened: records, family interview, visual inspection, radiography where the scan cannot wait.
- The full stop and final check happens at the Zone III–IV boundary; hearing aids, insulin pumps and glucose monitors come off there.
- Escorted and hospitalised patients get the augmented check, including screening of accompanying staff and inspection of transport equipment.
- MR Unsafe transport equipment may enter Zone III only when necessary, supervised and tethered — never Zone IV.
- Ear plugs plus ear muffs whenever possible; fit and function verified for every person, sedated patients included.
References
- highyieldimaging.com — MRI Screening, Patient Prep & Positioning https://highyieldimaging.com/mri-screening-positioning/
- highyieldimaging.com — MRI for Radiology Nurses https://highyieldimaging.com/mri-for-radiology-nurses/
- ACR Manual on MR Safety, 2026 edition https://www.acr.org/Clinical-Resources/Clinical-Tools-and-Reference/Radiology-Safety/MR-Safety
- highyieldimaging.com — MRI for Allied Health & Referrers https://highyieldimaging.com/mri-for-allied-health-and-referrers/