CT troubleshooting: What you see, what to try, when to escalate
For radiographers, technologists and students at the CT console. Image artefacts, equipment faults, patient problems and safety incidents, each with what you see, what to try first and when to stop and escalate.
On this page
Find your problemImage looks wrongEquipment and systemPatient and workflowSafety incidentsSourcesContrast reaction, arrest, extravasation, a fall or the wrong patient scanned: stop troubleshooting and follow your local emergency or incident procedure.
Contrast reactionExtravasationArrestWrong patient or wrong exposureFallAir embolism
- First decide whether the problem follows the patient or the scanner. An artefact on every patient, or on the daily water phantom, is likely an equipment or calibration fault: check the phantom and call service.
- Try the cheap fixes first: arms up before the localiser, the patient centred at the isocentre, metal removed from the scan range, a rehearsed breath-hold, and the protocol that matches the patient’s size.
- A pattern that could be disease is checked on another window, plane or reconstruction (thin slices, metal artefact reduction), and the radiologist is told before the study is reported.
- Before contrast, ask the screening questions (kidney function, metformin, previous reaction, thyroid) and secure a good cannula. Watch the site and the monitoring scans, and stop at once if there is pain or swelling.
- Wrong patient, wrong region, an unplanned repeat or a pregnancy found after the scan: tell the radiologist and report it. A medical physicist estimates the dose, and your regional rules decide who must be told.
Explain it simply
- If a strange pattern shows up on the test bottle of water as well as on people, the machine is at fault, not the patient.
- Start with the simple things: arms out of the way, the body in the middle of the scanner, metal off, and a practised breath-hold.
- A strange spot on one picture might be a trick of the scanner, so it is checked another way before anyone calls it disease.
- Before the dye, staff check the kidneys, medicines, allergies and thyroid, and watch the drip site closely during the injection.
- If the wrong person or the wrong part is scanned, it is reported openly, the dose is worked out, and the patient is told.
A guide only: follow your departmental protocols, your radiologist and medical physicist, and your regional radiation and contrast rules.
Find your problem
Steps in each card run from cheapest and lowest-risk first. Terms are vendor-neutral; your operator manual gives the names your system uses.
Image looks wrong
Artefacts and image quality. Careful positioning and the right scan parameters prevent many artefacts. If a pattern could be disease, check it on another window, plane or reconstruction and tell the radiologist before the study is reported. If the same artefact appears on every patient, or on the daily water phantom, it is likely an equipment or calibration fault (or a protocol change): go to Equipment and system.
Sources: Barrett 2004; Boas 2012; ACR CT QC 2017
Dark streaks between dense structures, or a darker centre (beam hardening) Can mimic pathology
- What you see
- Dark bands or streaks between two dense objects, such as across the posterior cranial fossa, between bones, or between pockets of dense contrast or barium, often with brighter streaks next to them. In a uniform object, a slight fall in CT numbers towards the centre (cupping).
- Why it happens
- The X-ray beam contains a range of energies. Dense material removes the low-energy photons first, so the beam that comes out is harder than the reconstruction assumes and the numbers along those paths come out too low. Scanners correct for an average amount of hardening, which works less well for dense bone, iodine and metal.
- Try first
- Keep dense material out of the slices you need: arms out of the scan range for the chest and abdomen, and patient position or gantry tilt to angle dense structures away from the area of interest where the protocol allows.
- A higher kV hardens the beam and reduces the artefact, at a cost in soft-tissue contrast. Use it only as your protocol allows.
- On a dual-energy scan, higher-energy virtual monoenergetic images reduce beam hardening.
- For streaks from dense contrast in the veins of the chest, see Dense contrast streaks.
- Stop and escalate when
- Cupping or dark bands appear on the daily water phantom: go to Daily water phantom out of limits.
- Can mimic pathology
- The artificially low numbers can hide or imitate low-density areas, for example in the brainstem. A related effect, incomplete correction for beam hardening, raises the numbers in small kidney cysts next to enhancing kidney (see Cyst looks enhanced).
Explain it simply
The X-ray beam mixes weak and strong rays. Thick bone soaks up the weak ones, so the beam coming out is tougher than the computer expects and it paints dark stripes.
Why it matters: A dark stripe across the back of the brain can hide detail the radiologist needs to see.
Sources: Boas 2012; Geleijns 2014; Kalisz 2016; Katsura 2018
Streaks from metal: implants, clips and dental work Can mimic pathology
- What you see
- Bright and dark streaks radiating from metal, such as joint replacements, spinal fixation, dental fillings, clips, wires or ECG leads, with dark bands between metal objects and lost detail around them.
- Why it happens
- Metal blocks most of the beam (photon starvation) and hardens what gets through (beam hardening); scatter and the sharp metal edges add more streaks. The size, shape, density and orientation of the metal all matter.
- Try first
- Remove external metal from the scan range: jewellery, clothing with metal, and ECG leads that can be moved.
- For dental metal on head scans, patient position or gantry tilt can angle the metal out of the slices you need, where the protocol allows.
- Apply the scanner’s metal artefact reduction (MAR) reconstruction if it has one. MAR can be applied after the scan; a dual-energy acquisition has to be chosen before scanning.
- On a dual-energy scan, high-energy virtual monoenergetic images reduce beam-hardening streaks.
- A higher kV reduces beam hardening and scatter, as your protocol allows.
- Send MAR images alongside the standard images, clearly labelled, not instead of them (a general step; follow your reporting protocol).
- Stop and escalate when
- Streaks still cover the area the clinical question is about: tell the radiologist, who may choose another test.
- Can mimic pathology
- Streaks can hide a collection, bleeding or a tumour next to the metal. MAR images have shown findings that were not visible on the originals.
- Good to know
- MAR can blur structures next to the metal and can create new artefacts, so MAR images are read next to the originals.
Explain it simply
Metal stops almost all the X-rays, so the computer has holes in its information and fills them with bright and dark stripes.
Why it matters: The area next to an implant is often exactly where the problem is, so reducing the stripes can change the diagnosis.
Sources: Boas 2012; Katsura 2018; Puvanasunthararajah 2021; Geleijns 2014
Coarse streaks across the shoulders, hips or a large patient (photon starvation) Can mimic pathology
- What you see
- Coarse, fine-grained streaks and noise across the widest part of the body, usually side to side through the shoulders or pelvis; worse in large patients and with the arms by the sides.
- Why it happens
- Along the most attenuating paths so few photons reach the detector that the signal is mostly noise, which the reconstruction spreads into streaks. The thickest path matters more than the average thickness, so arms at the sides make it much worse.
- Try first
- Raise the arms out of the scan range before the localiser (see Cannot raise arms).
- Use tube current modulation, which raises the current for the most attenuating projections; it is set from the localiser, so take the localiser in the final scan position.
- If the arms must stay down, arms resting across the front of the abdomen give less noise than arms at the sides.
- A higher kV or tube current for the problem region, as your protocol allows, reduces the streaks.
- For review, thicker slices or a softer kernel trade detail for less noise; iterative reconstruction and the scanner’s adaptive noise filtering reduce these streaks where available.
- Stop and escalate when
- Streaks still obscure the area of interest: ask the radiologist whether a limited repeat with adjusted settings is justified.
- Can mimic pathology
- Low-contrast edges, such as small liver lesions or bowel wall, can be lost in the noise.
Explain it simply
Through the widest part of the body, hardly any X-rays get to the other side. With so little information, the computer draws grainy stripes.
Why it matters: Getting the arms out of the way is often the cheapest way to a clear picture at a lower dose.
Sources: Boas 2012; Geleijns 2014; Puvanasunthararajah 2021; Kachelriess 2001; Hsieh 1998; Graef 2022
Blurring, double edges or streaks from movement Can mimic pathology
- What you see
- Blurred edges, doubled outlines, streaks from high-contrast edges, or steps on coronal and sagittal reformats; in the chest and abdomen, a blurred diaphragm and vessels from breathing.
- Why it happens
- The reconstruction assumes the patient is still during each rotation. Movement makes the projections disagree.
- Try first
- Explain the scan, rehearse the breath-hold, and give the same breath-hold instruction for the localiser and the scan.
- Make the patient comfortable and support the part with pads or straps so they do not have to work to keep still.
- Use the shortest rotation time and fastest acquisition the protocol allows.
- For patients who cannot hold their breath or keep still, see Cannot hold their breath and Children and immobilisation.
- Stop and escalate when
- Movement still spoils the images: ask the radiologist whether a repeat is justified and over which range.
- Can mimic pathology
- The IAEA handbook notes that aortic pulsation can cause artefacts that imitate an aortic dissection. At the aortic root, ECG gating reduces motion but does not remove it (Murphy and colleagues).
Explain it simply
A CT picture is built from many views taken as the tube spins. If you move halfway through, the views do not match and the picture blurs or doubles.
Why it matters: A blurred scan may need repeating, which means more radiation.
Sources: Boas 2012; Geleijns 2014; Kalisz 2016; Abbara 2016; Murphy 2018; CAMRT BPG
Blurred or stepped coronary arteries on gated scans Can mimic pathology
- What you see
- Blurred or doubled coronary arteries, steps between blocks of slices on reformats, or bands of different brightness across the heart.
- Why it happens
- Gated scans image the heart over one or more beats. A fast or irregular rhythm, ectopic beats, or breathing between blocks makes the data misalign.
- Try first
- Before scanning, check the ECG: clean the skin (shave if needed), keep the leads outside the scan field, and reposition them if the trace is noisy or the T wave is as tall as the R wave.
- Aim for the heart rate your protocol sets; the SCCT target for coronary CTA is below 60 beats per minute, using your department’s rate-control protocol.
- Rehearse the breath-hold and ask the patient not to strain (no Valsalva manoeuvre).
- For an irregular or fast rhythm, use the protocol’s retrospective gating or single-beat mode if the scanner has one.
- After the scan, review the ECG: editing trigger points and removing ectopic beats can often rescue the study.
- Stop and escalate when
- Steps or blur through the coronary arteries remain after ECG editing: the radiologist decides whether to repeat.
- Can mimic pathology
- Motion can imitate a narrowing or soft plaque, or hide one.
Explain it simply
The heart never stops, so the scanner takes pictures at the same moment in each beat. An uneven heartbeat or a breath between beats stops the pieces lining up.
Why it matters: Coronary arteries are only a few millimetres wide, so a small misalignment can look like disease.
Sources: Abbara 2016; Pack 2022; Moser 2024; Kalisz 2016
Rings or arcs centred on the middle of the image Can mimic pathology
- What you see
- Full or partial circles, darker or lighter, centred on the centre of rotation; clearest on uniform tissue and on the water phantom.
- Why it happens
- A miscalibrated or faulty detector element reads wrongly at every angle, which traces a circle. In one reported case a droplet of spilt contrast on the tube aperture caused a ring until it was cleaned.
- Try first
- Look for the same ring on the daily water phantom images.
- Run an air calibration as the manufacturer directs and repeat the phantom.
- Inspect the gantry window for spilt contrast and clean it as the manufacturer allows, then rescan.
- Stop and escalate when
- Rings persist after several air calibrations, or calibration is not a user task on your system: arrange service and tell the medical physicist. The QC team decides whether, and how, the scanner is used until it is fixed.
- Can mimic pathology
- A subtle ring has imitated a lesion in the pons.
- Good to know
- Rings are more visible in noisy images, for example in large patients.
Explain it simply
One sensor out of thousands is reading wrongly. Because the tube spins round, that one bad reading draws a circle.
Why it matters: A ring affects every patient on that scanner until it is fixed, which is why the daily phantom check exists.
Sources: Boas 2012; ACR CT QC 2017; Geleijns 2014; IAEA HHS 19; AAPM Report 74; Jha 2013; Uppot 2018
Small structures blurred or merged (partial volume) Can mimic pathology
- What you see
- Small or thin structures look larger and fainter than they are, edges blur, or a small cyst measures denser than water.
- Why it happens
- Each voxel shows the average attenuation of everything in it. Thick slices average tissues along the length of the patient.
- Try first
- Reconstruct thinner slices from the same acquisition for small structures; no extra exposure is needed.
- Review the area in a second plane.
- For calcium or stents, review with a wider window.
- Can mimic pathology
- A small kidney mass is too small to characterise when the slice is thicker than half its diameter. Calcium can make a narrowing look worse than it is.
Explain it simply
Each tiny box in the picture shows the average of what is inside it. If a thin structure only half fills the box, it looks fainter and wider.
Why it matters: Thin slices can turn an uncertain spot into a clear answer without another scan.
Sources: Geleijns 2014; Abbara 2016; Silverman 2019; Pack 2022
Spinning streaks near bone and blur at the edges of a wide detector
- What you see
- Alternating dark and light streaks fanning out from sharp, dense edges such as the skull base or ribs, which seem to rotate as you scroll through the slices; with wide detectors, streaks or blur at high-contrast edges along the patient’s length.
- Why it happens
- In helical scanning each image is interpolated between detector rows that cross the image plane at different angles, which near sharp edges gives the windmill pattern. With many rows, the outer rows see the patient along a cone rather than a flat fan, and the scanner’s cone-beam reconstruction corrects most but not all of this.
- Try first
- Review thicker reconstructions: windmill streaks are most visible on thin slices.
- Keep the area of interest near the centre of the field of view.
- Use the protocol’s standard reconstruction, which includes the scanner’s own corrections.
Explain it simply
In a spiral scan the computer has to fill in gaps between the rows of sensors. Next to bright bone, the guesses show up as spinning stripes.
Why it matters: Knowing the pattern stops it being mistaken for something real.
Sources: Boas 2012; Geleijns 2014
Steps or stripes on coronal and sagittal reformats
- What you see
- Serrated, staircase edges on reformats and 3D images, or alternating stripes of more and less noise (zebra), most marked away from the centre of the field of view; on gated cardiac scans, steps at the joins between blocks.
- Why it happens
- The same helical interpolation shows up in the reformat plane. On sequential cardiac scans, movement or a change in heart rate between beats misaligns the adjoining blocks.
- Try first
- Centre the area of interest in the field of view.
- Reconstruct thin source images for reformats.
- For cardiac scans, control the heart rate and rehearse the breath-hold; use single-beat or wide-coverage mode where available (see Cardiac motion).
Explain it simply
Side and top views are built by stacking the slices. If the slices do not join perfectly, edges look like a staircase.
Why it matters: Steps on a reformat can look like a fracture line or a narrowed vessel.
Sources: Boas 2012; Moser 2024; Abbara 2016; Pack 2022
Patient wider than the scan field (truncation) Can mimic pathology
- What you see
- A bright band or rim at the edge of the field of view, a missing or distorted body outline, or CT numbers that are wrong near the edge; arms or a large abdomen extend beyond the scan field.
- Why it happens
- The measured scan field is smaller than the bore on most scanners. Tissue outside it is measured on some projections and not others, and extended-field reconstructions estimate the missing data.
- Try first
- Centre the patient and bring the arms in or up so the body lies inside the scan field.
- Use the extended field-of-view reconstruction if the scanner has one, and tell the radiologist that the edge region is estimated.
- Check the reconstructed field of view has not cropped out anatomy the radiologist needs.
- Stop and escalate when
- The anatomy of interest lies outside the measured field: ask the radiologist about another position or a larger-bore scanner.
- Can mimic pathology
- Truncation or cropping can cut pathology out of the image.
- Where sources differ
- Boas and Fleischmann argue that tissue outside the field does not itself create artefacts inside it and attribute the bright rim to the reconstruction; Fonseca and colleagues found that errors in estimating the missing data can spread into the true scan field. Both agree that the extended region is estimated and its CT numbers are unreliable.
Explain it simply
The scanner can only measure a circle of a certain size. Anything outside that circle has to be guessed, so the edges look wrong.
Why it matters: A wrong outline or wrong numbers at the edge can mislead the radiologist, and in radiotherapy planning they change the dose calculation.
Sources: Boas 2012; Fonseca 2021; Uppot 2018; IAEA HHS 19
Images grainier than usual Can mimic pathology
- What you see
- Images look grainier than usual, across the whole scan or in one region, and low-contrast detail is hard to see.
- Why it happens
- Noise falls as more photons reach the detector. Common causes are a low-dose or wrong protocol, the wrong patient size or age group, an off-centre patient, a localiser taken in the wrong position, thin slices or a sharp kernel.
- Try first
- Check the protocol, the kV, the tube current or noise setting, and that tube current modulation is on.
- Check the patient is at the isocentre (see Patient off-centre).
- Check the slice thickness and kernel match the protocol; thicker slices or a softer kernel reduce noise for review.
- Apply the iterative or deep-learning reconstruction the protocol specifies.
- Stop and escalate when
- The daily water phantom noise is outside its limits, or every patient is noisy on an unchanged protocol: go to Daily water phantom out of limits.
- Can mimic pathology
- Noise hides low-contrast lesions, for example in the liver.
Explain it simply
A CT picture is built from counted X-rays. Too few counts and the picture looks speckled, like a photo taken in dim light.
Why it matters: The right protocol for the patient’s size gives a clear picture without extra radiation.
Sources: Boas 2012; Geleijns 2014; Kalisz 2016; Kaasalainen 2019; ACR CT QC 2017
Patient off-centre: dose modulation and noise
- What you see
- Higher dose or noise than expected, uneven noise across the image, or a tube current or kV that seems wrong for the patient’s size, when the patient lies above or below the isocentre. Studies find patients are most often positioned too low.
- Why it happens
- The bowtie filter and tube current modulation assume the patient is centred. On a front-to-back localiser, an off-centre patient is magnified or shrunk by the diverging beam, so the scanner misjudges their size and sets the current (and, with automatic kV selection, the kV) wrongly. Tissue away from the centre lies under a thicker part of the bowtie and is noisier.
- Try first
- Set the table height so the middle of the body part is at the isocentre, using the lateral light, before the localiser.
- Check centring on a lateral localiser, or with the scanner’s centring aid if it has one.
- Use the localiser direction your protocol specifies; changing it changes how modulation behaves.
- If the patient or table height moved after the localiser, repeat the localiser before scanning.
- Good to know
- In phantom studies, surface dose rose by about 18% and 41% when the phantom was 3 and 6 cm below the isocentre, with noise up by 6% and 22% (Toth and colleagues); the size of the effect varies with the scanner, the localiser direction and automatic kV selection.
Explain it simply
The scanner works out how big you are from the first quick picture. If you lie too low or too high, it guesses wrong and uses the wrong amount of radiation.
Why it matters: Good centring is free and lowers dose for every patient.
Sources: Toth 2007; Li 2007; Kaasalainen 2014; Kaasalainen 2019; Habibzadeh 2012; Graef 2022
Wrong reconstruction kernel or window
- What you see
- Bone looks blurred, or soft tissue is very grainy; lung or bone detail is missing; a series has been sent in only one kernel or window.
- Why it happens
- Sharp kernels show fine detail with more noise; soft kernels lower noise and blur detail. A bone kernel makes the brain unreadable through noise, and a soft-tissue kernel blurs bone. The window decides which range of CT numbers is visible.
- Try first
- Reconstruct the missing kernel from the raw data.
- Check the default window for each series; a narrow window exaggerates the size of calcium and stents.
Explain it simply
The same scan can be processed to look sharp, which suits bone, or smooth, which suits soft organs. Using the wrong one hides what the doctor is looking for.
Why it matters: Sending the right versions avoids recalling the patient.
Sources: Geleijns 2014; Abbara 2016; Boas 2012; Pack 2022
Wrong slice thickness or reconstruction field Can mimic pathology
- What you see
- Only thick slices were sent, reformats look blocky, small lesions cannot be characterised, or the reconstructed field of view is too small, off-centre or crops anatomy.
- Why it happens
- Thicker slices lower noise but average out small structures; reformats need thin source images. A reconstruction field that is too small or off-centre leaves anatomy out.
- Try first
- Reconstruct thin slices from the same acquisition for reformats and small structures.
- Check the reconstructed field of view covers all the anatomy and is centred on it.
- Compare the series sent with the protocol’s list before the patient leaves.
- Can mimic pathology
- A small kidney mass cannot be characterised when the slice is thicker than half its diameter.
Explain it simply
Thick slices are like looking through several pages at once: smooth, but small things blur together.
Why it matters: Thin slices let the radiologist answer the question without bringing the patient back.
Sources: Geleijns 2014; Abbara 2016; Silverman 2019; Uppot 2018
Contrast too early or too late Can mimic pathology
- What you see
- Arteries poorly opacified on an arterial study, veins already bright, or organs not enhanced in the phase requested.
- Why it happens
- When contrast arrives and peaks depends on cardiac output, body size, the injection site and rate, and disease, so a fixed delay cannot suit every patient. On fast scanners a short scan can outrun the bolus.
- Try first
- Use bolus tracking or a test bolus as your protocol sets.
- Check the injection actually ran as planned: the injector record, the cannula site and the volume given.
- Match the injection duration to the scan time; for very short scans, injection length matters.
- Use a saline chaser as the protocol sets.
- On a dual-energy scan, low-energy monoenergetic images can improve poor arterial opacification.
- Stop and escalate when
- Contrast is not arriving: stop and check for extravasation. A missed phase that could only be repeated with more contrast and dose: the radiologist decides.
- Can mimic pathology
- Poor opacification can leave a pulmonary embolus or a dissection impossible to exclude.
Explain it simply
The dye travels with the blood. The scan has to catch it in the right place, like taking a photo of a runner as they cross the line.
Why it matters: A badly timed scan may need repeating with more dye and more radiation.
Sources: Murphy 2018; Abbara 2016; Rau 2025; Bae 2010; ESUR 2025
Bolus tracking did not trigger, or triggered at the wrong time
- What you see
- Monitoring scans continue without reaching the threshold, the scan starts before contrast arrives, or the arteries are faint or the veins bright despite tracking.
- Why it happens
- Tracking measures the CT number in a region of interest and starts the scan after a set delay once a threshold is passed. A region on the wrong vessel or on the vessel wall, a threshold that does not suit the kV, an injection that did not run, or slow circulation can all upset it. Bolus tracking can itself be a source of error.
- Try first
- Place the region of interest in the centre of the target vessel, clear of the wall and of streaks (a general step; one study placed it centrally over a third to a half of the lumen).
- Watch the monitoring images: if no contrast arrives, check the injection site and the injector.
- Check the threshold suits the kV: a lower kV raises vessel CT numbers.
- The post-trigger delay must allow for table movement and the breath-hold instruction.
- If your protocol allows a manual start, start the scan when the vessel is clearly opacified (a general step; protocols differ).
- For repeated failures or complex anatomy, a test bolus is an alternative, and it also confirms the IV access.
- Stop and escalate when
- Contrast is not arriving at all: stop the injection and check for extravasation.
- Good to know
- Published thresholds and delays vary between protocols; use your own protocol’s values.
Explain it simply
The scanner watches one blood vessel and waits for the dye to make it bright. If it is watching the wrong spot, it waits for ever or starts too soon.
Why it matters: Watching the monitoring pictures yourself catches a failed injection before the patient has the full dose of radiation.
Sources: Abbara 2016; Murphy 2018; Wu 2024; Rau 2025; ESUR 2025
Dense contrast streaks from the injected vein or SVC Can mimic pathology
- What you see
- Bright and dark streaks from undiluted contrast in the subclavian or brachiocephalic vein or the superior vena cava, crossing the upper mediastinum, aortic arch or right pulmonary artery.
- Why it happens
- Undiluted contrast in the veins is dense enough to cause beam hardening and streaks.
- Try first
- Follow the contrast with a saline flush (chaser), which pushes the tail of the bolus on and reduces streaks.
- Inject through the right arm where the protocol allows: from the left arm, contrast crosses the mediastinum in the left brachiocephalic vein and causes more streaks.
- On a dual-energy scan, higher-energy monoenergetic images reduce these streaks.
- Can mimic pathology
- Streaks over the pulmonary arteries can look like an embolus or make the study indeterminate.
Explain it simply
Neat dye in the big vein near the heart is so dense that it throws stripes across the picture. A squirt of salt water behind it rinses the vein.
Why it matters: A stripe across a lung artery can look like a blood clot.
Sources: Murphy 2018; Abbara 2016; Wang 2025; Boas 2012
Kidney cyst looks enhanced (pseudoenhancement) Can mimic pathology
- What you see
- A small kidney cyst measures higher after contrast than before, enough to suggest enhancement; typically a small cyst inside the kidney, scanned at peak kidney enhancement.
- Why it happens
- Beam hardening and scatter from the brightly enhancing kidney around it, with partial volume, raise the cyst’s measured CT number.
- Try first
- Measure in the centre of the cyst, as far as possible from enhancing kidney.
- Use slices no thicker than half the lesion’s diameter, so partial volume is not added. Pseudoenhancement can still occur with thin slices.
- Tell the radiologist. The Bosniak 2019 update says masses inside the kidney under 1.5 cm next to avidly enhancing tissue can be affected, and may be too small to characterise.
- Dual-energy imaging reduces the effect but does not remove it.
- Can mimic pathology
- A simple cyst can look like a solid or cystic tumour.
- Where sources differ
- Mechanism: Boas and Fleischmann attribute it to beam hardening and scatter; Maki and colleagues found that beam hardening alone could not explain it and pointed to incomplete correction by the scanner. Reported rises are up to about 28 HU.
Explain it simply
The bright kidney around a small cyst throws off the scanner’s measurements, so the cyst seems to light up with dye when it has not.
Why it matters: A false reading could lead to unnecessary follow-up scans or worry.
Sources: Boas 2012; Maki 1999; Birnbaum 2002; Silverman 2019
Calcium or stents look bigger than they are (blooming) Can mimic pathology
- What you see
- Calcified plaque or stent struts look larger than they are, so the lumen looks narrower or darker.
- Why it happens
- Mainly partial volume from limited resolution, with motion adding to it; worse with denser calcium, smaller vessels, thicker struts and a narrow window.
- Try first
- Review with a wider window.
- Use the protocol’s sharp or dedicated kernel for stents or heavy calcium.
- On a dual-energy scan, high-energy monoenergetic images reduce blooming.
- A low kV exaggerates blooming; follow the protocol for patients with stents or heavy calcification.
- Can mimic pathology
- Blooming tends to overestimate a narrowing, though heavy calcium can also hide one.
- Where sources differ
- Why higher energy helps: Pack and colleagues argue that it works by compressing the CT number scale, much like widening the window, and that beam hardening adds little on modern scanners; Murphy and colleagues and the Kalisz review list higher energy as a remedy without that distinction.
Explain it simply
Bright specks of calcium look bigger than they are, like a torch that seems larger than its bulb.
Why it matters: An exaggerated narrowing could send a patient for an invasive test they do not need.
Sources: Pack 2022; Murphy 2018; Abbara 2016; Kalisz 2016
Smooth, blotchy or plastic look after iterative or deep-learning reconstruction Can mimic pathology
- What you see
- Images look unusually smooth or plastic, with a coarse, blotchy noise texture; small low-contrast lesions or small vessels look blurred.
- Why it happens
- Iterative reconstruction suppresses noise using a noise model and a preference for smooth images; at high strengths the remaining noise clumps into blotches. Deep-learning methods take their texture from their training data and can, rarely, add or remove small structures.
- Try first
- Check the reconstruction type and strength match the protocol.
- If the radiologist asks, reconstruct a lower strength or a standard reconstruction from the raw data.
- Do not lower the dose further because the images look smooth: at very low doses resolution falls and new artefacts appear.
- Can mimic pathology
- A blotchy texture can imitate pathology or hide small low-contrast lesions; deep-learning reconstruction can add structures that are not there or blur away small ones.
Explain it simply
Clever software can clean up a grainy picture, but if it cleans too hard everything looks like smooth plastic and small details get wiped away.
Why it matters: Smooth is not the same as accurate, and a missed small lesion matters more than a pretty picture.
Sources: Koetzier 2023; Boas 2012; Geleijns 2014; Geyer 2015
Anatomy cut off or wrong scan range
- What you see
- The organ of interest is not fully covered, for example the lung bases or apices missing, or the scan runs well beyond the requested range.
- Why it happens
- The range is planned on the localiser. A different breath between the localiser and the scan, and planning errors, move the boundaries.
- Try first
- Plan from the localiser with the margin your protocol specifies, and give the same breath-hold instruction for the localiser and the scan.
- Check coverage on the images before the patient leaves the table.
- If anatomy is missing, ask the radiologist whether to scan only the missing part.
- Check the reconstructed field of view has not cropped anatomy out.
- Good to know
- In a large study of low-dose chest CT, the lung bases were cut off in about 4% of scans, and scanning beyond the lungs was common.
Explain it simply
The scan’s start and end are set on a quick first picture. If the patient breathes differently for the real scan, the lungs may sit in a different place.
Why it matters: Missing a strip of lung means calling the patient back for more radiation; scanning too much gives radiation for nothing.
Sources: Geleijns 2014; Wienholt 2026; Uppot 2018; Abbara 2016
Equipment and system
Faults that follow the scanner rather than the patient. Record what you saw, the time, the message and what you were doing, in the fault log, and report a malfunction straight away. After a major repair, the scanner is tested before the next patient, at least with the daily water phantom test. Steps marked as general fault-finding come from no published source: your operator manual and service engineer take precedence.
Sources: IAEA SSG-46; HSE L121; ACR CT QC 2017
Daily water phantom out of limits
- What you see
- On the daily water phantom, the mean CT number or the noise is outside the limits set for your scanner, or the images show rings, streaks, lines or a darker or brighter centre.
- Why it happens
- The daily water test is sensitive to many problems: calibration drift, a faulty detector element, tube or software faults. The medical physicist sets the action limits for each scanner and protocol.
- Try first
- Double-check the basics: the right phantom, centred at the isocentre with the lights and phantom marks, the right image, region of interest and protocol.
- Run the air calibrations if the manufacturer recommends them, then repeat the phantom scan.
- Lines or streaks: inspect the gantry window for spilt contrast, clean it as the manufacturer allows and rescan.
- Record the results and every action in the QC log, where the QC team and service engineer can find them.
- Stop and escalate when
- Still outside the limits after a repeat, or rings persist after several air calibrations: tell the medical physicist, who helps decide whether service is needed before clinical scanning. The radiologist decides whether patient studies continue or are postponed.
- Where sources differ
- How often and how tight. The ACR manual and AAPM Report 74 make water CT number, noise and artefacts a daily technologist test. The IAEA programme (Human Health Series 19) lists them as a monthly radiographer test, with acceptable limits of ±5 HU for CT number, ±25% of baseline for noise and ±10 HU for uniformity. Follow your local QC programme.
- Good to know
- The ACR manual’s limit for the water mean on the manufacturer’s phantom is 0 ± 5 HU, or the manufacturer’s own specification; noise limits are set by the medical physicist. QC images are checked before the first clinical scan of the day.
Explain it simply
Every morning the scanner images a bottle of water. Water should always measure the same, so a wrong number warns that the machine has drifted.
Why it matters: A fault caught on the phantom is fixed before it affects the day’s patients.
Sources: ACR CT QC 2017; IAEA HHS 19; AAPM Report 74; Geleijns 2014
Warm-up or air calibration due, or calibration failed
- What you see
- The scanner asks for a tube warm-up or an air (detector) calibration, the calibration stops with an error, or the start-of-day routine has not been run.
- Why it happens
- Air calibrations correct for small differences in sensitivity between detector elements; without them, rings appear. They are usually needed once a day, sometimes more often or when the system flags them. Tube warm-up is run as the manufacturer recommends before the daily phantom test.
- Try first
- Run the warm-up and air calibrations as the manufacturer recommends, before the daily phantom and the first patient.
- Clear the gantry and table as the calibration requires, then repeat it once if it fails, following the on-screen prompts (a general fault-finding step).
- Check the daily phantom afterwards for rings or streaks.
- Stop and escalate when
- Calibration fails again, or rings persist after several calibrations, or calibration is not a user task on your system: call service and tell the medical physicist.
Explain it simply
Thousands of tiny sensors each see the X-rays slightly differently. Calibration measures the beam with nothing in it so the computer can even them out.
Why it matters: Skipping calibration can put rings on every patient’s scan.
Sources: IAEA HHS 19; ACR CT QC 2017; Geleijns 2014
Tube heat warning or cooling wait
- What you see
- The console warns that the tube is too hot, enforces a cooling wait, or will not allow the planned tube current, scan length or number of phases.
- Why it happens
- CT tubes run at high power for long periods, and the heat builds up in the anode and the tube housing. The generator calculates loading and the cooling time needed from the tube’s rating data. Long, high-current or repeated scans can exceed the heat capacity; the room and heat exchanger also need effective air conditioning.
- Try first
- Follow the console prompt and let the tube cool. Repeated high-power exposures without the required cooling time can damage the anode and bearings.
- If the patient cannot wait, ask the radiologist whether the protocol can be adapted: fewer images, thicker slices, lower tube current or a higher pitch reduce the heat load, at some cost in image quality.
- Check that the scanner room and equipment room air conditioning are working.
- Stop and escalate when
- Cooling waits on a normal workload, repeated tube errors, or a cooling-system message: call service and record it in the fault log.
Explain it simply
Making X-rays turns most of the electricity into heat. The tube has to cool down before it can work hard again, like an oven after a long bake.
Why it matters: Respecting the wait protects a very expensive tube and keeps the scanner available for emergencies.
Sources: Nowotny 2014; Geleijns 2014; IAEA HHS 19
Positioning lights misaligned
- What you see
- The alignment lights do not match the scan plane or the position set on the localiser, a marker placed at the light appears off the expected slice, or a light is dim or out.
- Why it happens
- The internal and external lights mark the scan plane for setting up. They are a rough guide; the localiser sets the exact range. Accurate lights matter for biopsy, for centring, for keeping the eyes out of brain scans and for radiotherapy planning.
- Try first
- Set the exact scan range on the localiser rather than relying on the light.
- Check the lights with a thin marker as your QC programme describes (the IAEA test uses a wire of about 1 mm).
- If the internal and external lights disagree, find which one matches the scan plane, and report the other.
- Keep the laser out of the patient’s eyes, and take particular care with unconscious or anaesthetised patients, who have no blink reflex.
- Stop and escalate when
- The error is beyond your programme’s limit, or the scanner is used for biopsy or radiotherapy planning: report it to service and the medical physicist and record it.
- Where sources differ
- Tolerance. The ACR manual’s annual physicist test calls for service if the light or localiser position is out by more than 2 mm; IEC uses ±2 mm (as reported by AAPM TG-233); the IAEA programme’s monthly radiographer test accepts ±5 mm, with ±1 mm achievable. Follow your local QC programme.
Explain it simply
The red lines on the patient show roughly where the scan will start. If they are out of line, a needle or a treatment plan could be in the wrong place.
Why it matters: For a biopsy or radiotherapy, a few millimetres can matter.
Sources: IAEA HHS 19; ACR CT QC 2017; AAPM TG-233; AAPM Report 74
Gantry, tilt or unusual noise
- What you see
- The gantry will not rotate or tilt, the tilt indicator reads wrongly, a gantry error appears, or there is a new grinding, knocking or whining noise.
- Why it happens
- The gantry rotates the tube and detectors at high speed under strong centrifugal force. Tilt is set and displayed by the gantry, and its indicator is checked in the QC programme. Most multislice scanners allow a tilted gantry only for axial scans, not helical ones (IAEA).
- Try first
- Stop the scan if it is running, note the message, the time and what was happening, and record it in the fault log.
- Report the fault straight away; do not keep scanning through a new mechanical noise (a general step; your operator manual takes precedence).
- Stop and escalate when
- Any mechanical fault, noise or error that recurs: report it and call service; service and the radiologist decide whether the scanner stays in use, and it is tested after repair. If the tilt indicator is in doubt, do not rely on tilted scans or plan radiotherapy until the physicist has confirmed the gantry returns to vertical.
Explain it simply
The ring inside the scanner spins several times a second with heavy parts on it. A new noise can be the first sign of a worn part.
Why it matters: Catching a mechanical fault early avoids a breakdown in the middle of an emergency scan.
Sources: Geleijns 2014; IAEA HHS 19; IAEA SSG-46; HSE L121; ACR CT QC 2017
Table will not move, moves inaccurately, or weight limit
- What you see
- The table will not move, stops or creeps, the position readout is wrong, or the patient may be near or above the table’s weight limit.
- Why it happens
- The table must carry heavy patients without bending, and its travel accuracy is tested in the QC programme. Weight and bore limits are specific to each system, and exceeding the manufacturer’s limit can damage the table mechanics.
- Try first
- Check nothing is caught: lines, injector tubing, cables or blankets (a general step). Leave enough tubing slack for the table to travel.
- Check the patient’s weight against the limit posted for your scanner, and the girth against the opening with the table at scanning height.
- Note any error message and record the fault.
- Stop and escalate when
- The table creeps, stops, or travels inaccurately: stop using it and call service, especially before biopsy or radiotherapy planning. A patient above the system’s limit: do not scan on that table; ask the radiologist about another scanner or test.
Explain it simply
The table moves the patient through the scanner to the millimetre. If it slips or is overloaded, the scan lands in the wrong place or the table is damaged.
Why it matters: Accurate table travel is what makes measurements and needle guidance trustworthy.
Sources: ACR CT QC 2017; Geleijns 2014; Uppot 2018; ACR Contrast Media 2026
Warning lights, door or emergency stop
- What you see
- The X-ray-on warning light at the room entrance does not light, a door is open while the beam is on, an emergency stop has been pressed or is blocked, or the intercom or view of the patient fails.
- Why it happens
- Warning lights and signs show when X-rays are being produced; they do not prevent exposure. Doors are kept shut while the beam is on, and the radiographer must be able to see and talk to the patient throughout.
- Try first
- Report a failed warning light, sign or intercom at once and follow your local rules until it is fixed.
- Keep doors shut when the beam is on, and keep emergency stops reachable and unobstructed.
- Do not test the emergency off button yourself; it is tested with the supplier’s engineer present.
- After an emergency stop, record what happened; the scanner needs resetting before the next scan.
- Stop and escalate when
- Warning lights, communication or the view of the patient do not work: tell your radiation protection adviser or medical physicist; local rules decide whether scanning continues.
- Where sources differ
- Door interlocks. The IAEA (SSG-46) says door interlocks are not appropriate in diagnostic X-ray rooms, because stopping the beam may mean repeating the procedure, and its CT QC programme says any door interlock should not shut down the scanner when someone enters. The UK Approved Code of Practice describes interlocks for high-dose-rate enclosures. Your local rules decide.
Explain it simply
The light outside the door tells people not to walk in while X-rays are on. If it is broken, someone could walk into the room during a scan.
Why it matters: Working warning systems protect staff and visitors as well as patients.
Sources: IAEA SSG-46; IAEA HHS 19; HSE L121; AAPM TG-233; 21 CFR 1020.33
Dose notification or alert value triggered
- What you see
- Before the scan, the console shows a notification or an alert: the planned dose index (CTDIvol or DLP) would exceed a value set on the scanner.
- Why it happens
- Scanners built to the NEMA XR-25 dose check standard compare the planned dose indices with values the site sets. A notification value applies to each scan in a protocol and flags a higher-than-usual dose; an alert value applies to the whole examination and flags when the accumulated dose index at any position would exceed it, even if no single scan triggered anything. These values are not limits, targets or diagnostic reference levels. The scanner cannot allow for patient size, so a large patient may appropriately exceed a notification value.
- Try first
- Stop and compare the kV, tube current or noise setting, pitch, detector configuration and scan range with the protocol, and correct anything wrong.
- Check that the protocol matches the patient (adult or child, body part) and that the patient is centred.
- Notification: if the settings are correct, confirm and proceed, entering a reason if your site asks for one (for example, patient size).
- Alert: this calls for a closer review. Your name is recorded, a password may be required, and your site procedure says who reviews before proceeding (for example the lead radiographer, the radiologist or the medical physicist).
- An alert should not stop a medically necessary scan; if a password is set, someone who knows it must always be available.
- Stop and escalate when
- You cannot find why the dose is higher than expected, or notifications keep firing on one protocol: tell the lead radiographer and the medical physicist; the event log is reviewed.
- Good to know
- The AAPM publishes starting values for notification values and says each facility adjusts them with a medical physicist; it reports an FDA-suggested alert value of 1000 mGy CTDIvol. In the United States, CT on equipment without the NEMA XR-29 dose-optimisation attributes attracts a Medicare payment reduction.
Explain it simply
Before the scan, the computer checks whether the planned radiation is higher than usual and asks you to double-check. It is a seatbelt warning, not a speed limit.
Why it matters: The check catches a wrong setting before the patient receives the dose, which is when it can still be fixed.
Sources: AAPM Dose Check 2011; AAPM Notification Slides 2014; CMS MM9250
Injector alarms or will not link to the scanner
- What you see
- The injector will not arm, stops on a pressure or air alarm, shows an error on the contrast or saline head, or will not link to the scanner so the injection and scan cannot start together.
- Why it happens
- Injectors stop at their pressure limit when something narrows the path: a kink, a closed clamp, a cannula too small for the flow rate, or viscous contrast.
- Try first
- Check the patient first: the cannula site for pain or swelling.
- Check the line for kinks and closed clamps, and that the cannula size suits the protocol’s flow rate.
- Re-prime and clear air from the syringe and tubing as the injector’s instructions describe, then turn the syringe so the tubing points downwards. Do not override an air or pressure alarm.
- Iodinated contrast warmed to body temperature is less viscous, which lowers injection pressure; follow your warming policy (sources differ, below).
- If the injector will not link to the scanner, follow the injector manual; start manually only as your protocol allows (a general step).
- Stop and escalate when
- The injector alarms with a clear line and good access, or a head or the link keeps failing: take it out of use and report it.
- Where sources differ
- Warming iodinated contrast. ESUR (2025) regards warming as widely accepted good practice; RANZCR says warming lowers viscosity and injection pressure but there is not enough data to say it reduces reactions; the ACR (2026) finds the evidence insufficient to recommend routine warming of any agent, noting some, inconsistent, evidence for higher-viscosity agents, and says warming may be considered to lower viscosity and injection pressure in certain higher-risk patients. These statements are about iodinated contrast only.
- Good to know
- Air detection and dual-head error handling are specific to each injector: your injector’s manual takes precedence.
Explain it simply
The injector is a pump that pushes dye into a vein. It stops on purpose when it meets resistance or air, to protect the patient.
Why it matters: Overriding an alarm could push air or dye where it should not go.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
ECG gating not triggering
- What you see
- The ECG trace is noisy, flat or low, the scanner misses R waves or triggers on T waves, or the heart rate is irregular so triggering is erratic.
- Why it happens
- Gated scans time the acquisition from the R wave. Poor skin contact, leads in the wrong place, a T wave nearly as tall as the R wave, or an irregular rhythm upset detection.
- Try first
- Clean the skin with alcohol, shave if needed, and use fresh electrodes.
- Reposition the leads if the baseline is noisy or the T wave is as tall as the R wave; keep them outside the scan field.
- Check the cable and connections to the monitor (a general step).
- For an irregular rhythm, use the protocol’s retrospective gating or single-beat option if the scanner has one.
- After the scan, ECG editing (moving trigger points, removing ectopic beats) can often rescue the study.
- Stop and escalate when
- The trace stays poor with new electrodes and leads on more than one patient: report the ECG module to service (a general step).
Explain it simply
The scanner listens to the heartbeat to know when to take each picture. If it cannot hear the beat clearly, it takes the pictures at the wrong moments.
Why it matters: A clean ECG trace is the difference between a usable heart scan and a repeat.
Sources: Abbara 2016; Kalisz 2016; Murphy 2018
Images stuck in reconstruction or not appearing
- What you see
- The scan finishes but images do not appear, the reconstruction queue stalls, or the console freezes.
- Why it happens
- Images are calculated from the raw projection data after acquisition, by reconstruction hardware and software that can stall.
- Try first
- Note the time, the series and any message; photograph the screen if useful.
- Check the reconstruction or job queue on the console.
- Ask service before restarting the console or deleting anything while data are held in the queue (a general step).
- If the patient is still on the table and the images are needed now, ask the radiologist whether to repeat the scan, as a repeat is another exposure.
- Stop and escalate when
- The queue does not clear: call service.
Explain it simply
The scanner records raw measurements first, then does a lot of maths to turn them into pictures. If the maths gets stuck, the pictures do not appear.
Why it matters: Clear notes help service find the fault, and waiting for advice can save the patient a repeat exposure.
Sources: Geleijns 2014; Clunie 2014; IAEA SSG-46
Images not on PACS, or patient details wrong
- What you see
- A study is complete on the scanner but missing or incomplete on PACS, the worklist does not show the patient, or details on the images are wrong.
- Why it happens
- The scanner sends images to a configured destination, which confirms storage (storage commitment) before the scanner deletes its copies. Patient details come from the modality worklist.
- Try first
- Check the send queue and the destination; resend failed series.
- Compare image counts on the scanner and on PACS.
- Do not delete the study from the scanner until PACS has confirmed storage.
- If the worklist is empty, widen the search (date, station) and check the booking; enter details by hand only under your downtime procedure.
- If details on the images are wrong, follow your correction process, so changes are made by authorised staff with an audit trail.
- Stop and escalate when
- Studies are still missing after a resend, or a patient identity error is found: contact your PACS or IT team, and report identity errors through your incident process.
Explain it simply
The scanner posts the pictures to the hospital’s picture library and waits for a receipt before throwing away its own copy.
Why it matters: Lost or mislabelled images can mean a missed result or a result given to the wrong person.
Sources: Clunie 2014; DICOM PS3.4
Patient and workflow
Most of these problems are solved before the scan: screening questions, a clear explanation, and a plan agreed with the radiologist when the answer is not routine. The patient and the radiographer must be able to communicate before, during and after any injection.
Sources: ACR Contrast Media 2026; RANZCR ICM 2018
Checking it is the right patient
- What you see
- The identity of the patient on the table needs confirming before scanning: a patient who cannot answer, a child, a common name, a language barrier, or a transfer from another area.
- Why it happens
- Scanning the wrong patient or the wrong body part is always possible, for example with similar names or patients unsure of their procedure. Several independent ways of confirming identity, and checking the request and the side, prevent it.
- Try first
- Ask the patient to state their full name, date of birth and what they are here for, including the site and side, rather than asking “Are you …?”. Match the answers to the request and the identity band.
- For a patient who cannot answer, use their representative, the identity band or the accompanying staff, as your policy describes.
- Immediately before scanning, stop and confirm the right patient, examination and site or side (an internal time-out when you work alone).
- Check the patient details and side markers on the images before they are released.
- Stop and escalate when
- Any mismatch: do not start until it is resolved, and report it through your incident process.
Explain it simply
Before a scan, staff check the name, date of birth and the reason for the scan, like checking a parcel’s address before delivering it.
Why it matters: Scanning the wrong person gives them radiation for nothing and can lead to the wrong treatment for two people.
Sources: ACSQHC CT/MRI protocol; IAEA SSG-46; ARPANSA RPS C-5; ORS C1 2024; IR(ME)R 2017
Patient position entered wrongly at the console
- What you see
- Images carry the wrong left and right, or reformats look mirrored, because the position selected at the console (head or feet first, supine or prone) does not match how the patient lies.
- Why it happens
- The patient can lie head first or feet first, supine or prone, and the position is usually recorded with the scan data. Schmidt and Odland describe two cases of mirror-image coronal scans: orientation depends on whether the patient is prone or supine, so it relies on the radiographer’s input, and in the largely symmetrical head and neck a mislabelling may not be noticed.
- Try first
- Before the localiser, check that the position selected at the console matches how the patient is lying (a general step).
- Before images are released, check the patient details and side markers against the patient and the request.
- If the position was entered wrongly, do not alter images yourself: tell the radiologist and correct the record through your authorised correction process.
- Stop and escalate when
- Images with wrong labels have already been released: tell the radiologist and the referrer straight away (a general step) and report it through your incident process.
Explain it simply
The scanner only knows which way round the patient is lying because someone tells it. If that is entered wrongly, left and right are swapped on every picture.
Why it matters: In one of the two cases Schmidt and Odland describe, a mirror-image scan led to an operation on the wrong side.
Sources: Geleijns 2014; Schmidt 2004; ACSQHC CT/MRI protocol; Clunie 2014
Pregnancy uncertain or confirmed
- What you see
- The patient might be pregnant, is not sure, or the screening answer is unclear before a CT that includes the abdomen or pelvis; or a pregnant patient is booked.
- Why it happens
- Pregnancy status is established before exposures that could give a significant dose to an embryo or fetus, mainly when the abdomen or pelvis is in the beam. A patient may not know she is pregnant. The ACR-SPR practice parameter says CT outside the abdomen and pelvis (with the possible exception of the hip) generally does not need verification.
- Try first
- Ask every patient of childbearing potential, privately where possible, using your department’s questions or form, immediately before the scan. A status written on the request is not enough on its own. Record the question and the answer.
- If pregnancy cannot be excluded for a scan that includes the abdomen or pelvis, tell the radiologist before scanning; a pregnancy test or a later appointment may be chosen.
- If the patient is pregnant and the scan is justified, the radiologist and medical physicist adapt the protocol: only the phases and range needed, at the lowest suitable dose.
- A critically urgent scan goes ahead without waiting; record the circumstances and who directed it.
- Iodinated contrast, if the radiologist decides it is needed: the ACR does not withhold it when needed, while ESUR gives it in pregnancy only in exceptional circumstances. ESUR and RANZCR ask for the newborn’s thyroid check, which routine newborn screening (the heel-prick test in Australia and New Zealand) provides; the ACR needs nothing beyond routine screening.
- Follow your local policy on shielding: the ACR-SPR no longer recommends shielding the pelvis during CT of other regions, as it can increase internal scatter; the IAEA (2018) still says to shield radiosensitive organs when appropriate.
- Stop and escalate when
- A positive test, or she says she is or may be pregnant: the radiologist decides before scanning.
- Where sources differ
- UK (IR(ME)R 2017; HPA 2009 and RCR 2026 guidance): the operator asks about pregnancy and breastfeeding as the employer’s procedure defines; for locally defined high-dose procedures (for example above about 10 mGy fetal dose, such as CT of the lower abdomen and pelvis), HPA gives two options: book within the first ten days of the cycle, or book normally and rebook patients in the second half of the cycle in whom pregnancy cannot be excluded. Australia (ARPANSA RPS C-5): establish status before any procedure likely to give more than 1 mSv to the uterus; if a pregnant patient proceeds, a medical physicist estimates the fetal dose. New Zealand (ORS C1): ascertain status before any procedure that could give a significant fetal dose; signs ask patients to tell staff.
Explain it simply
Babies in the womb are more sensitive to radiation, so staff ask about pregnancy before scanning the tummy. If the answer is yes or maybe, a doctor decides.
Why it matters: Asking every time, immediately before the scan, reduces unplanned exposures of an unborn baby.
Sources: ACR-SPR Pregnancy PP; HPA RCE-9 2009; RCR IR(ME)R 2026; IR(ME)R 2017; ARPANSA RPS C-5; ORS C1 2024; IAEA SSG-46; ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Pacemaker, neurostimulator, insulin pump or sensor in the scan
- What you see
- The patient has an implanted or worn electronic device in or near the scan range: pacemaker or ICD, neurostimulator, insulin pump, or a continuous glucose sensor.
- Why it happens
- When the X-ray beam passes directly over a device, brief interference is common in laboratory testing (transient oversensing, mostly inhibited pacing, in 17 of 20 pacemakers and defibrillators at typical doses), but the FDA says clinically significant problems, such as unintended neurostimulator shocks or insulin pump faults, are rare. Interference happens only when the device is in the direct beam; localisers are not linked to interference. A device should not stop a medically necessary CT.
- Try first
- Ask about devices, and use the localiser to see where the device sits relative to the scan range.
- Keep the device out of the scan range if possible, after confirming with the radiologist that the anatomy needed is still covered.
- Worn devices: most manufacturers recommend removing them and leaving them outside the room. Decide with the patient, as pausing an insulin pump has its own risk: check glucose before and after, and remind them to restart it.
- Neurostimulator: if your local policy and the device team allow it to be switched off, the patient switches it off before the scan and back on afterwards.
- If the device must be in the beam: use the lowest dose consistent with diagnostic quality and a lower dose rate (the FDA suggests a longer rotation time or lower pitch, if appropriate). Avoid dwelling over the device for more than several seconds (CT perfusion, CT fluoroscopy) unless it is off or the scan is urgent and monitored.
- Be ready to stop the scan for symptoms: light-headedness, palpitations, pain, shocks or device alarms.
- Stop and escalate when
- Symptoms or a device alarm during the scan: stop, assess the patient and get medical help. Afterwards the patient checks the device is on with the right settings; a suspected malfunction is reported, and the device team checks it at the next scheduled visit or sooner.
- Where sources differ
- Manufacturers’ labels differ. Medtronic (MiniMed 780G, US guide) says remove the pump and sensor before entering a CT room; Tandem says do not expose the pump or CGM to CT; Insulet says remove the Omnipod before CT; Dexcom says do not wear the G6 for CT and, for the G7, avoid the sensor in the scanned area; Abbott says FreeStyle Libre 2 and 3 sensors can be worn during CT in the United States, while some of its other regional pages still say remove. New Zealand and Australian labels for these devices could not be verified, so check the local labelling for the patient’s device.
Explain it simply
Strong X-rays hitting a tiny computer chip can occasionally confuse it. Keeping the device out of the beam, or switching it off, avoids the problem.
Why it matters: The scan is still important, so the aim is to protect the device without cancelling the test.
Sources: FDA CT devices; FDA PPHN 2008; McCollough 2007; Medtronic MiniMed 780G UG; Tandem ISI; Insulet Omnipod ISI; Dexcom FAQ; Abbott FreeStyle Libre
Poor IV access, cannula size and flow rate
- What you see
- Cannulation is difficult, only a small or distal vein is available, the protocol flow rate looks high for the cannula, or the patient has a central line, PICC or port.
- Why it happens
- Power injection needs a flexible plastic cannula in a vein large enough for the flow rate. Small, fragile or distal veins (hand, wrist, foot, ankle) and patients who cannot report pain carry a higher risk of extravasation.
- Try first
- Use a flexible plastic cannula rather than a metal needle or butterfly. An antecubital or large forearm vein is preferred.
- Match the flow rate to the cannula size as your protocol sets it, and reduce the rate for hand or wrist veins where feasible.
- Confirm placement with backflow (not always seen) and a saline test flush, watching the site where you can.
- Leave enough slack in the tubing for the table to move without pulling on the line.
- Central lines, PICCs and ports: power inject only devices labelled for it, within the manufacturer’s flow-rate limit. Small-bore lines inserted in the arm are not power injected unless the manufacturer permits it, because they can break.
- Ask the patient to tell you at once about pain or swelling during the injection.
- Stop and escalate when
- Pain or swelling on the test flush or during injection: stop and go to Contrast extravasation. No access suits the protocol’s flow rate: ask the radiologist whether to change the protocol.
- Where sources differ
- Cannula size and flow rate. The two relevant chapters of the ACR manual give different figures for the same small cannula; ESUR and RANZCR give no figures and say to match the cannula to the vein and the planned flow rate. Follow your department’s cannula and injector policy.
Explain it simply
The dye goes in fast through a soft plastic tube in a vein. A small or fragile vein may not cope with the speed and can leak.
Why it matters: The right vein and cannula prevent painful leaks and give the bright, well-timed pictures the radiologist needs.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Kidney function questions before IV contrast
- What you see
- Contrast is requested for a patient with kidney disease, diabetes, a single kidney, dialysis or metformin, or there is no recent kidney function result.
- Why it happens
- A fall in kidney function after contrast can have many causes; injury caused by intravenous contrast itself is now thought to be rare. The risk concentrates in patients with acute kidney injury or a low eGFR.
- Try first
- Ask your department’s screening questions. The ACR lists a history of kidney disease (including dialysis, kidney surgery or ablation) and metformin, with diabetes optional; RANZCR asks about kidney disease, diabetes and metformin. Age alone is not a reason to test.
- Check how recent the eGFR must be under your protocol (sources differ, below). In acute kidney injury the eGFR is unreliable.
- Do not delay emergency imaging, such as for stroke, bleeding or trauma, while waiting for a kidney result.
- eGFR below 30 mL/min/1.73 m² or acute kidney injury: tell the radiologist. The guidelines treat this as a relative, not absolute, contraindication; the decision weighs benefit against risk, and IV hydration may be ordered.
- Dialysis: patients who no longer pass urine and have no working transplant are not at risk of kidney injury from contrast; patients who still pass urine are treated as higher risk to protect what function remains. Dialysis is not started or rescheduled because of contrast.
- Give the dose the protocol sets (sources differ, below).
- Stop and escalate when
- eGFR below 30, acute kidney injury, a dialysis patient who still passes urine, or no result for a patient your protocol says must have one: the radiologist decides before contrast is given.
- Where sources differ
- ACR: with a stable eGFR of 45 or more, IV contrast is not an independent risk; at 30 to 44 the risk is rare; below 30 studies disagree, and 30 is the threshold with the best evidence; there is no agreed interval between test and scan. ESUR: risk below 30 for IV contrast, but below 45 for intra-arterial first-pass contrast and for intensive care patients; eGFR within 7 days for inpatients or acute illness and within 3 months for others, with a recheck at 48 hours in patients at risk. RANZCR: above 45 the risk is likely non-existent, 30 to 45 very likely low or non-existent, and the interval between test and scan is clinical judgement. Dose: the ACR advises against cutting the contrast dose to protect the kidneys; ESUR asks for the lowest dose consistent with a diagnostic result.
Explain it simply
The kidneys clear the dye from the blood. Most people’s kidneys cope easily, but very weak kidneys need a doctor to weigh the benefit of the scan first.
Why it matters: Asking the right questions protects the few patients at real risk without delaying everyone else’s scan.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018; ACR-NKF 2020
Patient takes metformin
- What you see
- A patient booked for IV contrast takes metformin, on its own or in a combination tablet.
- Why it happens
- Metformin does not raise the risk of kidney injury after contrast. The concern is a build-up of metformin if kidney function falls afterwards, so the rules depend on the eGFR and on acute kidney injury.
- Try first
- Check the latest eGFR and whether there is acute kidney injury.
- Follow your protocol’s metformin rule (the guidelines differ in detail, below).
- If metformin is to be stopped, make sure the patient and the referrer know who rechecks kidney function and when the tablets restart.
- Stop and escalate when
- eGFR below 30, acute kidney injury, an eGFR that is not known, or an intra-arterial study: ask the radiologist before contrast.
- Where sources differ
- ACR (chapter last revised 2014): eGFR 30 or more with no acute kidney injury, no need to stop; eGFR below 30, acute kidney injury, or arterial catheter studies that might send emboli to the kidneys, stop at or before the procedure for 48 hours and restart only once kidney function is re-checked and normal. ESUR (2025): continue if eGFR is above 30 with no acute kidney injury; stop from the time of contrast if eGFR is below 30, for any first-pass intra-arterial injection, or in acute kidney injury; check eGFR within 48 hours and restart if it has not changed significantly. RANZCR: for IV contrast, continue above 30; stop for at least 48 hours if eGFR is below 30, not known, or the patient is unwell, and check eGFR before restarting.
Explain it simply
Metformin is a diabetes tablet cleared by the kidneys. If the kidneys slow down after the scan, the tablet could build up, so some patients pause it for a couple of days.
Why it matters: Following the rule avoids a rare but serious problem while letting most patients carry on as normal.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018; ACR-NKF 2020
Previous contrast reaction or premedication question
- What you see
- The patient says they reacted to contrast before, reports an “iodine allergy”, or arrives asking about premedication.
- Why it happens
- A previous allergic-like reaction to the same class of contrast is the strongest risk factor for another one. Shellfish or skin iodine allergy is not a significant risk factor.
- Try first
- Find out what happened: the agent’s name, the date, the symptoms, how severe it was and how it was treated.
- Physiologic reactions (warmth, nausea, an altered taste, a vasovagal faint) are not allergic, and a prior physiologic reaction is not a reason for premedication.
- Tell the radiologist before contrast. Options include another test, a scan without contrast, a different agent, or premedication.
- If premedication was prescribed, check it was taken as ordered, including the timing your guideline requires (sources differ, below). Doses are on your departmental protocol.
- Be ready for a reaction whatever was given: premedication does not replace preparedness. Keep the cannula in and observe as your policy directs.
- Record the exact agent in the patient’s record; “iodine allergy” alone is not enough to guide the next scan.
- Stop and escalate when
- A previous severe reaction, or a reaction despite premedication (a breakthrough reaction): the radiologist decides whether and how to proceed, ideally with allergy advice.
- Where sources differ
- A reaction to the other class (gadolinium): the ACR finds no cross-reactivity and does not recommend routine premedication for it; CAR/CSACI counts only a previous reaction to the same type of agent. RANZCR’s screening asks about previous significant allergic reactions to other substances as a risk factor, and ESUR refers patients who have reacted to agents of both classes to an allergy specialist. Premedication. ACR (2026): may be considered after an allergic-like or unknown-type reaction to the same class; regimens shorter than 4 to 5 hours have no evidence of efficacy; switching to a different agent may help more, and both together seem to work best. ESUR (2025): switch agent and refer to an allergist; premedication only for an emergency examination after a severe reaction. CAR/CSACI (2025): recommends against routine steroid premedication; switching to a chemically different agent is the single best protective measure. RANZCR (2018): premedication reduces the likelihood and severity of reactions, but there is no evidence that it reduces deaths from a breakthrough reaction; oral corticosteroid premedication must start at least 6 hours before the contrast study; use a different agent from the one used before.
Explain it simply
Someone who reacted to the dye before is more likely to react again. Using a different dye, and being ready to treat a reaction, are the main protections.
Why it matters: Knowing exactly which dye caused the problem lets the next team choose a better one.
Sources: ACR Contrast Media 2026; ESUR 2025; CAR/CSACI 2025; RANZCR ICM 2018
Breastfeeding after iodinated contrast
- What you see
- A breastfeeding patient asks whether she needs to stop feeding after the scan.
- Why it happens
- Less than 1% of the injected contrast reaches breast milk in 24 hours, and less than 1% of that is absorbed by the baby.
- Try first
- Tell the patient that the ACR, ESUR and RANZCR all say breastfeeding can continue.
- If she still prefers to pause, the ACR says she may stop for 12 to 24 hours and express and discard the milk; there is little value in stopping for longer.
- Where sources differ
- The ACR offers the 12 to 24 hour pause as the mother’s informed choice; ESUR and RANZCR say feeding continues normally with no need to discard milk.
Explain it simply
Only a tiny trace of the dye gets into breast milk, and the baby’s gut absorbs almost none of it.
Why it matters: Clear, consistent advice avoids interrupting feeding for no benefit.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Thyroid disease or planned radioiodine
- What you see
- The patient has an overactive thyroid, a history of thyroid storm, a toxic nodule or goitre, or is booked for a thyroid scan or radioiodine treatment.
- Why it happens
- The iodine in contrast can rarely trigger an overactive thyroid in people at risk, and it blocks thyroid uptake of radioactive iodine for weeks.
- Try first
- Ask about thyroid disease and any planned thyroid scan or radioiodine treatment.
- Known or suspected untreated hyperthyroidism, or thyroid storm now: tell the radiologist before contrast.
- Radioiodine treatment or a thyroid scan coming up: tell the radiologist and the referrer, because the interference lasts weeks (below).
- After contrast in a patient at risk, the referrer arranges follow-up; RANZCR says an overactive thyroid usually appears three to six weeks later.
- Stop and escalate when
- Acute thyroid storm (ACR: avoid iodinated contrast) or manifest hyperthyroidism (ESUR: do not give it): the radiologist decides.
- Where sources differ
- ACR: a history of hyperthyroidism alone is not a reason to restrict contrast; avoid it in acute thyroid storm; before radioiodine or thyroid imaging, the suggested gap is ideally 3 to 4 weeks for hyperthyroid and 6 weeks for hypothyroid patients. ESUR: do not give iodinated contrast with manifest hyperthyroidism; untreated Graves’ disease and multinodular goitre with autonomy are at risk; it gives no gap before radioiodine. RANZCR: test and treat known or suspected hyperthyroidism before contrast; thyroid isotope uptake is blocked for 8 weeks.
Explain it simply
The thyroid uses iodine, and the dye contains a lot of it. An overactive gland can be pushed further, and a radioactive-iodine test or treatment will not work properly for a while.
Why it matters: A quick question avoids a delayed thyroid problem or a wasted thyroid treatment.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Cannot raise the arms
- What you see
- For a chest or abdomen scan, the patient cannot raise one or both arms above the head: injury, pain, a stiff shoulder, reduced consciousness, or lines and drains in the way.
- Why it happens
- Arms beside the body add the most attenuating paths through the scan, causing streaks and noise, and tube current modulation raises the dose to get through them. Modulation is set from the localiser, so the arm position at the localiser is what counts.
- Try first
- Raise the arms before the localiser, not after it: raising them after an arms-down localiser did not lower the dose in a trauma study.
- If both arms cannot go up, one arm up is better than none.
- If the arms must stay down, rest the forearms crossed over the front of the abdomen rather than at the sides.
- Keep the arms in the same position for the localiser and the scan; if they moved, repeat the localiser.
- With a suspected shoulder injury, or in an unstable patient where time matters, arms down is accepted (ESER).
- Good to know
- In one trauma study with automatic exposure control, effective dose was 18% higher with one arm down and 45% higher with both arms down than with both arms up (Brink and colleagues).
Explain it simply
Arms by the sides are extra thickness the X-rays must get through, so the scanner turns the power up and the picture gets streaky.
Why it matters: Getting the arms up before the first quick picture lowers the dose and gives a clearer scan.
Sources: Brink 2008; Graef 2022; ESER 2020; Boas 2012; IAEA SSG-46
Cannot hold their breath
- What you see
- The patient is breathless, cannot follow instructions or cannot hold their breath for the scan; images of the chest or abdomen show breathing motion.
- Why it happens
- Breathing during the acquisition blurs and doubles edges. Faster acquisitions leave less time for movement.
- Try first
- Explain and practise the breath-hold before the scan, using the same words for the localiser and the scan (a general step for routine scans; the SCCT recommends breath-hold practice for coronary CTA).
- Identify breathing limits before positioning and adapt where you can.
- Use the shortest rotation time and fastest acquisition the protocol allows.
- On a dual-source scanner, a high-pitch acquisition with free breathing has reduced breathing artefacts in breathless patients.
- If none of this works, scan during quiet breathing and tell the radiologist.
- Stop and escalate when
- Motion still spoils the region that answers the clinical question: ask the radiologist whether a repeat is justified.
Explain it simply
Holding your breath stops the chest and belly moving. If someone cannot, the scanner has to be fast enough to catch them between breaths.
Why it matters: A scan blurred by breathing may need repeating, which means more radiation.
Sources: Geleijns 2014; Abbara 2016; Bauer 2012; Varchetta 2021; Liu 2025; CAMRT BPG
Children and immobilisation
- What you see
- A baby or young child needs CT, or a child cannot keep still or follow breathing instructions.
- Why it happens
- Children are more sensitive to radiation and need protocols sized for them: adult protocols are inappropriate for children. Movement is the commonest reason a child’s scan fails.
- Try first
- Use the child-size protocol: kV and tube current adjusted to the child’s size, a single phase (usually enough), and only the region the question needs.
- Explain the scan to the child and the parent in words they understand, and coach older children through the breath-hold.
- Immobilise comfortably: straps, swaddling, head or body holders, foam pads, sponges, sandbags or pillows; vacuum mattresses help small children in chest CT.
- Use the fastest acquisition available; for chest CT in children under five, high-pitch fast protocols are preferred to reduce motion and dose.
- Avoid having anyone hold the child. If a carer must, they should not be pregnant and should not be a member of staff.
- Sedation or anaesthesia only under your local policy, with fasting instructions, a suitably trained team, and emergency equipment and drugs for the child’s size.
- Stop and escalate when
- The child cannot be kept still by these means: the radiologist and the referrer agree a plan, such as sedation, anaesthesia or another test.
Explain it simply
Children are smaller and more sensitive to radiation, so their scans use less of it. Keeping a small child still for a few seconds is the main challenge.
Why it matters: A child-sized scan done right the first time gives the lowest dose.
Sources: Image Gently CT; NCI Pediatric CT guide; ACR-SPR CT PP; IAEA SSG-46; ESR Essentials paediatric chest 2025
Large patient, table limit or scan field
- What you see
- The patient may be near or above the table’s weight limit, may not fit the bore, extends beyond the scan field, or the images are very noisy.
- Why it happens
- Weight and bore limits are specific to each system. Once the table is inside the gantry the vertical opening is smaller than the bore. The measured scan field is smaller than the bore, and extended-field reconstructions estimate what lies outside it. More tissue means fewer photons reach the detectors, so noise rises.
- Try first
- Confirm the weight and the girth, and compare them with the limits posted for your scanner; weight and girth matter more than body mass index.
- Check the fit with the table at scanning height, before the patient is moved onto it.
- Centre carefully and keep the body within the scan field; use the extended field-of-view reconstruction if available, and tell the radiologist the edge region is estimated.
- Use automatic tube current modulation; a higher kV, slower rotation or lower pitch may be used as your protocol allows; iterative reconstruction reduces noise.
- Record the weight and girth in the record so the next booking goes to a suitable scanner.
- Stop and escalate when
- The patient is above the system’s limits or does not fit: do not scan; ask the radiologist and referrer about a larger scanner or another test.
Explain it simply
Every scanner has a maximum weight and opening size. Bigger bodies also block more X-rays, so the picture needs more power and careful settings.
Why it matters: Checking limits first protects the patient, the staff and the scanner, and avoids a wasted trip.
Sources: Uppot 2018; Uppot 2007; IAEA HHS 19; Boas 2012
Language, hearing or other communication barriers
- What you see
- The patient does not share your language, is Deaf or hard of hearing, or cannot understand or respond to instructions such as breath-holds or reporting pain during the injection.
- Why it happens
- Screening, consent, breath-holds and the injection all depend on two-way communication. Patients who cannot report pain or swelling are at higher risk of extravasation.
- Try first
- Arrange a qualified interpreter before the appointment, in person or by a reliable remote service.
- Before the scan, agree how the patient will signal (call button, hand signal) and explain that they can be brought out at any time.
- Keep the patient in view and within talking distance from the control room throughout.
- For patients who cannot report pain, watch the injection site where you can and take extra care with cannula position.
- Stop and escalate when
- Screening or consent cannot be completed reliably: ask the radiologist before going ahead.
Explain it simply
A CT scan needs the patient to hold their breath on cue and to say if the injection hurts, so everyone needs a way to understand and be understood.
Why it matters: A patient who cannot tell you about pain at the cannula is more at risk of a leak.
Sources: CAMRT BPG; IAEA SSG-46; ACR Contrast Media 2026; HDC Code 1996; AAPM TG-233
Patient needs to stop mid-scan
- What you see
- The patient calls out, moves, sits up or asks to stop during the scan or injection.
- Why it happens
- Patients can be told before the scan that they can be brought out at any time. The operator can end the exposure at any point.
- Try first
- Stop the scan and the injection and talk to the patient straight away.
- Check the cannula site if the injection was running.
- Find the cause: pain, warmth from the contrast, nausea, anxiety, breathlessness or needing the toilet.
- Help the patient off the table only when they feel well, using the handling aids you need (see Fall).
- Agree with the patient and the radiologist whether to continue, and which part is still needed.
- Stop and escalate when
- Pain or swelling at the cannula, breathlessness, a rash or feeling faint after contrast: go to Extravasation or Contrast reaction.
Explain it simply
The patient can always ask to stop. Answering straight away keeps their trust and catches problems early.
Why it matters: Patients who trust you are more likely to finish the scan.
Sources: CAMRT BPG; 21 CFR 1020.33; IAEA SSG-46; ACR Contrast Media 2026
Safety incidents
These cards are reminders of what the guidelines say. In a real incident, follow your department’s emergency and incident procedures, your radiologist and medical physicist, and your regional rules, which take precedence over this page. Staff who give contrast are trained to recognise and treat reactions and in basic life support, and a defibrillator, oxygen and the minimum reaction drugs are kept in or near every room where contrast is injected.
Sources: ACR Contrast Media 2026; RANZCR ICM 2018; ESUR 2025
Acute contrast reaction
- What you see
- During or soon after iodinated contrast: hives or itching, flushing, swelling of the face or throat, hoarseness or stridor, wheeze or breathlessness, low blood pressure with a fast pulse, low blood pressure with a slow pulse, a seizure, or collapse.
- Do this now
- Stop the injection, keep the cannula in, and call for help early. Nearly all life-threatening reactions happen within 20 minutes of the injection.
- Assess quickly: general appearance, whether the patient can speak and how the voice sounds, breathing, pulse and blood pressure. Give oxygen by mask as your reaction card directs.
- For a severe reaction (anaphylaxis, stridor, wheeze with low oxygen, shock), call the resuscitation team. Adrenaline is the first-line drug; antihistamines and corticosteroids do not replace it.
- Lie the patient flat. Let them sit up if breathing is hard, but do not let them stand. Raise the legs if blood pressure is low. Low blood pressure with a slow pulse suggests a vasovagal reaction.
- A mild reaction can progress: observe the patient with the cannula in until it settles, for at least as long as your policy says.
- If your policy uses it, blood for serum tryptase is taken as soon as practical and within 4 hours of the start of the reaction (ESUR).
- Record the exact agent, the symptoms, severity, treatment and outcome; give the patient the agent’s name; report the reaction as your policy requires.
- Where sources differ
- Route of adrenaline. ESUR, RANZCR, CAR/CSACI and ASCIA give it into the muscle. The ACR also gives intramuscular doses, but its adult table prefers the intravenous route when blood pressure is low, because poorly perfused limbs may not absorb an intramuscular injection. Observation: after a mild reaction, the ACR says at least 20 to 30 minutes and ESUR at least 30 minutes; after a moderate reaction ESUR says 4 to 6 hours after treatment and admission after a life-threatening one; RANZCR says at least 4 to 6 hours after a severe reaction resolves, because biphasic reactions occur. After any injection, RANZCR keeps the cannula in and supervises for at least 15 minutes, or 30 minutes at increased risk. Route, dose and observation time come from your department’s card.
- Good to know
- Doses are on your department’s card and the site’s contrast reaction emergency card. Delayed skin reactions can appear from about an hour up to a week after the injection: tell the patient to contact their doctor if a rash appears.
Explain it simply
A few people react to contrast dye, from an itchy rash to a dangerous allergic reaction. Adrenaline is the key medicine for the serious kind, and help is called straight away.
Why it matters: The dangerous reactions come fast, usually within minutes, so staff stay close and ready.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018; CAR/CSACI 2025; ASCIA 2026
Contrast extravasation
- What you see
- Swelling, tightness, stinging or burning pain at the cannula site during or after injection; the injector alarms, or the monitoring scan shows no contrast arriving.
- Do this now
- Stop the injection. ESUR also says stop the scan.
- Tell the responsible doctor and mark the edge of the swelling (ESUR). A health care provider examines the limb: tenderness, swelling, redness, sensation, finger movement and circulation.
- Raise the limb above the level of the heart, and apply compresses as your policy directs (sources differ, below).
- Do not routinely aspirate through the cannula, inject hyaluronidase or steroids, or squeeze the area.
- Record the site, agent, volume, assessment and advice; tell the referrer about a symptomatic extravasation; record it in your incident system if your policy requires.
- Observe an outpatient until symptoms improve, and give clear written advice on where and when to seek help (ESUR: a patient leaflet), because a severe injury can develop several hours later.
- Stop and escalate when
- Get an urgent surgical opinion for severe pain, increasing swelling or pain, reduced capillary refill, a change in sensation, worsening movement, blistering or ulceration. ESUR also asks for a surgical opinion when a mild extravasation does not improve.
- Where sources differ
- Compresses. The ACR recommendation is cold compresses or ice packs at first rather than warm, although its text notes no clear evidence favouring either; ESUR (2025) uses ice packs; RANZCR allows cold or warm. Volume. The ACR says the volume alone should not trigger a surgical consultation; ESUR (2025) also recommends a surgical opinion for an extravasation of more than 150 mL. Follow your local policy.
Explain it simply
Sometimes the dye leaks into the skin instead of the vein. Most leaks settle, but some signs mean a doctor must check the arm urgently.
Why it matters: Serious damage can show up hours later, so patients go home knowing what to watch for.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Wrong patient, wrong region or wrong side scanned
- What you see
- You find that the person scanned was not the patient on the request, the wrong body part, side or examination was scanned, or images were stored under the wrong patient.
- Do this now
- Stop any further scanning, check on both patients and confirm who each one is, and tell the radiologist now.
- Keep the images and the dose report; do not delete anything. Write down what happened and when.
- Report it through your incident system as your local procedure directs. A medical physicist (medical physics expert) estimates the dose received.
- The referrer and the patient (or their representative) are told what happened, under your open disclosure or duty of candour policy (sources differ on when, below).
- Correct the patient records through your authorised correction process.
- Whether the regulator must be notified depends on your country’s criteria; your employer decides after the preliminary investigation. Do not guess thresholds yourself.
- Where sources differ
- The frameworks agree on investigating, estimating the dose and informing the referrer and patient; the detail differs. UK (IR(ME)R 2017 regulation 8, CQC guidance): the employer investigates and notifies the enforcing authority of exposures significantly greater than intended, with criteria that differ between England and the other nations; the duty to tell the patient covers clinically significant exposures (not defined in the regulations; RCR guidance helps employers decide), alongside the duty of candour. New Zealand (ORS C1): the licence holder notifies the Director of an overexposure as soon as practicable, investigates with dose estimates, and makes sure the referrer and patient are told of any unintended medical exposure; the HDC expects open disclosure usually within 24 hours. Australia (ARPANSA RPS C-5): the operator reports to the Responsible Person and the radiologist, a medical physicist estimates the dose, the radiologist informs the patient, and reports go to the state or territory regulator as it requires. IAEA SSG-46: investigation with a physicist’s dose estimate, the referrer notified, significant events reported to the regulator.
Explain it simply
If the wrong person or the wrong part is scanned, staff stop, look after both patients, work out how much radiation was given, and tell them openly.
Why it matters: Honest reporting finds the gap in the checks so it does not happen to the next patient.
Sources: IAEA SSG-46; IR(ME)R 2017; CQC SAUE 2024; RCR IR(ME)R 2026; ORS C1 2024; HDC Open Disclosure 2019; ARPANSA RPS C-5; ACSQHC CT/MRI protocol; Clunie 2014
Unplanned repeat or a higher exposure than intended
- What you see
- A scan needs repeating, or more was given than planned: the wrong protocol (such as an adult protocol on a child), wrong settings, scanning beyond the range, a failed injection or bolus tracking, movement, or an equipment failure.
- Do this now
- Before repeating, ask the radiologist whether a repeat is justified and over what range.
- Find and correct the cause first: settings, centring, protocol, injection or equipment.
- Record the reason and keep the dose report.
- Equipment fault, error or unusual behaviour: stop using the equipment until it is put right, record it and report it (see Scanner failure mid-scan).
- Report wrong settings, wrong protocols and exposures much greater than intended through your incident process; a medical physicist estimates the dose and your employer decides on notification.
- Good to know
- In the UK, a technical repeat with no procedural, human, systematic or equipment error (images spoiled by movement or contrast extravasation) is not a statutory notification, but events and near misses are still recorded and analysed locally (CQC). Any necessary repeat counts towards the total exposure from an incident.
Explain it simply
Every repeat is more radiation. A repeat is fine when it is needed, but the reason is fixed first and written down.
Why it matters: Recording repeats shows patterns, such as a protocol or a scanner that keeps causing them.
Sources: CQC SAUE 2024; IR(ME)R 2017; ARPANSA RPS C-5; ORS C1 2024; IAEA SSG-46
Pregnant patient found to have been scanned
- What you see
- After a CT, it emerges that the patient was pregnant at the time of the scan.
- Do this now
- Tell the radiologist and report it through your incident process: an inadvertent exposure of an embryo or fetus is investigated.
- A medical physicist estimates the dose to the embryo or fetus; it needs the scan details and the stage of pregnancy.
- The referrer and the patient are informed, and counselling is given by a radiologist or clinician with the right experience, using the dose estimate.
- Record how pregnancy status was checked before the scan, and the estimate.
- Your regional rules decide whether the regulator must be told.
- Good to know
- For the counselling clinician, not for the radiographer to give: the ACR-SPR practice parameter says a well-managed single-phase CT of the abdomen and pelvis gives the embryo or fetus less than 35 mGy, typically about 10 to 25 mGy, which does not warrant ending the pregnancy; CT away from the abdomen and pelvis gives much less. HPA (UK): after a higher-dose examination done inadvertently, termination would not be justified solely on the basis of the radiation risk to the unborn child.
Explain it simply
If someone turns out to have been pregnant during a scan, an expert works out how much radiation reached the baby, and a doctor explains what that means.
Why it matters: For almost all CT scans the risk is small, and a careful, honest explanation prevents needless distress.
Sources: IAEA SSG-46; ACR-SPR Pregnancy PP; HPA RCE-9 2009; RCR IR(ME)R 2026; ARPANSA RPS C-5; ORS C1 2024; IR(ME)R 2017
Air embolism during injection
- What you see
- Air on the images in the veins of the chest, the main pulmonary artery or the right ventricle; with a large volume, sudden breathlessness, air hunger, cough, chest pain, a fast pulse, low blood pressure or wheeze. Air reaching the arterial side can cause neurological signs.
- Why it happens
- Air left in the syringe or tubing is injected with the contrast. Small volumes are common and usually cause no symptoms; a large volume is rare but can be fatal. Patients with a right-to-left shunt or a pulmonary arteriovenous malformation are at higher risk even from small volumes.
- Do this now
- Stop the injection and call for help and the radiologist.
- Give 100% oxygen and turn the patient onto their left side (left lateral decubitus), as the ACR manual describes.
- If the patient arrests, start chest compressions at once and call the resuscitation team.
- Hyperbaric oxygen has been recommended; the medical team decides on referral.
- Report the incident and check the injector set-up before the next patient.
- Good to know
- Prevention, as the ACR manual describes it: clear air from the syringe and pressure tubing, then turn the syringe so the tubing points downwards.
Explain it simply
If air is left in the tubing, it can be pushed into the vein with the dye. A little is harmless; a lot can block blood flow through the heart.
Why it matters: Careful priming of the injector prevents a rare but dangerous event.
Cardiac or respiratory arrest on the CT table
- What you see
- The patient is unresponsive and not breathing normally, or only gasping.
- Do this now
- Call the resuscitation team on your emergency number and stop the scan.
- Start chest compressions at once: CPR is started for anyone unresponsive and not breathing normally, even with occasional gasps. Staff trained and willing to give rescue breaths do so.
- Attach a defibrillator or AED as soon as it arrives and follow its prompts. Minimise pauses in compressions.
- Bring the table out of the gantry and clear space so the team can reach the patient (a general step; follow your local procedure).
- If contrast was given and the cause of the arrest is unclear, the ACR suggests assuming it is an allergic-like reaction; if air embolism is possible, see Air embolism.
- In a visibly pregnant patient, the ACR advises compressions with the patient supine and the uterus displaced manually up and to the left, if enough staff are present.
- Afterwards, record the times and events and report the incident.
Explain it simply
If the heart or breathing stops, pressing hard and fast on the chest keeps blood moving to the brain until the team and a defibrillator arrive.
Why it matters: Every minute without compressions lowers the chance of survival, so they start straight away.
Sources: ANZCOR Guideline 8 2021; ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Injector over-pressure, tubing or line rupture
- What you see
- The injector stops on a pressure alarm, the tubing or a connector leaks or bursts, or a central line, PICC or port leaks or breaks during injection.
- Why it happens
- Pressure rises with a small cannula, a high flow rate, viscous contrast, a kink or a blocked line. Central lines that are not rated for power injection can break.
- Do this now
- Stop the injection and check the patient first: the cannula or line site, any swelling, and any symptoms.
- If the patient has signs of extravasation, go to Contrast extravasation.
- If a central line, PICC or port may be damaged, do not use it again and tell the radiologist and the team responsible for the line (a general step).
- Before restarting, find the cause: a kink or closed clamp, a cannula too small for the flow rate, or a central device not rated for power injection at that flow rate.
- Take a faulty injector or set out of use and report it under your equipment and incident procedures (a general step; your injector manual takes precedence).
- Good to know
- Warming and central-line rules: see Injector faults and Poor IV access.
Explain it simply
The injector is a pump. If something narrows or blocks the tube, the pressure climbs and a weak part can split.
Why it matters: Using the right line at the right speed stops the dye going somewhere it should not.
Sources: ACR Contrast Media 2026; ESUR 2025; RANZCR ICM 2018
Fall from the CT table or during a transfer
- What you see
- A patient falls, slips or is lowered to the floor while getting on or off the table, during a transfer, or while walking in the department.
- Why it happens
- Falls happen in imaging departments, often in CT and MRI areas, and most patients who fall have at least one risk factor; some are injured.
- Do this now
- Call for help, stay with the patient and check whether they are hurt; get a doctor to assess them (a general step; follow your falls procedure).
- Do not lift the patient alone: use the handling aids and the number of staff your assessment calls for.
- Record the fall and report it through your incident process, and follow up as your policy requires.
- Before the next transfer, reassess fall risk: it can change during the examination, for example after sedation.
- Good to know
- Prevention: assess each patient’s fall risk under your policy, help patients on and off the table, keep the call button within reach, and use slide sheets, slide boards or lifts as needed.
Explain it simply
Getting on and off a narrow, high table can be hard, especially after lying still or feeling unwell. A fall can cause a serious injury.
Why it matters: A quick risk check and the right help prevent most falls.
Sources: CAMRT BPG; Abujudeh 2011
Scanner stops or fails during a scan
- What you see
- The scan stops partway, an error appears, the tube fails, the table stops or the power goes off while the patient is on the table.
- Why it happens
- CT scanners end the exposure automatically when an equipment failure affects data collection, and show that this has happened; a manual reset is needed before another scan. A tube fault during acquisition can make the images unusable.
- Do this now
- Check the patient first, and stop any injection that is running.
- Bring the patient out of the gantry and off the table if the fault prevents scanning (a general step; follow your operator manual if the table is stuck).
- Note the time, the message, what was scanned and how much contrast was given; record it in the fault log and call service.
- Ask the radiologist whether a repeat is justified, and on which scanner; an unplanned repeat is a further exposure (see Unplanned repeat).
- After repair, the scanner is tested before it returns to clinical use.
- Good to know
- A power failure can lose work in progress unless an uninterruptible power supply is fitted; the IAEA advises one for CT, as a standby generator may not be stable enough.
Explain it simply
If a part fails in the middle of a scan, the machine stops the X-rays on its own. The pictures may be incomplete, so the doctor decides what happens next.
Why it matters: Clear records help service fix the fault and help the team account for any repeat dose.
Sources: 21 CFR 1020.33; IAEA SSG-46; Geleijns 2014; ACR CT QC 2017
Sources
Guidelines are cited in their current editions: the ACR Manual on Contrast Media (2026), the ESUR Contrast Media Safety Committee Guidelines (2025) and the RANZCR Iodinated Contrast Media Guideline (version 2.3). Radiation rules differ by country; each card names the frameworks it draws on. Sources were checked on 13 and 14 October 2026.
- AAPM Working Group on Standardization of CT Nomenclature and Protocols. CT Dose Notifications and Alerts: What are they, how do they work, and important information for successful implementation. Slide set, 16 April 2014. https://www.aapm.org/pubs/CTProtocols/documents/NotificationsAlertsSlides.pdf
- Abbara S, Blanke P, Maroules CD, Cheezum M, Choi AD, Han BK, et al. SCCT guidelines for the performance and acquisition of coronary computed tomographic angiography: a report of the Society of Cardiovascular Computed Tomography Guidelines Committee. J Cardiovasc Comput Tomogr 2016;10(6):435-449. doi:10.1016/j.jcct.2016.10.002
- Abbott. Can You Wear a CGM During an MRI? If It’s Ours, You Can (newsroom, US, 15 November 2024); FreeStyle Libre support FAQ (Philippines). https://www.abbott.com/en-us/corpnewsroom/diabetes-care/can-you-wear-a-cgm-during-an-mri–if-its-ours-you-can
- Abujudeh H, Kaewlai R, Shah B, Thrall J. Characteristics of falls in a large academic radiology department: occurrence, associated factors, outcomes, and quality improvement strategies. AJR Am J Roentgenol 2011;197(1):154-159. doi:10.2214/AJR.10.4994
- American Association of Physicists in Medicine, Working Group on Standardization of CT Nomenclature and Protocols. AAPM Recommendations Regarding Notification and Alert Values for CT Scanners: Guidelines for Use of the NEMA XR 25 CT Dose-Check Standard. Version 1.0, 27 April 2011. https://www.aapm.org/pubs/CTProtocols/documents/NotificationLevelsStatement.pdf
- American College of Radiology, Committee on Drugs and Contrast Media. ACR Manual on Contrast Media. 2026 edition. Reston, VA: ACR; 2026. https://edge.sitecorecloud.io/americancoldf5f-acrorgf92a-productioncb02-3650/media/ACR/Files/Clinical/Contrast-Manual/ACR-Manual-on-Contrast-Media.pdf
- American College of Radiology, Society for Pediatric Radiology. ACR-SPR Practice Parameter for Imaging Pregnant or Potentially Pregnant Patients with Ionizing Radiation. Revised 2023 (Resolution 31). https://gravitas.acr.org/PPTS/DownloadPreviewDocument?DocId=23
- American College of Radiology, Society for Pediatric Radiology. ACR-SPR Practice Parameter for Performing and Interpreting Diagnostic Computed Tomography (CT). Revised 2022, amended 2023. https://gravitas.acr.org/PPTS/DownloadPreviewDocument?DocId=132
- American College of Radiology. 2017 Computed Tomography Quality Control Manual. Reston, VA: ACR; 2017 (still the required edition per ACR Accreditation Support, revised 21 Mar 2025). https://edge.sitecorecloud.io/americancoldf5f-acrorgf92a-productioncb02-3650/media/ACR/Files/Clinical/Quality-Control-Manuals/CT-Quality-Control-Manual.pdf
- Australasian Society of Clinical Immunology and Allergy. ASCIA Guidelines: Acute Management of Anaphylaxis. June 2026. https://www.allergy.org.au/images/ASCIA_HP_Guidelines_Acute_Management_Anaphylaxis_2026_June.pdf
- Australian and New Zealand Committee on Resuscitation. Guideline 8: Cardiopulmonary Resuscitation (CPR). Approved April 2021. https://www.anzcor.org/home/basic-life-support/guideline-8-cardiopulmonary-resuscitation-cpr
- Australian Commission on Safety and Quality in Health Care. Ensuring correct patient, correct site, correct procedure in CT and MRI (protocol). Sydney: ACSQHC. https://www.safetyandquality.gov.au/sites/default/files/migrated/Protocol_CTMRI.pdf
- Australian Radiation Protection and Nuclear Safety Agency. Code for Radiation Protection in Medical Exposure. Radiation Protection Series C-5. Yallambie: ARPANSA; 2019. https://www.arpansa.gov.au/sites/default/files/medical-exposure-code-rps-c-5.pdf
- Bae KT. Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology 2010;256(1):32-61. doi:10.1148/radiol.10090908
- Barrett JF, Keat N. Artifacts in CT: recognition and avoidance. RadioGraphics 2004;24(6):1679-1691. doi:10.1148/rg.246045065
- Bauer RW, Schell B, Beeres M, Wichmann JL, Bodelle B, Vogl TJ, Kerl JM. High-pitch dual-source computed tomography pulmonary angiography in freely breathing patients. J Thorac Imaging 2012;27(6):376-381. doi:10.1097/RTI.0b013e318250067e
- Birnbaum BA, Maki DD, Chakraborty DP, Jacobs JE, Babb JS. Renal cyst pseudoenhancement: evaluation with an anthropomorphic body CT phantom. Radiology 2002;225(1):83-90. doi:10.1148/radiol.2251010930
- Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med 2012;4(2):229-240. doi:10.2217/iim.12.13
- Brink M, de Lange F, Oostveen LJ, Dekker HM, Kool DR, Deunk J, et al. Arm raising at exposure-controlled multidetector trauma CT of thoracoabdominal region: higher image quality, lower radiation dose. Radiology 2008;249(2):661-670. doi:10.1148/radiol.2492080169
- Byrne A, Macdonald DB, Kirkpatrick IDC, Pham M, Green CR, Copaescu AM, McInnes MDF, Ling L, Ellis A, Costa AF. CAR/CSACI practice guidance for contrast media hypersensitivity. Can Assoc Radiol J 2025;76(3):400-416. doi:10.1177/08465371241311253
- Canadian Association of Medical Radiation Technologists. Best Practice Guidelines: 01.01.06 Patient comfort and positioning; 02.01.04 Falls prevention; 05.03.01 Safe patient handling. Ottawa: CAMRT (validated April 2018). https://camrt-bpg.ca/
- Care Quality Commission (agreed by the four UK IR(ME)R enforcing authorities). IR(ME)R: Notify us about an exposure; Notification codes, categories and criteria; When to investigate and notify the enforcing authority; Keeping records of investigations. Web guidance, updated 28 August 2024. https://www.cqc.org.uk/guidance-providers/ionising-radiation/ionising-radiation-medical-exposure-regulations-irmer/notify-us-about-exposure
- Centers for Medicare & Medicaid Services. MLN Matters MM9250: Payment Reduction for Computed Tomography (CT) Diagnostic Imaging Services. Released 6 November 2015, effective 1 January 2016. https://www.cms.gov/Outreach-and-Education/Medicare-Learning-Network-MLN/MLNMattersArticles/downloads/MM9250.pdf
- Clunie D. Digital imaging. Chapter 16 in: Dance DR, Christofides S, Maidment ADA, McLean ID, Ng KH (technical eds). Diagnostic Radiology Physics: A Handbook for Teachers and Students. Vienna: IAEA; 2014. https://www-pub.iaea.org/MTCD/Publications/PDF/Pub1564webNew-74666420.pdf
- Davenport MS, Perazella MA, Yee J, Dillman JR, Fine D, McDonald RJ, Rodby RA, Wang CL, Weinreb JC. Use of intravenous iodinated contrast media in patients with kidney disease: consensus statements from the American College of Radiology and the National Kidney Foundation. Radiology 2020;294(3):660-668. doi:10.1148/radiol.2019192094
- Dexcom. Do I need to remove my Dexcom sensor for medical procedures? FAQ (UK site). https://www.dexcom.com/en-gb/faqs/medical-procedures
- ESUR Contrast Media Safety Committee. Contrast Media Safety Committee Guidelines 2025. European Society of Urogenital Radiology; November 2025. https://www.esur.org/wp-content/uploads/2025/12/Guidelines-2025-ESUR-vf-1.pdf
- Fonseca GP, Baer-Beck M, Fournie E, Hofmann C, Rinaldi I, Ollers MC, et al. Evaluation of novel AI-based extended field-of-view CT reconstructions. Med Phys 2021;48(7):3583-3594. doi:10.1002/mp.14937
- Geleijns J. Computed tomography. Chapter 11 in: Dance DR, Christofides S, Maidment ADA, McLean ID, Ng KH (technical eds). Diagnostic Radiology Physics: A Handbook for Teachers and Students. Vienna: IAEA; 2014. p. 257-290. https://www-pub.iaea.org/MTCD/Publications/PDF/Pub1564webNew-74666420.pdf
- Geyer LL, Schoepf UJ, Meinel FG, Nance JW Jr, Bastarrika G, Leipsic JA, et al. State of the art: iterative CT reconstruction techniques. Radiology 2015;276(2):339-357. doi:10.1148/radiol.2015132766
- Graef J, Bressem KK, Asbach P, Hamm B, Niehues SM. Changing the patient’s position: pitfalls and benefits for radiation dose and image quality of computed tomography in polytrauma. Diagnostics (Basel) 2022;12(11):2661. doi:10.3390/diagnostics12112661
- Habibzadeh MA, Ay MR, Asl AR, Ghadiri H, Zaidi H. Impact of miscentering on patient dose and image noise in x-ray CT imaging: phantom and clinical studies. Phys Med 2012;28(3):191-199. doi:10.1016/j.ejmp.2011.06.002
- Health and Disability Commissioner (New Zealand). Code of Health and Disability Services Consumers’ Rights, Right 5 (effective communication). https://www.hdc.org.nz/your-rights/about-the-code/code-of-health-and-disability-services-consumers-rights/
- Health and Disability Commissioner. Guidance on Open Disclosure Policies. Auckland: HDC; November 2019. https://www.hdc.org.nz/media/k2jjd3qj/guidance-on-open-disclosure-policies.pdf
- Health and Safety Executive. Work with ionising radiation. Ionising Radiations Regulations 2017. Approved Code of Practice and guidance. L121, second edition. HSE; 2018. https://www.hse.gov.uk/pubns/priced/l121.pdf
- Health Protection Agency, The Royal College of Radiologists, College of Radiographers. Protection of Pregnant Patients during Diagnostic Medical Exposures to Ionising Radiation. RCE-9. Chilton: HPA; 2009. https://assets.publishing.service.gov.uk/media/5a7d9efa40f0b635051d048c/RCE-9_for_web.pdf
- Hsieh J. Adaptive streak artifact reduction in computed tomography resulting from excessive x-ray photon noise. Med Phys 1998;25(11):2139-2147. doi:10.1118/1.598410
- Image Gently Alliance. Image Gently and CT Scans; CT Scans: What can I do as a Radiologist? Web pages. https://www.imagegently.org/Procedures/Computed-Tomography
- Insulet Corporation. Important Safety Information pages (US and UK). https://www.omnipod.com/safety
- International Atomic Energy Agency. Quality Assurance Programme for Computed Tomography: Diagnostic and Therapy Applications. IAEA Human Health Series No. 19. Vienna: IAEA; 2012. https://www-pub.iaea.org/mtcd/publications/pdf/pub1557_web.pdf
- International Atomic Energy Agency. Radiation Protection and Safety in Medical Uses of Ionizing Radiation. IAEA Safety Standards Series No. SSG-46. Vienna: IAEA; 2018. https://www-pub.iaea.org/MTCD/Publications/PDF/PUB1775_web.pdf
- Jha AK, Purandare NC, Shah S, Agrawal A, Puranik AD, Rangarajan V. Identification of a unique cause of ring artifact seen in computed tomography trans-axial images. Indian J Nucl Med 2013;28(4):232-233. doi:10.4103/0972-3919.121970
- Kaasalainen T, Mäkelä T, Kortesniemi M. The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments. Eur J Radiol Open 2019;6:24-32. doi:10.1016/j.ejro.2018.12.001
- Kaasalainen T, Palmu K, Reijonen V, Kortesniemi M. Effect of patient centering on patient dose and image noise in chest CT. AJR Am J Roentgenol 2014;203(1):123-130. doi:10.2214/AJR.13.12028
- Kachelriess M, Watzke O, Kalender WA. Generalized multi-dimensional adaptive filtering for conventional and spiral single-slice, multi-slice, and cone-beam CT. Med Phys 2001;28(4):475-490. doi:10.1118/1.1358303
- Kalisz K, Buethe J, Saboo SS, Abbara S, Halliburton S, Rajiah P. Artifacts at cardiac CT: physics and solutions. RadioGraphics 2016;36(7):2064-2083. doi:10.1148/rg.2016160079
- Katsura M, Sato J, Akahane M, Kunimatsu A, Abe O. Current and novel techniques for metal artifact reduction at CT: practical guide for radiologists. RadioGraphics 2018;38(2):450-461. doi:10.1148/rg.2018170102
- Koetzier LR, Mastrodicasa D, Szczykutowicz TP, van der Werf NR, Wang AS, Sandfort V, et al. Deep learning image reconstruction for CT: technical principles and clinical prospects. Radiology 2023;306(3):e221257. doi:10.1148/radiol.221257
- Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK. Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol 2007;188(2):547-552. doi:10.2214/AJR.06.0370
- Liu M, Zhang J, Deng X, et al. Thoracic-abdominopelvic contrast-enhanced dual-source computed tomography with a high-pitch acquisition protocol and free breathing for reducing motion artifacts and scan time in patients with cancer: a prospective study. Quant Imaging Med Surg 2025;15(11):10699-10713. doi:10.21037/qims-2025-1016
- Lovrenski J, Raissaki M, Plut D, Alexopoulou E, Görkem SB, Ozcan HN, et al. ESR Essentials: imaging of common paediatric pulmonary diseases. Practice recommendations by the European Society of Paediatric Radiology. Eur Radiol 2025;35(8):5037-5052. doi:10.1007/s00330-024-11268-4
- Maki DD, Birnbaum BA, Chakraborty DP, Jacobs JE, Carvalho BM, Herman GT. Renal cyst pseudoenhancement: beam-hardening effects on CT numbers. Radiology 1999;213(2):468-472. doi:10.1148/radiology.213.2.r99nv33468
- McCollough CH, Zhang J, Primak AN, Clement WJ, Buysman JR. Effects of CT irradiation on implantable cardiac rhythm management devices. Radiology 2007;243(3):766-774. doi:10.1148/radiol.2433060993
- Medtronic MiniMed. MiniMed 780G System User Guide (US), 2025. https://www.medtronicdiabetes.com/sites/default/files/library/download-library/user-guides/MiniMed-780G-system-user-guide.pdf
- Moser LJ, Mergen V, Allmendinger T, Manka R, Eberhard M, Alkadhi H. A novel reconstruction technique to reduce stair-step artifacts in sequential mode coronary CT angiography. Invest Radiol 2024;59(9):622-628. doi:10.1097/RLI.0000000000001066
- Murphy DJ, Aghayev A, Steigner ML. Vascular CT and MRI: a practical guide to imaging protocols. Insights Imaging 2018;9(2):215-236. doi:10.1007/s13244-018-0597-2
- National Cancer Institute. Radiation Risks and Pediatric Computed Tomography (CT): A Guide for Health Care Providers. Reviewed 4 September 2018. https://www.cancer.gov/about-cancer/causes-prevention/risk/radiation/pediatric-ct-scans
- NEMA. DICOM PS3.4 Service Class Specifications: Annex J Storage Commitment Service Class; Annex K Basic Worklist Management Service. https://dicom.nema.org/medical/dicom/current/output/chtml/part04/chapter_J.html
- Nowotny R. X ray production. Chapter 5 in: Dance DR, Christofides S, Maidment ADA, McLean ID, Ng KH (technical eds). Diagnostic Radiology Physics: A Handbook for Teachers and Students. Vienna: IAEA; 2014. https://www-pub.iaea.org/MTCD/Publications/PDF/Pub1564webNew-74666420.pdf
- Office of Radiation Safety, Ministry of Health. Code of Practice for Diagnostic and Interventional Radiology: ORS C1. Wellington: Ministry of Health; 1 December 2024. https://health.govt.nz/system/files/2024-11/Code-of-Practice-for-Diagnostic-and-Interventional-Radiology-ORS-C1.pdf
- Pack JD, Xu M, Wang G, Baskaran L, Min J, De Man B. Cardiac CT blooming artifacts: clinical significance, root causes and potential solutions. Vis Comput Ind Biomed Art 2022;5(1):29. doi:10.1186/s42492-022-00125-0
- Puvanasunthararajah S, Fontanarosa D, Wille ML, Camps SM. The application of metal artifact reduction methods on computed tomography scans for radiotherapy applications: a literature review. J Appl Clin Med Phys 2021;22(6):198-223. doi:10.1002/acm2.13255
- Rau A, Elsheikh S, Cimflova P, Stein T, Taschner CA, Hosp JA, et al. Contrast bolus timing in CT-angiography and CT-perfusion: insights from a large clinical dataset. Neuroradiology 2025;67(5):1171-1181. doi:10.1007/s00234-025-03558-5
- Royal Australian and New Zealand College of Radiologists. Iodinated Contrast Media Guideline, Version 2.3. Sydney: RANZCR; March 2018. https://www.ranzcr.com/wp-content/uploads/2018/03/Iodinated-Contrast-Media-Guideline-V2.3-2.pdf
- Samei E, Bakalyar D, Boedeker KL, Brady S, Fan J, Leng S, et al. Performance Evaluation of Computed Tomography Systems: The Report of AAPM Task Group 233. AAPM Report No. 233. 2019. https://www.aapm.org/pubs/reports/RPT_233.pdf
- Schmidt D, Odland R. Mirror-image reversal of coronal computed tomography scans. Laryngoscope 2004;114(9):1562-1565. doi:10.1097/00005537-200409000-00011
- Schultz DG. FDA Preliminary Public Health Notification: Possible Malfunction of Electronic Medical Devices Caused by Computed Tomography (CT) Scanning. FDA; 14 July 2008 (archived; superseded 2016). http://web.archive.org/web/20170118090844/http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/PublicHealthNotifications/ucm061994.htm
- Shepard SJ, Lin PJP, et al. Quality Control in Diagnostic Radiology. Report of Task Group 12. AAPM Report No. 74. Madison, WI: Medical Physics Publishing; 2002. https://www.aapm.org/pubs/reports/RPT_74.pdf
- Silverman SG, Pedrosa I, Ellis JH, Hindman NM, Schieda N, Smith AD, et al. Bosniak classification of cystic renal masses, version 2019: an update proposal and needs assessment. Radiology 2019;292(2):475-488. doi:10.1148/radiol.2019182646
- Tandem Diabetes Care. Important Safety Information pages for Tandem Mobi, t:slim X2 with Control-IQ and t:slim X2 (US). https://www.tandemdiabetes.com/legal/important-safety-information/tandem-mobi
- The Ionising Radiation (Medical Exposure) Regulations 2017 (SI 2017/1322), as amended by SI 2024/896 (England, Wales and Scotland). https://www.legislation.gov.uk/uksi/2017/1322/regulation/8
- The Royal College of Radiologists. IR(ME)R: Implications for clinical practice in diagnostic imaging, interventional radiology and diagnostic nuclear medicine. Second edition. London: RCR; August 2026. https://www.rcr.ac.uk/media/s1sdu4cq/rcr-publications-ir-me-r-implications-for-clinical-practice-in-diagnostic-imaging-interventional-radiology-and-diagnostic-nuclear-medicine-second-edition.pdf
- Toth T, Ge Z, Daly MP. The influence of patient centering on CT dose and image noise. Med Phys 2007;34(7):3093-3101. doi:10.1118/1.2748113
- U.S. Food and Drug Administration. Electronic Medical Devices, X-ray Imaging and Radiation Therapy: What to Know and How to Prevent Damage; with sub-pages on implantable heart devices, neurostimulators and wearable diabetes devices during CT. Content current as of 15 October 2024. https://www.fda.gov/radiation-emitting-products/radiation-safety/electronic-medical-devices-x-ray-imaging-and-radiation-therapy-what-know-and-how-prevent-damage
- Uppot RN, Sahani DV, Hahn PF, Gervais D, Mueller PR. Impact of obesity on medical imaging and image-guided intervention. AJR Am J Roentgenol 2007;188(2):433-440. doi:10.2214/AJR.06.0409
- Uppot RN. Technical challenges of imaging and image-guided interventions in obese patients. Br J Radiol 2018;91(1089):20170931. doi:10.1259/bjr.20170931
- US Code of Federal Regulations, Title 21, section 1020.33, Computed tomography (CT) equipment (current eCFR text). https://www.ecfr.gov/current/title-21/chapter-I/subchapter-J/part-1020/section-1020.33
- Varchetta F, Cosson P, Widdowfield M, Danzi R, Orlando G, Natale M, et al. Chest CT in patients with shortness of breath: comparing high pitch CT and conventional CT on respiratory artefacts and dose. Radiography (Lond) 2021;27(3):908-914. doi:10.1016/j.radi.2021.02.013
- Wang Q, Wang H, Lin C, Huang X, Chen S, Zhang J, et al. Differentiating pulmonary embolism from contrast accumulation-induced beam hardening artifacts using dual-energy CT quantitative parameters. Quant Imaging Med Surg 2025;15(6):5593-5603. doi:10.21037/qims-24-1385
- Wienholt P, Hermans A, Siepmann R, Kuhl C, Pinto Dos Santos D, Nebelung S, et al. Low-dose CT quality assurance at scale: automated detection of overscanning, underscanning, and image noise. Life (Basel) 2026;16(1):152. doi:10.3390/life16010152
- Wirth S, Hebebrand J, Basilico R, et al. European Society of Emergency Radiology: guideline on radiological polytrauma imaging and service (short version). Insights Imaging 2020;11:135. doi:10.1186/s13244-020-00947-7
- Wu H, Wang J, Zhou M, Wang Y, Cui C, Zhou C, et al. Application of bolus tracking: the effect of ROI positions on the images quality of cervicocerebral CT angiography. Heliyon 2024;10(7):e29260. doi:10.1016/j.heliyon.2024.e29260
Keep going
Everything on X-ray & CT: X-ray & CT hub. Or search the whole reference.
Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.