CT & Radiotherapy Calculators

Axial head CT labelled: frontal sinus, skull, lateral ventricle, brain. Handwritten note: bone white, air black.
CT: Mikael Häggström, Wikimedia Commons, CC0. Cropped; labels added.

CT & Radiotherapy Calculators

Dose and fractionation calculators for CT and radiotherapy

Updated September 2026

Estimation only. Effective dose from DLP is a population-averaged risk metric for a reference phantom. It is not a patient dose. Do not use it for individual risk counselling, and never use these outputs to plan a clinical exposure or a course of treatment.

Every coefficient below is a published value; sources are listed at the foot of the page. These are teaching and estimation tools, not a substitute for your department’s dosimetry or treatment-planning system.

CT dose

CT dose bench

Adult k-factors (AAPM Report 96, after Shrimpton 2004; ICRP 60 weighting). Paediatric k-factors are substantially higher and are not offered here.

Effective diameter = √(AP × LAT). Conversion factors are the AAPM Report 204 exponential fits.

Shared input

DLP and effective dose

DLP = CTDIvol × scan length. E ≈ DLP × k, where k depends on the body region.

350.0 mGy·cm Effective dose ≈ 4.90 mSv (k = 0.014)

k-factors (DLP → effective dose): AAPM Report No. 96, The Measurement, Reporting and Management of Radiation Dose in CT (2008), after Shrimpton PC et al., NRPB-W67 (2004). ICRP 60 tissue weighting. Values recomputed with ICRP 103 weighting (Deak PD et al., Radiology 2010;257:158–166) differ by up to ~30% for some regions.

SSDE (size-specific dose estimate)

Scanner CTDIvol assumes a fixed phantom. SSDE corrects it for the patient’s actual size.

15.07 mGy Effective diameter 24.5 cm · conversion factor 1.507 · 32 cm phantom

SSDE: AAPM Report No. 204, Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations (2011). Conversion factors use the published exponential fits: f32 = 3.704369·e−0.03671937·Deff, f16 = 1.874799·e−0.03871313·Deff.

Fractionation

BED and EQD2

Linear-quadratic model. BED = nd(1 + d/[α/β]); EQD2 = BED ÷ (1 + 2/[α/β]).

The LQ model is least reliable above roughly 8–10 Gy per fraction; SBRT comparisons made this way are indicative only.

BED 60.00 Gy EQD2 50.00 Gy · total dose 50.0 Gy in 25 fractions · α/β 10 Gy

Linear-quadratic model: Fowler JF, Br J Radiol 1989;62:679–694; Joiner MC & van der Kogel AJ, Basic Clinical Radiobiology, 5th ed. (2018).

Alternative fractionation

How many fractions of a new size deliver the same EQD2 as a reference schedule?

Iso-effective for the chosen α/β only. A schedule matched for tumour effect will not be matched for late normal tissue, and overall treatment time is ignored.

17.11 fractions of 2.75 Gy Rounded to 17 × 2.75 Gy = 46.8 Gy, EQD2 49.67 Gy (reference EQD2 50.00 Gy)

Linear-quadratic model: Fowler JF, Br J Radiol 1989;62:679–694; Joiner MC & van der Kogel AJ, Basic Clinical Radiobiology, 5th ed. (2018).

Contrast

Iodine delivery rate

IDR governs arterial enhancement far more than total contrast volume does.

Total iodine load = concentration × volume. Injection duration = volume ÷ flow rate.

1.40 g I/s Total iodine 28.0 g · injection duration 20.0 s

Iodine delivery rate: Bae KT, Intravenous contrast medium administration and scan timing at CT, Radiology 2010;256:32–61.

Contrast dose by weight

Weight-based iodine load, the approach recommended for body CT.

0.4–0.6 g I/kg is a common range for routine portal-venous body CT; departments differ, and your protocol governs.

107.1 mL Delivers 37.5 g iodine at 350 mg I/mL
Sources

k-factors (DLP → effective dose): AAPM Report No. 96, The Measurement, Reporting and Management of Radiation Dose in CT (2008), after Shrimpton PC et al., NRPB-W67 (2004). ICRP 60 tissue weighting. Values recomputed with ICRP 103 weighting (Deak PD et al., Radiology 2010;257:158–166) differ by up to ~30% for some regions.

SSDE: AAPM Report No. 204, Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations (2011). Conversion factors use the published exponential fits: f32 = 3.704369·e−0.03671937·Deff, f16 = 1.874799·e−0.03871313·Deff.

Linear-quadratic model: Fowler JF, Br J Radiol 1989;62:679–694; Joiner MC & van der Kogel AJ, Basic Clinical Radiobiology, 5th ed. (2018).

Iodine delivery rate: Bae KT, Intravenous contrast medium administration and scan timing at CT, Radiology 2010;256:32–61.

Related guides

More in Calculators

Everything on Tools: Tools hub. Or search the whole reference.

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.