X-ray and CT radiation safety

X-ray and CT radiation safety

The international system of radiological protection (ICRP 103) and how it is turned into law and practice in the UK, EU, US and worldwide: dose limits, diagnostic reference levels, occupational and patient protection, and pregnancy rules, with every figure traced to a primary source.

Related
On this pageFrameworkKey documents comparedOccupational protectionPatient protection & DRLsPregnancyQuick reference dosesSources

Figures verified against the primary sources listed in the References (UK NDRLs current as of the UKHSA list effective 11 December 2025). Educational summary. Always follow your local rules, employer’s procedures and radiation protection adviser/medical physicist.

Key points
  • Dose limits apply to workers and the public, never to patients: a patient’s medically justified exposure is controlled by justification and optimisation instead.
  • The occupational limit is 20 mSv per year under ICRP/IAEA/EU/UK but 50 mSv per year in US regulation, so quote the framework you are actually working under.
  • A DRL is an optimisation tool, not a limit: if your room’s median dose consistently exceeds it, investigate and optimise, and being under it does not mean optimisation is finished.
  • UK guidance: ask anyone of childbearing potential about pregnancy whenever the primary beam may irradiate the pelvis, meaning any exposure involving the area between the diaphragm and the knees; the 28-day rule covers most examinations, the 10-day rule higher-dose procedures.
  • Since 2019, major bodies recommend ending routine gonadal and fetal contact shielding in diagnostic X-ray imaging: shields frequently obscure anatomy and can trigger AEC to raise dose; follow local policy.
Explain it simply
  • Legal radiation caps are for staff and the public, not patients. A patient’s X-ray is controlled by making sure it is needed and that it uses as little radiation as is reasonable.
  • The yearly radiation cap for staff is higher under US rules than under the international, EU and UK rules. Always quote the rules you actually work under.
  • A reference level is a yardstick for a room’s usual dose, not a legal cap. If a room keeps going over it, find out why and fix it; being under it does not mean the work is done.
  • In the UK, anyone who could be pregnant is asked first if an X-ray beam may reach anywhere between the bottom of the chest and the knees. Higher-dose tests are booked early in the menstrual cycle, when pregnancy is very unlikely.
  • Experts now advise against routinely placing lead covers over the reproductive organs or an unborn baby. The covers often hide what needs to be seen and can make the machine use more radiation; follow your department’s rules.

A guide only: follow your departmental protocols and regional requirements.

The framework: how radiation protection is organised

Almost every national radiation-safety law in medicine descends from one document: ICRP Publication 103 (2007), the International Commission on Radiological Protection’s “Recommendations”.1 ICRP recommendations are not law; they flow downhill into the IAEA’s International Basic Safety Standards, the EU Basic Safety Standards Directive, and national regulations such as the UK’s IRR17/IR(ME)R 2017 and US federal and state rules.

1. Justification

Any decision that alters the radiation exposure situation should do more good than harm. In medicine: every exposure must be justified; the expected diagnostic benefit must outweigh the radiation detriment, taking alternatives (ultrasound, MRI) into account. No routine exposure “just in case”.

2. Optimisation (ALARA)

Doses should be kept As Low As Reasonably Achievable, economic and societal factors considered. This is where technique factors, collimation, filtration, exposure charts and diagnostic reference levels live. Optimisation ≠ minimisation: an image that cannot answer the clinical question is 100% wasted dose.

3. Dose limitation

Individual dose limits apply to workers and the public, never to patients. A patient’s medically justified exposure has no upper legal limit; it is controlled by justification and optimisation instead.

Safety-critical concept: dose limits do not apply to patients A common exam and real-world misconception. Patient exposures are controlled through justification (is this examination indicated?) and optimisation (is the dose as low as reasonably achievable for the diagnostic task?). Applying worker/public limits to patients could deny someone a life-saving examination.
The three exposure categories (ICRP 103)
  • Occupational exposure: staff exposed as a result of their work (radiographers, radiologists, IR nurses). Subject to dose limits and monitoring.
  • Public exposure: everyone else (people in the waiting room, ward visitors, workers not designated as radiation workers). Subject to the public dose limit.
  • Medical exposure: patients, plus carers/comforters and research volunteers. Controlled by justification, optimisation, DRLs and (for carers/research) dose constraints, not by limits.
Effective dose in 90 seconds

Absorbed dose (gray, Gy) is energy deposited per kilogram of tissue. Equivalent dose (sievert, Sv) weights that for radiation type; for X-rays the weighting factor is 1, so 1 mGy absorbed = 1 mSv equivalent. Effective dose (also Sv) additionally weights each organ’s dose by its relative sensitivity to stochastic harm (tissue weighting factors, ICRP 103) and sums across the body.1

Effective dose exists so that very different exposures (a chest X-ray vs. a CT vs. an ingested radionuclide) can be compared on one whole-body risk-related scale. It is a protection and comparison quantity for a reference person, not a measurement of an individual patient’s personal risk.

Units cheat-sheet
  • Gy (gray): absorbed dose. Used for organ/skin/fetal doses and deterministic (tissue-reaction) thresholds.
  • Sv (sievert): equivalent/effective dose. Used for stochastic risk, dose limits and comparisons. In diagnostic X-ray work you will mostly see millisievert (mSv) and microsievert (µSv).
  • KAP / DAP (Gy·cm²): kerma-area product (air kerma × field area). The standard machine-reported quantity for radiography/fluoroscopy DRLs.
  • Entrance surface / incident air kerma (mGy): dose metric at the skin entrance, used for some radiography DRLs and skin-injury assessment.
Stochastic vs. deterministic (tissue-reaction) effects
  • Stochastic effects (cancer, heritable effects): probability increases with dose; no assumed threshold (the linear no-threshold model underpins the protection system1); severity is independent of dose. This is what dose limits and optimisation target at diagnostic doses.
  • Deterministic effects / tissue reactions (skin erythema, epilation, cataract): occur above a threshold dose and worsen with dose. Essentially irrelevant for plain radiography; relevant for prolonged fluoroscopy/interventional work.

ICRP Publication 105 applies the ICRP 103 system specifically to medicine, covering justification at three levels (use of radiation in medicine generally; a defined procedure for a defined indication; the individual patient), optimisation via DRLs, and the special cases of pregnancy, carers and research.2

Back to contents

Key documents compared

How the main instruments relate: ICRP recommends → IAEA and the EU turn recommendations into standards/directives → nations legislate (UK IRR17 + IR(ME)R; EU member-state transpositions; US federal + state patchwork) → professional bodies (ACR, AAPM, NCRP, RCR, IPEM) issue practice guidance.

Show the full section
Instrument Scope & legal status Dose limits (occupational / public / eye lens / pregnancy) Who enforces Key operational requirements
ICRP 103 (2007) + ICRP 1051,2 International recommendations; the scientific/ethical basis of the whole system. Not legally binding anywhere. Workers: 20 mSv/yr averaged over 5 yr (≤50 mSv in any single year). Public: 1 mSv/yr. Eye lens: 20 mSv/yr averaged over 5 yr, ≤50 mSv in any year (2011 statement, ICRP 1183). Declared pregnancy: protect embryo/fetus broadly as a member of the public (~1 mSv for remainder of pregnancy). No one. Adopted via IAEA, EU and national law. Justification, optimisation, dose limitation; tissue weighting factors; DRL concept for medical exposure; categories of exposure.
IAEA BSS: GSR Part 3 (2014)4 International Basic Safety Standards, co-sponsored by WHO, ILO and others; template for national law in most of the world. Binding on states only insofar as adopted nationally. Workers: 20 mSv/yr averaged over 5 yr (≤50 mSv/yr). Public: 1 mSv/yr. Eye lens: 20 mSv/yr averaged over 5 yr. Pregnant worker: employment adapted so embryo/fetus gets the same level of protection as a member of the public (≈1 mSv). National regulatory bodies established under each state’s law. Graded regulatory control (notification→authorisation), qualified experts, DRLs required for medical exposures, radiological protection officer, monitoring, education & training, safety assessment.
EU BSS: Directive 2013/59/Euratom5 Legally binding on EU member states; each had to transpose into national law by Feb 2018. Consolidates five older directives incl. the Medical Exposures Directive. Workers: 20 mSv/yr. Public: 1 mSv/yr. Eye lens: 20 mSv/yr (or 100 mSv over 5 yr, ≤50 in one year, where national law allows). Declared pregnancy: embryo/fetus dose ≤1 mSv for the remainder of pregnancy. Member-state competent authorities (e.g. health ministries, nuclear/radiation authorities). Mandatory DRLs, clinical audit, dose recording in the report for CT/interventional, medical physics expert involvement, accidental-exposure reporting, patient information, controlled/supervised areas, RPE/RPO roles.
UK: IRR176 + IR(ME)R 20177 Two-law split: IRR17 protects workers and the public; IR(ME)R 2017 protects patients (post-Brexit retention of EU BSS structure). Both statutory instruments (criminal law). IRR17. Workers: 20 mSv/yr; public: 1 mSv/yr; eye lens: 20 mSv/yr; skin/extremities: 500 mSv/yr; declared pregnancy: ≤1 mSv to the fetus for the remainder of pregnancy. Trainees under 18: 6 mSv/yr. No patient limits (IR(ME)R uses justification/optimisation + DRLs). IRR17: Health and Safety Executive (HSE)8. IR(ME)R: CQC (England), Healthcare Improvement Scotland, HIW (Wales), RQIA (NI). IRR17: graded notification/registration/consent; RPA appointment; controlled areas + local rules + RPS; classified workers (>6 mSv/yr) with approved dosimetry. IR(ME)R: duty holders (employer, referrer, practitioner, operator), written employer’s procedures, entitlement, local DRLs informed by UK NDRLs9, clinical evaluation of every exposure, significant-accidental-exposure notification.
US: NRC 10 CFR 2010, state regs (CRCPD SSRs11), FDA 21 CFR 1020, NCRP, ACR–AAPM12 A patchwork: NRC regulates byproduct/reactor material; X-ray machines are regulated by the individual states, most following the CRCPD Suggested State Regulations. FDA sets equipment performance standards. NCRP reports and ACR–AAPM practice parameters are influential guidance; The Joint Commission accreditation embeds much of it. Occupational (10 CFR 20 & typical state regs): 50 mSv/yr (5 rem) TEDE; eye lens 150 mSv/yr; skin/extremities 500 mSv/yr; public 1 mSv/yr. Declared pregnancy: 5 mSv (0.5 rem) to embryo/fetus for the whole gestation; NCRP recommends ≤0.5 mSv/month. NCRP also recommends a cumulative lifetime guideline of 10 mSv × age in years (Report 11613). Note the US has not adopted the reduced 20 mSv/yr eye-lens limit. State radiation control programs (X-ray); NRC/agreement states (radioactive material); FDA (equipment); CMS/Joint Commission (accreditation leverage). State registration and inspection of X-ray tubes; qualified medical physicist surveys; ACR–AAPM DRLs/achievable doses drawn from NEXT and ACR Dose Index Registry data12; personnel monitoring above threshold fractions of limits; fluoroscopy-specific credentialing in many states.
The 20 vs 50 mSv trap The internationally recommended occupational limit (ICRP/IAEA/EU/UK) is 20 mSv per year; the US regulatory limit remains 50 mSv (5 rem) per year. Similarly the eye-lens limit is 20 mSv/yr internationally but still 150 mSv/yr in US regulation (NCRP has recommended lowering it). Quote the framework you are actually working under.
Memory hook Everywhere that follows ICRP: 20 / 1 / 20 (worker 20 mSv, public 1 mSv, lens 20 mSv per year). The US runs 50 / 1 / 150. Pregnancy after declaration: 1 mSv to the fetus (UK/EU/IAEA) or 5 mSv total gestation (US NRC).

Occupational protection

Dose limits at a glance
Quantity (annual)ICRP 103 / IAEA / EU / UK1,4–6US (10 CFR 20 / typical state regs)10
Effective dose (worker)20 mSv/yr (ICRP/IAEA: averaged over 5 yr, ≤50 mSv in any single year; UK IRR17: 20 mSv/yr)50 mSv (5 rem) TEDE
Effective dose (public)1 mSv/yr1 mSv (0.1 rem)
Equivalent dose (eye lens)20 mSv/yr (averaged over 5 yr where permitted, ≤50 in one year)150 mSv
Equivalent dose (skin)500 mSv (averaged over 1 cm²)500 mSv (shallow dose)
Equivalent dose (hands/feet)500 mSv500 mSv
Declared pregnant worker≤1 mSv to embryo/fetus for remainder of pregnancy5 mSv to embryo/fetus over gestation (NCRP: ≤0.5 mSv/month)
Trainees 16–18 (UK)6 mSv/yrMinors: 10% of adult limits
Personal monitoring
  • Who: anyone likely to receive a meaningful fraction of a limit. UK: workers likely to exceed 6 mSv/yr effective dose, 15 mSv/yr to the lens of the eye, or 150 mSv/yr to the skin or extremities must be classified and monitored by an HSE-approved dosimetry service with dose records kept long-term (IRR176). US state regs typically require monitoring when doses are likely to exceed 10% of a limit.
  • What: whole-body dosimeter (TLD/OSL badge) worn at trunk level, under the lead apron for whole-body dose; interventional staff often add a second badge at collar level over the apron (eye/thyroid estimate), extremity rings, and increasingly dedicated eye-lens dosimeters (Hp(3)) for high-workload fluoroscopy operators.
  • Wear rules: consistent position, never share, never deliberately expose, return on time, report lost badges and abnormal readings.
Lead PPE: what the numbers mean
Lead equivalenceTypical useAttenuation of scatter at 70–90 kV23
0.25 mm PbLight aprons for occasional/low-scatter work; some thyroid shields; paediatric holder apronsAbout 90–97%; less at higher kV
0.35 mm PbThe common default apron for fluoroscopy and general department workAbout 95–98%; less at higher kV
0.5 mm PbHigh-workload interventional/cardiac cath front panels; protective glasses commonly 0.5–0.75 mm Pb-equivalent lensesAbout 97–99.5%; less at higher kV

Some wraparound aprons are 0.25 mm per layer and reach 0.5 mm only where the front panels overlap; check whether a label gives the single-layer or the overlapped value.24 Attenuation percentages are approximate and beam-quality dependent; an apron that stops ~98% of a 70 kVp scatter spectrum stops less of a 120 kVp one. Aprons protect against scatter, not the primary beam. Check aprons at least annually (visual + fluoroscopic/radiographic screening) and store hung, never folded.

The three real controls: time, distance, shielding Distance costs nothing. At normal working distances, scatter dose falls roughly with the inverse square law: double your distance (from 1 m to 2 m, say) and your dose falls to about a quarter. One step back only does that if it doubles your distance, and very close to the patient the fall-off is gentler. Scatter is highest on the side where the beam enters the patient, which is why fluoroscopy with the tube under the couch, and standing on the image-receptor side, both lower staff dose.
Room shielding & designated areas
  • Room shielding is designed by a qualified expert (RPA/medical physicist) from workload, use factor, occupancy and design dose targets; typically lead-lined walls/doors (commonly ~1.5–2 mm Pb code equivalents in general rooms, more for CT), lead glass control windows, and warning lights interlocked with exposure. Methodologies: NCRP Report 147 (US)13, BIR/IPEM shielding guidance (UK).
  • Controlled area (UK IRR17): designated where special procedures are needed to restrict significant exposure, in practice every X-ray room when the tube is energised; entry restricted, physical demarcation, warning signs, local rules posted, overseen by a Radiation Protection Supervisor (RPS), with a consulted Radiation Protection Adviser (RPA). IAEA/EU use the same controlled/supervised area concept.
  • Supervised area: conditions kept under review but no special procedures needed (e.g. some control corridors).

Patient protection & diagnostic reference levels

DRLs explained

A diagnostic reference level is an optimisation tool, not a limit: typically the 75th percentile (third quartile) of the dose distribution from a national survey of standard-sized patients for a defined examination. If your room’s median dose for that exam consistently exceeds the DRL, you investigate and optimise; being under the DRL does not mean optimisation is finished (that’s what “achievable doses”, often the median/50th percentile, are for).2,12 UK law (IR(ME)R 2017) requires employers to set local DRLs having regard to national ones and to review when they are consistently exceeded.7

Current national DRL examples: plain radiography

UK national DRLs (UKHSA, effective 11 December 2025; DAP/KAP per complete examination, adult):9

Examination (adult)UK NDRL: DAP per exam (Gy·cm²)UK NDRL: DAP per single radiograph (Gy·cm²)
Chest0.12PA 0.077 · AP 0.092 · mobile AP 0.11 · lateral 0.34
Abdomen2.0AP 1.7
Lumbar spine3.9AP 1.4 · lateral 2.1
Pelvis1.9AP 1.6

Source: UKHSA “National Diagnostic Reference Levels (NDRLs) from 11 December 2025”, Tables 4 & 5 (values adopted 2025 from the UKHSA national patient dose audit).9 UK NDRLs are third-quartile values from national surveys.

US: there is no single federal DRL list for radiography. The ACR–AAPM Practice Parameter for Diagnostic Reference Levels and Achievable Doses in Medical X-ray Imaging defines US DRLs (75th percentile) and achievable doses (median) using national survey data (historically the FDA/CRCPD NEXT surveys and, increasingly, the ACR Dose Index Registry); NCRP Report 172 sets out the US DRL framework.12,13 Because US radiography DRLs are stated as entrance air kerma for specific NEXT protocols and are periodically revised, look values up in the current practice parameter rather than memorising them.

Everyday optimisation that moves dose
  • Collimation: the single most controllable factor. Tight collimation cuts the irradiated volume (dose and scatter, so image contrast improves too). DAP scales directly with field area.
  • Projection choice: PA instead of AP where possible (chest, scoliosis follow-up), which puts dose-sensitive anterior organs (breast, thyroid, gonads-anterior) on the exit side. Compare the UK NDRLs above: chest PA 0.077 vs AP 0.092 Gy·cm².9
  • kVp/mAs technique: higher kVp + lower mAs reduces patient dose for many exams at some contrast cost; correct AEC use; grid removal for small parts/paediatrics.
  • Filtration, SSD, compression, and, in digital radiography, watching the exposure index, because digital systems hide overexposure (“dose creep”).
  • No routine repeats; reject analysis is part of the QA programme.
Gonadal & fetal shielding: the guidance has CHANGED

Since 2019 the major bodies recommend discontinuing routine gonadal and fetal contact shielding for diagnostic X-ray imaging:

  • AAPM Position Statement PS 8-A (formerly PP 32-A) (approved 2 April 2019): “patient gonadal and fetal shielding during X-ray based diagnostic imaging should be discontinued as routine practice”, endorsed by the ACR and many other bodies.14
  • NCRP Statement No. 13 (January 2021) recommends ending routine gonadal shielding during abdominal and pelvic radiography.15
  • The British Institute of Radiology guidance (2020) and the Canadian Association of Radiologists position statement (2021) reach the same conclusion.16

Why: shields frequently obscure anatomy (causing repeat exposures), can trigger automatic exposure control to increase dose, are often mispositioned (ovarian position is variable and ovaries are frequently outside the shielded area anyway), and the heritable-effect risk they were meant to address is now understood to be far smaller than assumed in the 1950s, while most gonadal dose comes from internal scatter a shield cannot block. Follow your department’s policy and local regulations, communicate the change to patients sensitively, and note some jurisdictions have been updating older shielding laws to match.

Paediatric considerations
  • Children’s longer life expectancy and more radiosensitive tissues mean the lifetime stochastic risk per unit dose is higher; justification and optimisation are stricter, and paediatric DRLs are set separately by age or weight band (see UKHSA paediatric tables; Image Gently campaign resources17).
  • Size-based technique charts (weight/age/thickness), no grid for small bodies, tight collimation, immobilisation instead of repeats, and pausing to ask whether ultrasound or MRI answers the question.
  • Comforters/carers who hold children: provided with protection, positioned out of the primary beam, never pregnant, never routinely the same staff member.
Back to contents

Pregnancy

Pregnant patients: the 28-day and 10-day rules

UK joint guidance (HPA/RCR/College of Radiographers, “Protection of pregnant patients during diagnostic medical exposures to ionising radiation”) frames current practice:18

  • Always ask individuals of childbearing potential (commonly 12–55 years) whether they are or might be pregnant whenever the primary beam may irradiate the pelvis (UK guidance: any exposure involving the area between the diaphragm and the knees). Document the response.
  • 28-day rule (most examinations, i.e. low fetal dose): if the patient’s period is not overdue, proceed. If overdue or unsure, treat as potentially pregnant and re-justify.
  • 10-day rule (reserved for higher-dose procedures with fetal dose in the tens of mGy, e.g. some pelvic CT and abdominal/pelvic fluoroscopy): book within the first 10 days of the cycle when pregnancy is very unlikely, unless clinically urgent.
  • If pregnant and the examination is still justified: proceed with optimised technique. A justified examination should not be withheld because of pregnancy when it materially affects the mother’s care. Delaying a necessary diagnosis can harm mother and fetus more than the exposure.
Fetal dose context: plain films are very low dose
ExaminationTypical fetal dose (mGy)
Chest X-ray (any trimester)<0.01
Extremities, skull, dental<0.01
Abdomen AP radiograph0.1–1.0
Pelvis AP radiograph0.1–1.0
Lumbar spine series1.0–10
CT abdomen/pelvis~10–35

Representative values from UK HPA/RCR/CoR guidance and ACOG-cited literature;18,19 individual doses vary with technique and maternal size.

The 100 mGy perspective Deterministic fetal effects (malformation, growth restriction, severe intellectual disability) are associated with thresholds of the order of 100 mGy or more to the fetus, far above anything plain radiography can deliver, and above almost all single diagnostic examinations. ACOG: with few exceptions, radiation exposure from radiography, CT or nuclear medicine is at a dose much lower than that associated with fetal harm, and a clinically indicated examination “should not be withheld from a pregnant patient”.19 The residual stochastic (childhood-cancer) risk at plain-film doses is extremely small compared with the ~1-in-500 baseline childhood cancer incidence. An inadvertent early-pregnancy plain film is not grounds for anxiety about termination. Refer for a proper fetal dose assessment by a medical physicist and counselling.
Pregnant staff
  • Protection is triggered by written declaration of pregnancy to the employer. Declaring is the worker’s choice, but earlier declaration = earlier protection.
  • UK/EU/IAEA: after declaration, working conditions must ensure the dose to the embryo/fetus is unlikely to exceed 1 mSv for the remainder of the pregnancy (IRR17; EU BSS Art. 10; GSR Part 3).4–6
  • US (NRC/state): 5 mSv (0.5 rem) to the embryo/fetus for the whole gestation after declaration, with NCRP recommending ≤0.5 mSv per month.10,13
  • In practice a general radiographer’s fetal dose under an apron is usually far below these figures; pregnancy rarely requires redeployment from general radiography, though duties in high-scatter interventional work are commonly adjusted. A fetal/abdomen-level dosimeter is often issued for reassurance and records.
  • Breastfeeding staff need no X-ray-specific restriction (external X-ray exposure does not make anyone radioactive); restrictions apply only to unsealed radionuclide work.

Quick reference: typical effective doses

Typical adult effective doses, principally from the Mettler et al. catalogue of effective doses in radiology (Radiology 2008)20, the same dataset behind most patient-facing tables (e.g. RadiologyInfo21). Background-equivalent column uses the worldwide average natural background of ~2.4 mSv/year (UNSCEAR)22 ≈ 0.0066 mSv/day. (US average natural background is higher, ~3.1 mSv/yr per NCRP Report 160; UK average 2.1 mSv/yr per UKHSA.)

ExaminationTypical effective dose (mSv)≈ Equivalent natural background
Limb/extremity radiograph0.001a few hours
Intraoral dental radiograph0.005<1 day
Chest X-ray (PA)0.02~3 days
Chest X-ray (PA + lateral)0.1~2 weeks
Cervical spine series0.2~1 month
Mammography (bilateral)0.4~2 months
Pelvis (AP)0.6~3 months
Abdomen (AP)0.7~3–4 months
Hip0.7~3–4 months
Thoracic spine1.0~5 months
Lumbar spine series1.5~7 months
CT head2~10 months
CT chest7~3 years
CT abdomen–pelvis8–10~3–4 years
Barium enema (fluoroscopy)8~3 years

These are population-typical values for a standard adult; individual examinations vary severalfold with equipment, protocol and patient size. Mettler et al. 2008 average values;20 CT values in modern optimised practice are often lower.

Other useful anchors

  • One transatlantic flight (London–New York return): ~0.1 mSv of cosmic-ray dose (about five chest X-rays).
  • Worldwide average natural background: ~2.4 mSv/yr (range commonly 1–13 mSv/yr; some regions tens of mSv/yr).22
  • Radon is the biggest single natural contributor in most countries.
  • Public dose limit (non-medical, per year): 1 mSv, i.e. one year’s limit ≈ 50 chest X-rays’ worth of effective dose.
Talking to patients “A chest X-ray is about the same radiation as three days of the natural background we all live in” lands far better than millisieverts. Pair the comparison with the justification: the information from the image is expected to do you more good than that small exposure could do harm.
Don’t let comparisons breed complacency Background-equivalent framing is for communication, not for skipping optimisation. The system assumes every avoidable millisievert carries some risk (linear no-threshold model1). Justify, optimise, then reassure.
Back to contents

Sources

Sources (24)

24 sources listed

  1. ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007;37(2–4).
  2. ICRP. Radiological Protection in Medicine. ICRP Publication 105. Ann ICRP 2007;37(6).
  3. ICRP. ICRP Statement on Tissue Reactions / Early and Late Effects of Radiation in Normal Tissues and Organs. ICRP Publication 118. Ann ICRP 2012;41(1/2). (Basis of the 20 mSv/yr eye-lens limit.)
  4. IAEA. Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards. General Safety Requirements Part 3 (GSR Part 3). Vienna: IAEA; 2014.
  5. Council of the European Union. Council Directive 2013/59/Euratom laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation. OJ L13, 17.1.2014.
  6. UK Government. The Ionising Radiations Regulations 2017 (IRR17), SI 2017/1075. legislation.gov.uk.
  7. UK Government. The Ionising Radiation (Medical Exposure) Regulations 2017 (IR(ME)R), SI 2017/1322. legislation.gov.uk.
  8. Health and Safety Executive. Working with ionising radiation: Approved Code of Practice and guidance to IRR17 (L121, 2nd edition). HSE; 2018.
  9. UK Health Security Agency. National Diagnostic Reference Levels (NDRLs) from 11 December 2025. GOV.UK; updated 11 December 2025.
  10. US Nuclear Regulatory Commission. 10 CFR Part 20 — Standards for Protection Against Radiation. (Occupational and declared-pregnant-worker dose limits; state X-ray regulations generally mirror these values.)
  11. Conference of Radiation Control Program Directors. Suggested State Regulations for Control of Radiation (SSRCRs). CRCPD. (Model regulations adopted by US states, which regulate X-ray machines.)
  12. American College of Radiology / American Association of Physicists in Medicine. ACR–AAPM Practice Parameter for Diagnostic Reference Levels and Achievable Doses in Medical X-ray Imaging. ACR Practice Parameters and Technical Standards. See also Kanal KM et al., Diagnostic Reference Levels, AJR 2015.
  13. National Council on Radiation Protection and Measurements. NCRP Reports: incl. Report No. 116 (Limitation of Exposure to Ionizing Radiation), Report No. 147 (Structural Shielding Design for Medical X-Ray Imaging Facilities), Report No. 160 (Ionizing Radiation Exposure of the Population of the United States) and Report No. 172 (Reference Levels and Achievable Doses in Medical and Dental Imaging).
  14. American Association of Physicists in Medicine. AAPM Position Statement on the Use of Patient Gonadal and Fetal Shielding (PS 8-A, formerly PP 32-A), approved 2 April 2019. FAQ: Patient Gonadal and Fetal Shielding in Diagnostic Imaging (CARES FAQ).
  15. National Council on Radiation Protection and Measurements. NCRP Statement No. 13: Recommendations for Ending Routine Gonadal Shielding During Abdominal and Pelvic Radiography. January 2021.
  16. Canadian Association of Radiologists. Position Statement: Discontinuing the Use of Gonadal and Fetal Shielding for Patients. May 2021. (The British Institute of Radiology’s 2020 guidance on using shielding on patients for diagnostic radiology applications reaches the same conclusion: bir.org.uk.)
  17. Image Gently Alliance. Image Gently and Digital Radiography. Web page, copyright 2025 Image Gently and SPR.
  18. Health Protection Agency, Royal College of Radiologists, College of Radiographers. Protection of Pregnant Patients during Diagnostic Medical Exposures to Ionising Radiation (RCE-9). 2009 (GOV.UK).
  19. American College of Obstetricians and Gynecologists. Committee Opinion No. 723: Guidelines for Diagnostic Imaging During Pregnancy and Lactation. Obstet Gynecol 2017 (reaffirmed).
  20. Mettler FA, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 2008;248(1):254–263.
  21. RSNA/ACR. RadiologyInfo.org: Radiation Dose in X-Ray and CT Exams (patient-facing dose comparison tables).
  22. United Nations Scientific Committee on the Effects of Atomic Radiation. UNSCEAR 2008 Report: Sources and Effects of Ionizing Radiation, Vol. I (worldwide average natural background exposure ~2.4 mSv/yr).
  23. International Commission on Radiological Protection. ICRP Publication 120: Radiological Protection in Cardiology. Ann ICRP 2013;42(1). Table 6.3, apron transmission at 70–90 kVp.
  24. International Commission on Radiological Protection. ICRP Publication 139: Occupational Radiological Protection in Interventional Procedures. Ann ICRP 2018;47(2). ¶151 and ¶157, garments and lead equivalence.

This page is an educational summary for imaging professionals and students. It is not legal advice and does not replace the regulations, approved codes of practice, employer’s procedures, local rules or the advice of your Radiation Protection Adviser / Qualified Medical Physicist. Dose limits and DRLs are periodically revised. Always confirm against the current primary source before relying on a value in practice. © High Yield Imaging.

More in Reference

Next in Reference: X-ray and CT glossary

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.