MRI artifact recognition atlas

MRI artifact recognition atlas

What it looks like → why it happens → how to fix it. Equipment faults, patient problems and safety incidents are in MRI troubleshooting.

On this pageMotion / ghostingAliasing / wrap-around (fold-over)Chemical shift (type 1)Chemical shift (type 2 / India-ink)Susceptibility / metal bloomGibbs / truncation ringingZipper artifactDielectric (standing-wave) shadingFlow-related (pulsatile ghosting)Herringbone / spikeEPI distortion & N/2 ghostFailed / inhomogeneous fat suppressionResidual unfolding (parallel imaging)Quick rule: which direction does the artifact travel?Troubleshooting: patient, sequence, or hardware?Sources
Related

Motion / ghosting

Motion
Looks like
Repeated faint copies of the anatomy smeared along the phase-encode direction; blurring with breathing/swallowing.
Cause
Position changes between phase-encode steps. Ghosts propagate along phase regardless of true motion direction.
Fix
Immobilise/coach; breath-hold or navigator/triggering; swap phase & frequency directions to move ghosts off the ROI; motion-correction (PROPELLER/BLADE); shorter scans; saturation bands over moving fat.

Aliasing / wrap-around (fold-over)

Encoding
Looks like
Anatomy outside the FOV folds onto the opposite side of the image, along the phase direction.
Cause
FOV smaller than the object in the phase direction → under-sampling; frequencies alias back into the image.
Fix
Enlarge FOV; oversample phase / “no phase wrap”; swap phase-encode direction to the shorter body axis; saturation bands outside FOV; surface-coil roll-off.

Chemical shift (type 1)

Chem shift
Looks like
Bright/dark rim at fat–water interfaces (kidney, orbit, vertebrae) along the frequency direction; worse at higher field.
Cause
Fat resonates ~3.4 to 3.5 ppm below water; scanner mis-maps fat spatially by that frequency offset.
Fix
Increase receiver bandwidth; lower field; fat suppression; swap frequency direction; smaller voxel.
Back to contents

Chemical shift (type 2 / India-ink)

Chem shift
Looks like
Black outline etched around organs on opposed-phase GRE (adrenal, kidney) where fat & water share a voxel.
Cause
At opposed-phase TE, fat & water signals cancel within the same voxel at their interface.
Fix
Usually diagnostic (used for microscopic fat). Compare in/opposed-phase; if unwanted use in-phase TE or Dixon.

Susceptibility / metal bloom

Susceptibility
Looks like
Signal void with bright rim, geometric distortion and dropout around metal, air-tissue, or haemorrhage; worse on GRE/EPI.
Cause
Local B0 inhomogeneity from differing magnetic susceptibilities → rapid dephasing and mis-mapping.
Fix
Use SE/FSE over GRE; short TE; higher bandwidth; thin slices; smaller voxels; MARS/VAT/SEMAC/MAVRIC; lower field. STIR over spectral fat-sat near metal (not after gadolinium).
Back to contents

Gibbs / truncation ringing

Truncation
Looks like
Regular parallel bright/dark lines paralleling high-contrast edges; classic pseudo-syrinx in cord.
Cause
Finite k-space sampling can’t represent a sharp edge (truncated Fourier series).
Fix
Increase matrix/resolution in the affected direction; apodisation filters; smaller voxel. Recognise it to avoid a false syrinx.

Zipper artifact

RF / hardware
Looks like
A bright or noisy line running across the image at a fixed location, often through the FOV centre.
Cause
Stray RF entering the room (failed shield, open door, faulty electronics) or RF hardware/feedback.
Fix
Close the door fully; remove offending electronics; check RF shielding/penetration panel; service if hardware.
Back to contents

Dielectric (standing-wave) shading

RF / field
Looks like
Regional signal drop / central darkening, typically 3 T abdomen; worse with ascites, large habitus, pregnancy.
Cause
RF wavelength approaches body dimensions at 3 T → standing waves & B1 inhomogeneity.
Fix
Dielectric/RF pads; B1-shimming or dual-transmit (not for a pregnant patient unless your MR safety lead advises it; MHRA 2026); lower field; reposition coil.
Back to contents

Herringbone / spike

Hardware
Looks like
Criss-cross corduroy/herringbone striping across the whole image.
Cause
A single bad (“spike”) data point in k-space; electrical arcing/static discharge, faulty component.
Fix
Repeat the acquisition; check for loose/faulty hardware, humidity, gradient coil issues; service engineer.

EPI distortion & N/2 ghost

EPI
Looks like
Stretched/pinched geometry near sinuses (DWI, fMRI); a faint copy shifted by half the FOV (Nyquist ghost).
Cause
EPI’s long readout is very sensitive to B0 offsets (distortion) and odd/even echo mismatch (N/2 ghost).
Fix
Shim well, higher bandwidth, reduced-FOV or readout-segmented EPI, field-map/reverse-PE correction; phase-correction calibration for the ghost.
Back to contents

Failed / inhomogeneous fat suppression

Fat-sat
Looks like
Patchy bright fat where it should be dark, often at edges of FOV or near metal/air.
Cause
B0 inhomogeneity throws fat off its expected frequency so spectral fat-sat mis-hits.
Fix
Localised shimming; centre ROI at isocentre; use STIR (B0-insensitive; not after gadolinium) or Dixon over spectral fat-sat, especially near metal.

Residual unfolding (parallel imaging)

Parallel imaging
Looks like
Central residual fold-over band (SENSE/ASSET) or blotchy noise amplification (g-factor) with high acceleration.
Cause
Errors in the coil-sensitivity/calibration reference or too-high acceleration factor.
Fix
Reduce acceleration (R); re-acquire calibration/ACS lines; ensure FOV covers the object; correct coil geometry.
Back to contents

Quick rule: which direction does the artifact travel?

Phase-encode direction: motion ghosts, flow/pulsation ghosts, aliasing/wrap-around, Gibbs ringing (can occur in either direction; often more obvious along phase). Swapping phase & frequency encoding moves the artifact off your ROI.

Frequency-encode direction: chemical-shift type 1.

Worse at 3 T: chemical shift, susceptibility, dielectric shading, B1 inhomogeneity.

Sequence-dependent: susceptibility & geometric distortion worst on GRE/EPI, mildest on FSE/TSE.

Troubleshooting: patient, sequence, or hardware?

Geometric pattern that does not change with breathing or pulsation (line, herringbone, N/2 ghost, zipper): hardware/RF; check door, stray electronics, RF shield, then service.

Changes with breathing/pulsation: physiologic; coach, immobilise, gate, sat bands, swap phase.

Tied to a tissue interface, metal, or field strength: physics; adjust bandwidth, TE, sequence family, fat-sat method, shimming.

Back to contents

Sources

Key references: Westbrook & Talbot, MRI in Practice, 5th ed (Wiley, 2018). Morelli et al., “An Image-based Approach to Understanding the Physics of MR Artifacts,” RadioGraphics 2011;31:849–866. Zhuo & Gullapalli, “MR Artifacts, Safety, and Quality Control,” RadioGraphics 2006;26:275–297. Elster, Questions & Answers in MRI (mriquestions.com). MHRA, Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, 5th edition, 2026, sections 2.6.3 and 2.6.6 (RF shimming in pregnancy).

More in Physics and foundations

Next in Physics and foundations: Normal MRI anatomy

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.