Siemens Console Parameter Guide

Axial T1 brain MRI (MNI ICBM152 template) labelled: skull, white matter, grey matter, scalp fat: bright. Handwritten note: CSF: dark on T1.
MRI: MNI ICBM152 2009b template. Copyright (C) 1993-2004 Louis Collins, McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University. Labels added.

Siemens Console Parameter Guide

What every setting on the syngo protocol card does, with the physics, the trade-offs, and the practical picks. Focused on MAGNETOM (VE/XA line).

Updated September 2026 · 8 sources listed

On this pageHow to read the consoleContrast cardResolution cardGeometry & SystemFlip-angle modes ★RF pulse types ★HASTE playbook ★Sequence cheat-sheet ★Vendor terms & refsSources
Key points
  • On the Siemens syngo protocol, averages raise SNR and fight motion, while concatenations split slices across multiple TR passes to lift the slice ceiling or drop SAR per pass, at a linear time cost.
  • Siemens RF pulse type: Fast uses short, high-amplitude pulses for the shortest echo spacing and TE but higher per-pulse SAR; Low SAR uses longer, lower-amplitude pulses for the lowest SAR but longer echo spacing and time; Normal sits between.
  • On the Siemens Sequence card, higher bandwidth (Hz/px) shortens echo spacing and reduces chemical-shift and susceptibility artefact but lowers SNR (proportional to 1/√BW); echo spacing is set by bandwidth together with RF pulse type and matrix.
  • Siemens iPAT (GRAPPA) parallel imaging cuts scan time, or the echo-train length in single-shot sequences such as HASTE and EPI, but lowers SNR and adds a risk of reconstruction artefact.
  • Interpolation is not resolution: 2× interpolation doubles the displayed matrix without adding true detail, which comes from base and phase resolution and voxel size.
Explain it simply
  • On Siemens, averages repeat the scan for a cleaner image and fewer motion problems. Concatenations spread slices over several passes to fit more slices or lower heating per pass, each pass adding time.
  • Siemens has three radio-pulse settings. ‘Fast’ uses short, strong pulses: quickest echoes, more heating per pulse. ‘Low SAR’ uses longer, weaker pulses: least heating, but slower. ‘Normal’ sits between.
  • Raising bandwidth on Siemens shortens the gap between echoes and reduces fat-shift and distortion near metal or air, but makes images grainier. The radio-pulse setting and image matrix also set that gap.
  • Siemens’ speed-up setting (iPAT) shortens the scan, or the echo train in single-shot scans, but makes the image grainier and can add artefacts when the image is rebuilt.
  • Upscaling the image (interpolation) makes it look finer but adds no real detail. Real detail comes from the actual sampling settings and the size of each image block.

A guide only: follow your departmental protocols and regional requirements.

How the syngo protocol is organised

Every Siemens protocol opens as a stack of parameter cards (tabs) across the top of the exam card. The settings you touch daily live in five of them.

CardOwnsDaily levers
RoutineThe everyday knobsSlices, TR, TE, averages, concatenations, distance factor
ContrastSignal weightingTE, TR, flip angle, TI, fat sat, magn. prep, turbo/echo train
ResolutionSampling matrixBase/phase resolution, phase FoV, PAT/iPAT, partial Fourier, filters
GeometryWhere & how thickFoV, slice thickness, orientation, saturation regions, tilt
Sequence (Part 1 & 2)The engine roomBandwidth, RF pulse type, flip-angle mode, echo spacing, RF spoiling, asymmetric echo, reordering
PhysioMotion/timingTriggering, gating, dark-blood, navigators
Two settings, flip-angle mode and RF pulse type, both live in Sequence → Part 2. They are the most under-used levers on the console and the ones that most change SAR, echo spacing and blurring.

The three quantities everything trades against

SNR

Grows with voxel volume, √(averages), √(1/bandwidth) and lower acceleration.

SAR / B1+rms

Rises with B0², flip angle², pulses-per-second and shorter TR. The main brake at 3T.

Echo spacing (ESP)

Time between refocusing echoes in TSE/HASTE. Short ESP = less blurring & distortion. Driven by bandwidth and RF pulse duration.

Scan time

TR × phase lines × averages ÷ (turbo factor × acceleration).

Contrast card

SettingWhat it doesPush up →Push down →
TRTime between excitations; T1 weighting & slice countMore slices, more T2/PD, more SNR, longerMore T1, fewer slices, faster
TEEcho time; T2/T2* weightingMore T2, less signalLess T2 (PD/T1), more signal
TIInversion time; nulls a tissue (STIR/FLAIR/MP-RAGE)Set to 0.69×T1 to null a tissue at long TR
Flip angleExcitation (GRE) / refocusing (TSE) angleMore T1 & signal (GRE, to Ernst); more SARLess SAR, flatter contrast, less TSE blurring
Turbo factor / ETLEchoes per TR in TSEFaster, more blurring, longer trainSharper, truer T2, slower
Fat satFat suppression methodSPAIR’s adiabatic inversion tolerates 3T B1 inhomogeneity better than conventional spectral fat-sat (both stay B0-sensitive); Dixon copes well with B0 inhomogeneity, useful for large or off-centre FoV (Del Grande et al. 2014)
Magn. preparationDark-blood, T2-prep, MTC, inversionAdds a prep train before readout (extra SAR & time)
Averages vs. concatenations: averages raise SNR by √N and fight motion; concatenations split slices across multiple TR passes to lift the slice ceiling or drop SAR per pass, at a linear time cost.

Resolution card

SettingWhat it doesTrade-off
Base resolutionReadout (frequency) matrix↑ sharpness, ↓ SNR, cheap in time
Phase resolution %Fraction of phase lines sampled↑ sharpness, ↑ time linearly
Phase FoV %Phase FoV as % of read FoVShrink to save time; too small → wrap
PAT / iPAT (GRAPPA)Parallel-imaging acceleration↓ time (fewer shots in multi-shot TSE; a shorter echo train in single-shot HASTE/EPI), but ↓ SNR by √R×g & recon artifact risk
Partial Fourier (phase)Samples >½ k-space, synthesises rest↓ time (and ↓ effective TE in single-shot EPI/HASTE), ↓ SNR & some blurring. HASTE is partial-Fourier (~60%)
Asymmetric echo (read)Partial Fourier in readoutShortens min TE (short-TE/in-phase GRE)
FiltersNormalise, elliptical, distortion-corrPrescan Normalize evens surface-coil shading; 2D/3D distortion correction fixes gradient-non-linearity geometry (keep it on for measurements, large FoV and radiotherapy planning); elliptical/raw filters change k-space weighting, trading ringing against sharpness
Interpolation ≠ resolution. 2× interpolation doubles the displayed matrix without adding true detail. Real resolution comes from base + phase resolution and voxel size.

Geometry & System / Sequence cards

Geometry

SettingNotes
FoV read / phaseSmaller FoV → higher resolution at fixed matrix, less SNR. Keep anatomy inside to avoid wrap.
Slice thickness / gapThinner = more resolution, less SNR, more slices/SAR. Distance factor = % of slice thickness.
Saturation regionsPre-sat slabs to kill flow/motion/fat ghosts; each adds SAR & time.
Orientation / tilt / phase dirSwap phase & read to move wrap/ghosts off the anatomy of interest.

Sequence: Part 1 (bandwidth & echo)

SettingNotes
Bandwidth (Hz/px)Higher BW → shorter ESP, less chemical-shift & susceptibility, lower SNR (∝1/√BW). Shortens echo spacing, along with RF pulse type and matrix.
Echo spacingOften read-only (result of bandwidth + RF pulse type + matrix). Minimise for HASTE/EPI/diffusion.
Flow compRephases flow: fewer flow ghosts, longer TE/ESP.

Sequence: Part 2 (the engine room)

SettingNotes
RF pulse typeLow SAR / Normal / Fast (see dedicated tab). Trades SAR against echo spacing and minimum timing.
Flip-angle modeConstant / T2-var / hyperecho (see dedicated tab). Trades SAR & blurring against pure T2 contrast.
RF spoilingKills residual transverse coherence in spoiled GRE (FLASH). Leave on for T1 GRE.
Reordering / turbo directionLinear vs centric vs elliptical. Sets which k-space lines land at the effective TE.
Optimisation (SAR)Lets the sequence relax pulses to stay under the SAR limit at a small IQ cost.

Flip-angle modes ★

In a TSE/HASTE echo train, each echo is refocused by an RF pulse. The flip-angle mode decides whether every refocusing pulse is the same angle, or whether the angle is swept along the train. This one setting governs SAR, through-train signal stability and blurring.

ModeRefocusing anglesSARBlurringT2 contrastBest for
Constant (180°)Flat, high (150–180°)HighestRises as the train lengthensTruestThin-slice / high-res, short trains, 1.5T
T2-variable (VFA)Swept, tuned to flatten T2-weighted signalLowLower for long trainsSlightly reducedLong-train 3D SPACE, routine 2D T2 at 3T; variable refocusing angles can also help single-shot HASTE (see note below).
HyperechoSymmetric sweep around a central full-angle hyperechoLowNot ranked here (depends on the angle schedule)ReducedSAR-limited long trains
Echo-train decay and blurring (the k-space reason)

Signal surviving to each echo depends on the refocusing history. With variable/hyperecho low angles the echo amplitudes evolve differently across the train: variable-flip-angle schedules designed to hold the signal steady (T2-var, SPACE) slow the signal decay, so a long train blurs less than a constant high-angle train of the same length, at some cost in T2 contrast and SNR (Busse et al. 2006; Mugler 2014). Both schemes cut SAR well below constant 180° trains; the exact saving depends on the prescribed angles (Hennig & Scheffler 2001; Hennig et al. 2003). Echo spacing is set mainly by RF pulse type, bandwidth and matrix; check the displayed echo spacing after changing flip-angle mode. A longer echo train (in real time) means k-space lines late in the train are more T2-decayed than early ones. That amplitude taper across k-space acts like a filter, broadening the point-spread function → blurring in the phase-encode direction. So the blurring penalty stacks two things: the shape of the decay across the train, and any increase in echo spacing that stretches that decay out.

Well-designed VFA schemes actually stabilise the signal envelope and can reduce blurring versus a naive constant-angle single-shot readout, while slashing SAR, which is why VFA/hyperecho is the default for long trains. The blurring caveat bites hardest when echo spacing is forced up (thin slices, high matrix); see the HASTE section. (Weigel & Hennig; Hyperecho-TSE 3T neuro study; refs tab.)
Cost of low angles: reduced T2 contrast

Because hyperTSE runs low refocusing angles, it delivers less T2 contrast than a constant-180° train at the same TE. On heavily T2-weighted work where lesion conspicuity is everything, nudge the angle up or fall back to constant.

Rule of thumb

Long train, SAR-bound, contrast-tolerant → hyperecho / T2-var. Short train, thin-slice, contrast-critical → constant. The flip-angle value is a separate knob layered on top (see the HASTE section).

T2-var in HASTE: compare before you choose

Variable refocusing flip angles can improve single-shot FSE: in Loening et al. (2015) they stabilised the signal across the train, reduced T2-decay blurring, lowered SAR and increased speed at 3T. Whether Siemens’ T2-var or hyperecho mode suits a given HASTE protocol depends on the design, so compare both on your scanner against the manufacturer’s default protocol and your site protocols. More on the HASTE tab.

RF pulse types ★

The syngo RF pulse type selector (Sequence → Part 2) offers three designs for most sequences. They trade SAR against echo spacing and minimum timing.

TypePulseESP / min TESlice profileSAR / B1+rmsUse when…
Low SARLong duration, low B1 amplitudeLongestNot ranked here: Siemens describes Low SAR pulses as longer, lower-power pulses that may need slightly longer TE and TR; compare slice profile and cross-talk on your own systemLowestSAR-limited and can afford slightly longer TE, TR and echo spacing
NormalBalanced defaultBetween Fast and Low SARNot ranked hereModerateEveryday choice for routine TSE/GRE
FastHigh bandwidth, short durationShortest ESP & TENot ranked here: Fast pulses are the shortest; compare slice profile and cross-talk on your own systemHighest per pulseSequences where short echo spacing makes image quality: HASTE, EPI, TrueFISP, TurboFLASH
The core trade: “Fast” buys the shortest echo spacing (less blurring & distortion in single-shot) at the cost of higher per-pulse SAR. “Low SAR” buys the lowest SAR at the cost of longer echo spacing and time. “Normal” sits between.

How this interacts with flip-angle mode

  • Single-shot, blurring-sensitive (HASTE): priority is minimum echo spacing → favour Fast, and check the displayed echo spacing.
  • Thin-slice high-res (MRCP thin-slab, small-part): train is short → Normal (or Low SAR if SAR-bound), with a constant flip angle for the truest T2 contrast; check slice profile and echo spacing on your scanner.
  • Long routine TSE / 3D SPACE: SAR dominates → Normal pulse + hyperecho/T2-var flip mode.
Watch

Stacking Low SAR pulse + hyperecho mode minimises SAR, but the longer Low SAR pulse lengthens echo spacing. Great for a SAR-crashed sequence, worst for blurring. Don’t reach for both on a sharpness-critical scan.

HASTE playbook ★

HASTE (Half-Fourier Acquisition Single-shot Turbo spin Echo) grabs ~60% of 2D k-space in one very long echo train after a single excitation. Tuning it is a fight against T2 decay across that train.

Why the long train hurts

Because k-space is filled during one long echo train, later lines are acquired at longer effective echo times and are more T2-attenuated, especially short/intermediate-T2 tissue. That amplitude taper across k-space is a low-pass filter → T2 blurring in the phase-encode direction. Half-Fourier worsens it because sampled lines sit further from centre. Minimising echo spacing and stabilising the flip-angle envelope is how you claw sharpness back.

Practical recipe (routine body/abdomen HASTE)
SettingPickWhy
Flip-angle modeHyperecho or T2-var (compare)Long single-shot train is deeply SAR-bound, esp. at 3T; both schemes cut SAR, so compare them on your scanner
Refocusing flip angleReduced (well below 180°)Starting point only: balances T2 contrast, SNR and SAR; values vary by vendor protocol, field strength and body region, so confirm against your scanner’s default HASTE protocol and SAR limits. Lower loses contrast; 180° blows SAR
RF pulse typeFastShortest echo spacing → least T2 blurring & distortion, the dominant single-shot failure mode
BandwidthHighFurther shortens ESP and cuts chemical-shift
PAT / partial FourierOn (GRAPPA 2–3)Shortens effective train → less decay across k-space → less blurring
Starting point: routine HASTE → a reduced refocusing angle with a Fast RF pulse, in hyperecho or T2-var mode (compare both); confirm against your scanner’s default protocol and SAR limits.
T2-variable (VFA) or hyperecho for HASTE?

T2-var holds the echo envelope flat for one prescribed target T2; hyperecho is symmetric around a central full-angle hyperecho at the k-space centre. Both cut SAR well below constant 180°. In single-shot FSE, variable refocusing flip angles stabilised the signal across the train, reduced T2-decay blurring, lowered SAR and increased speed at 3T (Loening et al. 2015). Which scheme suits a given HASTE protocol depends on the vendor’s implementation, so compare both on your scanner against the manufacturer’s default protocol and your site protocols.

The exception: thin-slice / high-res HASTE (e.g. MRCP thin-slab)
Switch to constant flip angle

On thin-slice, high-matrix HASTE the pulses and geometry force echo spacing up. Longer echo spacing stretches the T2 decay across an already-long train → more blurring exactly where you need crisp thin-slice detail. Echo spacing is set mainly by RF pulse type, bandwidth and matrix, so keep it tight with a Fast/Normal pulse and check the displayed value after any change; a constant flip angle gives the truest T2 contrast. You pay in SAR; manage it with concatenations, TR, fewer slices per pass or PAT rather than by softening the flip-angle scheme.

So the decision rule is: routine thick-slice HASTE → a reduced refocusing angle (hyperecho or T2-var, compared on your scanner); thin-slice / resolution-critical HASTE (MRCP thin slabs, small-part) → constant flip angle, protect echo spacing.

3T localiser caveat
HASTE localisers at 3T

At 3T the shorter RF wavelength drives B1 inhomogeneity and dielectric (standing-wave) shading across the body, and the long HASTE train is SAR-heavy and contrast-soft. On a fast localiser that shows up as uneven brightness, signal dropout and poor tissue definition, so a HASTE localiser can be degraded at 3T. A fast GRE or TrueFISP-type localiser is a common alternative for planning at 3T (local practice, not a published standard), and use dielectric pads + patient-adaptive B1 shimming when body HASTE itself is needed.

RF pulse type summary for HASTE

HASTE is the textbook home of the Fast RF pulse; measurement speed and short echo spacing are exactly what it needs. Only drop to Normal/Low-SAR if SAR or another constraint forces it, accepting the echo-spacing (blurring) penalty.

Sequence cheat-sheet ★

Starting-point picks for the two levers that matter most, flip-angle mode and RF pulse type, plus the one thing to protect on each sequence. Values are defaults to reason from, not site protocols.

SequenceFlip-angle modeRF pulseProtectNotes
2D T2 TSE (routine)T2-var / hyperecho at 3T; constant OK at 1.5TNormalTurbo factor not too highVFA cuts SAR at 3T; keep ETL modest to limit blurring
2D T1 TSEConstant (short train)NormalShort TE, low ETLShort echo train → constant angle keeps contrast crisp
PD/T2 dual-echo TSEConstant / T2-varNormalEcho spacingBalance the two effective TEs
FLAIRT2-var / hyperechoNormalTI (CSF null), SARLong TR/TE/TI is SAR-heavy → VFA helps a lot at 3T
STIRT2-var / hyperechoNormalTI (fat null ~0.69×T1fat)Adiabatic inversion adds SAR; VFA offsets it
HASTE (thick-slice)Hyperecho or T2-var, reduced angle (confirm locally)FastEcho spacing (blurring)The workhorse setting; short ESP is everything
HASTE / thin-slab MRCPConstantFast / NormalEcho spacingLong echo spacing blurs thin slices (set by pulse type, bandwidth and matrix); keep angle flat
3D MRCP (SPACE, resp-trig)T2-var (SPACE VFA)NormalSAR, navigator efficiencyVery long 3D train; VFA mandatory for SAR & SNR
3D T2/PD SPACET2-var (VFA core of SPACE)NormalSAR, blurringSPACE is a VFA sequence by design
3D T1 MP-RAGEn/a (GRE readout)NormalTI, spoilingContrast set by prep inversion, not refocus angles
DWI / EPIn/a (single-shot EPI)FastEcho spacing, distortionShort ESP + high BW + PAT to cut distortion/blur
TrueFISP / bSSFPn/a (balanced GRE)FastShim (banding), TRFast pulse for short TR; good 3T localiser alternative
FLASH / VIBE (T1 GRE)n/a (spoiled GRE)NormalRF spoiling on, Ernst angleFlip value near Ernst for T1; keep spoiling on
TOF MRAn/a (spoiled GRE)NormalFlip for inflow, MTC SARHigher flip boosts background suppression at SAR cost
Two quick heuristics. (1) Long train + SAR-bound + can tolerate slightly softer T2 → variable/hyperecho. (2) Single-shot or thin-slice where sharpness is king → protect echo spacing: Fast RF pulse and high bandwidth; in single-shot readouts, PAT also shortens the echo train.

See the Flip-angle, RF-pulse and HASTE tabs for the reasoning behind each pick.

Vendor terminology (Siemens → others)

SiemensGEPhilipsCanon
TSEFSETSEFSE
HASTESSFSESSH-TSE / Ultra-fast SEFASE
Turbo factorETLTurbo factorETL / shot factor
SPACE (3D TSE, variable flip angle); hyperecho is a separate Siemens refocusing schemeCUBE (VFA)VISTA (VFA)3D MVOX
iPAT (GRAPPA)ASSET / ARCSENSESPEEDER
TrueFISPFIESTAbFFETrue SSFP
FLASH (spoiled GRE)SPGRT1-FFEField Echo
TurboFLASHFastSPGR / FGRETFEFast FE
SPAIRASPIRSPAIRSPAIR

Sources

  1. Hennig J, Weigel M, Scheffler K. Hyperecho / variable flip angle refocusing theory; Development & optimization of Hyperecho-TSE for reduced RF power. ScienceDirect
  2. Hyperecho-Turbo Spin-Echo Sequences at 3T: Clinical Application in Neuroradiology. AJNR 2008;29(5):956. AJNR · PMC
  3. Increased speed and image quality in single-shot FSE via variable refocusing flip angles. PMC4684814
  4. HASTE with parallel acquisition and T2-decay compensation. PMC2630054
  5. HASTE / SS-FSE physics: blurring, effective TE, half-Fourier. Questions & Answers in MRI · Hyperechoes
  6. RF pulse types (Low SAR / Normal / Fast) & SAR reduction. How to reduce SAR
  7. Siemens MRI Acronyms: Cross-Vendor Comparison. Siemens Healthineers PDF
  8. Ultrafast T2-weighted HASTE at 3T (blurring / DL acceleration). PMC12265358

Parameter behaviour reflects the syngo MR VE/XA interface; exact labels vary by software line and sequence. Recipe values are starting points. Confirm against your site’s protocols and SAR limits.

More in Protocols and technique

Everything on MRI: MRI hub. Or search the whole reference.

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.