MRI Sequence Types: The Full Catalogue

Sagittal T1 brain MRI near the midline (MNI ICBM152 template) labelled: corpus callosum, pons, fourth ventricle, cerebellum. Handwritten note: face on the left.
MRI: MNI ICBM152 2009b template. Copyright (C) 1993-2004 Louis Collins, McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University. Labels added.

MRI Sequence Types: The Full Catalogue

Every major sequence family and named technique you will meet on the console and in reports, grouped by how they generate signal, with the physical principle, the clinical job it does, and the equivalent name on each vendor. For the parameter physics behind these families (TR/TE/flip, k-space), see the Sequence Physics Explainer; for tuning them, see the Sequence Optimiser.

Updated September 2026

On this pageMaster comparisonSpin-echo familyInversion-recovery familyGradient-echo familyEcho-planar (EPI) readoutDiffusion, perfusion & advanced contrastVascular & flow: MRA and MRVFat–water separation & suppressionVendor name Rosetta tableSequence decision guide: choosing between close alternativesSources
Key points
  • Spin echo’s 180° refocusing pulse cancels static field inhomogeneity, so spin echo is robust to it; gradient echo replaces the 180° with a gradient reversal, so it is very fast but susceptibility-sensitive.
  • In inversion recovery, STIR nulls fat and FLAIR nulls CSF; do not use STIR after gadolinium, because it can also null enhancing tissue.
  • Always read DWI with the ADC map: restricted water stays bright on DWI, and the ADC map removes T2 shine-through.
  • Vendors give the same physics different names: fast spin echo is TSE on Siemens and Philips but FSE on GE and Canon.
  • When renal function rules out gadolinium, use non-contrast MRA (TOF, phase contrast, QISS or FBI); phase contrast also answers questions about flow direction or velocity.
Explain it simply
  • Spin echo scans use an extra radio pulse that cancels uneven magnetism, so they resist the faults it causes. Gradient echo scans skip it: much faster, but more easily disturbed by small magnetic differences.
  • STIR scans make fat go dark and FLAIR scans make the fluid around the brain and spine go dark. Do not use STIR after contrast: it can also darken the parts the contrast lights up.
  • Diffusion scans (DWI) show water that cannot move freely as bright. Always read them with the matching ADC map, which removes a misleading brightness called T2 shine-through.
  • Scanner makers give the same type of scan different brand names. The same fast spin echo scan is called TSE on Siemens and Philips scanners and FSE on GE and Canon ones.
  • Blood vessels can be scanned without a contrast injection when kidney problems rule out the dye. One method, phase contrast, can also show which way and how fast blood flows.

A guide only: follow your departmental protocols and regional requirements.

1 · Master comparison

The 20-second orientation. Detailed cards for each follow below.

FamilyCore mechanismTypical weighting/outputBread-and-butter use
Spin echo (SE / FSE·TSE)90° then 180° refocusing pulse(s)T1, PD, T2: the reference standardAlmost everything; robust to field inhomogeneity
Inversion recovery (STIR, FLAIR, PSIR, MP-RAGE)180° inversion + tailored TI to null a tissueFat-null, fluid-null, or high-contrast T1 3DOedema, MS plaques, fat suppression, volumetric brain
Gradient echo: spoiled (FLASH/SPGR/T1-FFE)Gradient reversal, no 180°, spoiled residual signalT1 / T2* fast imagingBreath-hold T1, dynamic post-contrast, 3D VIBE
Gradient echo: balanced SSFP (TrueFISP/FIESTA/bFFE)Fully rewound gradients, coherent steady stateHigh-SNR T2/T1 ratio, bright fluid & bloodCardiac cine, fetal, MRCP, inner ear
Echo-planar (EPI)Single/multi-shot readout of whole k-space per excitationVery fast; carries DWI, perfusion, BOLDDiffusion, DSC perfusion, functional MRI
Diffusion / functionalMotion-probing gradients or physiologic contrastDWI/ADC, DTI, ASL, BOLD, SWI, MRSStroke, tractography, tumour, microbleeds, spectra
Vascular / flow (MRA·MRV)Inflow, phase shift, or contrast bolusArteriograms, venograms, flow quantificationCircle of Willis, carotids, aorta, DVT/sinus thrombosis
Fat–water / suppressionChemical-shift separation or selective saturationFat-only, water-only, in/out-of-phaseAdrenal/liver fat, marrow, uniform fat-sat

2 · Spin-echo family

A 90° excitation followed by one or more 180° refocusing pulses. The 180° cancels static field inhomogeneity, so SE is the family most resistant to susceptibility and field-inhomogeneity artefact and the benchmark for tissue contrast.

Spin echo

Conventional SE

Principle
One 180° refocus per TR. True T1, PD or T2 depending on TR/TE.
Use
Reference T1 (short TR/TE). Largely replaced by FSE for T2/PD because it is slow.

▲ Clean contrast, low artefact · ▼ Long acquisition

Spin echo

Fast / Turbo SE (FSE · TSE)

Principle
An echo train: multiple 180° pulses per TR fill several k-space lines (echo-train length/turbo factor), cutting scan time in proportion.
Use
The workhorse for T2 and PD everywhere. Fat stays brighter than on true SE (reduced J-coupling), so often paired with fat-sat.

Vendor: Siemens TSE · GE FSE · Philips TSE · Canon FSE

Spin echo

Single-shot FSE (SSFSE · HASTE)

Principle
All (or half, with half-Fourier) k-space lines in one very long echo train: a whole slice in under a second.
Use
Motion-insensitive T2: fetal, uncooperative patients, MRCP, myelography.

Vendor: Siemens HASTE · GE SSFSE · Philips SSh-TSE · Canon FASE

Spin echo

3D FSE volumetric (SPACE·CUBE·VISTA)

Principle
3D FSE with variable-flip refocusing pulses to sustain a long echo train; isotropic sub-mm volume, reformat in any plane.
Use
3D T2/FLAIR/STIR of brain, spine, MSK; one acquisition reformatted instead of three 2D stacks.

Vendor: Siemens SPACE · GE CUBE · Philips VISTA · Canon 3D MVOX

3 · Inversion-recovery family

A 180° inversion pulse precedes the imaging readout. By choosing the inversion time (TI), the signal from one tissue is caught at its null (zero crossing) and disappears. See the TI-to-null calculator for the numbers.

Inversion recovery

STIR (Short TI IR)

Principle
Short TI (~150–170 ms at 1.5T) nulls fat by its short T1. Suppression is B0/B1-insensitive, so it works where chemical fat-sat fails.
Use
Marrow/soft-tissue oedema, MSK, whole-body. Do not use with gadolinium; it can also null enhancing tissue.
Inversion recovery

FLAIR (Fluid-Attenuated IR)

Principle
Long TI (~2000–2500 ms) nulls CSF while keeping heavy T2 weighting: bright lesions on a dark-fluid background.
Use
Periventricular MS plaques, gliosis, SAH detection. 3D FLAIR (SPACE/CUBE) for small cortical/juxtacortical lesions.
Inversion recovery

PSIR (Phase-Sensitive IR)

Principle
Recovers the sign of the inverted signal, doubling the effective T1 dynamic range and giving crisp grey–white contrast.
Use
Cortical/spinal-cord MS lesions, late-gadolinium cardiac viability imaging.
Inversion recovery

MP-RAGE / 3D IR-SPGR

Principle
A single inversion prepares magnetisation, then a fast spoiled-GRE readout: isotropic, high grey–white-contrast T1 volume.
Use
Volumetric brain, neuro-navigation, volumetry. MP2RAGE variant removes B1 bias for quantitative T1.

Vendor: Siemens MPRAGE · GE BRAVO/3D IR-SPGR · Philips 3D TFE · Canon 3D FFE

Inversion recovery

DIR (Double IR)

Principle
Two inversion pulses null two tissues at once (e.g. CSF and white matter).
Use
Cortical/juxtacortical MS lesions; black-blood cardiac and vessel-wall imaging.

4 · Gradient-echo family

A gradient reversal replaces the 180°, so flip angles are small and TR/TE short; it is very fast, but sensitive to susceptibility (T2*). Two branches: spoiled (residual transverse signal destroyed each TR) and balanced steady-state (residual signal preserved and reused).

Spoiled GRE

Spoiled GRE (FLASH·SPGR·T1-FFE)

Principle
RF or gradient spoiling removes leftover transverse magnetisation → clean T1 (or PD/T2* by TE).
Use
Breath-hold T1, dynamic contrast, angio source images.

Vendor: Siemens FLASH · GE SPGR · Philips T1-FFE · Canon FastFE

Spoiled GRE 3D

3D fat-sat spoiled GRE (VIBE·LAVA·THRIVE)

Principle
3D spoiled GRE with Dixon/spectral fat-sat: thin isotropic T1 volume in a breath-hold.
Use
Dynamic liver, pre/post-contrast abdomen & pelvis, breast.

Vendor: Siemens VIBE · GE LAVA · Philips THRIVE · Canon Quick3D/LAVA-like

Balanced SSFP

Balanced SSFP (TrueFISP·FIESTA·bFFE)

Principle
All gradients rewound each TR → coherent steady state with very high SNR and a T2/T1 contrast; fluid and flowing blood are bright.
Use
Cardiac cine, fetal, MRCP, cranial nerves/inner ear, bowel. Prone to dark banding artefact at field inhomogeneity (needs good shim/short TR).

Vendor: Siemens TrueFISP · GE FIESTA · Philips Balanced-FFE · Canon True SSFP

Non-balanced SSFP

Coherent/steady-state variants (FISP·GRASS · PSIF·SSFP-echo · DESS·MERGE)

Principle
Partially rewound (FISP/GRASS) or time-reversed (PSIF) steady states; DESS/MEDIC combine both for T2*-like cartilage contrast.
Use
Cartilage (DESS), spine/nerve (MEDIC/MERGE), flow-tolerant fast imaging.

Vendor: Siemens FISP/PSIF/DESS/MEDIC · GE GRASS/SSFP/MERGE · Philips FFE/T2-FFE

SWI is a GRE too. Susceptibility-weighted imaging is a high-resolution flow-compensated 3D spoiled GRE that combines magnitude and phase to exaggerate susceptibility. It is covered under functional/advanced below.

5 · Echo-planar (EPI) readout

Not a contrast in itself but an ultra-fast readout: an oscillating gradient traverses much or all of k-space after a single excitation. It is the delivery vehicle for diffusion, perfusion and BOLD, at the cost of geometric distortion and susceptibility drop-out near air/bone.

EPI

Single-shot EPI

Principle
Whole slice in one shot (~50–100 ms). Freezes motion but distorts at air interfaces.
Use
Standard DWI, DSC perfusion, fMRI.
EPI

Multi-shot / readout-segmented EPI

Principle
k-space split across shots (e.g. RESOLVE) → less distortion, higher resolution.
Use
High-resolution DWI of orbit, skull base, spine, breast.

Vendor: Siemens RESOLVE · GE MUSE · Philips multi-shot DWI

6 · Diffusion, perfusion & advanced contrast

Sequences whose contrast comes from tissue physiology or microstructure rather than plain T1/T2.

Diffusion

DWI + ADC

Principle
Paired motion-probing gradients dephase moving water; restricted water stays bright. Acquire ≥2 b-values → ADC map removes T2 shine-through.
Use
Hyperacute stroke, abscess vs necrosis, cellular tumour, cholesteatoma. Always read DWI with the ADC map.
Diffusion

DTI / DKI / tractography

Principle
Diffusion sampled in ≥6 directions → anisotropy (FA) and fibre orientation; DKI adds non-Gaussian kurtosis.
Use
White-matter tract mapping for surgical planning, microstructural research.
Perfusion

DSC perfusion

Principle
Dynamic T2*-EPI tracks a gadolinium bolus first pass → CBV, CBF, MTT.
Use
Tumour grading, recurrence vs radiation necrosis, stroke penumbra.
Perfusion

DCE perfusion

Principle
Dynamic T1 spoiled-GRE tracks contrast leakage → permeability (Ktrans).
Use
Prostate (mpMRI), breast, tumour angiogenesis.
Perfusion

ASL (Arterial Spin Labelling)

Principle
Magnetically labels arterial blood water as an endogenous tracer (no contrast agent).
Use
Cerebral perfusion in children/renal-impaired patients, tumour, ischaemia.
Functional

BOLD fMRI

Principle
T2*-EPI detects deoxy/oxy-haemoglobin ratio changes with neuronal activity.
Use
Pre-surgical eloquent-cortex mapping, research.
Susceptibility

SWI / SWAN / T2*-GRE

Principle
Flow-compensated 3D spoiled GRE combining magnitude + phase to amplify susceptibility from blood products, iron, calcium.
Use
Microbleeds, DAI, cavernomas, mineralisation, vein mapping.

Vendor: Siemens SWI · GE SWAN · Philips SWIp · Canon SWI/FSBB

Spectroscopy

MRS (single-voxel / CSI)

Principle
Resolves metabolite chemical shifts (NAA, choline, creatine, lactate, lipid) instead of an image.
Use
Tumour vs necrosis, metabolic disease, prostate.

7 · Vascular & flow: MRA and MRV

Three physical routes to a vessel image: inflow (TOF), velocity-encoded phase (PC), and contrast bolus (CE-MRA), plus time-resolved and non-contrast variants built on them. Venography (MRV) uses the same physics tuned to slow venous flow.

Show the full section
Inflow

Time-of-Flight MRA (2D & 3D TOF)

Principle
Stationary tissue is saturated by rapid RF; fresh unsaturated blood flowing in is bright. No contrast. A travelling saturation band suppresses either arteries or veins by inflow direction.
Use
3D TOF for Circle of Willis/intracranial arteries; 2D TOF for slower/longer runs (carotids, peripheral, and TOF venography).

▲ No contrast, great for COW · ▼ Slow flow saturates, in-plane flow lost, overestimates stenosis

Phase

Phase-Contrast MRA (PC)

Principle
Bipolar gradients give moving spins a phase shift proportional to velocity. You set a VENC (encoding velocity); aliasing occurs if true velocity exceeds it.
Use
Directional & quantitative flow, slow venous flow (PC-MRV, dural sinuses), CSF flow, distinguishing thrombus from slow flow.

▲ Quantitative, direction-encoded, background fully suppressed · ▼ Slow; VENC must match expected velocity

Contrast

Contrast-Enhanced MRA (CE-MRA)

Principle
Fast 3D spoiled GRE during a timed gadolinium bolus shortens blood T1 directly; signal comes from contrast, not flow, so flow artefacts largely vanish. Timed by test-bolus or bolus-tracking (SmartPrep/CARE-Bolus/BolusTrak).
Use
Aorta, renal, carotid, run-off/peripheral, pulmonary arteries.

▲ Fast, large FOV, few flow artefacts · ▼ Needs gadolinium & correct bolus timing (venous contamination if late)

Time-resolved

Time-resolved CE-MRA (TWIST·TRICKS·4D-TRAK·DISCO)

Principle
Repeatedly under-samples k-space (sampling the centre often, the periphery sparsely) to make a rapid 3D dynamic series that captures arterial → venous phases. Removes the need for precise bolus timing and adds temporal information.
Use
AVMs/fistulae, distal run-off, hand/foot, congenital and dialysis-access vessels where flow dynamics matter.

Vendor: Siemens TWIST · GE TRICKS · Philips 4D-TRAK · Canon DISCO-like/Freeze-Frame

4D flow

4D flow MRI (PC-VIPR)

Principle
Time-resolved 3-directional velocity encoding over a 3D volume: flow through the whole territory across the cardiac cycle.
Use
Aortic & congenital-heart haemodynamics, wall shear stress, shunt/valve quantification (research → clinical).
Non-contrast

Non-contrast MRA: ECG-gated FSE (FBI/NATIVE-SPACE)

Principle
3D FSE subtraction of bright-artery systolic and dark-artery diastolic images (fresh-blood imaging). No gadolinium.
Use
Peripheral run-off in renal-impaired patients (NSF avoidance).

Vendor: Canon FBI · Siemens NATIVE SPACE · Philips TRANCE · GE Delta Flow

Non-contrast

QISS (Quiescent-Interval Single-Shot MRA)

Principle
2D single-shot balanced-SSFP acquired station-by-station, triggered off the cardiac cycle, either ECG-gated (leads on the chest) or peripheral-pulse-gated (a finger/toe pulse-oximetry trigger), the latter often more reliable at distal stations or when ECG signal is degraded by gradient/RF noise or an irregular rhythm. A slice-selective saturation pulse is followed by a “quiescent interval” (~150–200 ms, timed to diastole) that lets fresh unsaturated arterial blood flow into the section while a travelling venous saturation band keeps veins dark: bright artery on a suppressed background, no contrast.
Use
Peripheral run-off (aorta to foot); the best-validated non-contrast option for lower-extremity peripheral arterial disease, particularly when gadolinium is contraindicated (severe renal impairment / NSF risk).

▲ No contrast, robust venous suppression, station-by-station coverage, gating flexible (ECG or pulse) · ▼ Slower than CE-MRA; needs a regular, reliably detected cardiac trigger; arrhythmia, poor pulse signal, or a mistimed saturation band degrades quality

Non-contrast

Other inflow / labelling non-contrast MRA (NATIVE TrueFISP · Inhance · Time-SLIP)

Principle
Balanced-SSFP inflow techniques (NATIVE TrueFISP, Inhance) brighten arterial inflow the same way as QISS. Time-SLIP instead uses a spatially selective ASL-style tag placed proximal or distal to the target vessel; imaging after a chosen delay shows only the labelled blood that has flowed into the region: a selective, direction-controlled single-vessel “roadmap” angiogram.
Use
Renal arteries, portal/hepatic and other abdominal vessels, paediatric imaging, and any renal-impaired patient where gadolinium should be avoided.

Vendor: Siemens NATIVE TrueFISP · GE Inhance · Philips TRANCE/b-TRANCE · Canon Time-SLIP

Venography

MR venography (MRV)

Principle
Same three routes, tuned to slow venous flow: 2D TOF venography (dural sinuses/legs), PC-MRV (direction & quantification, thrombus vs slow flow), and CE-MRV (large-territory venous mapping).
Use
Cerebral venous sinus thrombosis, DVT, venous mapping. SWI complements MRV for slow/occluded veins.
Quick chooser. Suspected sinus thrombosis → 2D TOF or PC-MRV (± CE-MRV). Renal impairment, want no gadolinium → TOF, PC, QISS or FBI. Need flow numbers or direction → phase contrast / 4D flow. Big arterial territory, fast → CE-MRA. AVM / dynamic filling → time-resolved (TWIST/TRICKS).

8 · Fat–water separation & suppression

Not standalone families but techniques layered onto SE/GRE to control fat signal; a frequent source of confusion, so grouped here.

Chemical shift

In-/opposed-phase GRE (chemical shift)

Principle
Two TEs where fat & water are in phase vs opposed; voxels with both drop signal on opposed-phase.
Use
Adrenal adenoma, hepatic steatosis, marrow.
Dixon

Dixon (multi-echo)

Principle
Acquires multiple echoes and computes four sets (in-phase, opposed-phase, water-only, fat-only) in one pass. Robust uniform fat-sat + fat-fraction maps.
Use
Uniform fat suppression over large/off-isocentre FOV, PDFF quantification.

Vendor: Siemens Dixon · GE IDEAL · Philips mDixon · Canon WFS/Dixon

Spectral sat

SPAIR · SPIR · fat-sat · water excitation

Principle
Frequency-selective pulses saturate (SPIR) or invert-and-null (SPAIR) fat, or excite only water. SPAIR is more B1-robust than plain fat-sat.
Use
Post-contrast fat-sat (unlike STIR, compatible with gadolinium), cartilage, breast.

9 · Vendor name Rosetta table

Same physics, different marketing. Read across the row.

GenericSiemensGEPhilipsCanon
Fast/Turbo SETSEFSETSEFSE
Single-shot FSEHASTESSFSESSh-TSEFASE
3D FSE volumeSPACECUBEVISTA3D MVOX
Spoiled GRE T1FLASHSPGRT1-FFEFastFE
3D fat-sat spoiled GREVIBELAVATHRIVE3D Quick
Balanced SSFPTrueFISPFIESTAbFFETrue SSFP
3D IR T1 (volumetric brain)MPRAGEBRAVO3D TFE3D FFE
Susceptibility-weightedSWISWANSWIpSWI/FSBB
Time-resolved CE-MRATWISTTRICKS4D-TRAKFreeze-Frame/DISCO-like
Non-contrast inflow MRA (QISS-type)NATIVE TrueFISPInhanceTRANCE/b-TRANCETime-SLIP
Dixon fat-waterDixonIDEALmDixonWFS/Dixon

Sequence decision guide: choosing between close alternatives

Practical “this vs that” calls that come up when building or tuning protocols: what each option buys you, what it costs, and when to pick which. Educational reference; always validate against your local protocol and radiologist preference.

1 · T1 MPRAGE brain post-contrast: fat-sat vs non-fat-sat

Post-gadolinium 3D T1 MPRAGE of the brain can be run with or without fat suppression. The choice depends on what you’re looking for and where.

 Non-fat-sat MPRAGEFat-sat MPRAGE
Best forParenchymal / intra-axial enhancement, general post-Gd brain, volumetricsSkull base, orbits, cavernous sinus, IACs, scalp/calvarium, post-op beds, and anywhere marrow/orbital fat masks enhancement
WhyBright T1 fat is a familiar anatomical backdrop; maximal SNR & grey–white contrast for brainSuppressing bright fat unmasks subtly enhancing lesions next to fat (dural tails, perineural spread, orbital/skull-base disease) that would otherwise be iso-intense to fat
CostEnhancement adjacent to fat can be hiddenLower SNR; fat-sat failure/inhomogeneity at skull base & air interfaces; slightly longer
Rule of thumb: routine brain / neuro-oncology parenchymal work → non-fat-sat (or add a Dixon that gives both). Skull base, orbit, IAC, cranial-nerve, meningeal or extra-axial questions → fat-sat post-contrast. Many centres run a Dixon MPRAGE/VIBE to get in-phase and fat-suppressed T1 from one acquisition.
2 · Whole-spine imaging: MSK vs neuro emphasis, and FS vs non-FS post-contrast

“Whole spine” means different things to a neuro service (cord, canal, leptomeninges) and an MSK/oncology service (marrow, bones, soft tissue). That drives sequence weighting and fat-sat choice.

 Neuro emphasisMSK / marrow emphasis
QuestionCord signal, compression, MS/myelitis, leptomeningeal/drop mets, CSFMarrow infiltration, mets, discitis-osteomyelitis, fractures, soft-tissue mass
Core sequencesSag T2 & T1, STIR, axials through lesions; post-Gd T1 for tumour/infection/inflammationSag T1 (marrow), STIR/T2 fat-sat (oedema), Dixon; post-Gd T1 fat-sat
Post-contrast fat-sat?Yes, generally fat-sat: spine marrow & epidural fat are bright on T1; unsuppressed, enhancing leptomeningeal/epidural/marrow disease blends into fat. Fat-sat (or subtraction) makes enhancement conspicuousYes, fat-sat: essential to see marrow & soft-tissue enhancement against bright fatty marrow
When non-fat-sat T1 post-Gd helpsAs an anatomical companion, or via subtraction (pre/post non-fat-sat) where fat-sat is inhomogeneous over a long FOVRarely alone; pairs with pre-contrast T1 for subtraction
Long-FOV caveat: spectral fat-sat is often inhomogeneous across a whole-spine FOV (B0 variation top-to-bottom). Use STIR (fluid-sensitive) for the pre-contrast oedema screen and Dixon or subtraction for reliable post-contrast fat suppression over long coverage.
Post-contrast whole-spine is almost always fat-suppressed (Dixon or spectral) or read with subtraction; unsuppressed T1 post-Gd hides enhancement in fatty marrow/epidural fat. Choose neuro vs MSK weighting by the clinical question, but the fat-sat answer is the same.
3 · Black-blood imaging: pre- vs post-contrast

Black-blood techniques (double/triple inversion recovery, or T1 SPACE/CUBE/VISTA with flow-suppression) null flowing blood so the vessel wall, meninges or myocardium stand out against a dark lumen.

 Pre-contrast black-bloodPost-contrast black-blood
ShowsBaseline wall thickness, morphology, intramural haematoma/plaque, oedemaWall/leptomeningeal/plaque enhancement, the key marker of active inflammation, vulnerable plaque, or leptomeningeal/dural disease
RoleReference for comparison; without it you can’t call enhancementThe diagnostic series in vessel-wall imaging, black-blood brain-met detection & leptomeningeal disease
  • Vessel-wall MRI (intracranial vasculitis, dissection, plaque): pre and post black-blood T1 are both needed; enhancement is only meaningful against the matched pre-contrast image.
  • Black-blood post-Gd T1 for brain metastases / leptomeningeal disease: suppressing bright vascular signal stops small enhancing vessels mimicking or hiding tiny enhancing mets and improves small-lesion conspicuity vs bright-blood post-Gd.
Rule: for wall/leptomeningeal enhancement questions, acquire matched pre- and post-contrast black-blood with identical parameters. Post alone answers “does it enhance?” only if you trust a prior baseline; for plaque/vasculitis you want both.
4 · T2 SPACE (3D TSE) vs conventional 2D T2 TSE

SPACE (Siemens) / CUBE (GE) / VISTA (Philips) / 3D MVOX (Canon) is a single-slab 3D turbo-spin-echo with a long, variable-flip echo train. How it compares to standard 2D T2 TSE:

 2D T2 TSET2 SPACE / 3D TSE
ResolutionGood in-plane; thick slices with gapsIsotropic sub-mm: reformat in any plane from one acquisition
Coverage/planesOne plane per acquisition; repeat for eachOne slab → axial/sag/cor reformats; replaces several 2D series
SNR per voxelHigher per voxel (thicker slices)Lower per (tiny) voxel, offset by 3D averaging & thin slices
Blurring/contrastCrisp; conventional T2 contrastVariable-flip train → some T2 blurring & slightly altered contrast; excellent for CSF/fluid & small structures
Time & motionFast per plane; robustLonger single acquisition; more motion-sensitive (whole slab degrades)
Flow/pulsationCan have CSF flow artefactBetter flow suppression; good for cisternal nerves, internal auditory canal, spine canal
Pick SPACE/3D TSE when you need multiplanar reformats, thin isotropic slices, small-structure or cisternal/nerve detail, or want to consolidate planes. Pick 2D T2 TSE for speed, higher per-voxel SNR, motion-prone patients, or when conventional crisp T2 contrast in a single plane is all that’s needed.
5 · STEALTH T1 MPRAGE neuronavigation: 0.9 mm vs 1.0 mm isotropic

For a StealthStation / neuronavigation 3D T1 (typically post-contrast MPRAGE covering the whole head), the isotropic voxel choice trades registration/detail against SNR and time.

 1.0 mm isotropic0.9 mm isotropic
DetailStandard for navigation; adequate for most cranial cases~30% smaller voxel volume → finer surface/lesion detail, marginally better small-target localisation
SNRHigher per voxel (voxel is ~37% larger)Lower SNR per voxel; needs recovering (averages, BW, DL-recon) or images look noisier
TimeShorterLonger (more encoding steps for full-head coverage)
Navigation uploadUniversally accepted; whole-head coverage & contiguous slices matter more than 0.1 mmAccepted; benefit is real only if SNR is maintained and the patient holds still for the longer scan
Practical call: 1.0 mm isotropic is the robust default: reliable SNR, full-head coverage, contiguous no-gap slices, faster (less motion risk), and fully adequate for frameless navigation. Go 0.9 mm only when the surgical target genuinely needs the extra detail and you can defend SNR (DL-recon or an extra average) without an unacceptable time/motion penalty. For neuronav, whole-head coverage, no slice gap, minimal motion and geometric fidelity matter far more than the 0.1 mm.
Non-negotiables for STEALTH regardless of voxel: isotropic, contiguous (no gap), full scalp-to-vertex coverage including fiducials/skin surface, no distortion correction that alters geometry beyond what’s validated, and the exact series the surgical planning system expects.
6 · T1 MPRAGE: axial vs sagittal acquisition

Because MPRAGE is a 3D isotropic acquisition, you can reformat any plane afterwards, so the acquired plane matters less for viewing and more for artefact direction, wrap and motion behaviour.

 Sagittal-acquired MPRAGEAxial-acquired MPRAGE
Typical useThe common default: efficient whole-head slab, natural for midline structures (corpus callosum, brainstem, ventricles), volumetricsChosen when axial is the primary read plane or to orient wrap/phase artefact away from a region of interest
Coverage efficiencyFewer slices to cover the head L–R; often fastest for whole brainMore slices to cover vertex–foramen; can be longer
Wrap/foldPhase wrap tends to be S–I or A–P depending on setup; midline geometry cleanWrap can come off shoulders/neck; watch phase direction
ReformatsIsotropic → axial & coronal reformats are equivalent in qualityIsotropic → sagittal & coronal reformats equivalent
For isotropic MPRAGE the acquired plane is largely a matter of efficiency and artefact management, not final image quality; you reformat the rest. Sagittal is the usual, efficient default (great for midline & volumetrics). Acquire axial when it’s your main plane, when a navigation/planning system expects it, or to steer wrap/motion artefact away from the target.
7 · “True axial” vs “normal axial”

These terms describe how the slice/plane is aligned relative to the patient’s true anatomical axes versus the scanner/table axes.

  • True axial (orthogonal to the anatomy): the slice is angled to the patient’s actual anatomical planes, e.g. perpendicular to the long axis of the structure or aligned to a defined anatomical landmark line (AC–PC for brain, disc space for spine, the specific joint axis for MSK). It corrects for how the patient is actually lying or how the structure is oriented.
  • Normal / straight axial (orthogonal to the scanner): the slice is aligned to the magnet’s z-axis / table: straight transverse “as the patient lies,” without angling to anatomy.
 True axialNormal axial
Aligned toPatient anatomy / a defined landmark lineScanner/table (straight transverse)
ProsSymmetric, reproducible planes; correct comparison L vs R; matches reporting conventions & prior studies; essential for AC–PC brain, orbits, IACs, cardiac, angled jointsFast to prescribe; fine when the patient is well-positioned and the structure is already aligned to the bore
ConsTakes a moment more to plan off the localiserAsymmetric/oblique through the target if the patient is rotated or the structure is angled → harder to compare sides & to priors
Use true axial whenever symmetry and reproducibility matter: brain (AC–PC line), orbits, internal auditory canals, cardiac (double-oblique), and any joint that sits obliquely. Normal axial is acceptable when the anatomy is already square to the bore and speed matters. The habit worth building: plan every axial off the localiser to the anatomy, not the table.
8 · Contrast-enhanced vs non-contrast MRA/MRV: when to skip gadolinium

Both arteriography and venography can be built on a contrast platform (CE-MRA / CE-MRV) or a non-contrast one (TOF, PC, QISS, FBI/NATIVE-SPACE). The choice turns on renal function, the vascular territory, and whether the question is anatomy alone or flow direction/quantification too.

 Contrast-enhanced (CE-MRA / CE-MRV)Non-contrast (TOF · PC · QISS · FBI)
Signal sourceShortened blood T1 from a timed gadolinium bolus, independent of flow patternFlow itself: inflow saturation (TOF/QISS), velocity-induced phase shift (PC), or ECG-gated subtraction (FBI); signal depends on flow velocity, direction and rhythm
Best forLarge-territory, single fast-acquisition coverage: aorta, renal, carotid, pulmonary, run-off; time-resolved variants (TWIST/TRICKS) for AVMs and dynamic arterial→venous fillingRenal impairment / NSF risk, paediatric or contrast-averse patients, screening exams, and any case where flow direction or velocity is itself the answer (PC)
Flow-pattern sensitivityLargely avoided; the vessel is bright from contrast regardless of flow pattern, so slow, turbulent or in-plane flow is not a problemA real limitation: TOF saturates slow flow and can overestimate stenosis; PC needs a correctly chosen VENC; QISS/FBI need a regular, reliably detected cardiac trigger (ECG or peripheral pulse)
Venous sideCE-MRV maps large venous territories quickly but needs correct bolus timing to avoid arterial contamination2D TOF and PC-MRV are the default for dural-sinus thrombosis / DVT screening without contrast; PC-MRV additionally distinguishes thrombus from slow flow by direction
Main costGadolinium exposure and bolus-timing risk (too early = arterial-only, too late = venous contamination); agent and time costLonger acquisition; more sensitive to motion, arrhythmia and technique; less forgiving of a poorly planned saturation band or VENC
Rule of thumb: reach for contrast-enhanced MRA/MRV when you need one fast, large-territory acquisition or dynamic time-resolved filling information. Reach for non-contrast when renal function rules out gadolinium, when the clinical question is flow direction or velocity itself (PC), or for a quick screening study (3D TOF Circle of Willis, 2D TOF/PC dural sinuses, QISS peripheral run-off). For suspected cerebral venous sinus thrombosis specifically, most protocols start non-contrast (2D TOF or PC-MRV) and add CE-MRV only if the non-contrast study is equivocal.

See the full sequence catalogue above for the underlying families (MPRAGE, SPACE/3D TSE, IR, black-blood, MRA/MRV) and the Sequence & Protocol Optimiser for how to defend SNR when you push resolution.

Teaching reference. Parameter ranges, vendor implementations and clinical indications vary by scanner, field strength and local protocol; always confirm against your site’s protocols and manufacturer documentation.

Related on High Yield Imaging: Sequence Physics Explainer (TR/TE/flip & k-space) · Sequence Optimiser (problem-solving & tuning) · Protocols & Parameters explorer · MR Calculators (VENC, TI-to-null, Ernst angle).

Sources

Key references: Bitar R et al. MR pulse sequences: what every radiologist wants to know but is afraid to ask. RadioGraphics 2006. · Hartung MP et al. Magnetic resonance angiography: current status and future directions. J Cardiovasc Magn Reson 2011. · Miyazaki M, Lee VS. Nonenhanced MR angiography. Radiology 2008. · Edelman RR et al. Quiescent-interval single-shot unenhanced magnetic resonance angiography of peripheral vascular disease. AJR 2010. · Markl M et al. 4D flow MRI. JMRI 2012. · Haacke EM et al. Susceptibility-weighted imaging. AJNR 2009. · ISMRM/ACR white papers & manufacturer sequence documentation.

More in Protocols and technique

Everything on MRI: MRI hub. Or search the whole reference.

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.

Contents