Nuclear medicine physics and artifacts
The physics you use at the console: decay and detection, how the gamma camera and PET scanner form an image, and the artifacts that fool the unwary, each with a fix.
Related
- Collimator choice trades spatial resolution against sensitivity and must suit the isotope energy; a very hot focus of a high-energy isotope can penetrate the septa and cause star streaks.
- PET detects both 511 keV annihilation photons in coincidence, so needs no physical collimator, giving higher sensitivity and better resolution than collimated gamma-camera imaging; time-of-flight narrows the annihilation’s position along the line, improving signal-to-noise.
- On cardiac SPECT, breast or diaphragm attenuation can mimic an infarct; fix with attenuation correction or prone or upright imaging, and check gated wall motion is normal over the ‘defect’.
- Corrections can mislead: moving or breathing between CT and emission scans misaligns the attenuation map, creating false defects; check fusion alignment before correcting, and review non-corrected images alongside corrected.
- A detector uniformity defect reconstructs as rings centred on the axis of rotation: acquire a new high-count uniformity correction or have a faulty PMT re-tuned or replaced; centre-of-rotation errors need routine COR QC and recalibration.
Explain it simply
- The collimator, which lets only straight-on rays reach the camera, trades sharpness against sensitivity and must match the tracer’s energy. A very bright spot of high-energy tracer can leak through, causing star streaks.
- PET detects pairs of rays flying off in opposite directions at the same moment, so it needs no collimator and catches more signal, with sharper detail than a gamma camera. Timing their arrival reduces noise.
- In heart scans, breast tissue or the diaphragm can block rays and mimic an old heart attack. Correcting for this, scanning face-down or upright, and checking wall movement help show whether it is real.
- If the patient moves or breathes between the CT and the main scan, the correction can misalign and fake problems. Check the two line up first, and look at uncorrected pictures too.
- An uneven spot on the detector shows up as rings on 3-D (SPECT) scans; a fresh uniformity calibration, or repairing a faulty detector part, fixes it. Regular checks and recalibration fix an off-centre rotation.
A guide only: follow your departmental protocols and regional requirements.
Decay & detection
Radioactive decay
Activity A = A₀·e^(−λt), where λ = ln2 / T½. After one half-life half remains; after ten, ~0.1%. The becquerel (Bq) = 1 disintegration/s; 1 mCi = 37 MBq.
Decay modes that matter in NM
| Mode | Emission | Used for |
|---|---|---|
| Isomeric transition | mainly γ (e.g. Tc-99m 140 keV), with a small internal-conversion and Auger electron component | SPECT imaging (low particulate dose) |
| β⁺ (positron) | positron → 2× 511 keV annihilation γ | PET |
| β⁻ (electron) | β⁻ (± γ) | Therapy (I-131, Lu-177, Y-90) |
| α | helium nucleus, short range, high LET | Targeted α-therapy (Ra-223) |
| Electron capture | characteristic X-rays ± γ (Tl-201 imaged via Hg X-rays; I-123 via its 159 keV γ) | SPECT |
Interaction with matter
Photoelectric (dominant at low keV, ∝ Z³; the basis of collimation + shielding) and Compton scatter (dominant at 140–511 keV; degrades contrast, rejected by the energy window). Attenuation follows I = I₀·e^(−µx); half-value layer (HVL) is the thickness halving intensity. 511 keV photons are penetrating → PET rooms + syringe shields need more lead than Tc-99m.
Gamma camera & PET
The gamma (Anger) camera
Photons from the patient pass through a collimator (only near-perpendicular photons reach the crystal; defines spatial resolution + rejects scatter), strike a NaI(Tl) scintillation crystal (γ → light), detected by an array of photomultiplier tubes; Anger logic computes the X,Y position and the energy (Z). Events inside the energy window (e.g. 140 keV ±10%) are accepted.
Collimator choice = resolution vs sensitivity trade-off
| Collimator | Use |
|---|---|
| LEHR (low-energy high-resolution) | Tc-99m general imaging |
| LEAP / general-purpose | Higher count-rate, dynamic studies |
| Medium-energy | Ga-67, In-111 |
| High-energy | I-131 |
| Pinhole | Thyroid, paediatric kidneys, small parts |
SPECT and PET
SPECT rotates one or more heads around the patient and reconstructs a 3-D volume from projections. PET exploits annihilation: two 511 keV photons fly ~180° apart, detected in coincidence along a line of response; no physical collimator needed, giving higher sensitivity and better resolution than collimated gamma-camera imaging; time-of-flight (TOF) localises the annihilation along the line, improving signal-to-noise. Modern systems are hybrid SPECT/CT and PET/CT (or PET/MR) for attenuation correction + anatomical localisation.
Reconstruction & corrections
Reconstruction
Filtered back-projection (FBP) is fast but noisy and streaky. Iterative reconstruction (OSEM) (model the acquisition, compare, correct, repeat) dominates modern practice: less noise, allows resolution recovery + corrections, at higher compute cost. Filters (e.g. Butterworth) trade noise against resolution.
The corrections
| Correction | Why | How |
|---|---|---|
| Attenuation | Deep photons under-counted → false central defects | CT map (hybrid) or transmission source |
| Scatter | Mispositioned Compton photons blur contrast | Energy-window / model-based |
| Resolution recovery | Depth-dependent blur | Model collimator response in OSEM |
| Randoms + dead-time (PET) | False coincidences / high count-rate loss | Delayed-window, dead-time model |
| Motion | Respiratory/cardiac blur | Gating (respiratory, ECG) |
| Decay + partial volume | Quantitation (SUV) | Decay-correct; PVC for small lesions |
Artifact atlas
Attenuation artifact
Breast tissue (anterior) or the diaphragm (inferior) absorbs photons → apparent fixed perfusion defect on cardiac SPECT; mimics infarct.
Fix: CT/line-source attenuation correction; prone or upright imaging; check gated wall motion is normal over the “defect”.
Misregistration
Patient moves (or breathes) between the CT and the emission scan → attenuation map misaligned, creating false defects or shifting uptake off the anatomy.
Fix: check fusion alignment before correcting; re-acquire CT; respiratory-averaged CT for the diaphragm.
Star / septal penetration
Very hot focus (e.g. injection site, bladder) with a high-energy isotope → photons penetrate collimator septa producing radiating “star” streaks.
Fix: correct collimator for the isotope energy; shield/void the hot source; reposition FOV.
Halo / over-correction
Aggressive scatter/attenuation correction near a hot–cold boundary (e.g. bladder edge, lung–liver) can create a dark halo or negative rim.
Fix: review non-corrected images alongside corrected; adjust reconstruction parameters.
Ring artifact (flood non-uniformity)
A uniformity defect at a fixed point on the detector (a faulty or drifting PMT, or an out-of-date or low-count uniformity correction) lies in the same place in every SPECT view, so reconstruction turns it into a ring, or concentric rings, centred on the axis of rotation, and amplifies it most near the axis.
Fix: acquire a new high-count uniformity correction; physics or service to re-tune or replace a faulty PMT.
Centre-of-rotation (COR) error
When the electronic centre of the projections does not match the mechanical axis of rotation, each view is back-projected slightly off position, and the tomographic image blurs and loses contrast. On a point-source check the point reconstructs as a small ring around its own position (a “doughnut”, 360° orbit) or a “tuning fork” (180° orbit).
Fix: routine COR QC and recalibration.
Motion / respiratory blur
Breathing smears lung-base and liver-dome lesions on PET; patient motion during SPECT causes streaks and doubling.
Fix: respiratory gating; coach breathing; re-acquire; motion-correction software.
Truncation
Arms-down or large patient extends beyond the CT FOV → truncated attenuation map and edge artifacts on PET.
Fix: arms-up positioning where possible; extended-FOV reconstruction.
Contamination / free pertechnetate
Urine/sweat contamination or free Tc-99m pertechnetate → spurious hot spots (skin) or thyroid/stomach/salivary uptake mimicking pathology.
Fix: inspect skin/clothing, re-image after cleaning; check radiochemical purity (poor tagging).
Brown fat & muscle (FDG)
Cold, anxious or recently-exercised patients show symmetric neck/supraclavicular brown-fat and muscle FDG uptake that can obscure or mimic nodal disease.
Fix: keep patient warm + rested pre-injection; correlate with CT (fat density); benzodiazepine/beta-blocker per protocol.
Sources
Sources (5)
5 sources listed
- Cherry, Sorenson & Phelps, Physics in Nuclear Medicine (decay, detection, reconstruction).
- IAEA. Nuclear Medicine Physics: A Handbook for Teachers and Students. iaea.org
- SNMMI/EANM instrumentation & QC guidance; NEMA standards. snmmi.org
- Zanzonico P. Routine quality control of clinical nuclear medicine instrumentation: a brief review. J Nucl Med. 2008;49(7):1114–1131. doi.org/10.2967/jnumed.107.050203
- Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control. J Nucl Med Technol. 2001;29(1):12–18. tech.snmjournals.org
High Yield Imaging · Physics & Artifact Atlas. Educational use only.
More in Reference
Next in Reference: Nuclear medicine radiation safety
Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.