Nuclear medicine QC toolkit
The tests that keep the numbers trustworthy: what to run, how often, what “pass” looks like, and what a failure means for patient imaging. Covers the gamma camera, SPECT, PET and the dose calibrator.
Related
Planar gamma camera
| Test | Frequency | What it checks | Fail looks like |
|---|---|---|---|
| Energy peaking | Daily | Photopeak centred in the window | Off-peak → scatter/loss, poor uniformity |
| Uniformity (flood) | Daily | Even response across the field (intrinsic or extrinsic) | Focal spots (PMT/crystal), gradients |
| Spatial resolution + linearity | Weekly | Bar phantom: resolvable bars + straight lines | Blurring, bowed lines |
| Sensitivity | Periodic | Counts per unit activity | Drift → wrong quantitation |
| Collimator integrity | Periodic / on damage | No dents/holes | Focal defects mimicking pathology |
SPECT
| Test | Frequency | What it checks | Fail looks like |
|---|---|---|---|
| Centre of rotation (COR) | Weekly/periodic | Alignment of mechanical + electronic centre | Blurring and loss of contrast; a point source reconstructs as a small ring (360° orbit) or a “tuning fork” (180° orbit) |
| High-count uniformity | Weekly | Uniformity correction map for tomography | Concentric ring artifact |
| Multi-head registration | Periodic | Heads agree in x,y,z | Misregistered reconstruction |
| Overall SPECT performance (phantom) | Quarterly/annual | Resolution, contrast, uniformity (e.g. Jaszczak) | Loss of cold-rod/sphere contrast |
| SPECT/CT registration | Periodic | Emission–CT alignment for attenuation correction | False defects from misalignment |
PET / CT
| Test | Frequency | What it checks | Fail looks like |
|---|---|---|---|
| Daily QC (blank/normalisation) | Daily | Detector block stability, coincidence timing | Failed blocks → streaks, quantitation error |
| Normalisation | Periodic | Detector-to-detector sensitivity correction | Non-uniform sensitivity |
| Well-counter / SUV calibration | Periodic | Cross-calibration of scanner ↔ dose calibrator | Systematically wrong SUV |
| NEMA performance | Acceptance + annual | Sensitivity, resolution, NECR, scatter fraction | Below-spec image quality |
| CT QC | Per CT programme | Number/CT-value accuracy, alignment | Wrong attenuation map → SUV error |
Dose calibrator
The dose calibrator (re-entrant ionisation chamber) is the legal measurement of the activity a patient receives; its QC underpins every administration.
| Test | Frequency | What it checks |
|---|---|---|
| Constancy | Daily | Reproducibility using a long-lived reference source (e.g. Cs-137, Co-57) |
| Accuracy | Annual (+ install) | Reading vs certified reference source activity |
| Linearity | Quarterly | Correct across the activity range (decay or shield method) |
| Geometry | At setup / new container | Reading independent of volume/container |
QC schedule at a glance
| Daily | Weekly | Quarterly / periodic | Annual / acceptance |
|---|---|---|---|
| Camera energy peak + flood uniformity · PET daily QC · dose-calibrator constancy | Bar-phantom resolution/linearity · SPECT COR + high-count uniformity | Dose-calibrator linearity · SPECT phantom · PET normalisation · SUV cross-cal | Dose-calibrator accuracy · NEMA PET performance · CT QC · collimator survey |
Troubleshooting quick-map
| You see… | Likely cause | Do |
|---|---|---|
| Focal cold/hot spot on flood | Faulty PMT, crystal damage, off-peak | Re-peak; call physics; camera down until cleared |
| Ring artifact on SPECT | Flood-field non-uniformity (faulty or drifting PMT, out-of-date or low-count uniformity correction) | Acquire a new high-count uniformity correction; call physics if it persists |
| Blurred tomographic images | COR miscalibration, motion | Re-run COR; check patient motion |
| SUV looks systematically off | Cross-calibration / clock / residual activity error | Check well-counter cal, injection + scan clocks, residual dose |
| Poor tagging / free pertechnetate | Radiochemical-purity failure | Re-check ITLC; re-prepare kit |
| Streak artifacts on PET | Failed detector block | Repeat daily QC; service block |
Sources
Sources (7)
7 sources listed
- IPEM Report 86 / national QC guidance for gamma cameras + SPECT. ipem.ac.uk
- IAEA: Quality Assurance for SPECT/PET systems (Human Health series). iaea.org
- NEMA NU-1 (gamma camera) & NU-2 (PET) performance standards. nema.org
- AAPM task-group reports on dose-calibrator + PET/CT QC. aapm.org
- SNMMI/EANM instrumentation QC guidance. snmmi.org
- Zanzonico P. Routine quality control of clinical nuclear medicine instrumentation: a brief review. J Nucl Med. 2008;49(7):1114–1131. doi.org/10.2967/jnumed.107.050203
- Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control. J Nucl Med Technol. 2001;29(1):12–18. tech.snmjournals.org
High Yield Imaging · QC & Instrumentation Toolkit. Educational reference. Your medical-physics QC programme, manufacturer spec and national standards are authoritative. Do not use these values as device acceptance criteria.