Nuclear medicine QC toolkit

Nuclear medicine QC toolkit

The tests that keep the numbers trustworthy: what to run, how often, what “pass” looks like, and what a failure means for patient imaging. Covers the gamma camera, SPECT, PET and the dose calibrator.

On this pageGamma cameraSPECTPET/CTDose calibratorSchedule & troubleshootingSources
Frequencies + tolerances are jurisdiction- and vendor-specific. Follow your medical-physics QC programme, manufacturer specification and national standard (IPEM, AAPM, IAEA, NEMA). Values here are typical orientation, not acceptance criteria for your device.
Related

Planar gamma camera

TestFrequencyWhat it checksFail looks like
Energy peakingDailyPhotopeak centred in the windowOff-peak → scatter/loss, poor uniformity
Uniformity (flood)DailyEven response across the field (intrinsic or extrinsic)Focal spots (PMT/crystal), gradients
Spatial resolution + linearityWeeklyBar phantom: resolvable bars + straight linesBlurring, bowed lines
SensitivityPeriodicCounts per unit activityDrift → wrong quantitation
Collimator integrityPeriodic / on damageNo dents/holesFocal defects mimicking pathology
Intrinsic vs extrinsic flood: intrinsic = bare crystal with a point source (no collimator); extrinsic = with collimator + sheet source. Extrinsic also catches collimator faults.

SPECT

TestFrequencyWhat it checksFail looks like
Centre of rotation (COR)Weekly/periodicAlignment of mechanical + electronic centreBlurring and loss of contrast; a point source reconstructs as a small ring (360° orbit) or a “tuning fork” (180° orbit)
High-count uniformityWeeklyUniformity correction map for tomographyConcentric ring artifact
Multi-head registrationPeriodicHeads agree in x,y,zMisregistered reconstruction
Overall SPECT performance (phantom)Quarterly/annualResolution, contrast, uniformity (e.g. Jaszczak)Loss of cold-rod/sphere contrast
SPECT/CT registrationPeriodicEmission–CT alignment for attenuation correctionFalse defects from misalignment
A ring artifact centred on the axis of rotation is the classic sign of flood-field non-uniformity: the defect lies at the same place on the detector in every view, so reconstruction draws it as a ring and amplifies it, most of all near the axis. A COR error blurs the image instead.
Back to contents

PET / CT

TestFrequencyWhat it checksFail looks like
Daily QC (blank/normalisation)DailyDetector block stability, coincidence timingFailed blocks → streaks, quantitation error
NormalisationPeriodicDetector-to-detector sensitivity correctionNon-uniform sensitivity
Well-counter / SUV calibrationPeriodicCross-calibration of scanner ↔ dose calibratorSystematically wrong SUV
NEMA performanceAcceptance + annualSensitivity, resolution, NECR, scatter fractionBelow-spec image quality
CT QCPer CT programmeNumber/CT-value accuracy, alignmentWrong attenuation map → SUV error
SUV depends on cross-calibration. If the scanner and the dose calibrator disagree, every SUV you report is wrong even when the image looks fine. This is why the well-counter calibration matters for quantitative PET.
Back to contents

Dose calibrator

The dose calibrator (re-entrant ionisation chamber) is the legal measurement of the activity a patient receives; its QC underpins every administration.

TestFrequencyWhat it checks
ConstancyDailyReproducibility using a long-lived reference source (e.g. Cs-137, Co-57)
AccuracyAnnual (+ install)Reading vs certified reference source activity
LinearityQuarterlyCorrect across the activity range (decay or shield method)
GeometryAt setup / new containerReading independent of volume/container
Wrong isotope setting = wrong dose. Always select the correct radionuclide button; a mis-set calibration factor silently changes the measured activity.

QC schedule at a glance

DailyWeeklyQuarterly / periodicAnnual / acceptance
Camera energy peak + flood uniformity · PET daily QC · dose-calibrator constancy Bar-phantom resolution/linearity · SPECT COR + high-count uniformity Dose-calibrator linearity · SPECT phantom · PET normalisation · SUV cross-cal Dose-calibrator accuracy · NEMA PET performance · CT QC · collimator survey

Troubleshooting quick-map

You see…Likely causeDo
Focal cold/hot spot on floodFaulty PMT, crystal damage, off-peakRe-peak; call physics; camera down until cleared
Ring artifact on SPECTFlood-field non-uniformity (faulty or drifting PMT, out-of-date or low-count uniformity correction)Acquire a new high-count uniformity correction; call physics if it persists
Blurred tomographic imagesCOR miscalibration, motionRe-run COR; check patient motion
SUV looks systematically offCross-calibration / clock / residual activity errorCheck well-counter cal, injection + scan clocks, residual dose
Poor tagging / free pertechnetateRadiochemical-purity failureRe-check ITLC; re-prepare kit
Streak artifacts on PETFailed detector blockRepeat daily QC; service block
Back to contents

Sources

Sources (7)

7 sources listed

  1. IPEM Report 86 / national QC guidance for gamma cameras + SPECT. ipem.ac.uk
  2. IAEA: Quality Assurance for SPECT/PET systems (Human Health series). iaea.org
  3. NEMA NU-1 (gamma camera) & NU-2 (PET) performance standards. nema.org
  4. AAPM task-group reports on dose-calibrator + PET/CT QC. aapm.org
  5. SNMMI/EANM instrumentation QC guidance. snmmi.org
  6. Zanzonico P. Routine quality control of clinical nuclear medicine instrumentation: a brief review. J Nucl Med. 2008;49(7):1114–1131. doi.org/10.2967/jnumed.107.050203
  7. Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control. J Nucl Med Technol. 2001;29(1):12–18. tech.snmjournals.org

High Yield Imaging · QC & Instrumentation Toolkit. Educational reference. Your medical-physics QC programme, manufacturer spec and national standards are authoritative. Do not use these values as device acceptance criteria.

More in Reference

Next in Reference: Nuclear medicine glossary

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.