Nuclear medicine and PET calculators

Nuclear medicine and PET calculators

Activity, decay, paediatric dosing, SUV and dose-rate calculators; values update as you type

Scope

Educational tool, not a substitute for a measured activity, a written directive, or your local protocol. Half-lives and dose coefficients below are nominal reference values. Always confirm the coefficient for the actual radiopharmaceutical, route and patient age.

Decay and timing

Radioactive decay

A = A₀ × 2^(−t / T½)

Activity remaining

Enter A₀ and elapsed time.

SourceHalf-lives: standard reference values; confirm against your local isotope certificate.

Decay to a target activity

t = T½ × log₂(A_current / A_target)

Time to target

Enter half-life, current and target.

SourceHalf-lives: standard reference values; confirm against your local isotope certificate.

Net injected activity

Net = pre-injection − residual (both decay-corrected to the same instant)

Net injected

Enter both activities.

Patient dose

Weight-based dosing

Shared input

Paediatric activity: EANM dosage card

A = baseline × multiple (by weight and class A, B or C). If the result is below the minimum recommended activity, give the minimum. EANM Dosage Card, version 5.7.2016.

The class decides the multiple: at 20 kg it is 2.88 for class A, 4.86 for class B and 8.33 for class C. The card prints rows for set weights and does not say what to do between them: for a weight between two rows this calculator shows the figure for each row and leaves the choice to your department’s protocol. The card covers 3 to 68 kg and says the national diagnostic reference levels should not be exceeded. Its minimum activities assume commonly used cameras; lower activities could be given on systems with higher counting efficiency (card footnote 1). For Technegas the figure is the activity loaded into the device; the inhaled activity will be lower (card footnote 3). This is a card look-up to check, not a prescription: the activity is set by your department’s protocol and the practitioner who authorises the study.

Administered activity

Enter the weight, then choose the radiopharmaceutical.

SourcePaediatric dosing: EANM Dosage Card, version 5.7.2016.

SUV (body weight)

SUVₘ = ROI concentration (kBq/mL) ÷ [ injected activity (MBq) ÷ weight (kg) ]

SUV

Decay-correct the activity to scan time first.

SourceSUV convention: body-weight normalisation (SUVbw); lean-body-mass (SUL) differs.

Effective dose estimate

E (mSv) ≈ activity (MBq) × dose coefficient (mSv/MBq), adult values: F-18 FDG and Tc-99m MDP from ICRP 128; Tc-99m sestamibi (rest) and MAG3 from ARSAC 2026 (7.2 mSv for 800 MBq; 0.7 mSv for 100 MBq); Ga-68 DOTATATE: Sandström 2013 (ARSAC 2026 lists a higher figure, 6.4 mSv for 250 MBq)

Effective dose

Enter an activity.

SourceDose coefficients: ICRP Publication 128, Radiation Dose to Patients from Radiopharmaceuticals (2015): adult values, nominal (F-18 FDG, Tc-99m MDP). Tc-99m sestamibi (rest) and MAG3: ARSAC, Notes for guidance on the clinical administration of radiopharmaceuticals (July 2026), effective dose divided by reference activity. Ga-68 DOTATATE: Sandström M et al., J Nucl Med 2013;54(10):1755-1759.

Standalone

Activity units

1 mCi = 37 MBq (exact, by definition)

Converted

Type in either box.

Inverse-square dose rate

D₂ = D₁ × (r₁ / r₂)²; point source, no shielding, no attenuation

Dose rate at r₂

Enter dose rate and both distances.

Reference half-lives

IsotopeHalf-lifePrincipal use
F-18109.77 minFDG PET, NaF bone PET
Ga-6867.71 minDOTATATE, PSMA
Tc-99m6.01 hBone, renal, cardiac, sentinel node
I-12313.2 hThyroid uptake, DaTscan
Tl-20173 hMyocardial perfusion (legacy)
In-1112.80 dLabelled leucocytes, Octreoscan
Lu-1776.65 dPRRT, PSMA therapy
I-1318.02 dThyroid ablation, therapy
Sources

Dose coefficients: ICRP Publication 128, Radiation Dose to Patients from Radiopharmaceuticals (2015): adult values, nominal (F-18 FDG, Tc-99m MDP). Tc-99m sestamibi (rest) and MAG3: ARSAC, Notes for guidance on the clinical administration of radiopharmaceuticals (July 2026), effective dose divided by reference activity. Ga-68 DOTATATE: Sandström M et al., J Nucl Med 2013;54(10):1755-1759. Paediatric dosing: EANM Dosage Card, version 5.7.2016. Half-lives: standard reference values; confirm against your local isotope certificate. SUV convention: body-weight normalisation (SUVbw); lean-body-mass (SUL) differs.
Educational content only, not medical or dosimetric advice. Always follow the written directive and local protocol.

Related guides

More in Calculators

Next in Calculators: Mammography calculators

Everything on Tools: Tools hub. Or search the whole reference.

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.