Veterinary MR Calculators & Reference

Small-animal ready-reckoners for dog & cat MRI: formulae, worked examples and pre-computed lookup tables.

Updated September 2026

Before you dose: all contrast doses are weight-based; always confirm against the current product labelling and with the attending clinician. General anaesthesia is required for small-animal MRI; follow your practice’s anaesthetic protocols. Values shown are typical; verify against local protocol.

Contrast dosing

Gadolinium dose

Standard extracellular gadolinium-based contrast agent (GBCA) dose is 0.1 mmol/kg body weight. The injected volume depends on the agent’s molar concentration.

Dose (mmol) = weight (kg) × 0.1
Volume (mL) = Dose (mmol) ÷ concentration (mmol/mL)

Most agents (gadoterate / gadoteridol / gadobenate) are 0.5 mmol/mL → 0.2 mL/kg. Gadobutrol is concentrated at 1.0 mmol/mL → 0.1 mL/kg.

Worked example (10 kg dog): 10 × 0.1 = 1.0 mmol. At 0.5 mmol/mL → 1.0 ÷ 0.5 = 2.0 mL. At 1.0 mmol/mL (gadobutrol) → 1.0 ÷ 1.0 = 1.0 mL.

Lookup table (0.1 mmol/kg)

Weight (kg)Dose (mmol)mL @ 0.5 mmol/mLmL @ 1.0 mmol/mL
20.20.40.2
40.40.80.4
60.61.20.6
101.02.01.0
151.53.01.5
202.04.02.0
303.06.03.0
404.08.04.0
505.010.05.0

Cats receive the same weight-based dose (small cats typically 2–6 kg). Reduce or avoid GBCA in renal disease. Typically 0.1 mmol/kg; verify against local protocol & product label.

Contrast dosing

Contrast for MR angiography & portosystemic shunt studies

Contrast-enhanced MRA and portosystemic (liver) shunt studies use the same GBCA dose class as routine imaging (≈0.1 mmol/kg); use the Gadolinium dose table for volumes. The critical variable is timing, not dose.

Dose ≈ 0.1 mmol/kg (as per standard table)
Success = correct bolus timing relative to vessel of interest

Key timing points:

  • Arterial-phase MRA: image during first pass of the bolus; a rapid power-injected or hand bolus followed by saline flush, with bolus-tracking or a test bolus/fluoro-trigger where available.
  • Portosystemic shunt: dynamic multi-phase acquisition (arterial → portal venous → delayed) to catch the anomalous vessel; portal venous phase is usually most informative.
  • A short, compact bolus and reliable saline flush matter more than the absolute millimolar amount.

Typically same 0.1 mmol/kg dose class with phase-optimised timing; verify against local protocol & product label.

Workflow

Anaesthesia-adjacent quick refs & scan-time planning

Scope note: this panel gives no anaesthetic drug doses; those are out of scope. All anaesthetic agents, doses and monitoring are the responsibility of the attending anaesthetist / clinician and your practice protocol.

General anaesthesia is required for diagnostic small-animal MRI (motion-free, immobile patient). Plan total table/anaesthesia time from the sequence list, not just the “scan” time.

Total anaesthetic time ≈ induction + positioning + Σ(sequence times) + repeats/contrast + recovery handling

Rough sequence-time planning guide

Study elementTypical planning time
Positioning & coil set-up / localisers5–10 min
Single anatomical sequence3–8 min each
Typical brain / spine protocol (multi-sequence)25–45 min
Add post-contrast series+5–15 min
Low-field magnet (longer averaging)often 1.5–2× longer

Plan generously and build in repeat-sequence contingency. Follow your practice anaesthetic protocol and direct all drug dosing to the anaesthetist. Typical planning ranges only; verify against local protocol.

Safety concept

SAR / RF heating

Specific Absorption Rate (SAR) quantifies radiofrequency (RF) energy deposited in tissue, expressed in watts per kilogram (W/kg). RF energy from the transmit coil is absorbed and turned into heat.

SAR (W/kg) = RF power absorbed ÷ tissue mass
SAR ↑ with B₀², flip angle, echo-train length & RF duty cycle

Why it matters in small-animal MRI

  • Small patients: a cat or toy-breed dog has little tissue mass to distribute and dissipate heat; localised heating can be relatively greater than in a large patient, and thermoregulation under anaesthesia is impaired.
  • Implants & foreign bodies: conductive materials (microchips, leads, some plates) can concentrate RF energy and heat locally; the transmit coil drives this, not the receive coil.
  • Field strength: RF heating rises roughly with the square of B₀, so higher-field magnets deposit more energy for the same sequence.
  • Mitigation: lower flip angles, fewer/less-dense RF pulses, longer TR, and monitoring anaesthetised patient temperature.

No numeric per-patient calculation is required; the scanner reports and limits SAR. Understand the principle and monitor the anaesthetised patient. Verify against local protocol & scanner safety limits.

Sequence parameter

TI-to-null (STIR / FLAIR)

To null a tissue in an inversion-recovery sequence, set the inversion time (TI) to the zero-crossing of that tissue’s longitudinal recovery:

TI ≈ 0.69 × T1  (= ln 2 × T1)

STIR nulls fat (short T1); FLAIR nulls fluid / CSF (long T1). T1 lengthens with field strength, so the required TI changes with B₀.

Worked example (fat at 1.5 T): T1(fat) ≈ 260 ms → TI = 0.69 × 260 ≈ 179 ms (STIR).

Approximate T1 & resulting TI (ms) by field strength

Tissue0.2 T0.5 T1.0 T1.5 T
Fat: T1~140~180~230~260
Fat: TI (STIR)~97~124~159~179
CSF / fluid: T1~2400~2900~3800~4300
CSF: TI (FLAIR)~1656~2001~2622~2967

STIR fat suppression is relatively field-independent and robust at low field; it relies on T1 nulling rather than chemical-shift, making it the workhorse for vet magnets where spectral fat-sat is unreliable. T1 values are approximate and tissue-/vendor-dependent. Typical values; verify against local protocol.

Image quality

SNR & field-strength ready-reckoner

Signal-to-noise ratio scales (qualitatively) with field strength and with voxel volume:

SNR ∝ B₀  ×  voxel volume  ×  √(averages)

Relative SNR & practical implication

FieldRelative SNR (approx.)Practical implication
0.2 T~0.13×Thicker slices / larger voxels, more averages, longer scans
0.5 T~0.33×Compromise resolution; favour STIR over spectral fat-sat
1.0 T~0.67×Good balance for most small-animal work
1.5 T1.0× (reference)Thinner slices / higher resolution achievable in less time

To recover SNR lost at low field you can increase voxel size, add signal averages (SNR ∝ √N, so 4× averages → 2× SNR at 4× the time), or lengthen the scan, all with anaesthetic-time consequences.

Relative SNR figures scale linearly with B₀ and are indicative only; real SNR also depends on coil, sequence and tissue. Verify against local protocol.

Image quality

Voxel volume & scan-time trade-off

Voxel volume is the product of the in-plane pixel dimensions and slice thickness:

Voxel = (FOV_x / matrix_x) × (FOV_y / matrix_y) × slice thickness
Worked example: FOV 160 mm, matrix 256×256, slice 3 mm.
In-plane = 160 ÷ 256 = 0.625 mm per pixel → voxel = 0.625 × 0.625 × 3 = 1.17 mm³.
Coarsen matrix to 128×128 → 1.25 × 1.25 × 3 = 4.69 mm³ (4× the volume → ~4× the SNR, at the cost of resolution).

Trade-off cheat sheet

ChangeEffect on SNREffect on scan time / detail
Double voxel volume~doubles SNRLower spatial detail
Halve slice thickness~halves SNRFiner detail, more slices
Double averages (NEX)×√2 (~1.4)Doubles scan time
Finer matrixLower SNRHigher resolution, longer scan

Note: doubling voxel volume approximately doubles SNR, the primary lever for low-field vet magnets. Typical relationships; verify against local protocol.

Handy conversions

Unit conversions for vet MRI

kg = lb ÷ 2.2046  |  lb = kg × 2.2046
mL = mmol ÷ concentration (mmol/mL)

Body weight: pounds ↔ kilograms

Pounds (lb)Kilograms (kg)
20.9
52.3
104.5
156.8
2210.0
3315.0
4420.0
6629.9
8839.9
11049.9

Gadolinium: mmol ↔ mL for the two concentrations

Dose (mmol)mL @ 0.5 mmol/mLmL @ 1.0 mmol/mL
0.250.50.25
0.51.00.5
1.02.01.0
2.04.02.0
3.06.03.0
5.010.05.0

Typical conversions using 1 kg = 2.2046 lb; verify weights on a calibrated scale and doses against the product label.

Educational reference only. This page is not a substitute for the product label, the scanner’s safety documentation, or veterinary and anaesthetic clinical judgement. All contrast and anaesthetic decisions rest with the attending clinician.

Related guides

More in Veterinary imaging

Everything on Veterinary: Veterinary hub. Or search the whole reference.

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.