What you’ll be able to do after this lesson

  • Calculate voxel size from FOV, matrix and slice thickness
  • Explain why SNR is proportional to voxel volume, and what that means for a “small” parameter change
  • Predict the SNR, resolution and time consequence of any of the five geometry parameters
  • Explain why halving the FOV costs so much more signal than it appears to

The three numbers that define a voxel

Everything in this lesson comes from one relationship:

Voxel dimensions = (FOV_freq ÷ matrix_freq) × (FOV_phase ÷ matrix_phase) × slice thickness

A 240 mm FOV with a 256 × 256 matrix and 4 mm slices gives voxels of 0.94 × 0.94 × 4 mm — a volume of about 3.5 mm³.

Two observations before the trade-offs.

In-plane pixel size is FOV divided by matrix. Smaller FOV or larger matrix both give smaller pixels, but they are not equivalent — smaller FOV risks wrap, larger matrix costs time (along phase) and both cost signal.

Voxels are usually not cubes. In a typical 2D sequence the slice dimension is three to five times the in-plane dimension. That anisotropy is why 2D data reformats badly and why partial volume is a through-plane problem far more often than an in-plane one.

SNR follows voxel volume

The core relationship:

SNR ∝ voxel volume × √(number of samples averaged)

Signal comes from the protons inside the voxel, so it scales with the volume of tissue the voxel contains. Noise, by contrast, is roughly constant per measurement — it comes from the patient and the electronics, not from the voxel.

The word doing the work here is volume, and it is the reason that innocuous-looking parameter changes are not innocuous:

  • Halve the slice thickness → half the volume → half the SNR. One parameter, one dimension, linear.
  • Double the matrix in both in-plane directions at fixed FOV → each pixel dimension halves → volume falls to a quarter → SNR falls to a quarter. And phase-encode steps have doubled, so the scan takes twice as long. Both axes of the trade move against you at once.
  • Halve the FOV at fixed matrix → both in-plane dimensions halve → SNR falls to a quarter. This is the one that surprises people, because the change looks like a single adjustment. It is two.

To recover the SNR lost by a quarter you would need sixteen times the averaging (√16 = 4), which is not a trade anyone makes. In practice you recover some of it elsewhere: a lower receiver bandwidth, a thicker slice, a better coil, or by accepting the noise.

Averaging is the expensive fix. NSA (Siemens/Philips: averages/NSA; GE: NEX) improves SNR as √NSA, so doubling averages buys 41% more SNR for 100% more time. It is the least efficient lever available and should generally be the last one reached for — although it does have a second benefit worth remembering, which is that averaging suppresses inconsistent signal and therefore reduces motion ghosting.

Receiver bandwidth: the quiet lever

The one geometry-adjacent parameter that is not about voxel size:

SNR ∝ 1 / √(receiver bandwidth)

Bandwidth is the range of frequencies sampled during the readout. A narrower bandwidth admits less noise per sample, so lower bandwidth means more SNR — a useful lever precisely because it costs no resolution and no phase steps.

It is not free. Lower bandwidth means a longer readout, which lengthens the minimum TE (constraining short-TE sequences), leaves more time for chemical shift misregistration and susceptibility distortion to develop, and can restrict how many slices fit within a TR. Conversely, raising bandwidth reduces chemical shift and distortion at the cost of SNR — which is exactly why high-bandwidth acquisitions are part of imaging near metal.

Putting it together

You change Resolution SNR Time
↑ matrix (phase) ↑ ↓ ↑
↑ matrix (frequency) ↑ ↓ ≈ unchanged
↓ FOV (fixed matrix) ↑ ↓↓ unchanged (wrap risk ↑)
↑ slice thickness ↓ (through-plane) ↑ unchanged
↑ NSA unchanged ↑ as √NSA ↑ proportionally
↑ receiver bandwidth unchanged ↓ min TE ↓, slices ↑
Rectangular FOV / partial phase FOV unchanged ≈ ↓

Note the two genuinely efficient entries. Frequency-matrix increases buy resolution at almost no time cost, because readout samples are free — the SNR cost remains. And rectangular FOV (phase FOV reduced to match a narrow body part) removes phase steps that were only encoding empty air, so it buys time with no resolution penalty at all. Both deserve to be reached for before averaging.

At the console

A lumbar spine axial T2 is reported as too noisy. The sequence is 200 mm FOV, 320 × 256 matrix, 3 mm slices, NSA 1.

Walk the levers in order of efficiency rather than reaching for the obvious one.

Is the FOV doing useful work? At 200 mm with phase left–right, a large part of the phase FOV is air lateral to the patient. A rectangular FOV cuts those steps and buys time you can spend elsewhere at no cost in resolution.

Is the bandwidth unnecessarily high? Dropping it recovers SNR without touching resolution or phase steps — check first that the minimum TE still allows the intended weighting, and that slice coverage survives.

Is 0.63 mm in-plane resolution (200/320) actually needed for this question? Coarsening the frequency matrix to 256 gives 0.78 mm pixels and about 25% more in-plane voxel area. On a spine axial, that is often an acceptable trade and the largest single SNR gain available.

Would 4 mm slices do? A 33% thicker slice is 33% more signal, at the cost of more partial volume through the exiting nerve roots — a clinical judgement, not a physics one.

Only if none of those work should you double NSA, and you should say out loud what you are buying: 41% more SNR for 100% more time.

Second case. A colleague reduces the FOV on a wrist from 120 mm to 90 mm “to see the scapholunate ligament better”, keeps the matrix, and is surprised the images are considerably noisier. Both in-plane dimensions shrank by 25%, so voxel area fell to about 56% of its former value — nearly half the signal, from one adjustment. That is the trade, and it may still be the right one; the point is to make it knowingly.

Check yourself

Work these out before checking: a sequence has FOV 220 mm, matrix 256, slices 5 mm. What is the voxel volume? Now the radiologist asks for 3 mm slices and a 320 matrix at the same FOV. What is the new voxel volume, and by what factor has SNR changed? What would you need to change to get it back — and is any of it worth it?

This is a private prompt. Nothing you write is stored or assessed.

Key takeaways

  • Voxel size = (FOV ÷ matrix) in each in-plane direction × slice thickness. Most 2D voxels are strongly anisotropic.
  • SNR is proportional to voxel volume. Changes that alter two dimensions at once — FOV, or a matrix change in both directions — hit SNR quadratically.
  • NSA improves SNR only as √NSA, so it is the least time-efficient lever. It does, however, also suppress motion ghosting.
  • Lower receiver bandwidth raises SNR at the cost of longer readout, higher chemical shift and more distortion. Raising it is part of imaging near metal.
  • Reach for rectangular FOV and frequency-matrix changes before averaging: they are the efficient trades.

References

  1. McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from Picture to Proton, 3rd ed. Cambridge University Press, 2017 — ch. 5, ‘What you set is what you get’
  2. Elster AD. Questions and Answers in MRI — ‘Signal-to-noise ratio’ and ‘Voxel size’ https://mriquestions.com/signal-to-noise-ratio.html
  3. Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. Elsevier Academic Press, 2004 — §11.1, sampling and field of view
  4. HighYield Imaging. MRI Sequence Physics Explainer — the four-way trade-off https://highyieldimaging.com/mri-sequence-physics/