What you’ll be able to do after this lesson

  • Explain resonance in one sentence, and why an RF pulse must be at the Larmor frequency to do anything
  • Describe what a 90° and a 180° pulse do to the magnetisation vector, and what flip angle really means
  • Say precisely where the detected signal comes from, and why the free induction decay decays
  • Explain why transmit and receive are separate jobs, and what that means for coil choice

Resonance in one sentence

Resonance is energy transfer that only works at a matched frequency. Push a swing at its natural rhythm and small pushes accumulate into a large arc; push at the wrong rhythm and you fight the swing. Sing at a wine glass at its resonant pitch and it responds; sing a tone away and it does not.

Hydrogen’s natural rhythm in the bore is the Larmor frequency, ω₀ = γB₀ — about 63.9 MHz at 1.5 T. Transmit a radiofrequency pulse at exactly that frequency and the protons absorb energy. Transmit at 60 MHz and essentially nothing happens: the pulse and the precession fall out of step, and the pushes cancel.

That specificity is not an inconvenience — it is the tool. Because precession frequency depends on the field a proton actually sits in, and because gradient coils let you make the field vary across the patient, “which frequency” becomes “which position”. The whole of Topic 3 is built on this. For now, hold on to the narrower point: the RF pulse works because it is tuned.

What the pulse does to M0

At equilibrium the net magnetisation M0 lies along z. The transmit coil produces a second, much weaker oscillating field — B1 — applied perpendicular to B0 and rotating at the Larmor frequency.

The cleanest way to see the result is to imagine standing on a turntable spinning at ω₀ — the rotating frame. From there, precession about B0 disappears from view, and all that remains is B1, apparently stationary. M0 simply tips about B1, steadily, for as long as B1 is on.

How far it tips is the flip angle, and it depends on only two things: how strong B1 is, and how long it is applied. Double the amplitude or double the duration and you double the angle.

  • A 90° pulse tips M0 completely into the transverse plane. Longitudinal magnetisation is momentarily zero; transverse magnetisation is maximal.
  • A 180° pulse carries it all the way over to −z. Applied to equilibrium magnetisation this is an inversion pulse (the basis of STIR and FLAIR, lesson 06); applied to magnetisation already in the transverse plane it is a refocusing pulse (the basis of spin echo, lesson 05).
  • A small flip angle — 5° to 30° — tips only a fraction. Less transverse signal per pulse, but most of the longitudinal magnetisation is left intact, so the next pulse can follow almost immediately. That is the trick that makes fast gradient echo imaging possible, and it is lesson 07.

There is a second, equally important consequence. Before the pulse, protons precess with random relative phase. The RF pulse brings them into step, so they precess coherently — pointing the same way at the same moment. Coherence is what makes the transverse magnetisation add up rather than cancel. Losing it is what makes the signal decay.

Where the signal actually comes from

Faraday’s law of induction: a changing magnetic flux through a loop of wire induces a voltage in it. Nothing more exotic than that is going on.

After a 90° pulse, the transverse magnetisation is a real magnetic vector rotating in the xy plane at ~64 MHz. Place a coil of wire nearby and that rotating vector sweeps flux through it, inducing an oscillating voltage of a few microvolts. That voltage, amplified and digitised, is the MRI signal. There is no other source. Every image you have ever produced is a reconstruction of induced voltages.

Two consequences follow immediately and matter at the console:

Only transverse magnetisation is detectable. Longitudinal magnetisation is static; it induces nothing. This is why a sequence that leaves magnetisation along z gives no signal from that tissue — the mechanism behind nulling in inversion recovery.

Signal falls off steeply with distance from the coil. Flux through the loop drops rapidly as the source moves away. This is the physics behind every positioning rule you have been taught: the smallest coil that covers the anatomy wins, and the coil should sit as close to the anatomy as comfort and safety allow.

The free induction decay — and why it dies so fast

Switch the RF off and watch the signal on an oscilloscope. You see an oscillation at ω₀ whose envelope collapses within tens of milliseconds. This is the free induction decay (FID): free (no driving field), induction (Faraday), decay (loss of coherence).

The decay is not protons losing their magnetism. It is protons losing agreement. Two mechanisms run in parallel:

  1. Genuine spin–spin interaction (T2). Neighbouring protons perturb each other’s local field microscopically and randomly. Phase drifts apart irreversibly.
  2. Static field inhomogeneity (the extra bit in T2*). No magnet is perfectly uniform, and tissue itself distorts the field — air–tissue boundaries, iron, metal. Protons a few millimetres apart genuinely sit in slightly different fields, so they precess at slightly different rates and fan out. This loss is systematic, not random — which means, crucially, it can be undone. That is what the 180° refocusing pulse of spin echo does, and why spin echo images survive near metal when gradient echo does not.

Together these give T2*, always shorter than T2. The FID therefore dies faster than tissue physics alone demands. Lesson 03 takes this apart properly.

Transmit and receive are different jobs

Transmitting needs a large, uniform B1 across the whole imaging volume — typically the body coil built into the bore. Receiving needs sensitivity and proximity — small elements placed close to the anatomy, combined into arrays.

Because the requirements conflict, most modern systems transmit with the body coil and receive with a dedicated array. The practical rule follows directly: selecting the wrong receive coil elements, or leaving a coil unplugged or badly seated, costs you signal you can never recover — no reconstruction step can restore a voltage that was never induced.

Anything involving RF energy deposition in the patient — SAR limits, heating, coil and cable positioning, burns — is a safety topic, governed by local policy and covered in the site’s MRI safety guides. It is deliberately out of scope here.

At the console

A knee comes back grainy and dim. The images are diagnostic-ish, but plainly worse than yesterday’s.

Work backwards through this lesson. Signal is induced voltage, so ask what could have reduced the induced voltage. Was the correct coil selected in the protocol, and are all its elements showing as active on the coil display? Is the coil seated and plugged? Is the knee inside the coil, or sitting proud of it on a pad? Is the knee at isocentre, where B0 is most uniform and B1 best behaved? Only after those four does it make sense to reach for parameters.

Now a second case. A shoulder patient has a small titanium plate in the humerus. Your gradient echo sequence shows a large black bloom; your TSE sequence, same region, shows a much smaller void. Nothing has changed about the metal. What changed is whether the sequence undoes the static inhomogeneity the metal creates. Gradient echo does not, and reads out a T2*-decayed signal. TSE inserts a 180° refocusing pulse and recovers the coherence that the field distortion scrambled — the mechanism you just met in the FID section, doing real clinical work.

Check yourself

Explain to an imaginary first-week student, without using the word “energy”, why a 90° pulse produces signal and a 0° pulse does not. Then: your FID decays in 30 ms but the tissue’s true T2 is 90 ms — where did the other 60 ms go, and which sequence family gets it back?

This is a private prompt. Nothing you write is stored or assessed.

Key takeaways

  • Resonance means matched frequency. The RF pulse must be at the Larmor frequency, and that same selectivity later becomes spatial encoding.
  • Flip angle is set by B1 amplitude and duration. 90° maximises transverse signal; 180° inverts or refocuses; small angles preserve longitudinal magnetisation for speed.
  • The signal is a voltage induced by rotating transverse magnetisation (Faraday). Longitudinal magnetisation is invisible, and signal falls off steeply with coil distance.
  • The FID decays because protons lose phase coherence — irreversibly through T2, and reversibly through static field inhomogeneity. The reversible part is what spin echo recovers.
  • Transmit and receive are separate jobs with conflicting requirements. Coil selection and positioning are signal decisions, not comfort decisions.

References

  1. McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from Picture to Proton, 3rd ed. Cambridge University Press, 2017 — ch. 8
  2. Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. Elsevier Academic Press, 2004 — ch. 3, RF pulse shapes and flip angle
  3. Elster AD. Questions and Answers in MRI — ‘Free induction decay’ and ‘What is resonance?’ https://mriquestions.com/free-induction-decay.html
  4. Radiopaedia. Free induction decay (physics article) https://radiopaedia.org/articles/free-induction-decay