What you’ll be able to do after this lesson
- Explain why hydrogen is the nucleus we image, and what “spin” actually means for a proton
- Describe how net magnetisation forms in B0 — and why it is tiny, yet enough
- Calculate the Larmor frequency for a given field strength, and say why that number matters at the console
- Name three things that change when you move a patient from 1.5 T to 3 T, from first principles
Why hydrogen
Your patient is roughly 60% water by mass, and every water molecule carries two hydrogen nuclei. Fat adds many more. A single hydrogen nucleus is one proton — the simplest nucleus there is, and by a wide margin the most abundant magnetically useful one in the body.
“Magnetically useful” is the operative phrase. A nucleus only responds to a magnetic field if it has a property called spin, and spin only exists when the nucleus has an odd number of protons or neutrons. Hydrogen-1 qualifies with a single proton. Carbon-12 and oxygen-16, the other bulk constituents of tissue, have even numbers of both and are magnetically invisible. So the entire clinical enterprise of MRI rests on one nucleus, present in enormous numbers, in a molecule that also happens to differ usefully between tissues.
Spin is an intrinsic quantum property, not a literal rotation, but the classical picture is a fair working model: treat each proton as a tiny bar magnet with a magnetic moment. Ordinarily those moments point in every direction at random, and the tissue has no net magnetism. Nothing you could detect.
What B0 actually does
Put that tissue in the bore, and each proton’s magnetic moment now experiences the main field, B0. Two things follow.
First, a slight population bias. Protons settle into one of two energy states — loosely, aligned with B0 or against it. Aligned is very slightly lower in energy, so very slightly more protons end up there. “Slightly” is not a figure of speech. At 1.5 T and body temperature the excess is on the order of a few protons per million. Everything MRI does is built on that surplus.
It works only because Avogadro is generous. A single millilitre of water contains about 6.7 × 10²² hydrogen nuclei, so a few per million is still an enormous absolute number of unpaired moments. Summed, they produce a small but real magnetisation vector pointing along B0. This is M0, the net magnetisation — the thing you will spend the rest of this course manipulating. Individual protons are never observed in clinical MRI. Only M0 is.
Net magnetisation
In the scanner’s field, hydrogen protons settle into two groups: slightly more line up with the field than against it. That tiny surplus adds up to one small magnet, the net magnetisation M0, which points along B0 and, once an RF pulse tips it over, gives the MRI signal.

The surplus, and so M0, grows in proportion to B0: one reason higher-field scanners give more signal.
Explain it simply
- Inside the magnet, the tiny proton magnets point either with the big magnet or against it. It is almost a tie.
- Almost: for every million protons, only about 5 more point with the magnet than against it (at 1.5 T).
- But even one drop of water holds thousands of billions of billions of protons, so those few extras add up to one small magnet. It is called M0, and it is what the scanner works with.
- Why it matters: a stronger magnet makes that surplus bigger. That is one reason 3 T scanners get more signal than 1.5 T scanners.
A simplified picture, useful for intuition: strictly, a single proton need not be purely “parallel” or “antiparallel”; the two groups describe the average over very many protons, which is what M0 measures. Surplus from the Boltzmann distribution, ΔN/N ≈ γħB0/(2kT), where ΔN is parallel minus antiparallel and N is all protons, at 310 K with γp/2π = 42.577 478 461 MHz T−1 (CODATA 2022, NIST): 4.9 per million at 1.5 T, 9.9 per million at 3 T. Books that divide by one level’s population quote twice this (about 10 per million at 1.5 T). The balance illustration is only the analogy; the labelled diagram carries the physics.
Second, precession. A magnetic moment placed at an angle to a field does not simply swing into line. It precesses around the field direction, sweeping out a cone, exactly as a spinning top precesses about gravity. The rate of that precession is what makes MRI selective, and it obeys one equation you should know by heart:
ω₀ = γB₀ — the Larmor equation.
For hydrogen, γ/2π = 42.58 MHz per tesla.
So at 1.5 T, hydrogen precesses at 42.58 × 1.5 ≈ 63.9 MHz. At 3 T, ≈ 127.7 MHz. At 0.55 T, ≈ 23.4 MHz. These are radio frequencies — the FM broadcast band sits at 88–108 MHz for comparison — which is why the room is a Faraday cage and why a badly seated door panel puts a zipper artefact across your image. Your scanner is a very good radio receiver listening for a very small signal.
Note what γ does not depend on: the tissue, the patient, or the sequence. It is a constant of the nucleus. Sodium-23 and phosphorus-31 have different gyromagnetic ratios, which is precisely why they need different hardware and are not imaged incidentally.
Precession
A spinning top turns about its own axis; give it a lean and that axis slowly sweeps around the vertical. A proton in a magnetic field does the same, far faster: its spin axis sweeps around the field direction millions of times a second.

For hydrogen, γ/2π ≈ 42.58 MHz per tesla, so f0 ≈ 63.9 MHz at 1.5 T and ≈ 127.7 MHz at 3 T.
Explain it simply
- Your body is mostly water, and every water molecule contains hydrogen. The centre of each hydrogen atom, a proton, acts like a tiny spinning magnet.
- Inside the scanner’s big magnet, each proton’s spin wobbles round in a circle, like a spinning top that is leaning over.
- The stronger the magnet, the faster the wobble. In a 1.5 T scanner it is about 64 million times a second, a radio frequency just below the FM radio band. In a 3 T scanner it is twice as fast.
- Why it matters: the scanner talks to your protons with radio waves at that same speed. That is why the scan room is shielded, like a cage that keeps outside radio signals out.
A classical picture, useful for intuition; strictly it is the net magnetisation of many protons (and the average of each spin) that precesses like this. B0 is drawn vertical by convention; in a closed-bore scanner it runs along the bore. The cone angle is drawn for clarity, not to scale. Proton gyromagnetic ratio γp/2π = 42.577 478 461 MHz T−1, CODATA 2022 (NIST, physics.nist.gov/cgi-bin/cuu/Value?gammapbar). The top illustration is only the analogy; the labelled diagram carries the physics.
Longitudinal, transverse, and why you can’t detect M0 directly
Convention: B0 defines the z axis, running along the bore. Magnetisation along z is longitudinal. Magnetisation in the perpendicular xy plane is transverse.
At equilibrium, M0 lies entirely along z. And here is the frustration that makes the next lesson necessary: you cannot detect it there. A receiver coil detects changing magnetic flux — a moving magnet induces a current, a stationary one does not. M0 sitting quietly along z, parallel to a field a million times stronger, induces nothing.
To get a signal you must tip M0 into the transverse plane, where its precession sweeps past your coil and induces a voltage. That is the entire purpose of the RF pulse, and it is the subject of lesson 02.
What changes at 3 T
Because M0 scales with B0, doubling the field roughly doubles the available magnetisation — the usual headline reason for 3 T. But three other consequences follow from the same first principles, and they are the ones that bite in practice:
- Larmor frequency doubles. The scanner transmits and receives at ~128 MHz instead of ~64 MHz. Shorter RF wavelength in tissue leads to standing-wave and dielectric shading effects, which is why abdominal 3 T images can show central brightness or darkening that 1.5 T does not.
- Chemical shift doubles in hertz. Fat and water differ by a fixed ~3.5 ppm, so the gap widens from about 220 Hz at 1.5 T to about 440 Hz at 3 T. Spectral fat suppression gets easier; chemical shift misregistration gets worse. You will meet both again in Topic 4.
- Relaxation times change. T1 lengthens appreciably at 3 T. T2 in brain changes comparatively little — measured values move by only a few milliseconds either way — so the T1 change is the one that reshapes contrast. A protocol copied unchanged from a 1.5 T scanner will not produce the same contrast.
None of this makes 3 T better or worse in the abstract. It makes it different, in ways that are predictable from ω₀ = γB₀ and from M0 ∝ B0.
At the console
You are on a 3 T scanner and the routine brain protocol has just been imported from the department’s 1.5 T system. The T1-weighted sequence comes back flat: grey and white matter are far less distinguishable than the registrar expects, and everything looks a bit washed out.
Nothing is broken. T1 values lengthen at 3 T. Published measurements disagree on the absolute numbers more than most textbooks admit: white matter is quoted at roughly 600–900 ms at 1.5 T rising to roughly 830–1100 ms at 3 T, and grey matter at roughly 950–1150 ms rising to roughly 1300–1800 ms, depending on the measurement method (Wansapura 1999 gives 832 and 1331 ms at 3 T; Stanisz 2005 gives 1084 and 1820 ms at 3 T, against 884 and 1124 ms at 1.5 T). Treat the direction of travel as the fact and the digits as approximate. A TR that was short relative to T1 at 1.5 T is no longer short relative to the longer 3 T T1, so less T1 contrast is generated. The fix lives in the next two lessons — the TR and TI have to be re-tuned for the field, not copied. What this lesson gives you is the reason to suspect the field strength rather than the scanner.
Second scenario, same day: a colleague asks why the system won’t let a sodium research sequence run on the standard body coil. Larmor again. At 3 T, sodium precesses near 33.8 MHz, nowhere near hydrogen’s 127.7 MHz. The coil is tuned to a frequency sodium simply does not sing at.
Check yourself
Without looking back: what is the Larmor frequency of hydrogen at 1.5 T, and what physically would happen to the received signal if a scanner transmitted 1 MHz off that value? Then answer the harder one — if only a few protons per million contribute, why is MRI’s signal-to-noise ratio workable at all?
This is a private prompt. Nothing you write is stored or assessed.
Key takeaways
- MRI images hydrogen because it has spin, is overwhelmingly abundant in tissue, and sits in molecules that differ between tissues.
- B0 creates a tiny population excess — a few protons per million — that sums over enormous numbers into the net magnetisation M0.
- ω₀ = γB₀, with γ/2π = 42.58 MHz/T for hydrogen: 63.9 MHz at 1.5 T, 127.7 MHz at 3 T. This selectivity is the whole basis of excitation and encoding.
- M0 along z is undetectable. Signal requires transverse magnetisation — which is what the RF pulse is for.
- Field strength changes M0, Larmor frequency, chemical shift in hertz, and relaxation times together. Protocols do not port between field strengths unchanged.
References
- McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from Picture to Proton, 3rd ed. Cambridge University Press, 2017 — ch. 8 ‘Getting in tune’
- Elster AD. Questions and Answers in MRI — ‘Net magnetization’ and ‘Larmor frequency’ https://mriquestions.com/net-magnetization.html
- Radiopaedia. Nuclear magnetic resonance (physics article) https://radiopaedia.org/articles/nuclear-magnetic-resonance
- Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. Elsevier Academic Press, 2004 — §1.1
- Stanisz GJ, Odrobina EE, Pun J, et al. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magnetic Resonance in Medicine. 2005;54(3):507–512 https://doi.org/10.1002/mrm.20605
- Wansapura JP, Holland SK, Dunn RS, Ball WS. NMR relaxation times in the human brain at 3.0 tesla. Journal of Magnetic Resonance Imaging. 1999;9(4):531–538 https://doi.org/10.1002/(SICI)1522-2586(199904)9:4<531::AID-JMRI4>3.0.CO;2-L