Generic MRI departmental protocols: Patient safety, staff safety, emergencies, QA

Generic MRI departmental protocols: Patient safety, staff safety, emergencies, QA

A ready-to-adapt set of standard operating procedures for an MRI department: patient safety checks, staff safety and zoning, emergency procedures, daily quality assurance, cleaning and infection control, and safety governance. Written to align with the ACR Manual on MR Safety (2026), MHRA Safety Guidelines for MRI Equipment in Clinical Use, 5th edition (2026), and RANZCR MRI Safety Guidelines v3.0 (2021).

On this page1 · Patient Safety2 · Staff Safety & Zoning3 · Emergencies4 · Daily QA5 · Cleaning & IC6 · GovernanceSources
Template notice: these are generic templates for adaptation. Every department must localise them to its own scanners, site layout, staffing, and national regulations, and have them approved by the MR Medical Director / MR Safety Officer before use. They do not replace local policy or vendor operator manuals.
Key points
  • Screen patients at booking and again face to face on arrival; any unresolved item keeps them out of the magnet room (Zone IV) until the MRSO or supervising radiologist clears it.
  • Identify an implant’s exact make and model and check it against the manufacturer’s current MR labelling; if labelling cannot be found or its conditions cannot be met, defer and escalate.
  • On a superconducting scanner the field is always on: the emergency stop cuts power, not the field; among emergency controls only a completed quench removes it, and even then verify before ferromagnetic equipment enters.
  • In a cardiac arrest in the magnet room (Zone IV), evacuate the patient to the designated resuscitation point in Zone II/III and keep the arrest team out of the magnet room.
  • Quench when someone pinned or trapped by a ferromagnetic object cannot be freed quickly and safely, or emergency services must enter with ferromagnetic equipment; cardiac arrest is not a reason, as the patient is evacuated.
Explain it simply
  • Patients answer safety questions when booking and again in person on arrival. Any unresolved answer keeps them out of the magnet room until a senior safety lead or radiologist clears them.
  • Staff find out exactly which implant a patient has and check the maker’s current scanning instructions, not memory. If they can’t find them or can’t meet them, the scan waits for a senior decision.
  • On helium-cooled scanners the magnet stays on even when the power is cut. Of the emergency buttons, only a full shutdown (a quench) removes its pull, and staff still check before magnetic metal goes in.
  • If a patient’s heart stops during a scan, staff move them out of the magnet room to a set resuscitation point straight away, and keep the emergency team out of the magnet room.
  • Staff shut the magnet down if someone trapped by a magnetic object can’t be freed quickly and safely, or rescuers must bring magnetic equipment in. A stopped heart isn’t a reason: move the patient out.

A guide only: follow your departmental protocols and regional requirements.

Related

Patient Safety Protocols

Screening, implants, pregnancy, burns prevention, hearing protection, sedation

P1: Two-Stage Patient Screening

Responsibility: Booking staff (stage 1) · MR radiographer / technologist (stage 2)

  1. Stage 1 (at booking): screening questionnaire completed by phone or referral triage. Any implant, device, injury with retained metal, or previous surgery flagged for follow-up before the appointment day.
  2. Stage 2 (on arrival): written screening form completed and signed by the patient (or guardian/next of kin), then reviewed face-to-face by a Level 2 MR radiographer, who countersigns.
  3. Immediately before entry to Zone IV, the radiographer performs a final verbal check: “Do you have anything in or on your body that we haven’t discussed: pacemaker, pumps, patches, hearing aids, keys, phone, cards?”
  4. Patients change into metal-free gowns for all examinations unless a documented local exception applies. Medication patches with metallic backing: removal or repositioning is reviewed by a Level 2 MR Physician, with the prescriber for an important drug, and a plan to replace the dose is in place.
  5. Any unresolved screening item = the patient does not enter Zone IV until cleared by the MRSO or supervising radiologist. Document the decision.
Unconscious / non-communicative patients: obtain history from family, referrer, and notes; examine for scars and devices; obtain orbits/chest radiographs where indicated per local policy; final risk–benefit decision rests with the MR Medical Director or delegate.
P2: Implant & Device Verification

Responsibility: Level 2 MR radiographer · escalation to MRSO

  1. Identify the exact make and model (operative note, device card, radiograph, manufacturer registry). “Probably titanium” is not identification.
  2. Verify MR labelling (MR Safe / MR Conditional / MR Unsafe) against the manufacturer’s current labelling, not memory, not a colleague’s recollection.
  3. For MR Conditional devices, confirm every condition is met on your scanner: field strength, spatial gradient, dB/dt, SAR/B1+rms limits, positioning, waiting period, and any programming/interrogation requirements (e.g. CIEDs scanned per cardiology-supported pathway).
  4. Record the device, labelling source, and conditions applied in the patient record.
  5. If labelling cannot be found or conditions cannot be met: defer and escalate to the MRSO/MRMD for a documented risk–benefit decision.
P3: Pregnancy (Patients)

Responsibility: Referrer · supervising radiologist · MR radiographer

  • Ask all patients of childbearing potential about possible pregnancy at screening; document the response.
  • MRI may be performed in any trimester when the clinical question cannot wait and cannot be answered by non-ionising alternatives; the risk–benefit decision is made by the supervising radiologist and documented.
  • Gadolinium-based contrast is avoided in pregnancy unless essential; if used, guidelines differ on the agent: the ACR Manual on Contrast Media (2026 edition, US) specifies an agent at low risk of NSF (ACR Group II, which includes the linear agents gadobenate and gadoxetate), while ESUR (2026, Europe) specifies a macrocyclic agent. Either way, use the lowest diagnostic dose with documented justification and follow your departmental protocol.
  • Keep the examination within normal operating mode where practicable and minimise acoustic exposure and scan time.
P4: Burns Prevention

Responsibility: MR radiographer performing positioning

  • Use only pads and positioning aids approved for MR; place insulation pads between the patient and bore wall and cables.
  • No skin-to-skin loops: separate hands from thighs, ankles from each other (pads or sheets).
  • Cables run straight down the centre of the table where possible: no loops, no crossing, nothing trapped under the patient.
  • Remove leads/electrodes not required for the exam; only MR Conditional monitoring equipment inside Zone IV.
  • Check for tattoos, permanent make-up, drug patches, and metallic clothing fibres (athletic wear); apply local mitigation (cold compress availability, gown change).
  • Instruct the patient to report any warming or burning immediately via the call bell, and confirm the call bell works before starting.
P5: Hearing Protection

Responsibility: MR radiographer

  • All patients (and anyone remaining in Zone IV during scanning) receive hearing protection (earplugs correctly fitted, plus headphones or ear muffs whenever possible), targeting exposure below 99 dB(A) at the ear.
  • For neonates and sedated/anaesthetised patients, use appropriately sized protection (mini-muffs) and document application.
P6: Sedation, Anaesthesia & Monitoring

Responsibility: Anaesthetic/sedation team + MR radiographer jointly

  • Sedated or anaesthetised patients are monitored continuously with MR Conditional physiological monitoring (SpO2, ECG, NIBP, capnography as indicated), used within its conditions.
  • All anaesthesia equipment entering Zone IV is verified MR Conditional (or MR Safe where it contains no metal) and labelled; gas cylinders and their regulators enter Zone IV only if labelled MR Conditional, and are used within their conditions.
  • The anaesthesia team completes staff screening and induction before working in the unit; induction and recovery occur outside Zone IV where practicable.
  • A pre-scan WHO-style pause confirms: patient screened, monitoring conditional, emergency plan understood (arrest = evacuate, see Emergency Procedures).
Back to contents

Staff Safety & Zoning Protocols

Four-zone model, personnel levels, access control, staff screening, pregnant staff, cryogens

S1: Four-Zone Access Model (ACR)

Responsibility: MRSO (design & enforcement) · all MR personnel (daily enforcement)

ZoneDefinitionAccess
Zone IGeneral public areas (entrance, corridor)Unrestricted
Zone IIReception/interview area between public space and controlled zones; screening happens herePatients under general supervision of MR personnel
Zone IIIControl room and adjacent areas; free access could result in injury from the magnetic fieldScreened MR personnel; screened patients and other non-MR personnel only under the direct supervision of Level 2 MR personnel; physically restricted (locked/badge access)
Zone IVThe magnet room itselfScreened MR personnel; screened patients and other non-MR personnel only under the direct supervision of Level 2 MR personnel; door demarcated with lighted “Magnet On” signage
The magnet is always on. Zone IV signage, door control, and supervision apply 24/7, including nights, weekends, power failures, and fire alarms.

S2: MR Personnel Levels & Training

Responsibility: MRSO maintains the register; managers ensure compliance

CategoryDefinitionTraining requirement
Non-MR personnelAnyone not trained within the last 12 monthsTreated as visitors: screened and escorted, never unsupervised in Zone III/IV
Level 1Staff who have passed the MR safety training your MR medical director sets, so they are not a danger to themselves or others in the MR environment, and who work there regularly (e.g. patient aides, technologist assistants, some nursing roles); they do not supervise non-MR personnelAnnual MR safety training; after screening, may have unaccompanied access to Zones III and IV (local policy may be stricter)
Level 2Staff with extensive training who supervise Zone III/IV (MR radiographers, MR radiologists, MRSO)Annual advanced training: RF burns, projectiles, quench, screening decisions
  • Ancillary staff who may enter under supervision (cleaners, orderlies, security, biomed, fire wardens) receive a documented site induction annually.
  • Local fire and ambulance services are offered periodic familiarisation visits; this is recorded in the safety file.

S3: Staff & Visitor Screening

Responsibility: Each staff member · verified by MRSO

  • All MR personnel complete a screening form at appointment and re-attest annually, and immediately after any new implant, injury, or procedure.
  • Visitors, relatives, and accompanying staff complete the same written screening as patients before entering Zone III/IV and remove all ferromagnetic items.
  • Contractors and service engineers sign in, are screened, and work under a permit-to-work system agreed with the MRSO.

S4: Pregnant Staff

Responsibility: Staff member + manager + MRSO

  • Pregnant MR personnel may continue to work in and around the MR environment, including Zone IV positioning duties, throughout pregnancy.
  • Pregnant staff should not remain in Zone IV (the MR examination room) while scanning is in progress (ACR Manual on MR Safety, 2026 edition; MHRA, 5th edition, 2026).
  • The staff member notifies their manager as early as they are comfortable so duties can be reviewed and documented.

S5: Ferromagnetic Object Control

Responsibility: All Level 2 personnel

  • All equipment entering Zone IV carries a visible label: MR Safe (green), MR Conditional (yellow, conditions attached), or MR Unsafe (red, never enters).
  • Unlabelled equipment is treated as MR Unsafe until tested/verified.
  • Zone IV equipment inventory is audited quarterly; anything unaccounted for is investigated.
  • Trolleys, beds, wheelchairs, IV poles, oxygen cylinders: only MR Conditional versions cross the Zone IV threshold. Patient transfer to an MR Conditional trolley occurs in Zone III.
  • Ferromagnetic detection systems (if installed) are an adjunct, never a substitute for screening.

S6: Cryogens & Oxygen Monitoring

Responsibility: MRSO + engineering + vendor service

  • An oxygen depletion monitor is fitted in the magnet room (and equipment room where applicable), alarming to the control room; it is tested per the planned maintenance schedule.
  • Helium fills and cryogen handling are performed only by trained service personnel with the room ventilation confirmed operational.
  • The quench pipe outlet is inspected per the maintenance schedule (unobstructed, weatherproof, away from public areas).
  • If the O2 alarm activates: evacuate the magnet room, do not re-enter, ventilate, and follow the quench/cryogen incident procedure (Emergency Procedures, E2).
Back to contents

Emergency Procedures

Cardiac arrest, quench, projectile, contrast reaction, fire, power failure, claustrophobia

E0: Know Your Three Buttons

All Level 2 personnel: verify locations at every site induction

ControlWhat it doesWhen to use
Table stop / releaseStops table motion, allows manual releasePatient entrapment by the table, positioning emergency
Emergency stop (electrical)Cuts scanner electrical power. Does NOT remove the static fieldElectrical fault, smoke from equipment, RF/gradient emergency
Quench button (magnet stop)Shuts the magnetic field down fast by boiling off helium, but not instantly; the time varies between systems and should be shown on a notice by the button. Scanner out of action afterwardsPerson pinned or trapped by a ferromagnetic object who cannot be freed quickly and safely, or fire crews must enter with ferromagnetic equipment. Never for a cardiac arrest
Never assume the field is off after any button press. Of the emergency controls, only a completed quench removes the field of a superconducting magnet, and even then, verify before allowing ferromagnetic equipment in.

E1: Cardiac Arrest / Medical Emergency in Zone IV

First responder: MR radiographer · code team responds to Zone II/III

  1. Call the arrest team, stating clearly: “MRI unit: respond to the MRI control room, do NOT enter the magnet room.”
  2. Start basic life support in the magnet room only as far as needed to move the patient: immediately evacuate the patient on the (MR Conditional) table/trolley to the designated resuscitation point in Zone II/III.
  3. Close and secure the Zone IV door behind you. A Level 2 staff member guards the door for the entire event; arriving code teams will try to enter with steel cylinders and laryngoscopes.
  4. Full ACLS proceeds at the resuscitation point with standard (non-MR) equipment.
  5. Debrief and incident report afterwards; rehearse this scenario at least annually with the arrest team.

E2: Quench (Spontaneous or Deliberate)

Responsibility: Senior radiographer on duty · MRSO notified immediately

  1. Deliberate quench criteria: a person is pinned or trapped by a ferromagnetic object and cannot be freed quickly and safely, or emergency services must enter the magnet room with ferromagnetic equipment (for example, a fire that cannot be contained). When in doubt, quench (ACR Manual on MR Safety, 2026 edition). Any Level 1 or Level 2 MR staff member present should activate it when these criteria are met. Do not wait for approval.
  2. On any quench: evacuate the patient and all staff from the magnet room immediately. No re-entry until a qualified person confirms the field has fallen (MHRA 5th edition (2026), s. 4.4.6; RANZCR v3.0, s. 5.2.1).
  3. If helium vents into the room (quench pipe failure: misting, hissing, cold fog, O2 alarm): do not re-enter; be aware the door may be pressurised (open cautiously or use the vent hatch per vendor guidance); ventilate the area; keep everyone clear until O2 levels normalise.
  4. Asphyxiation and cold-burn risk: anyone exposed to venting helium gets a medical review.
  5. Notify MRSO, engineering, and the vendor; quarantine the room; complete an incident report. The magnet is re-energised only after vendor inspection.

E3: Projectile / Ferromagnetic Incident

Responsibility: All staff (stop the line)

  1. Object attracted to the magnet with no one injured and no equipment/patient at risk: do not attempt to pull large objects off the magnet. Clear the room, secure the door, call the MRSO and vendor.
  2. Person pinned: if they cannot be freed quickly and safely, quench (E2). Do not try to pull the object off them.
  3. Small object incidents (scissors, coins, pens) with no harm are still reportable near-misses; they reveal a screening failure.
  4. After any incident: review screening/access controls before resuming scanning; complete the incident report and notify per Safety Governance.

E4: Acute Contrast Reaction

Responsibility: MR radiographer + supervising doctor per local contrast policy

  1. Stop the injection; maintain IV access; call for help per local escalation.
  2. Remove the patient from Zone IV to the designated treatment area; resuscitation equipment stays outside the magnet room.
  3. Grade severity (mild / moderate / severe, ACR/ESUR categories) and treat per the department’s contrast reaction algorithm; adrenaline IM early for anaphylaxis.
  4. Observe post-reaction per policy; document the reaction, drugs given, and batch details; report to the national pharmacovigilance scheme.
  5. Use a structured documentation form; see our acute contrast reaction documentation form and the emergency treatment card.

E5: Fire & Evacuation

Responsibility: All staff · fire warden liaison pre-planned with MRSO

  • Only non-ferromagnetic (MR Conditional CO2) extinguishers are stored in/near Zone IV; standard steel extinguishers never cross the threshold.
  • Small, contained equipment fire: emergency electrical stop, evacuate the patient, attempt first-aid firefighting only with the MR Conditional extinguisher and only if safe.
  • Structural fire / fire service entry required into Zone IV: quench before entry (E2), and confirm field decay to the incident commander. A Level 2 staff member remains the fire crew’s MR safety contact.
  • On any alarm activation: Zone IV door control is maintained: a designated staff member prevents well-meaning responders entering with ferromagnetic kit.
  • Evacuation route, assembly point, and “magnet always on” briefing are part of the annual fire training for the unit.

E6: Power Failure & System Faults

Responsibility: Radiographer on duty + engineering

  • The static field persists during power failure. Zone IV discipline continues; door access control reverts to the local failsafe (know whether yours fails locked or unlocked).
  • Remove the patient using the manual table release; check patient wellbeing; log the event.
  • On restoration, complete daily QA checks (Daily QA & Equipment Checks) before clinical scanning resumes; check helium level/cold head function.

E7: Patient Distress / Claustrophobia / Emergency Removal

Responsibility: MR radiographer

  • Confirm the call bell before every scan; respond immediately when pressed.
  • Practise the fastest patient extraction: stop scan, table out (or manual release), assist patient, reassure, reassess.
  • Document premature terminations and offer rebooking strategies (open-bore, feet-first, anxiolysis via referrer).
Back to contents

Daily QA & Equipment Checks

Start-of-day, weekly and periodic quality assurance

Q1: Start-of-Day Checklist

Responsibility: First radiographer on duty · logged daily, initialled

  • Helium level / cold-head running (listen for pulse tube or compressor); chiller status normal; log the reading.
  • Oxygen depletion monitor: powered, no alarms.
  • Zone IV door closes, latches, and “Magnet On” light functioning; badge/lock access working.
  • Emergency controls visually checked: table release, emergency stop, quench button covers intact.
  • Intercom and patient call bell tested end-to-end.
  • Emergency trolley (Zone III side) checked and sealed; MR Conditional extinguisher in place.
  • Coils inventoried, connectors undamaged, pads/positioning aids clean and intact (frayed cables = withdraw from use).
  • MR Conditional ancillary kit present: trolley, wheelchair, IV pole, aluminium O2 cylinder (contents adequate).
  • Room temperature/humidity within vendor spec.
  • Any faults escalated before first patient; scanner not used clinically if a safety-critical item fails.

Q2: Weekly Image Quality Check

Responsibility: Designated QA radiographer · results trended

  • Phantom check per local/ACR QC program: centre frequency, transmit gain, SNR, uniformity, geometric accuracy, logged and compared against baseline action limits.
  • Artifact review: any new recurring artifact (spikes, zippers, shading) reported to engineering with example images.
  • Drift outside action limits → service call; document scanner status decision (continue/restrict/stop).

Q3: Periodic & Annual

Responsibility: MRSO + medical physicist + vendor

  • Planned preventive maintenance per vendor contract; records retained in the safety file.
  • Annual medical physics performance survey (ACR MR QC or equivalent national program).
  • Annual test of O2 monitor, door interlocks, emergency stops (with vendor), and RF door seal integrity (waveguide/penetration check after any building work).
  • Quarterly Zone IV equipment audit; annual review of all signage.
Back to contents

Cleaning & Infection Control

Routine cleaning, transmissible precautions, equipment handling

C1: Routine Cleaning

Responsibility: MR radiographers (between patients) · trained cleaners (scheduled, supervised)

  • Between every patient: wipe table, coils in contact with the patient, pads, headphones/earphone covers, call bell, and positioning aids with a vendor-approved disinfectant wipe (check compatibility; some agents craze coil housings).
  • Single-use items (ear plugs, couch paper, some pads) discarded after each patient.
  • Daily: floors and surfaces in Zones III–IV cleaned with MR Conditional-labelled equipment only (no steel buckets/mop handles); cleaners are inducted, screened, and supervised by Level 2 staff, and clean Zone IV only with the door controlled.
  • Weekly/monthly deep clean schedule documented, including bore liner, cable channels, and storage areas.

C2: Transmissible Infection Precautions

Responsibility: MR radiographer + infection prevention team

  • Schedule known contact/droplet/airborne-precaution patients at the end of a list where practicable.
  • Standard precautions plus the transmission-based precautions specified by the IPC team; PPE donned/doffed per policy. A mask worn by anyone being scanned has no metal nose strip or metallic fibres; staff masks, spectacles and visors are checked for ferromagnetic parts.
  • Cover the table with impervious sheets; strip and disinfect all contact surfaces and the bore entrance after the exam; allow the room air-change/settle time specified by IPC for airborne cases before the next patient.
  • MDRO/C. difficile: use sporicidal agents approved by both IPC and the scanner vendor.

C3: Ancillary Equipment & Linen

Responsibility: All staff

  • Only designated MR Conditional pumps/ventilators/monitors enter Zone IV for isolation or ICU patients; they are cleaned per both IPC and vendor instructions before and after.
  • Linen: metal-free; laundered per hospital policy; no patient-owned blankets in the bore without checking for metallic threads/heat packs.
  • Sharps and clinical waste containers in Zone III (non-ferromagnetic if kept in Zone IV).
Back to contents

Safety Governance

Roles, incident reporting, audit, document control, references

G1: Safety Roles

Appointments documented in writing with defined responsibilities

RoleWhoCore responsibility
MRMD (MR Medical Director)RadiologistOverall clinical MR safety authority; final risk–benefit decisions; approves policies
MRSO (MR Safety Officer)Senior MR radiographerDay-to-day safety operations, screening standards, training register, incident follow-up
MRSE (MR Safety Expert)MR physicist (may be external)Advice on complex implants, sequences/exposure limits, site planning, QA oversight

A named deputy exists for each role. Contact details are posted in the control room.

G2: Incident & Near-Miss Reporting

Responsibility: Any staff member reports · MRSO investigates · MRMD reviews

  1. Report all incidents and near-misses (screening failures, ferromagnetic items reaching Zone III/IV, burns of any severity, hearing complaints, quenches, device malfunctions) in the local incident system within 24 h.
  2. MRSO investigates using a no-blame, systems approach; corrective actions tracked to closure.
  3. External reporting where required: device-related harm to the national regulator (e.g. Medsafe NZ, MHRA UK, TGA AU, FDA MedWatch US) and to the manufacturer.
  4. Incidents and trends are a standing agenda item at the departmental safety meeting (at least quarterly).

G3: Training Matrix & Drills

Responsibility: MRSO

  • Annual MR safety update for all Level 1/2 personnel; records kept with expiry alerts.
  • Annual practical drills: Zone IV cardiac arrest evacuation (with the arrest team), fire response, and quench tabletop exercise.
  • New starters: no unsupervised Zone III/IV access until induction and competency sign-off.

G4: Document Control & Audit

Responsibility: MRSO + quality lead

  • All protocols carry: version, date, author, approver (MRMD), and review date (max 2-yearly, or after any relevant incident, new scanner, or guideline update).
  • Annual safety audit against the ACR Manual on MR Safety / MHRA guidelines checklist: zoning, signage, screening compliance sample, training currency, QA logs, incident closure (or use our interactive MR safety self-audit checklist).
  • The MR safety file (policies, training records, maintenance, incident log, audit results) is kept current and available for inspection.
Back to contents

Sources

Sources (6)

6 sources listed

  1. ACR Committee on MR Safety. ACR Manual on MR Safety, 2026 edition.
  2. MHRA. Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, 5th edition, 2026.
  3. RANZCR. MRI Safety Guidelines, version 3.0 (2021).
  4. IEC 60601-2-33 Ed 4.0 (2022): particular requirements for the safety of MR equipment.
  5. ACR Manual on Contrast Media, 2026 edition, and ESUR Contrast Media Safety Committee Guidelines 2026, for reaction management.
  6. ACR MRI Quality Control Manual for QA methodology.

More in Running the department

Next in Running the department: MRI QA and audit toolkit

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.