MRI in radiotherapy

MRI in radiotherapy

How magnetic resonance imaging is used to plan and guide radiation treatment, explained for patients and for radiation therapists.

For patientsFor professionals

On this pageFor PatientsWhy an MRI?What to expectRisksMRI-guided treatmentFor Radiation TherapistsMR simulationPlanning sequencesDistortion & QARegistration & contouringMR-LinacSafety in RTBrachytherapyMR-simulation patient positioning & immobilisation
Key points
  • MR simulation aims for geometric fidelity and a reproducible treatment position, not maximum diagnostic detail: a flat indexed couch, treatment immobilisation, and coils on bridges so the skin contour is not deformed.
  • In a CT-based workflow, CT stays the geometric and dosimetric reference and MRI adds soft-tissue definition; prefer rigid registration for rigid anatomy and verify the fusion at the target.
  • MRI geometry can be spatially inaccurate: use high bandwidth, 3D acquisitions and vendor distortion correction, keep the target near isocentre, and constrain the FOV to the QA-characterised region.
  • DWI/EPI can distort by several millimetres, so use it only as a biological overlay and delineate targets on distortion-corrected T1/T2 images, never on EPI geometry.
  • Verify that every RT accessory is MR Safe or MR Conditional before it enters Zone IV, and build documented MR status for each immobilisation and QA item into device commissioning.
Explain it simply
  • Planning MRI scans aim to show the body in the treatment position with accurate shapes, using a flat table, the same supports as treatment, and coils that don’t press on the skin.
  • The CT scan stays the reference for position and dose, while MRI shows soft tissue such as the tumour’s edges. The two are lined up, and staff check the match at the tumour itself.
  • MRI images can be distorted, especially away from the centre. Planning scans use settings and corrections that reduce this, and keep the tumour near the middle of the scanner.
  • Diffusion scans can be distorted by several millimetres, so they help show what the tumour is like, but its outline is drawn on other, corrected scans.
  • Radiotherapy staff check that every frame, bar, clamp and support is labelled MR Safe or MR Conditional (allowed under set conditions) before it enters the magnet room, and record each item’s status.

A guide only: follow your departmental protocols and regional requirements.

Related

For Patients

If you’re having radiotherapy, you may be sent for an MRI as part of planning your treatment.

Why do I need an MRI before radiotherapy?

Radiotherapy uses carefully aimed radiation to treat a tumour while sparing the healthy tissue around it. To aim accurately, your team needs a very clear picture of exactly where the tumour is and where the nearby organs sit.

The main planning scan for radiotherapy is a CT scan, because CT measures tissue density, which is what the radiation dose calculation needs. But CT is not always good at showing the edges of a tumour, especially in the brain, head and neck, prostate, and pelvis.

MRI shows soft tissue (brain, muscle, glands, tumour margins) far more clearly than CT. Your team overlays (“fuses”) the MRI onto the planning CT so they can draw the treatment target more precisely.

This gives a smaller, more accurate target, which usually means less healthy tissue is exposed to radiation.

What happens at the MRI appointment

Before

You’ll fill in a safety questionnaire about any metal implants, pacemakers, or previous surgery. Tell staff about everything, even old or minor implants.

Positioning

For planning scans you may be positioned exactly as you will be for treatment, sometimes on a flat couch, with the same headrest, mask, or supports. This helps the images line up with your treatment.

During

The scan is painless but noisy (knocking and buzzing). You’ll get ear protection, usually earplugs with headphones or ear muffs over them. Staying still is important. Ask your department how long your scan will take.

Contrast

Some scans use an injection of contrast dye (gadolinium) to make tumours stand out. A small cannula goes in your arm.

You can usually eat, drink and take your medication normally. Bring your appointment letter and arrive a little early for the safety check.

What are the risks of MRI?

MRI uses a strong magnet and radio waves, not X-rays, so there is no ionising radiation.

The main precaution is the magnet: it’s always on, so certain metal implants and devices can be a problem.

Always tell staff if you have: a pacemaker or defibrillator, cochlear (ear) implant, aneurysm clips, an implanted pump or stimulator, metal fragments in the eye, or any recent surgery. Many of these can still be scanned, but only after checking.

Reactions to the gadolinium contrast injection are uncommon and usually mild, and staff are trained to treat them. Let them know if you have kidney problems or have reacted to contrast before.

Want to look up an implant? Try the MRI Implant Safety Checker.

MRI-guided radiotherapy (MR-Linac)

Some centres now have a machine that combines an MRI scanner with a radiotherapy machine, called an MR-Linac. It takes live MRI pictures while you’re being treated.

Because the team can see soft tissue in real time, they can check the tumour hasn’t moved or changed shape that day and, in some cases, adjust the plan on the spot (“adaptive” radiotherapy). This is especially useful for tumours that move, like those in the abdomen and pelvis.

Treatment on an MR-Linac can take a bit longer per session, and you’ll go through the same magnet safety checks as a normal MRI. Not every patient or cancer type needs it; your team will advise.

Back to contents

For Radiation Therapists

A clinical reference on the role of MRI across the radiotherapy pathway: simulation, planning, distortion & QA, MR-Linac, contouring and safety.

MR simulation (MR-sim)

MR-sim is imaging performed specifically for treatment planning, distinct from diagnostic MRI. The goal is geometric fidelity and reproducibility of the treatment position, not maximum diagnostic detail.

Set-up requirements

ElementRationale
Flat (indexed) tabletop overlayReproduces the Linac/CT couch geometry; curved diagnostic couches change patient contour and position.
RT immobilisation (masks, boards, knee/ankle fix)Same devices as treatment; must be MR Conditional and coil-compatible.
External laser bridge / MR Conditional lasersMarks a reproducible reference/isocentre point.
Large / open coils, coil bridges, body arrays on standsAvoid deforming the patient contour; keep coils off the skin.
Wide-bore magnet (≥70 cm)Accommodates immobilisation devices and treatment position.
Reproducing the treatment position is a key requirement of MR-sim over diagnostic MRI. If the patient is positioned differently, downstream registration accuracy is compromised.
Planning sequences

Sequence choice balances tumour/OAR conspicuity against geometric accuracy and, for MR-only workflows, the ability to derive electron density.

PurposeTypical sequenceNotes
GTV delineation (brain, H&N)3D T1 post-Gd (MPRAGE/BRAVO)Isotropic, thin slices for fusion; enhancing margin.
Oedema / non-enhancing tumour3D T2 FLAIRGlioma infiltrative margin, CTV concepts.
Prostate target/zonal anatomyHigh-res T2 TSE (axial)Small FOV; correlate with mpMRI lesion.
Cellularity / responseDWI + ADCEPI: high distortion; use for biology, not geometry.
MR-only planningDixon / UTE / dedicated sCT sequencesFeeds synthetic-CT (pseudo-CT) electron density.

Representative parameters by site (1.5 T & 3 T)

Indicative values only; optimise on your own scanner/coil and to local protocol. TR/TE/TI in ms; BW = receiver bandwidth (Hz/px). 3 T gains SNR/resolution but has more susceptibility & chemical-shift distortion, hence higher BW.

Brain (SRS / glioma)

SequenceFieldTRTETIVoxel (mm)FOVBW
3D T1 MPRAGE post-Gd (Siemens convention: TR = time between inversions; GE BRAVO shows the gradient-echo TR and a separate prep time instead)1.5 T1900–23002.3–3.09001×1×1 iso256~200
3D T1 MPRAGE post-Gd (Siemens convention)3 T2100–25002.0–2.7900–11000.8–1 iso256~240
3D T2 FLAIR1.5 T4800–6000350–3901650–18001×1×1 iso256~750
3D T2 FLAIR3 T5000–6000380–4201700–19001 iso256~780
3D T2 SPACE/CUBE1.5 T3200400—1×1×1 iso256~590
3D T2 SPACE/CUBE3 T2500–3200400–560—0.8–1 iso256~650

Head & neck

SequenceFieldTRTESlice (mm)In-plane (mm)FOVBW
Axial T2 TSE fat-sat1.5 T4000–500080–10030.6–0.8230–250~300
Axial T2 TSE fat-sat3 T4000–500080–1002.5–30.5–0.7230–250~400
Axial T1 TSE post-Gd fat-sat1.5 T600–7508–1230.6–0.8230–250~250
Axial T1 TSE post-Gd fat-sat3 T600–8008–122.5–30.5–0.7230–250~350
3D T1 post-Gd (fusion; MPRAGE-type, Siemens convention)1.5 T19002.51 iso1 iso256~200
3D T1 post-Gd (fusion; MPRAGE-type, Siemens convention)3 T2000–23002.20.9–1 iso0.9–1 iso256~240

Prostate

SequenceFieldTRTESlice (mm)In-plane (mm)FOVBW
Axial high-res T2 TSE1.5 T4000–600090–1103 (no gap)0.5–0.7180–200~200
Axial high-res T2 TSE3 T3000–500090–1203 (no gap)0.4–0.6180–200~220
Sagittal / coronal T2 TSE1.5 T400010030.6200~200
Sagittal / coronal T2 TSE3 T3500–450010030.5200~220
DWI + ADC (biology only)1.5 / 3 T3000+min3–41.5–2200–260high (EPI)

Pelvis (rectum / gynae / cervix)

SequenceFieldTRTESlice (mm)In-plane (mm)FOVBW
Oblique small-FOV T2 TSE (⊥ to tumour axis)1.5 T4000–500090–11030.6–0.8200–240~200
Oblique small-FOV T2 TSE (⊥ to tumour axis)3 T3500–500090–11030.5–0.7200–240~240
Sagittal T2 TSE1.5 T40001003–40.7240~200
Sagittal T2 TSE3 T3500–45001003–40.6240~240
Large-FOV T1/T2 for fusion1.5 / 3 T——31350–400~300

Geometric-fidelity settings (all sites)

Prefer high receiver bandwidth (reduces chemical-shift and B0 distortion), 3D acquisitions, isotropic voxels, and vendor 2D/3D distortion-correction (gradient non-linearity) filters ON. Keep the target near isocentre and constrain FOV to the QA-characterised region. Avoid EPI-based sequences for the geometry that defines the target.

DWI/EPI can distort by several millimetres along the phase-encode direction, often most at air–tissue interfaces. Never contour a target on EPI geometry; use it as biological overlay only.

Higher field (3 T) improves SNR/resolution but increases susceptibility and chemical-shift distortion, so raise bandwidth accordingly. MR-only workflows add a dedicated Dixon/UTE sCT sequence for electron density.

Geometric distortion & QA

Unlike CT, MRI geometry can be spatially inaccurate. Two error classes matter for RT:

System-related

Gradient non-linearity (grows toward periphery of FOV) and B0 inhomogeneity. Largely correctable with vendor distortion-correction and shimming.

Patient/sequence-related

Susceptibility (air–tissue, implants) and chemical shift. Increase at higher field strength and lower receiver bandwidth; mitigated by higher bandwidth.

QA programme

Use a large-FOV geometric-distortion phantom to map residual distortion across the bore; define a clinically acceptable tolerance (commonly <1–2 mm within the relevant FOV). Include daily/weekly geometric checks, coil and SNR checks, and end-to-end tests. Frameworks: AAPM TG-284 (MR-sim), ACR MRI QA, IPEM guidance.

Keep the target near isocentre where gradient non-linearity is smallest, and constrain FOV to the region where distortion is characterised and within tolerance.
Registration & contouring

In a CT-based workflow, MRI is fused to the planning CT. The CT remains the geometric and dosimetric reference; MRI contributes soft-tissue definition.

CT gives the electron density the dose engine needs; MRI shows where the tumour actually is. Fusing them lets the target be drawn on soft-tissue MRI while dose is still computed on CT geometry.

Registration

Prefer rigid registration for intracranial and other rigid anatomy; it’s verifiable and traceable. Use deformable registration cautiously (moving anatomy, filling changes); always validate on bony/anatomical landmarks near the target. Match the imaging position to the planning-CT position to minimise registration error.

Contouring pitfalls

PitfallMitigation
Contouring on distorted (EPI) geometryDelineate on distortion-corrected T1/T2 images, using 3D sequences with 3D distortion correction where available, never on EPI geometry; follow your department’s MR-sim protocol and your physicist’s distortion QA.
Position mismatch CT vs MRReproduce treatment position at MR-sim.
Registration error propagated to GTVVerify fusion at the target, not globally.
Slice thickness too coarseIsotropic thin slices for fusion.
MR-Linac & MR-guided adaptive RT

Integrated MRI + linear accelerator systems (e.g. 1.5 T and 0.35 T platforms) provide on-table soft-tissue imaging for daily set-up and online adaptive replanning.

Capabilities

Daily soft-tissue set-up without added imaging dose; real-time cine imaging for motion monitoring and gating; online adaptive re-contouring and re-optimisation to the anatomy of the day. Strong candidates: pancreas, liver, adrenal, prostate, and other sites with day-to-day variation or nearby serial OARs.

Practical considerations

FactorImplication
Electron Return Effect (ERE)Magnetic field perturbs dose at tissue–air interfaces; accounted for in the TPS.
Longer on-table timeAdaptive workflow adds re-contour/re-plan/QA steps.
MR Conditional everythingImmobilisation, monitoring, and QA equipment must be labelled MR Safe or MR Conditional.
Team workflowPhysicist, RTT and oncologist often present per fraction for adaptive.

See also the dedicated MR-LINAC page for a deeper dive into MRI-guided adaptive radiotherapy.

Safety in the RT setting

MR-sim and MR-Linac bring RT staff and equipment into the magnetic environment; many are less familiar with MR hazards than diagnostic MR staff.

Projectile risk: RT accessories (bolus frames, indexing bars, clamps, tools) must be verified MR Safe or MR Conditional before entering Zone IV. Ferromagnetic immobilisation hardware is a recurring hazard.

Maintain the four-zone access model, formal screening for patients and staff, and hearing protection. Confirm implant conditions (field strength, SAR, gradient limits) against the device labelling. For contrast, follow local gadolinium and renal-screening policy.

Build MR safety sign-off into the RT device commissioning process: every immobilisation and QA item used at MR-sim/MR-Linac should carry documented MR status.

Verify individual devices with the MRI Implant Safety Checker. Frameworks: ACR Manual on MR Safety, 2026 edition; AAPM TG-284; MHRA Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, 5th edition (2026).

MRI in brachytherapy

MRI is the reference imaging standard for image-guided adaptive brachytherapy (IGABT), most established in locally advanced cervical cancer. It defines the residual tumour at the time of each application, so the high-dose region tracks tumour regression.

Cervix (GEC-ESTRO framework)

With the applicator in situ, MRI is used to delineate the GTV-res (residual gross tumour), the high-risk CTV (HR-CTV) and intermediate-risk CTV, plus OARs (bladder, rectum, sigmoid, bowel). Dose is reported to HR-CTV D90 and OAR D2cc per the ICRU-89 / GEC-ESTRO recommendations.

SequenceFieldTRTESlice (mm)In-plane (mm)Notes
Para-transverse T2 TSE (⊥ to uterine axis)1.5 T4000–500090–1103–4 (no gap)0.5–0.8Primary delineation series
Para-transverse T2 TSE3 T3500–500090–1103 (no gap)0.4–0.6Higher SNR; watch susceptibility near applicator
Sagittal & coronal T2 TSE1.5 / 3 T40001003–40.6–0.8Applicator geometry, uterine axis
Scan with the applicator in place and the patient in the treatment position. Use MR Conditional applicators and follow the vendor’s field/SAR conditions; titanium/plastic applicators reduce artefact vs steel.
Applicator-tip and dwell-position identification drive dosimetry. Confirm reconstruction on the T2 series (or a dedicated applicator-visualisation sequence / CT-MR fusion); susceptibility can obscure the tip.

Other sites using MR-guided brachytherapy include prostate (HDR boost) and, less commonly, vaginal and other gynae presentations. Principles are the same: MR for soft-tissue target/OAR definition, applicator reconstruction, and D90/D2cc reporting.

Frameworks: GEC-ESTRO (GYN) recommendations; ICRU Report 89; EMBRACE-II protocol.

MR-simulation patient positioning & immobilisation

Setup on an MR-sim (or MR-Linac) mirrors the treatment position exactly: a flat RT couch overlay replaces the curved diagnostic table, immobilisation is indexed to the couch, and coils are supported on bridges/standoffs so they never touch and deform the skin contour. Keep the target isocentre near bore centre to minimise gradient non-linearity, and apply large-FOV 3D distortion correction for planning geometry.

Brain / CNS
Position
Supine, head first, chin tucked to neutral; arms at sides.
Immobilise
Thermoplastic head mask over an indexed headrest (e.g. type S); locked to couch.
Coil setup
Flex/head array on a coil bridge clearing the mask, with no compression of contour.
Landmarks
External lasers to mask marks; reference point near lesion.
Motion
Free-breathing; motion negligible; mask fixation is key.
QA note
Brain sits near isocentre; still apply 3D distortion correction for stereotactic accuracy.

Tip: For SRS, verify residual distortion at field edge; small mask marks double as fusion checks.

Head & neck
Position
Supine, head first, hyperextended per plan; shoulders pulled down.
Immobilise
5-point thermoplastic mask (head + shoulders) on indexed headrest; shoulder retraction straps.
Coil setup
Anterior flex coil on bridge/standoffs off the mask; posterior in-couch array.
Landmarks
Lasers to mask; bite-block/reference marks for reproducibility.
Motion
Free-breathing; swallowing minimised by fast sequences.
QA note
Long neck FOV: distortion correction essential; keep larynx toward bore centre.

Tip: Reproduce the treatment shoulder position (shoulders down) exactly, so the lower neck anatomy matches the planning CT and treatment.

Breast
Position
Supine on inclined breast board, both arms up in arm troughs; head turned away.
Immobilise
Angled breast board indexed to couch; wrist/hand grips; incline flattens breast.
Coil setup
Body flex coil on bridge over sternum, never resting on breast tissue.
Landmarks
Lasers to sternal/lateral tattoos or surface-guidance (SGRT) reference.
Motion
DIBH for left breast to spare heart; free-breathing or breath-hold otherwise.
QA note
Large lateral extent; apply distortion correction; keep chest wall near iso.

Tip: A coil bridge is essential here; any contact deforms the breast and invalidates the planning contour.

Thorax / lung
Position
Supine, arms up over head on a wing board; head first.
Immobilise
Wing board with indexed arm/hand grips; optional vac bag for stability.
Coil setup
Anterior body array on bridge/standoffs; no chest compression that alters breathing.
Landmarks
Lasers to thoracic tattoos; SGRT for gating surrogate.
Motion
4D/free-breathing for tumour trajectory; DIBH or abdominal compression to limit excursion.
QA note
Flat couch + iso near carina; distortion correction over wide thorax FOV.

Tip: On MR-Linac, real-time cine tracks the target, so reproduce the exact breathing regime used at sim.

Abdomen
Position
Supine, arms up on wing board; head first, straight alignment.
Immobilise
Wing board + vac bag; abdominal compression plate if used per protocol.
Coil setup
Body array on bridge; standoffs prevent contour deformation of soft abdomen.
Landmarks
Lasers to abdominal tattoos; reference near target (e.g. pancreas, adrenal).
Motion
Breath-hold (exhale) preferred for upper-abdomen SBRT; 4D or compression as alternative.
QA note
Keep target near bore centre; large-FOV distortion correction for geometric fidelity.

Tip: Standardise breath-hold phase between sim and treatment.

Pelvis / prostate
Position
Supine, arms on chest, legs straight; head first or feet first per plan.
Immobilise
Indexed knee + ankle fix (foam lock); optional vac bag.
Coil setup
Body array on bridge over pelvis; no pressure on lower abdomen/bladder.
Landmarks
Lasers to pelvic tattoos; bladder-full / rectum-empty prep protocol.
Motion
Free-breathing; organ motion managed by bladder/rectum prep, not breathing.
QA note
Prostate at bore centre; endorectal coils/balloons displace anatomy (usually avoided for planning).

Tip: Reproduce bladder/rectum status at every fraction; an endorectal balloon changes the planning contour.

Rectum / anus
Position
Prone on belly-board (displaces small bowel) or supine per protocol; arms up/forward.
Immobilise
Indexed belly-board with abdominal aperture; foam/vac bag support.
Coil setup
Posterior in-couch + anterior flex on bridge; avoid compressing belly-board aperture.
Landmarks
Lasers to sacral/gluteal tattoos; reference at target level.
Motion
Free-breathing; bladder-full prep spares small bowel.
QA note
Verify couch/board flatness; distortion correction over pelvic FOV.

Tip: Belly-board only helps if bladder-filling and hip flexion are reproduced identically each day.

Gynae
Position
Supine, arms on chest, legs straight and neutral; head first.
Immobilise
Indexed knee/ankle fix or vac bag; consistent pelvic tilt.
Coil setup
Body array on bridge over pelvis; no compression of lower abdomen.
Landmarks
Lasers to pelvic tattoos; bladder-filling protocol (full or comfortably full).
Motion
Free-breathing; uterine/target motion managed by bladder/rectum prep.
QA note
Uterus/cervix near iso; large-FOV distortion correction; note brachytherapy applicators separately.

Tip: Bladder volume strongly shifts the uterus; standardise the fill and time it consistently.

Cervical / spine SBRT
Position
Supine, head first; neutral spine alignment along couch long axis.
Immobilise
Full-length vac bag; wing board for arms; thermoplastic mask for cervical/upper-thoracic levels.
Coil setup
Posterior in-couch spine array + anterior flex on bridge; no coil weight on the spine contour.
Landmarks
Lasers to spinal-level tattoos; vertebral reference for fusion.
Motion
Free-breathing; rigid immobilisation critical for sub-mm SBRT accuracy.
QA note
Distortion is a major SBRT risk; verify geometric accuracy at treated level, keep it near iso.

Tip: For cord-adjacent SBRT, confirm residual distortion at the target level is within your local QA tolerance before planning.

Extremity / sarcoma
Position
Limb-dependent; affected extremity positioned/abducted to spare adjacent tissue and skin folds.
Immobilise
Custom vac bag moulded to the limb; indexed to couch; consistent joint angle.
Coil setup
Flex coil wrapped loosely on a standoff/bridge, never strapped tight enough to compress the limb.
Landmarks
Lasers to limb tattoos; reference marks proximal and distal for long targets.
Motion
Free-breathing; limb SBRT relies on rigid, reproducible vac-bag fixation.
QA note
Off-centre anatomy suffers most distortion; position the target as close to bore centre as feasible and correct.

Tip: Off-axis limbs are the worst case for gradient non-linearity; centre the tumour as close to the bore centre as feasible as well as applying distortion correction.

General MR-sim principles: flat RT couch overlay indexed to immobilisation; external lasers/tattoos or surface guidance for setup; coils supported on bridges/standoffs so the skin contour used for planning is never deformed; target isocentre near bore centre with large-FOV 3D distortion correction to preserve planning geometry. Reproduce the treatment breathing regime (DIBH / 4D / breath-hold / compression) exactly between simulation and delivery.

Educational reference only. Follow your local departmental protocols and current professional guidance. Not a substitute for clinical judgement.

More in Running the department

Next in Running the department: MRI for allied health and referrers

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.