MRI in radiotherapy
How magnetic resonance imaging is used to plan and guide radiation treatment, explained for patients and for radiation therapists.
On this page
For PatientsWhy an MRI?What to expectRisksMRI-guided treatmentFor Radiation TherapistsMR simulationPlanning sequencesDistortion & QARegistration & contouringMR-LinacSafety in RTBrachytherapyMR-simulation patient positioning & immobilisation- MR simulation aims for geometric fidelity and a reproducible treatment position, not maximum diagnostic detail: a flat indexed couch, treatment immobilisation, and coils on bridges so the skin contour is not deformed.
- In a CT-based workflow, CT stays the geometric and dosimetric reference and MRI adds soft-tissue definition; prefer rigid registration for rigid anatomy and verify the fusion at the target.
- MRI geometry can be spatially inaccurate: use high bandwidth, 3D acquisitions and vendor distortion correction, keep the target near isocentre, and constrain the FOV to the QA-characterised region.
- DWI/EPI can distort by several millimetres, so use it only as a biological overlay and delineate targets on distortion-corrected T1/T2 images, never on EPI geometry.
- Verify that every RT accessory is MR Safe or MR Conditional before it enters Zone IV, and build documented MR status for each immobilisation and QA item into device commissioning.
Explain it simply
- Planning MRI scans aim to show the body in the treatment position with accurate shapes, using a flat table, the same supports as treatment, and coils that don’t press on the skin.
- The CT scan stays the reference for position and dose, while MRI shows soft tissue such as the tumour’s edges. The two are lined up, and staff check the match at the tumour itself.
- MRI images can be distorted, especially away from the centre. Planning scans use settings and corrections that reduce this, and keep the tumour near the middle of the scanner.
- Diffusion scans can be distorted by several millimetres, so they help show what the tumour is like, but its outline is drawn on other, corrected scans.
- Radiotherapy staff check that every frame, bar, clamp and support is labelled MR Safe or MR Conditional (allowed under set conditions) before it enters the magnet room, and record each item’s status.
A guide only: follow your departmental protocols and regional requirements.
For Patients
If you’re having radiotherapy, you may be sent for an MRI as part of planning your treatment.
Why do I need an MRI before radiotherapy?
Radiotherapy uses carefully aimed radiation to treat a tumour while sparing the healthy tissue around it. To aim accurately, your team needs a very clear picture of exactly where the tumour is and where the nearby organs sit.
The main planning scan for radiotherapy is a CT scan, because CT measures tissue density, which is what the radiation dose calculation needs. But CT is not always good at showing the edges of a tumour, especially in the brain, head and neck, prostate, and pelvis.
This gives a smaller, more accurate target, which usually means less healthy tissue is exposed to radiation.
What happens at the MRI appointment
Before
You’ll fill in a safety questionnaire about any metal implants, pacemakers, or previous surgery. Tell staff about everything, even old or minor implants.
Positioning
For planning scans you may be positioned exactly as you will be for treatment, sometimes on a flat couch, with the same headrest, mask, or supports. This helps the images line up with your treatment.
During
The scan is painless but noisy (knocking and buzzing). You’ll get ear protection, usually earplugs with headphones or ear muffs over them. Staying still is important. Ask your department how long your scan will take.
Contrast
Some scans use an injection of contrast dye (gadolinium) to make tumours stand out. A small cannula goes in your arm.
What are the risks of MRI?
MRI uses a strong magnet and radio waves, not X-rays, so there is no ionising radiation.
The main precaution is the magnet: it’s always on, so certain metal implants and devices can be a problem.
Reactions to the gadolinium contrast injection are uncommon and usually mild, and staff are trained to treat them. Let them know if you have kidney problems or have reacted to contrast before.
Want to look up an implant? Try the MRI Implant Safety Checker.
MRI-guided radiotherapy (MR-Linac)
Some centres now have a machine that combines an MRI scanner with a radiotherapy machine, called an MR-Linac. It takes live MRI pictures while you’re being treated.
Treatment on an MR-Linac can take a bit longer per session, and you’ll go through the same magnet safety checks as a normal MRI. Not every patient or cancer type needs it; your team will advise.
For Radiation Therapists
A clinical reference on the role of MRI across the radiotherapy pathway: simulation, planning, distortion & QA, MR-Linac, contouring and safety.
MR simulation (MR-sim)
MR-sim is imaging performed specifically for treatment planning, distinct from diagnostic MRI. The goal is geometric fidelity and reproducibility of the treatment position, not maximum diagnostic detail.
Set-up requirements
| Element | Rationale |
|---|---|
| Flat (indexed) tabletop overlay | Reproduces the Linac/CT couch geometry; curved diagnostic couches change patient contour and position. |
| RT immobilisation (masks, boards, knee/ankle fix) | Same devices as treatment; must be MR Conditional and coil-compatible. |
| External laser bridge / MR Conditional lasers | Marks a reproducible reference/isocentre point. |
| Large / open coils, coil bridges, body arrays on stands | Avoid deforming the patient contour; keep coils off the skin. |
| Wide-bore magnet (≥70 cm) | Accommodates immobilisation devices and treatment position. |
Planning sequences
Sequence choice balances tumour/OAR conspicuity against geometric accuracy and, for MR-only workflows, the ability to derive electron density.
| Purpose | Typical sequence | Notes |
|---|---|---|
| GTV delineation (brain, H&N) | 3D T1 post-Gd (MPRAGE/BRAVO) | Isotropic, thin slices for fusion; enhancing margin. |
| Oedema / non-enhancing tumour | 3D T2 FLAIR | Glioma infiltrative margin, CTV concepts. |
| Prostate target/zonal anatomy | High-res T2 TSE (axial) | Small FOV; correlate with mpMRI lesion. |
| Cellularity / response | DWI + ADC | EPI: high distortion; use for biology, not geometry. |
| MR-only planning | Dixon / UTE / dedicated sCT sequences | Feeds synthetic-CT (pseudo-CT) electron density. |
Representative parameters by site (1.5 T & 3 T)
Indicative values only; optimise on your own scanner/coil and to local protocol. TR/TE/TI in ms; BW = receiver bandwidth (Hz/px). 3 T gains SNR/resolution but has more susceptibility & chemical-shift distortion, hence higher BW.
Brain (SRS / glioma)
| Sequence | Field | TR | TE | TI | Voxel (mm) | FOV | BW |
|---|---|---|---|---|---|---|---|
| 3D T1 MPRAGE post-Gd (Siemens convention: TR = time between inversions; GE BRAVO shows the gradient-echo TR and a separate prep time instead) | 1.5 T | 1900–2300 | 2.3–3.0 | 900 | 1×1×1 iso | 256 | ~200 |
| 3D T1 MPRAGE post-Gd (Siemens convention) | 3 T | 2100–2500 | 2.0–2.7 | 900–1100 | 0.8–1 iso | 256 | ~240 |
| 3D T2 FLAIR | 1.5 T | 4800–6000 | 350–390 | 1650–1800 | 1×1×1 iso | 256 | ~750 |
| 3D T2 FLAIR | 3 T | 5000–6000 | 380–420 | 1700–1900 | 1 iso | 256 | ~780 |
| 3D T2 SPACE/CUBE | 1.5 T | 3200 | 400 | — | 1×1×1 iso | 256 | ~590 |
| 3D T2 SPACE/CUBE | 3 T | 2500–3200 | 400–560 | — | 0.8–1 iso | 256 | ~650 |
Head & neck
| Sequence | Field | TR | TE | Slice (mm) | In-plane (mm) | FOV | BW |
|---|---|---|---|---|---|---|---|
| Axial T2 TSE fat-sat | 1.5 T | 4000–5000 | 80–100 | 3 | 0.6–0.8 | 230–250 | ~300 |
| Axial T2 TSE fat-sat | 3 T | 4000–5000 | 80–100 | 2.5–3 | 0.5–0.7 | 230–250 | ~400 |
| Axial T1 TSE post-Gd fat-sat | 1.5 T | 600–750 | 8–12 | 3 | 0.6–0.8 | 230–250 | ~250 |
| Axial T1 TSE post-Gd fat-sat | 3 T | 600–800 | 8–12 | 2.5–3 | 0.5–0.7 | 230–250 | ~350 |
| 3D T1 post-Gd (fusion; MPRAGE-type, Siemens convention) | 1.5 T | 1900 | 2.5 | 1 iso | 1 iso | 256 | ~200 |
| 3D T1 post-Gd (fusion; MPRAGE-type, Siemens convention) | 3 T | 2000–2300 | 2.2 | 0.9–1 iso | 0.9–1 iso | 256 | ~240 |
Prostate
| Sequence | Field | TR | TE | Slice (mm) | In-plane (mm) | FOV | BW |
|---|---|---|---|---|---|---|---|
| Axial high-res T2 TSE | 1.5 T | 4000–6000 | 90–110 | 3 (no gap) | 0.5–0.7 | 180–200 | ~200 |
| Axial high-res T2 TSE | 3 T | 3000–5000 | 90–120 | 3 (no gap) | 0.4–0.6 | 180–200 | ~220 |
| Sagittal / coronal T2 TSE | 1.5 T | 4000 | 100 | 3 | 0.6 | 200 | ~200 |
| Sagittal / coronal T2 TSE | 3 T | 3500–4500 | 100 | 3 | 0.5 | 200 | ~220 |
| DWI + ADC (biology only) | 1.5 / 3 T | 3000+ | min | 3–4 | 1.5–2 | 200–260 | high (EPI) |
Pelvis (rectum / gynae / cervix)
| Sequence | Field | TR | TE | Slice (mm) | In-plane (mm) | FOV | BW |
|---|---|---|---|---|---|---|---|
| Oblique small-FOV T2 TSE (⊥ to tumour axis) | 1.5 T | 4000–5000 | 90–110 | 3 | 0.6–0.8 | 200–240 | ~200 |
| Oblique small-FOV T2 TSE (⊥ to tumour axis) | 3 T | 3500–5000 | 90–110 | 3 | 0.5–0.7 | 200–240 | ~240 |
| Sagittal T2 TSE | 1.5 T | 4000 | 100 | 3–4 | 0.7 | 240 | ~200 |
| Sagittal T2 TSE | 3 T | 3500–4500 | 100 | 3–4 | 0.6 | 240 | ~240 |
| Large-FOV T1/T2 for fusion | 1.5 / 3 T | — | — | 3 | 1 | 350–400 | ~300 |
Geometric-fidelity settings (all sites)
Prefer high receiver bandwidth (reduces chemical-shift and B0 distortion), 3D acquisitions, isotropic voxels, and vendor 2D/3D distortion-correction (gradient non-linearity) filters ON. Keep the target near isocentre and constrain FOV to the QA-characterised region. Avoid EPI-based sequences for the geometry that defines the target.
Higher field (3 T) improves SNR/resolution but increases susceptibility and chemical-shift distortion, so raise bandwidth accordingly. MR-only workflows add a dedicated Dixon/UTE sCT sequence for electron density.
Geometric distortion & QA
Unlike CT, MRI geometry can be spatially inaccurate. Two error classes matter for RT:
System-related
Gradient non-linearity (grows toward periphery of FOV) and B0 inhomogeneity. Largely correctable with vendor distortion-correction and shimming.
Patient/sequence-related
Susceptibility (air–tissue, implants) and chemical shift. Increase at higher field strength and lower receiver bandwidth; mitigated by higher bandwidth.
QA programme
Use a large-FOV geometric-distortion phantom to map residual distortion across the bore; define a clinically acceptable tolerance (commonly <1–2 mm within the relevant FOV). Include daily/weekly geometric checks, coil and SNR checks, and end-to-end tests. Frameworks: AAPM TG-284 (MR-sim), ACR MRI QA, IPEM guidance.
Registration & contouring
In a CT-based workflow, MRI is fused to the planning CT. The CT remains the geometric and dosimetric reference; MRI contributes soft-tissue definition.
CT gives the electron density the dose engine needs; MRI shows where the tumour actually is. Fusing them lets the target be drawn on soft-tissue MRI while dose is still computed on CT geometry.
Registration
Prefer rigid registration for intracranial and other rigid anatomy; it’s verifiable and traceable. Use deformable registration cautiously (moving anatomy, filling changes); always validate on bony/anatomical landmarks near the target. Match the imaging position to the planning-CT position to minimise registration error.
Contouring pitfalls
| Pitfall | Mitigation |
|---|---|
| Contouring on distorted (EPI) geometry | Delineate on distortion-corrected T1/T2 images, using 3D sequences with 3D distortion correction where available, never on EPI geometry; follow your department’s MR-sim protocol and your physicist’s distortion QA. |
| Position mismatch CT vs MR | Reproduce treatment position at MR-sim. |
| Registration error propagated to GTV | Verify fusion at the target, not globally. |
| Slice thickness too coarse | Isotropic thin slices for fusion. |
MR-Linac & MR-guided adaptive RT
Integrated MRI + linear accelerator systems (e.g. 1.5 T and 0.35 T platforms) provide on-table soft-tissue imaging for daily set-up and online adaptive replanning.
Capabilities
Daily soft-tissue set-up without added imaging dose; real-time cine imaging for motion monitoring and gating; online adaptive re-contouring and re-optimisation to the anatomy of the day. Strong candidates: pancreas, liver, adrenal, prostate, and other sites with day-to-day variation or nearby serial OARs.
Practical considerations
| Factor | Implication |
|---|---|
| Electron Return Effect (ERE) | Magnetic field perturbs dose at tissue–air interfaces; accounted for in the TPS. |
| Longer on-table time | Adaptive workflow adds re-contour/re-plan/QA steps. |
| MR Conditional everything | Immobilisation, monitoring, and QA equipment must be labelled MR Safe or MR Conditional. |
| Team workflow | Physicist, RTT and oncologist often present per fraction for adaptive. |
See also the dedicated MR-LINAC page for a deeper dive into MRI-guided adaptive radiotherapy.
Safety in the RT setting
MR-sim and MR-Linac bring RT staff and equipment into the magnetic environment; many are less familiar with MR hazards than diagnostic MR staff.
Maintain the four-zone access model, formal screening for patients and staff, and hearing protection. Confirm implant conditions (field strength, SAR, gradient limits) against the device labelling. For contrast, follow local gadolinium and renal-screening policy.
Verify individual devices with the MRI Implant Safety Checker. Frameworks: ACR Manual on MR Safety, 2026 edition; AAPM TG-284; MHRA Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use, 5th edition (2026).
MRI in brachytherapy
MRI is the reference imaging standard for image-guided adaptive brachytherapy (IGABT), most established in locally advanced cervical cancer. It defines the residual tumour at the time of each application, so the high-dose region tracks tumour regression.
Cervix (GEC-ESTRO framework)
With the applicator in situ, MRI is used to delineate the GTV-res (residual gross tumour), the high-risk CTV (HR-CTV) and intermediate-risk CTV, plus OARs (bladder, rectum, sigmoid, bowel). Dose is reported to HR-CTV D90 and OAR D2cc per the ICRU-89 / GEC-ESTRO recommendations.
| Sequence | Field | TR | TE | Slice (mm) | In-plane (mm) | Notes |
|---|---|---|---|---|---|---|
| Para-transverse T2 TSE (⊥ to uterine axis) | 1.5 T | 4000–5000 | 90–110 | 3–4 (no gap) | 0.5–0.8 | Primary delineation series |
| Para-transverse T2 TSE | 3 T | 3500–5000 | 90–110 | 3 (no gap) | 0.4–0.6 | Higher SNR; watch susceptibility near applicator |
| Sagittal & coronal T2 TSE | 1.5 / 3 T | 4000 | 100 | 3–4 | 0.6–0.8 | Applicator geometry, uterine axis |
Other sites using MR-guided brachytherapy include prostate (HDR boost) and, less commonly, vaginal and other gynae presentations. Principles are the same: MR for soft-tissue target/OAR definition, applicator reconstruction, and D90/D2cc reporting.
Frameworks: GEC-ESTRO (GYN) recommendations; ICRU Report 89; EMBRACE-II protocol.
MR-simulation patient positioning & immobilisation
Setup on an MR-sim (or MR-Linac) mirrors the treatment position exactly: a flat RT couch overlay replaces the curved diagnostic table, immobilisation is indexed to the couch, and coils are supported on bridges/standoffs so they never touch and deform the skin contour. Keep the target isocentre near bore centre to minimise gradient non-linearity, and apply large-FOV 3D distortion correction for planning geometry.
Brain / CNS
- Position
- Supine, head first, chin tucked to neutral; arms at sides.
- Immobilise
- Thermoplastic head mask over an indexed headrest (e.g. type S); locked to couch.
- Coil setup
- Flex/head array on a coil bridge clearing the mask, with no compression of contour.
- Landmarks
- External lasers to mask marks; reference point near lesion.
- Motion
- Free-breathing; motion negligible; mask fixation is key.
- QA note
- Brain sits near isocentre; still apply 3D distortion correction for stereotactic accuracy.
Tip: For SRS, verify residual distortion at field edge; small mask marks double as fusion checks.
Head & neck
- Position
- Supine, head first, hyperextended per plan; shoulders pulled down.
- Immobilise
- 5-point thermoplastic mask (head + shoulders) on indexed headrest; shoulder retraction straps.
- Coil setup
- Anterior flex coil on bridge/standoffs off the mask; posterior in-couch array.
- Landmarks
- Lasers to mask; bite-block/reference marks for reproducibility.
- Motion
- Free-breathing; swallowing minimised by fast sequences.
- QA note
- Long neck FOV: distortion correction essential; keep larynx toward bore centre.
Tip: Reproduce the treatment shoulder position (shoulders down) exactly, so the lower neck anatomy matches the planning CT and treatment.
Breast
- Position
- Supine on inclined breast board, both arms up in arm troughs; head turned away.
- Immobilise
- Angled breast board indexed to couch; wrist/hand grips; incline flattens breast.
- Coil setup
- Body flex coil on bridge over sternum, never resting on breast tissue.
- Landmarks
- Lasers to sternal/lateral tattoos or surface-guidance (SGRT) reference.
- Motion
- DIBH for left breast to spare heart; free-breathing or breath-hold otherwise.
- QA note
- Large lateral extent; apply distortion correction; keep chest wall near iso.
Tip: A coil bridge is essential here; any contact deforms the breast and invalidates the planning contour.
Thorax / lung
- Position
- Supine, arms up over head on a wing board; head first.
- Immobilise
- Wing board with indexed arm/hand grips; optional vac bag for stability.
- Coil setup
- Anterior body array on bridge/standoffs; no chest compression that alters breathing.
- Landmarks
- Lasers to thoracic tattoos; SGRT for gating surrogate.
- Motion
- 4D/free-breathing for tumour trajectory; DIBH or abdominal compression to limit excursion.
- QA note
- Flat couch + iso near carina; distortion correction over wide thorax FOV.
Tip: On MR-Linac, real-time cine tracks the target, so reproduce the exact breathing regime used at sim.
Abdomen
- Position
- Supine, arms up on wing board; head first, straight alignment.
- Immobilise
- Wing board + vac bag; abdominal compression plate if used per protocol.
- Coil setup
- Body array on bridge; standoffs prevent contour deformation of soft abdomen.
- Landmarks
- Lasers to abdominal tattoos; reference near target (e.g. pancreas, adrenal).
- Motion
- Breath-hold (exhale) preferred for upper-abdomen SBRT; 4D or compression as alternative.
- QA note
- Keep target near bore centre; large-FOV distortion correction for geometric fidelity.
Tip: Standardise breath-hold phase between sim and treatment.
Pelvis / prostate
- Position
- Supine, arms on chest, legs straight; head first or feet first per plan.
- Immobilise
- Indexed knee + ankle fix (foam lock); optional vac bag.
- Coil setup
- Body array on bridge over pelvis; no pressure on lower abdomen/bladder.
- Landmarks
- Lasers to pelvic tattoos; bladder-full / rectum-empty prep protocol.
- Motion
- Free-breathing; organ motion managed by bladder/rectum prep, not breathing.
- QA note
- Prostate at bore centre; endorectal coils/balloons displace anatomy (usually avoided for planning).
Tip: Reproduce bladder/rectum status at every fraction; an endorectal balloon changes the planning contour.
Rectum / anus
- Position
- Prone on belly-board (displaces small bowel) or supine per protocol; arms up/forward.
- Immobilise
- Indexed belly-board with abdominal aperture; foam/vac bag support.
- Coil setup
- Posterior in-couch + anterior flex on bridge; avoid compressing belly-board aperture.
- Landmarks
- Lasers to sacral/gluteal tattoos; reference at target level.
- Motion
- Free-breathing; bladder-full prep spares small bowel.
- QA note
- Verify couch/board flatness; distortion correction over pelvic FOV.
Tip: Belly-board only helps if bladder-filling and hip flexion are reproduced identically each day.
Gynae
- Position
- Supine, arms on chest, legs straight and neutral; head first.
- Immobilise
- Indexed knee/ankle fix or vac bag; consistent pelvic tilt.
- Coil setup
- Body array on bridge over pelvis; no compression of lower abdomen.
- Landmarks
- Lasers to pelvic tattoos; bladder-filling protocol (full or comfortably full).
- Motion
- Free-breathing; uterine/target motion managed by bladder/rectum prep.
- QA note
- Uterus/cervix near iso; large-FOV distortion correction; note brachytherapy applicators separately.
Tip: Bladder volume strongly shifts the uterus; standardise the fill and time it consistently.
Cervical / spine SBRT
- Position
- Supine, head first; neutral spine alignment along couch long axis.
- Immobilise
- Full-length vac bag; wing board for arms; thermoplastic mask for cervical/upper-thoracic levels.
- Coil setup
- Posterior in-couch spine array + anterior flex on bridge; no coil weight on the spine contour.
- Landmarks
- Lasers to spinal-level tattoos; vertebral reference for fusion.
- Motion
- Free-breathing; rigid immobilisation critical for sub-mm SBRT accuracy.
- QA note
- Distortion is a major SBRT risk; verify geometric accuracy at treated level, keep it near iso.
Tip: For cord-adjacent SBRT, confirm residual distortion at the target level is within your local QA tolerance before planning.
Extremity / sarcoma
- Position
- Limb-dependent; affected extremity positioned/abducted to spare adjacent tissue and skin folds.
- Immobilise
- Custom vac bag moulded to the limb; indexed to couch; consistent joint angle.
- Coil setup
- Flex coil wrapped loosely on a standoff/bridge, never strapped tight enough to compress the limb.
- Landmarks
- Lasers to limb tattoos; reference marks proximal and distal for long targets.
- Motion
- Free-breathing; limb SBRT relies on rigid, reproducible vac-bag fixation.
- QA note
- Off-centre anatomy suffers most distortion; position the target as close to bore centre as feasible and correct.
Tip: Off-axis limbs are the worst case for gradient non-linearity; centre the tumour as close to the bore centre as feasible as well as applying distortion correction.
Educational reference only. Follow your local departmental protocols and current professional guidance. Not a substitute for clinical judgement.
More in Running the department
Next in Running the department: MRI for allied health and referrers
Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.