Nuclear medicine exam prep
Single-best-answer practice for nuclear-medicine certification: technologist boards (NMTCB, ARRT(N), CAMRT) and the physics/safety knowledge physician trainees need. This free sampler shows the format and depth; the full bank is coming with Pro.
Physics & instrumentation
1 Which gamma-ray energy from Tc-99m does the gamma camera use for imaging?
A. Tc-99m emits a 140 keV gamma ray as it decays to Tc-99, and the gamma camera sets its photopeak window around that energy. 159 keV is the main I-123 photopeak, 364 keV is the I-131 gamma ray, and 511 keV is the annihilation photon energy imaged in PET.
2 A SPECT slice shows a ring artefact centred on the axis of rotation. What is the most likely cause?
C. A non-uniformity sits at the same detector position in every projection, so reconstruction turns it into a ring around the axis of rotation. IAEA and Zanzonico both link SPECT ring artefacts to non-uniformity, and a fresh high-count uniformity correction is the fix. Motion, attenuation and sparse angular sampling cause other artefact patterns.
3 Why does PET have much higher sensitivity than SPECT?
D. PET detects the two 511 keV annihilation photons in coincidence, a form of electronic collimation, so it needs no absorptive collimator. SPECT depends on a lead collimator, which gives up sensitivity for resolution. PET photons carry more energy than the 140 keV of Tc-99m, not less, and a PET ring uses no parallel-hole collimator.
4 After four physical half-lives, what percentage of the starting activity remains?
B. Each half-life halves the activity: 50%, 25%, 12.5%, then 6.25% (one-sixteenth) after the fourth. 12.5% is what remains after three half-lives and 3.1% after five.
More physics & instrumentation questions, coming with Pro
Full coverage of decay, detectors, collimation, SPECT/PET reconstruction, corrections and QC, with worked explanations and difficulty tiers.
Radiopharmacy
5 Which parent radionuclide is loaded on the column of the generator that supplies Tc-99m?
C. Mo-99, with a half-life of about 2.7 days, is adsorbed on an alumina column and decays to Tc-99m, which is eluted with saline as pertechnetate. Ge-68 and Sr-82 are the parents in the Ga-68 and Rb-82 PET generators, and Tc-99 is the long-lived ground state that Tc-99m decays to.
6 Free Tc-99m pertechnetate in a labelled preparation would show unexpected uptake in which sites?
A. Pertechnetate is trapped by the thyroid and taken up by gastric mucosa and the salivary glands. Activity in these organs on a scan with a labelled agent therefore points to free pertechnetate from poor labelling or breakdown of the preparation.
7 Which renal tracer is cleared mainly by tubular secretion and is used for dynamic renography?
D. MAG3 is removed from plasma mainly by proximal tubular uptake and secretion, and its extraction fraction is about twice that of DTPA. DTPA is purely filtered by the glomerulus, DMSA is retained in the renal cortex for static images, and MDP is a bone agent.
8 Which therapeutic radiopharmaceutical is paired with Ga-68 somatostatin-receptor PET for neuroendocrine tumours?
B. Ga-68 somatostatin analogue PET shows somatostatin receptor status, which selects patients for peptide receptor radionuclide therapy with Lu-177 DOTATATE; the higher the PET uptake, the better the expected response. The other options do not target somatostatin receptors.
More radiopharmacy questions, coming with Pro
Every kit, generator, radiochemical-purity/ITLC scenario and theranostic pair, with tagging-troubleshooting questions.
Safety & regulation
9 Which description best fits diagnostic reference levels (DRLs) in nuclear medicine?
A. ICRP Publication 135 describes DRLs as an optimisation tool: when a facility’s median value is above the DRL, practice is investigated. DRLs are not limits and do not apply to individual patients, and ICRP dose limits exclude medical exposure of patients.
10 Which preparation do procedure guidelines recommend before I-123 mIBG imaging?
C. Radioactive iodide can dissociate from mIBG, so guidance recommends blocking the thyroid with stable iodine, or perchlorate when iodine is contraindicated, starting before injection. Fasting, furosemide and caffeine are not part of mIBG preparation.
11 Under US NRC rule 10 CFR 35.75, a patient given radioactive material may be released when the dose to any other person is unlikely to exceed what?
B. 10 CFR 35.75 permits release when the total effective dose equivalent to any other individual is not likely to exceed 5 mSv (0.5 rem). Written instructions are needed if it is likely to exceed 1 mSv. The other values are not release criteria.
12 After which radiopharmaceutical should a patient stop breastfeeding altogether?
D. Radioiodine is excreted in breast milk, and ARSAC advises stopping breastfeeding altogether after I-131 iodide; after I-131 therapy it is completely contraindicated. FDG and MAG3 need no interruption or a short one, and pertechnetate needs a longer but time-limited break.
More safety & regulation questions, coming with Pro
ICRP principles, dose limits, DRLs, pregnancy/breastfeeding, release criteria, contamination + spill scenarios, and jurisdiction comparisons.
Procedures
13 On hepatobiliary scintigraphy the gallbladder still fails to fill after morphine augmentation. What does this most likely indicate?
C. Morphine contracts the sphincter of Oddi and raises biliary pressure, pushing bile into a patent cystic duct. Non-filling 30 minutes after morphine, with good liver uptake and clearance, indicates cystic duct obstruction, the hallmark of acute cholecystitis. In chronic cholecystitis the gallbladder usually fills after morphine or on delayed images.
14 On a ventilation-perfusion study, a segmental perfusion defect has normal ventilation. How is this described?
A. A segmental perfusion defect with preserved ventilation is a mismatch, the pattern of pulmonary embolism. EANM lists matched defects in atelectasis, tumour or empyema, and reverse mismatch, ventilation reduced more than perfusion, in early pneumonia and COPD. The stripe sign is preserved perfusion along the pleural edge.
15 On myocardial perfusion imaging, a defect seen on stress images fills in on rest images. What does this represent?
D. A stress defect that fills in at rest is reversible, the sign of inducible ischaemia. A defect on both stress and rest images is fixed: usually infarction, though hibernating myocardium can look the same. Gated images help, since a fixed defect with normal wall motion is typically attenuation artefact.
16 What uptake period do EANM guidelines recommend before whole-body FDG PET for tumour imaging?
B. EANM version 2.0 recommends 60 minutes, with an accepted range of 55 to 75 minutes. SNMMI guidance asks for at least 45 minutes and a constant uptake time, because SUV depends on it. Keep the interval the same when two scans are compared.
More procedure questions and rapid-fire cases, coming with Pro
Study set-up, timing, interpretation and pitfalls across all SPECT, PET and therapy procedures, plus timed mock papers mapped to NMTCB / ARRT(N) blueprints.
Study smart
Pair these questions with the reference pages: shore up weak spots with the Physics Atlas, Radiopharmaceutical Reference, Radiation Safety and Study Explorer. Then re-test until the explanations feel obvious.
Related
Sources
Sources (4)
4 sources listed
- NMTCB exam content outlines. nmtcb.org/exams
- ARRT(N) Nuclear Medicine Technology content specifications. arrt.org
- SNMMI / EANM procedure guidelines (clinical items). eanm.org
- Zanzonico P. Routine quality control of clinical nuclear medicine instrumentation: a brief review. J Nucl Med. 2008;49(7):1114–1131. doi.org/10.2967/jnumed.107.050203
High Yield Imaging · Exam Prep. Original teaching questions mapped to public blueprints, not reproduced exam content. Educational use only; verify clinical facts against current guidance.
More in Question banks and mocks
Where next: MRI interview guide
Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.