X-ray and CT question bank

X-ray and CT question bank

Twenty exam-standard single-best-answer questions across the four pillars of plain radiography: physics & equipment, dose & radiation protection, positioning & technique, and image interpretation & artifacts. Work through each question before revealing the answer. Every explanation tells you why the right option is right and why the others are wrong.

On this pagePhysics & equipmentDose & radiation protectionPositioning & techniqueInterpretation & artifactsSources

Physics & equipment

Beam production, exposure factors, geometry and the numbers examiners love: the 15% rule, penumbra arithmetic and filtration standards.

Start the questions (5)

1 An abdomen radiograph is taken at 70 kVp and 20 mAs. To reduce patient dose, the radiographer raises the kVp by about 15%, to about 80 kVp. What mAs keeps the receptor exposure about the same?

Correct answer: B

A rise of about 15% in kVp roughly doubles receptor exposure, so the mAs is halved: 20 to 10 mAs. 5 mAs would suit two 15% steps, 20 mAs ignores the change, and 40 mAs would roughly quadruple receptor exposure. The trade is slightly lower subject contrast.

Source: Chusin et al., J Assoc Med Sci 2024 (15% rule phantom study); Martin, Biomed Imaging Interv J 2007.

2 For an AP thoracic spine radiograph, how should the X-ray tube be turned to make use of the anode heel effect?

Correct answer: A

Beam intensity is higher on the cathode side, because photons heading towards the anode side pass through more of the target. With the cathode towards the feet, the stronger part of the beam covers the thicker lower thoracic spine, evening out exposure. Turning the tube across the patient moves the gradient side to side.

Source: Seibert and Boone, J Nucl Med Technol 2005 (tube orientation); Seibert, J Nucl Med Technol 2004.

3 A tungsten-target tube is operated at 60 kVp. Which best describes the X-ray beam it produces?

Correct answer: E

Tungsten’s K-shell binding energy is about 69.5 keV, so K-characteristic X-rays need a tube voltage above 69.5 kVp. At 60 kVp no electron can eject a K-shell electron, so the useful beam is bremsstrahlung. Option C describes a beam above 69.5 kVp.

Source: Seibert, J Nucl Med Technol 2004 (x-ray production); X-Ray Data Booklet, Table 1-1 (tungsten K-shell 69,525 eV).

4 What is the main purpose of the aluminium filtration in a general radiography X-ray tube?

Correct answer: D

Filtration removes low-energy photons that would be absorbed in the skin without reaching the detector, lowering entrance surface dose. It hardens the beam, raising its mean energy, which slightly lowers contrast. Grids deal with scatter, and because filtration reduces tube output it does not lower the mAs needed.

Source: Martin, Biomed Imaging Interv J 2007; Seibert, J Nucl Med Technol 2004.

5 A 0.6 mm focal spot is used at 100 cm SID, with the anatomy of interest 25 cm from the image receptor. What is the approximate geometric unsharpness at the image receptor?

Correct answer: C

By similar triangles, geometric unsharpness at the receptor = focal spot size × OID ÷ SOD, which equals focal spot size × (M − 1) because M = SID ÷ SOD. SOD is 100 − 25 = 75 cm, so 0.6 × 25 ÷ 75 = 0.20 mm. Dividing by the magnification (M = 100 ÷ 75) gives 0.15 mm, the blur referred back to object size; 0.60 mm ignores the geometry. A smaller focal spot, longer SID or shorter OID reduces it.

Source: Seibert and Boone, J Nucl Med Technol 2005 (geometric blur, M = SID/SOD).

Dose & radiation protection

Inverse square arithmetic, effective doses you must know cold, deviation index interpretation and the practical decisions that cut dose.

Start the questions (5)

6 Scatter air kerma is 1 µGy per exposure 2 m from a patient during a mobile examination. Treating the patient as a point source, what would it be at 1 m?

Correct answer: D

Intensity varies as 1/d². Halving the distance multiplies it by 2² = 4, so 1 µGy becomes 4 µGy. 2 µGy assumes a linear change, 8 µGy a cube law, and 0.25 µGy applies the law the wrong way round. The same law makes stepping back such effective protection.

Source: Rivera and Ahn, Cureus 2025.

7 A patient asks how the effective dose of an adult AP abdomen radiograph compares with a single PA chest radiograph. Which answer fits published typical values?

Correct answer: A

Mettler and colleagues give about 0.02 mSv for a PA chest and 0.7 mSv for an abdomen radiograph; a Korean dosimetry study gives 0.029 and 0.44 mSv. The abdomen is many times higher but still well below an abdomen and pelvis CT, about 7.7 mSv on RadiologyInfo.

Source: Mettler et al., Radiology 2008; Kim et al., J Radiat Prot 2002; RadiologyInfo.org, Radiation dose (2025).

8 A digital pelvis radiograph shows a deviation index (DI) of +3.0, as defined by AAPM Task Group 116. What does this mean?

Correct answer: D

DI = 10 × log10(indicated exposure ÷ target). A DI of +3 gives 10^0.3 ≈ 2, so the detector received about twice the target; each +1 is about 25% more. The image may still look good; TG-116 advises a repeat only if relevant anatomy is clipped or burned out.

Source: AAPM Report No. 116 (2009); Seibert and Morin, Pediatr Radiol 2011.

9 A small child’s abdomen measures about 9 cm thick, and the table Bucky has a 12:1 grid. Which single change most reduces the child’s dose for an AP abdomen?

Correct answer: C

A thin abdomen produces little scatter, so the grid adds little contrast but needs much more exposure; Bucky factors are typically 2 to 4. Image Gently advises avoiding grids for body parts under 10 to 12 cm thick. Lower kVp with more mAs raises dose, and focal spot size does not change it.

Source: Image Gently, Digital radiography; Fritz and Jones, Pediatr Radiol 2014; Seibert and Boone, J Nucl Med Technol 2005.

10 Skin erythema after a long interventional fluoroscopy procedure is best described as which type of radiation effect?

Correct answer: B

ICRP now calls deterministic effects tissue reactions: they appear only above a threshold dose, and severity increases with dose. Skin injury after long fluoroscopy is the classic example. For stochastic effects such as cancer, probability rises with dose with no threshold assumed, and severity does not depend on dose.

Source: ICRP Publication 103; ICRP Publication 118; AAPM MPPG 12.a, Fluoroscopy dose management.

Positioning & technique

Grid conversion arithmetic, magnification, projection criteria and the views that catch the fractures everyone else misses.

Start the questions (5)

11 A non-grid knee radiograph is satisfactory at 12 mAs. It is repeated in the Bucky with a 12:1 grid at the same kVp, and the department’s chart gives this grid a conversion factor of 5. What mAs keeps receptor exposure the same?

Correct answer: E

Grid conversion factor = mAs with grid ÷ mAs without grid, so the new mAs is 12 × 5 = 60. Dividing gives 2.4 mAs, adding gives 17 mAs and a factor of 4 gives 48 mAs. Published factors vary with grid ratio, kVp and source, so use the figure for your own equipment.

Source: Martin, Biomed Imaging Interv J 2007 (grid dose penalty); Seibert and Boone, J Nucl Med Technol 2005 (Bucky factor).

12 On a PA chest radiograph, which finding most reliably shows that the patient was rotated?

Correct answer: A

On a true PA chest the medial ends of the clavicles are equidistant from the spinous processes; rotation makes them unequal and can change lung density and the mediastinal outline. Rib count reflects inspiration, scapulae over the lungs reflect shoulder position, blurring suggests movement, and missing angles reflect field coverage.

Source: Kelly, Ulster Med J 2012.

13 A trauma lateral cervical spine radiograph shows C1 to C6 clearly, but the shoulders hide the C7–T1 junction. Which projection is designed to show this junction?

Correct answer: D

The swimmer’s lateral raises one arm above the head and lowers the other shoulder, separating the shoulder shadows so C7–T1 can be seen. Many trauma pathways go straight to CT when the junction stays hidden, and positioning must respect spinal precautions. The other options are frontal, soft-tissue or movement views, not designed to show this junction from the side.

Source: Rethnam et al., BMC Med Imaging 2008; Toksoy et al., Int J Emerg Med 2010.

14 Ulnar deviation is added to the PA wrist projection when a scaphoid fracture is suspected. What does it achieve?

Correct answer: B

Ulnar deviation rotates the scaphoid so it appears elongated, making a fracture line easier to see; radial deviation foreshortens it. Positioning the wrist does not change field size or dose. A normal first series does not exclude a scaphoid fracture, so follow the local pathway for further imaging.

Source: Bhat et al., Indian J Plast Surg 2011; Cheung et al., J Hand Surg Br 2006; Carpenter et al., Acad Emerg Med 2014.

15 An object 6 cm wide lies 25 cm from the image receptor and is radiographed at 100 cm SID. How wide is its image on the receptor?

Correct answer: D

Magnification = SID ÷ SOD. SOD is 100 − 25 = 75 cm, so M = 1.33 and the image is 6 × 1.33 = 8.0 cm. 7.5 cm comes from using 1 + OID/SID, 4.5 cm inverts the ratio, and 6.0 cm assumes no magnification. A long SID and short OID keep M close to 1.

Source: Seibert and Boone, J Nucl Med Technol 2005 (M = SID/SOD).

Image interpretation & artifacts

Measurements with hard numbers, the mimics that generate call-backs, and what to do with the incidental finding.

Start the questions (5)

16 An adult’s lateral cervical spine radiograph after a fall shows normal alignment, but the prevertebral soft tissue at C2 measures 10 mm. What is the best interpretation?

Correct answer: C

Traditional radiograph limits are about 6 to 7 mm at C2 and 22 mm at C6, so 10 mm at C2 suggests swelling such as haematoma and needs further assessment, usually CT, under the local pathway. These limits are specific but not sensitive: a normal measurement does not exclude injury.

Source: Savakus et al., JAAOS Glob Res Rev 2020; DeBehnke and Havel, Ann Emerg Med 1994; Matar and Doyle, Australas Radiol 1997.

17 A curvilinear line over the right hemithorax on a frontal chest radiograph raises the question of pneumothorax. Which feature best indicates a skin fold instead?

Correct answer: A

In a pneumothorax the space beyond the visceral pleural line holds air, with no vessels. Vessels crossing or beyond the line mean lung is still there, favouring a skin fold. A fine, sharp line favours pneumothorax, and folds often extend beyond the chest wall. If doubt remains, repeat the film after removing clothing and repositioning the arm.

Source: Kelly, Ulster Med J 2012; O’Connor and Morgan, BMJ 2005.

18 CT shows a single solid 5 mm lung nodule in a 45-year-old never-smoker with no cancer history and no immunosuppression. Under the Fleischner Society 2017 guidelines, what follow-up is advised?

Correct answer: E

Fleischner 2017 advises no routine follow-up for a solid nodule under 6 mm in a low-risk patient, as cancer risk is below 1%. CT at 6 to 12 months applies to 6 to 8 mm nodules, and the 3-month, PET/CT or sampling route to nodules over 8 mm. Screening programmes follow separate rules.

Source: MacMahon et al., Radiology 2017 (Fleischner Society).

19 A digital radiograph taken with a stationary grid shows a fine wavy pattern across the whole image. What is the most likely cause?

Correct answer: B

When the grid’s line frequency is close to the detector’s sampling frequency, the two patterns beat to form a moiré pattern. Fixes include a moving grid, a suitable line rate and orientation, or grid-line suppression software. Cut-off lightens the image, erasure faults leave a ghost image, and movement blurs.

Source: Lin et al., J Digit Imaging 2006; Shetty et al., AJR 2011.

20 Perforated viscus is suspected. Of these plain radiographic approaches, which is most sensitive for a small amount of free intraperitoneal gas?

Correct answer: C

Free gas rises under the diaphragm, where an erect chest radiograph shows it best, but small amounts need about 10 minutes upright to collect there. Supine films show free gas less often and only indirectly. A left lateral decubitus view suits patients who cannot sit up.

Source: Lee et al., J Korean Surg Soc 2012; Chiu et al., Am J Emerg Med 2009; Kelly, Ulster Med J 2012.

PRO

High Yield Pro

You’ve finished the free sampler. When Pro opens, it will add full timed mock examinations mapped to registration-exam blueprints, rapid-reporting stations with normal/abnormal image banks, and detailed performance breakdowns by topic.

  • The full single-best-answer question bank
  • Timed mock exams
  • Rapid-reporting stations
  • Full teaching explanations
  • Topic-by-topic score tracking

What Pro will include

Related
Back to contents

Sources

32 sources listed

  1. Shepard SJ, et al. An exposure indicator for digital radiography: AAPM Task Group 116. AAPM Report 116 (2009).
  2. Dave JK, et al. Current state of practice regarding digital radiography exposure indicators and deviation indices: AAPM TG-232. Med Phys 2018;45:e1146–e1160.
  3. Seibert JA. X-ray imaging physics for nuclear medicine technologists. Part 1: basic principles of x-ray production. J Nucl Med Technol 2004;32(3):139–147. tech.snmjournals.org
  4. Seibert JA, Boone JM. X-ray imaging physics for nuclear medicine technologists. Part 2: X-ray interactions and image formation. J Nucl Med Technol 2005;33(1):3–18. tech.snmjournals.org
  5. Martin CJ. Optimisation in general radiography. Biomed Imaging Interv J 2007;3(2):e18. doi:10.2349/biij.3.2.e18
  6. Kelly B. The chest radiograph. Ulster Med J 2012;81(3):143–148. PMC3632825
  7. Bhat AK, Kumar B, Acharya A. Radiographic imaging of the wrist. Indian J Plast Surg 2011;44(2):186–196. doi:10.4103/0970-0358.85339
  8. Mettler FA, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 2008;248:254–263.
  9. Health Physics Society. Doses from Medical X-Ray Procedures.
  10. ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007;37(2–4).
  11. MacMahon H, et al. Guidelines for management of incidental pulmonary nodules detected on CT images: Fleischner Society 2017. Radiology 2017;284:228–243.
  12. Shetty CM, et al. Computed radiography image artifacts revisited. AJR Am J Roentgenol 2011;196:W37–W47.
  13. O’Connor AR, Morgan WE. Radiological review of pneumothorax. BMJ 2005;330(7506):1493–1497. doi:10.1136/bmj.330.7506.1493
  14. DeBehnke DJ, Havel CJ. Utility of prevertebral soft tissue measurements in identifying patients with cervical spine fractures. Ann Emerg Med 1994;24(6):1119–1124. doi:10.1016/s0196-0644(94)70242-x
  15. Miller RE, Nelson SW. The roentgenologic demonstration of tiny amounts of free intraperitoneal gas. AJR 1971;112:574–585.
  16. Chusin T, Sudchai W, Jitnarin N, et al. The influence of 10 kVp and 15% rule applications on patient dose and image quality in extremities radiography: a phantom study. J Assoc Med Sci 2024;57(3):166–176. doi:10.12982/jams.2024.058
  17. Rivera K, Ahn SS. Radiation physics and safety in fluoroscopy: a clinician’s guide to principles and practice. Cureus 2025;17(7):e87245. doi:10.7759/cureus.87245
  18. Kim WR, Lee CS, Lee JK. Assessment of effective dose from diagnostic X-ray examinations of adult. J Radiat Prot 2002;27(3):155. jrpr.org
  19. Seibert JA, Morin RL. The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric population. Pediatr Radiol 2011;41(5):573–581. doi:10.1007/s00247-010-1954-6
  20. Fritz S, Jones AK. Guidelines for anti-scatter grid use in pediatric digital radiography. Pediatr Radiol 2014;44(3):313–321. doi:10.1007/s00247-013-2824-9
  21. ICRP; Stewart FA, Akleyev AV, Hauer-Jensen M, et al. ICRP Publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs — threshold doses for tissue reactions in a radiation protection context. Ann ICRP 2012;41(1–2):1–322. doi:10.1016/j.icrp.2012.02.001
  22. Fisher RF, Applegate KE, Berkowitz LK, et al. AAPM Medical Physics Practice Guideline 12.a: fluoroscopy dose management. J Appl Clin Med Phys 2022;23(3):e13526. doi:10.1002/acm2.13526
  23. Rethnam U, Yesupalan RS, Bastawrous SS. The Swimmer’s view: does it really show what it is supposed to show? A retrospective study. BMC Med Imaging 2008;8:2. doi:10.1186/1471-2342-8-2
  24. Toksoy A, Bektas F, Eken C, et al. Value of the swimming position and arm traction in visualizing the cervicothoracic junction over the standard lateral cervical X-ray. Int J Emerg Med 2010;3(2):85–90. doi:10.1007/s12245-010-0159-y
  25. Cheung GC, Lever CJ, Morris AD. X-ray diagnosis of acute scaphoid fractures. J Hand Surg Br 2006;31(1):104–109. doi:10.1016/j.jhsb.2005.09.001
  26. Carpenter CR, Pines JM, Schuur JD, et al. Adult scaphoid fracture. Acad Emerg Med 2014;21(2):101–121. doi:10.1111/acem.12317
  27. Savakus JC, Weinberg DS, Moore TA, et al. Prevertebral soft-tissue swelling at C7 is highly sensitive for cervical spine ligamentous injury. J Am Acad Orthop Surg Glob Res Rev 2020;4(4):e19.00093. doi:10.5435/JAAOSGlobal-D-19-00093
  28. Matar LD, Doyle AJ. Prevertebral soft-tissue measurements in cervical spine injury. Australas Radiol 1997;41(3):229–237. doi:10.1111/j.1440-1673.1997.tb00665.x
  29. Lin CY, Lee WJ, Chen SJ, et al. A study of grid artifacts formation and elimination in computed radiographic images. J Digit Imaging 2006;19(4):351–361. doi:10.1007/s10278-006-0630-8
  30. Lee CH, Kim JH, Lee MR. Postoperative pneumoperitoneum: guilty or not guilty? J Korean Surg Soc 2012;82(4):227–231. doi:10.4174/jkss.2012.82.4.227
  31. Chiu YH, Chen JD, Tiu CM, et al. Reappraisal of radiographic signs of pneumoperitoneum at emergency department. Am J Emerg Med 2009;27(3):320–327. doi:10.1016/j.ajem.2008.03.004
  32. Image Gently Alliance. Image Gently and Digital Radiography. Web page, copyright 2025 Image Gently and SPR.

Educational content for exam preparation only, not a substitute for local protocols, statutory dose regulations or clinical judgement. Numeric values (dose figures, grid factors, measurement thresholds) are representative values from the cited studies and guidelines; local diagnostic reference levels and departmental standards take precedence.

More in Question banks and mocks

Next in Question banks and mocks: Radiography exams and registration

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.

Contents