X-ray and radiation dose calculators
Exposure factors, geometry, attenuation and dose estimation for general radiography, fluoroscopy and dental imaging
Teaching aids, not QC instruments and not clinical advice. Output values (µGy/mAs), grid conversion factors and backscatter factors vary by tube, filtration, kVp and geometry; use your own measured values. Effective dose is a population risk metric for a reference phantom, never an individual patient’s dose.
Technique factors
Enter mAs once for inverse square and grid conversion. The 15% rule and entrance surface dose take their own mAs.
Inverse square & mAs distance compensation
I₂ = I₁(d₁/d₂)² · mAs₂ = mAs₁(d₂/d₁)²
Intensity falls with the inverse square; to hold receptor dose constant, mAs rises with the direct square. The two exponents point opposite ways. That’s the classic exam trap.
Grid conversion
mAs₂ = mAs₁ × (GCF₂ ÷ GCF₁)
Grid conversion factors are nominal and kVp-dependent; these are the conventional teaching values. Adding a grid raises patient dose in direct proportion to the mAs increase.
15% rule: kVp ↔ mAs
steps = ln(kVp₂/kVp₁) ÷ ln(1.15) · mAs₂ = mAs₁ ÷ 2^steps
A 15% kVp rise roughly doubles receptor exposure, so mAs is halved to compensate. Handles fractional steps. Raising kVp lowers patient dose but also lowers subject contrast.
Reciprocity & exposure time
mAs = mA × t · t = mAs ÷ mA
Reciprocity holds for film-screen and digital receptors across normal clinical ranges. The shortest time that gives the required mAs is the one that best controls motion blur.
Entrance surface dose
ESD = output × mAs × (100/SSD)² × BSF
Output must be your own measured value for that tube, kVp and filtration. BSF is typically 1.25–1.45 for diagnostic beams and field sizes.
Dose in context
Effective dose is entered in mSv and cumulative dose in µSv.
Effective dose in context
CXR equivalents = E ÷ 0.02 mSv · background days = E ÷ 2.4 × 365
A PA chest radiograph is taken as 0.02 mSv, world-average natural background as 2.4 mSv/year (UNSCEAR). Both are population averages, useful for explaining scale, not for individual risk.
Cumulative dose & background equivalent
total = Σ(n × E) · background days = total ÷ 2.4 × 365
Useful for explaining a bitewing series to a patient. Doses add; risk from repeated small exposures is assumed to add linearly, which is a conservative modelling choice, not an observed fact.
Dental: DAP to effective dose
E = DAP × conversion coefficient
Conversion coefficients are strongly technique- and field-specific: an intraoral coefficient does not apply to a panoramic or CBCT exposure. Supply the coefficient published for your own examination type.
Standalone calculators
Magnification
MF = SID ÷ SOD · object = image ÷ MF
SOD = SID − OID. Magnification is why cardiothoracic ratio is only assessed on a PA film at a long SID; an AP mobile at 100 cm magnifies the heart.
Half-value layer attenuation
transmitted = 0.5^(thickness ÷ HVL)
Strictly true only for a monoenergetic beam. A real polyenergetic beam hardens as it passes through the absorber, so the second HVL is thicker than the first; this underestimates transmission at depth.
Exposure index & deviation index
DI = 10 · log₁₀(EI ÷ EIₜ)
DI 0 = target. Roughly: ±0.5 ideal, ±1 acceptable, +3 means about twice the intended dose, −3 means half. IEC 62494-1 / AAPM TG-116.
Fetal dose in context
fraction of the 100 mGy threshold = dose ÷ 100
ACOG and ACR state that fetal doses below 50 mGy have not been associated with an increased risk of fetal anomalies or loss; 100 mGy is the level below which termination is not justified on radiation grounds (ICRP 84). A diagnostic radiograph almost never approaches these numbers. Any actual fetal dose estimate must come from a medical physicist, not this page.
Sources
Exposure index / deviation index: IEC 62494-1; AAPM Task Group 116, An Exposure Indicator for Digital Radiography (2009).
Background radiation: UNSCEAR 2008 Report, world-average natural background 2.4 mSv/y. Chest radiograph dose: nominal 0.02 mSv (PA projection).
Fetal dose: ICRP Publication 84, Pregnancy and Medical Radiation; ACOG Committee Opinion 723 (2017); ACR–SPR Practice Parameter for imaging pregnant patients.
Dental dose: conversion coefficients are examination-specific; see the UK HPA/SEDENTEXCT guidance and ICRP 103 tissue weighting. No coefficient is supplied by default because none is universal.
Educational content only. Backscatter factors, grid conversion factors and tube outputs must come from your own measurements.