Mammography calculators

Mammography calculators

Mean glandular dose, screening audit metrics, compression force conversion and a BI-RADS density reference

Teaching aids, not equipment QC, and not a substitute for a qualified medical physicist or your accreditation manual. Dose limits are quoted for the phantom and geometry in which they were defined; a per-view figure computed here does not demonstrate compliance. Audit thresholds are population benchmarks, not targets for an individual reader.

Dose

One bench · mean glandular dose

Mean glandular dose

AGD = Kair × conversion factor · per breast = AGD × views

The conversion factor bundles the g, c and s factors and depends on HVL, target/filter, breast thickness and glandularity. It must come from your physicist’s report for that unit; the values filled in here are examples: enter your own. MQSA limits dose to 3.0 mGy per view for a 4.2 cm compressed breast; EUREF (4th edition) sets acceptable and achievable levels per PMMA thickness: for 4.5 cm PMMA (a 5.3 cm equivalent breast), below 2.5 mGy acceptable and 2.0 mGy achievable.

Per view 1.68 mGy per view Dance DR et al. Phys Med Biol 2000;45:3225–3240
Per breast 3.36 mGy per breast over 2 views MQSA 21 CFR 900.12(e)(5)(vi) · EUREF 4th ed.

Screening audit

One bench · six counts, seven metrics

Screening audit metrics

recall = recalls ÷ exams · CDR = cancers ÷ exams × 1000 · PPV₁ = cancers ÷ recalls

Specificity uses TN = exams − recalls − FN and FP = recalls − TP. Sensitivity and specificity here depend on the follow-up interval used to define an interval cancer, so they are only comparable within one audit’s own definitions.

Recall and cancer detection Recall 9.5% · CDR 5.0 per 1000 Lehman CD et al. Radiology 2017;283:49–58 (BCSC)
Positive predictive value PPV₁ 5.3% · PPV₂ 16.7% · PPV₃ 17.9% ACR BI-RADS Atlas 5th ed., follow-up and outcome monitoring
Sensitivity and specificity Sensitivity 83.3% · Specificity 90.9% ACR BI-RADS Atlas 5th ed., follow-up and outcome monitoring
Benchmark check Outside typical benchmarks: PPV₂ below 20%. Lehman CD et al. Radiology 2017;283:49–58 (BCSC)

Reference

Standalone · one plate each

Compression force conversion

1 daN = 10 N · 1 lbf = 4.4482216 N

EUREF (4th edition) gives no optimal compression force: compression should be firm but tolerable. Its equipment check sets the maximum automatically applied force at 130–200 N, held for at least 1 minute, with the displayed force within ± 20 N of the measured value. Force is not pressure: the same force over a smaller contact area produces more pressure and more discomfort.

Converted 15.0 daN · 33.7 lbf Unit definitions · 1 daN = 10 N · 1 lbf = 4.4482216 N
EUREF equipment range 150 N · within 130–200 N, the EUREF test range for the maximum force a unit applies automatically (an equipment check, not a clinical target)

BI-RADS breast density

Reference: assessed visually, not computed

BI-RADS 5th edition deliberately removed the percentage ranges that the 4th edition used. Density is a qualitative visual assessment of the potential for masking, not a measured proportion. Categories c and d are the “dense” categories for notification purposes in many jurisdictions.

Category Category c: Heterogeneously dense ACR BI-RADS Atlas 5th ed., breast composition
Masking risk Moderate masking risk; small masses may be obscured. A dense category; supplemental screening may be discussed, particularly with elevated risk.

Sources

Dose limits: MQSA, 21 CFR 900.12(e)(5)(vi): 3.0 mGy per view for a 4.2 cm compressed breast phantom. European guidelines for quality assurance in breast cancer screening and diagnosis, 4th edition (EUREF, 2006), acceptable and achievable AGD levels per PMMA thickness, s. 2a.2.5.1 (p. 83) and Table 2b.2 (p. 149); compression, s. 2a.2.1.4 (p. 76).

Dose formalism: Dance DR et al. Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys Med Biol 2000;45:3225–3240.

Audit metrics and benchmarks: Lehman CD et al. National performance benchmarks for modern screening digital mammography. Radiology 2017;283:49–58 (BCSC). ACR BI-RADS Atlas, 5th edition: follow-up and outcome monitoring.

Density: ACR BI-RADS Atlas, 5th edition: mammography, breast composition.

Educational content only. Conversion factors and dose limits must be applied by a qualified medical physicist.

Related guides

More in Calculators

Next in Calculators: DXA calculators

Everything on Tools: Tools hub. Or search the whole reference.

Educational reference for imaging professionals. It does not replace local policy, manufacturer labelling or clinical judgement.