Mammography calculators
Mean glandular dose, screening audit metrics, compression force conversion and a BI-RADS density reference
Teaching aids, not equipment QC, and not a substitute for a qualified medical physicist or your accreditation manual. Dose limits are quoted for the phantom and geometry in which they were defined; a per-view figure computed here does not demonstrate compliance. Audit thresholds are population benchmarks, not targets for an individual reader.
Dose
One bench · mean glandular dose
Mean glandular dose
AGD = Kair × conversion factor · per breast = AGD × views
The conversion factor bundles the g, c and s factors and depends on HVL, target/filter, breast thickness and glandularity. It must come from your physicist’s report for that unit; the values filled in here are examples: enter your own. MQSA limits dose to 3.0 mGy per view for a 4.2 cm compressed breast; EUREF (4th edition) sets acceptable and achievable levels per PMMA thickness: for 4.5 cm PMMA (a 5.3 cm equivalent breast), below 2.5 mGy acceptable and 2.0 mGy achievable.
Screening audit
One bench · six counts, seven metrics
Screening audit metrics
recall = recalls ÷ exams · CDR = cancers ÷ exams × 1000 · PPV₁ = cancers ÷ recalls
Specificity uses TN = exams − recalls − FN and FP = recalls − TP. Sensitivity and specificity here depend on the follow-up interval used to define an interval cancer, so they are only comparable within one audit’s own definitions.
Reference
Standalone · one plate each
Compression force conversion
1 daN = 10 N · 1 lbf = 4.4482216 N
EUREF (4th edition) gives no optimal compression force: compression should be firm but tolerable. Its equipment check sets the maximum automatically applied force at 130–200 N, held for at least 1 minute, with the displayed force within ± 20 N of the measured value. Force is not pressure: the same force over a smaller contact area produces more pressure and more discomfort.
BI-RADS breast density
Reference: assessed visually, not computed
BI-RADS 5th edition deliberately removed the percentage ranges that the 4th edition used. Density is a qualitative visual assessment of the potential for masking, not a measured proportion. Categories c and d are the “dense” categories for notification purposes in many jurisdictions.
Sources
Dose limits: MQSA, 21 CFR 900.12(e)(5)(vi): 3.0 mGy per view for a 4.2 cm compressed breast phantom. European guidelines for quality assurance in breast cancer screening and diagnosis, 4th edition (EUREF, 2006), acceptable and achievable AGD levels per PMMA thickness, s. 2a.2.5.1 (p. 83) and Table 2b.2 (p. 149); compression, s. 2a.2.1.4 (p. 76).
Dose formalism: Dance DR et al. Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys Med Biol 2000;45:3225–3240.
Audit metrics and benchmarks: Lehman CD et al. National performance benchmarks for modern screening digital mammography. Radiology 2017;283:49–58 (BCSC). ACR BI-RADS Atlas, 5th edition: follow-up and outcome monitoring.
Density: ACR BI-RADS Atlas, 5th edition: mammography, breast composition.
Educational content only. Conversion factors and dose limits must be applied by a qualified medical physicist.